Reports until 14:18, Wednesday 25 June 2025
H1 SQZ
camilla.compton@LIGO.ORG - posted 14:18, Wednesday 25 June 2025 (85336)
Adjusted SQZ BLRMS #2 to be 90 to 250Hz

Sheila, Camilla 

After we found that yellow BLRMs wasn't the best measure of the range 84997, I edited SQZ BLRMs #2 to be a 90 - 250Hz BLRMS, see in the attached plot that it is a measure of SQZ angle, but doesn't seem any better than yellow (#3) right now. If we do PSAMs scans in the future, we should look at the orange BLRMs as another measure of low freq range.

Images attached to this report