H1 GRD (SQZ)
matthewrichard.todd@LIGO.ORG - posted 17:07, Friday 27 June 2025 (85400)
Guardian Change to choose turn on SQZ_ASC after thermalizing

M. Todd, S. Dwyer


This morning we made a change to the thermalization guardian (managed by ISC) with flags that allow the SQZ guardian to turn on the SQZ_ASC in the unthermalized state or this guardian can turn on SQZ_ASC after 75 minutes (thermalized-ish). We also turned the monitoring off so that this switch does not knock out observing.

For more info see:

/opt/rtcds/userapps/release/isc/h1/guardian/THERMALIZATION.py