Reports until 16:42, Monday 30 June 2025
H1 General
anthony.sanchez@LIGO.ORG - posted 16:42, Monday 30 June 2025 - last comment - 17:19, Monday 30 June 2025(85444)
Monday Ops Eve Shift report.

TITLE: 06/30 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 10mph Gusts, 4mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.08 μm/s
QUICK SUMMARY:

H1 Is Currently Locked at NLN {600} But not Observing due to a SQZ_SHG issue:
Error SQZ_SHG [LOCKED.run] USERMSG 0: PZT voltage limits exceeded.
Sheila is currently working on the issue.
 

Comments related to this report
anthony.sanchez@LIGO.ORG - 17:19, Monday 30 June 2025 (85445)OpsInfo, SQZ

Sheila Fixed the issue, I have attached soem pictures to help future Ops that find this issue:
Sitemap -> SQZ -> SQZ_OVERVIEW -> SQZT0 ->  PZT -> Move H1:SQZ-SHG_PZT_OFFSET slider.

We Now have a "pump fiber rej power in ham7, nominal 35e-3, align fiber pol on sqzt0" error message on SQZ-OPO_LR. nopted will do tomorrow.

Then we took SQZ_ANG_ADJUST Guardian to ADJUST_SQZ_ANG_ADF [10]  which kinda worked but didn't  improve sqzing and reverted back.

I then accepted the SDFs and now we're back in Observing at 00:07:25 UTC.

An NDScope of what the GRD-SQZ_SHG Guardian was doing.

Images attached to this comment