Reports until 11:19, Thursday 11 September 2025
H1 ISC (PSL)
camilla.compton@LIGO.ORG - posted 11:19, Thursday 11 September 2025 - last comment - 14:01, Thursday 11 September 2025(86855)
Looking into if ISS 2nd Loop is causing IFO Glitches

Sheila, Ryan S, Tony, Oli, Elenna, Camilla, TJ ...

From Derek's analysis 86848, Sheila was suspicious that the ISS second loop was causing the glitches, see attached. The ISS turning on on a normal lock vs in this low-range glitchy lock attached, it's slightly more glitchy in this lock.

The IMC OLG was checked 86852.

Sheila and Ryan unlocked the ISS 2nd loop at 16:55UTC. This did not cause a lockloss although Shiela found that the IMC WFS sensors saw a shift in alignment which is unexpected.

See the attached spectrum of IMC_WFS_A_DC_YAW and LSC_REFL_SERVO (identified in Derek's search), red is with the PSL 2nd loop ON, blue is off. There is no large differences so maybe the ISS 2nd loop isn't to blame.

We lost lock at 17:42UTC, unknown why but the buildups starting decreasing 3 minutes before the LL. It could have been from Sheila changing alignments of the IMC PZT.

Sheila found IMC REFL DC channels are glitching whether the IMC is locked or unlocked, plot attached. But this seems to be the case even before the power outage.

Once we unlocked, Oli out the IMC PZTs back to their location before the power outage, attached is what IMC cameras looks like locked at 2W after WFS converged for a few minutes.

Images attached to this report
Comments related to this report
elenna.capote@LIGO.ORG - 11:26, Thursday 11 September 2025 (86856)

Here are three spectrograms of the IMC WFS DC YAW signal: before the power outage, after the power outage with the ISS on, after the power outage with the ISS off.

I don't see glitches in the before outage spectrogram, but I do see glitches in BOTH the ISS on and ISS off spectrograms. The ISS off spectrogram shows more noise in general but it appears that it has the same amount of nonstationary noise.

Images attached to this comment
camilla.compton@LIGO.ORG - 12:05, Thursday 11 September 2025 (86858)

IM4 trans with IMC locked at 2W before the power outage was PIT, YAW: 0.35,-0.11. After the outage PIT, YAW: 0.39,-0.08. Plot attached. These are changes of 0.04 in PIT, 0.03 in YAW. 

Oli and Ryan found that the alignment on IMC REFL WFS was the same but PSL-ISS_SEONDLOOP different, this makes us think the alignment change is in the MC or IMs. The IMs osems haven't changed considerably.

Images attached to this comment
camilla.compton@LIGO.ORG - 14:01, Thursday 11 September 2025 (86862)PSL

Ryan, Sheila

Ryan and Sheila noticed that the PSL-PMC_TEMP changed ~0.5deg without the set point changing. After looking into his we compared to the April power outage where it came back 1deg cooler. We therefore don't think this is causing us any issues. Both bower outages plotted in attached.

We did notice tough that some glitches in PMC-TEMP_OUT started ~ 31st May 2025 and have been present since, plot attached. Tagging PSL.

Images attached to this comment