Compaing now to a time before the power outage, the ISS diffracted power is now the same (3.6%). The PMC transmission has dropped by 2%, and the PMC reflection has increased by 3.5%.
Looking at IMC refl, before the power outage when the IMC was locked with 2W input power.
time | IMC refl | refl percent of before outage | MC2 trans | MC2 trans % of before outage |
9/10 8:12 UTC (before power outage, 2W IMC locked) | 0.74 | 317 | ||
9/11 00:22 UTC (IMC relocked at 2W after outage) | 1.15 | 155% | 312 | 98% |
9/11/1:50 UTC (after first 60 W and quick lockloss, IMC relocked at 2W) | 1.27 | 171% | 310 | 97% |
9/11 18:17 UTC (after overnight 60W lock)** had one IMC ASC loop off | 2.06 | 278% | 278 | 87% |
9/11 19:22 UTC (after all IMC ASC on) | 2.13 | 287% | 301 | 95% |
9/11 21:17 UTC (after sitting at 2W for 3 hours) | 2.03 | 274% | 303 | 95% |
The attached screenshot shows that the ISS second loop increased the IMC input power during the overnight 60W lock to keep the IMC circulating power constant.