david.barker@LIGO.ORG - posted 10:16, Monday 24 March 2025 - last comment - 14:52, Monday 24 March 2025(83522)
BSC3 sensor glitch caused multiple VACSTAT alarms with new trip levels

This morning at 09:04:42 Mon 24mar2025 we had a BSC3 sensor gitch, which VACSTAT normally logs as a single gauge event and does not send alarms. Today however many LVEA gauges also tripped and alarms were sent to the Vacuum Team.

The reason is that on 13mar2025 we had a lock-loss due to an LVEA vacuum event which was below VACSTAT's nominal trip level. To counter this some LVEA slope trip levels were reduced from 1.0e-10 to 3.0e-12 (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=83366)

Fast forward to today and the sequence was:

09:04:42 standard BSC3 sensor glitch, num_glitched = 1, all other gauges set to sensitive mode, meaning 3.0e-13 for most LVEA gauges

09:05:03 2nd LVEA sensor glitched on noise while in sensitive mode, num_glitched = 2, Alarms sent to Vacuum Team

09:11:07 3rd LVEA sensor glitched on noise while in sensitive mode

09:12:37 4th LVEA sensor glitched on noise while in sensitive mode

09:18:35 5th LVEA sensor glitched on noise while in sensitive mode

 

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 10:42, Monday 24 March 2025 (83524)
david.barker@LIGO.ORG - 14:52, Monday 24 March 2025 (83535)

Vacstat code was changed to permit a sensitivity multiplier to be applied on a gauge-by-gauge basis. The multiplier for those LVEA gauges which have a 3.0e-12 slope trip was set to 1.0 (i.e. no additional sensitivity following a BSC3 trip). The increased sensitivity for the Delta-P trip level was retained.

prod.yaml

---
ifo: H1
glitch_monitor:
  lookback_times: [60, 300, 600]
  proc_period_secs: 10
  vacuum_gauges:
    default:
      glitch_press_rate: [1.0e-10, 1.0e-10, 1.0e-10]
      glitch_delta_press: [1.0e-08, 1.0e-08, 1.0e-08]
      valid_value_min: 1.0e-10
      valid_value_max: 1.0e-04
      sensitivity_press_multiplier: 10.0
      sensitivity_deltap_multiplier: 10.0
    H0:VAC-LY_X0_PT100B_PRESS_TORR:
      description: "Corner Station HAM1"
    H0:VAC-LY_Y1_PT120B_PRESS_TORR:
      description: "Corner Station BSC2"
      glitch_press_rate: [3.0e-12, 3.0e-12, 3.0e-12]
      sensitivity_press_multiplier: 1.0
    H0:VAC-LX_Y8_PT132_MOD2_PRESS_TORR:
      description: "Corner Station BSC3"
    H0:VAC-LX_Y0_PT110_MOD1_PRESS_TORR:
      description: "Corner Station HAM6"
    H0:VAC-LY_Y3_PT114B_PRESS_TORR:
      description: "Corner Station CP1"
      glitch_press_rate: [3.0e-12, 3.0e-12, 3.0e-12]
      sensitivity_press_multiplier: 1.0
    H0:VAC-LX_X3_PT134B_PRESS_TORR:
      description: "Corner Station CP2"
      glitch_press_rate: [3.0e-12, 3.0e-12, 3.0e-12]
      sensitivity_press_multiplier: 1.0
    H0:VAC-LY_Y4_PT124B_PRESS_TORR:
      description: "Corner Station Y-Arm"
      glitch_press_rate: [3.0e-12, 3.0e-12, 3.0e-12]
      sensitivity_press_multiplier: 1.0
    H0:VAC-LX_X4_PT144B_PRESS_TORR:
      description: "Corner Station X-Arm"
      glitch_press_rate: [3.0e-12, 3.0e-12, 3.0e-12]
      sensitivity_press_multiplier: 1.0
    H0:VAC-MY_Y1_PT243B_PRESS_TORR:
      description: "Mid-Y Y1 Beam Tube"
    H0:VAC-MY_Y5_PT246B_PRESS_TORR:
      description: "Mid-Y Y2 Beam Tube"
    H0:VAC-EY_Y6_PT427_MOD1_PRESS_TORR:
      description: "End-Y Y6 BT Ion Pump"
    H0:VAC-EY_Y1_PT423B_PRESS_TORR:
      description: "End-Y Beam Tube"
    H0:VAC-EY_Y3_PT410B_PRESS_TORR:
      description: "End-Y BSC10"
    H0:VAC-MX_X1_PT343B_PRESS_TORR:
      description: "Mid-X X1 Beam Tube"
    H0:VAC-MX_X5_PT346B_PRESS_TORR:
      description: "Mid-X X2 Beam Tube"
    H0:VAC-EX_X6_PT527_MOD1_PRESS_TORR:
      description: "End-X X6 BT Ion Pump"
    H0:VAC-EX_X1_PT523B_PRESS_TORR:
      description: "End-X Beam Tube"
    H0:VAC-EX_X3_PT510B_PRESS_TORR:
      description: "End-X BSC9"
 