Reports until 07:34, Tuesday 21 October 2025
H1 General
oli.patane@LIGO.ORG - posted 07:34, Tuesday 21 October 2025 - last comment - 08:00, Tuesday 21 October 2025(87601)
Ops Day Shift Start

TITLE: 10/21 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Microseism
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: USEISM
    Wind: 8mph Gusts, 5mph 3min avg
    Primary useism: 0.08 μm/s
    Secondary useism: 0.71 μm/s 
QUICK SUMMARY:

Detector was put in DOWN earlier due to very high secondary ground motion. Maintenance day is today.

Comments related to this report
david.barker@LIGO.ORG - 08:00, Tuesday 21 October 2025 (87602)

Several Guardian nodes reloaded at 07:45 this morning, modifed sqzparams.py

sqzparams.py was modifed yesterday evening at 19:40 and SQZ_ANG_ADJUST was reloaded at that time to apply the code change. Five other Guardian nodes import sqzparams.py and were in CFC state this morning.

I took the opportunity to test the automatic loading of these nodes after verifying none use the modified lines in sqzparams.py. The command is

    guardian_modified_not_loaded --load-nodes

a confirmation is required before proceeding with the load. The nodes loaded were (pre-load params shown):

======================================================================================
Node            Status      File          Source Date            Running Date
======================================================================================
SQZ_CLF_LR      NOT LOADED  sqzparams.py  19:40 Mon 20oct2025    08:49 Tue 14oct2025
SQZ_FC          NOT LOADED  sqzparams.py  19:40 Mon 20oct2025    10:19 Mon 20oct2025
SQZ_MANAGER     NOT LOADED  sqzparams.py  19:40 Mon 20oct2025    10:19 Mon 20oct2025
SQZ_OPO_LR      NOT LOADED  sqzparams.py  19:40 Mon 20oct2025    19:29 Mon 20oct2025
THERMALIZATION  NOT LOADED  sqzparams.py  19:40 Mon 20oct2025    10:18 Mon 20oct2025
======================================================================================
GRD CFC is now green.