Single Bounce OMC scans taken - process notes

I took two single bounce OMC scans today with the help of TJ and Tony. Here are some notes to future me and others to reference if we want to do single bounce scans:

Edit to add: unfortunately the scan results from today look pretty bad. In short, the peaks look "lopsided" somehow, and so I'm not sure the results are usable. Looking back at Jennie W's previous scans, it looks like she had to slow them down to 200 second scans. I only did a 100 second scan with amplitude 105 so maybe I scanned too fast. I'm not sure what the correct resolution of this is, because the scans I did in 2022 were 100 second scans and the results were fine. Adding this note here for reference in the future to think about the appropriate scan length and amplitude.

Comments related to this report
elenna.capote@LIGO.ORG - 10:00, Friday 16 May 2025 (84429)

Jennie, Sheila, and I ran OMC scans this morning and realized that the proper way to slow down the scan to avoid weird saturation effects is to reduce the excitation frequency in the template. The nominal templates have excitation frequencies of 0.01 Hz, so sweeping over 200 seconds just sweeps at the same speed twice. To sweep once, slower, you have to increase the sweep time to 200 seconds AND reduce the sweep frequency to 0.005 Hz.

elenna.capote@LIGO.ORG - 10:05, Friday 16 May 2025 (84430)

Sheila and I want to note some things that are "obvious" but easily forgotten:

  • high voltage must be on
  • sensor correction must be on