Reports until 12:33, Friday 28 February 2025
H1 CDS
david.barker@LIGO.ORG - posted 12:33, Friday 28 February 2025 (83112)
VACSTAT detected BSC3 sensor glitch, restarted VACSTAT to clear

At 12:17:36 PST we had a sensor glitch on the BSC3 gauge, causing a VACSTAT alarm. This was another of the periodic glitches, the pressure stepped up from 3.2e-08 to 4.9e-08 for 4 seconds and then immediately stepped back.

vacstat_ioc.service was restarted at 12:26.