Since it was installed, HAM7 ISI has been prone to trips if people are working nearby. Often walking by the chamber is enough, but work in the squeeze racks, HAM6 ISC racks or unplugging electronics from nearby outlets has caused CPS saturations. After talking to Jeff, we decided to try disconnecting HAM7 from the corner timing system and try running the ISI using it's own local 10khz timing clock signal. I pulled the P3 jumper from the H1 CPS and disconnected the timing DB9 that connects HAM7 to the timing system. Position readout seems unchanged before vs. after the change. This shouldn't affect anything else.