I needed to retake measurements for the SR3 M1 to M1 L estimator in the L and P degrees of freedom because the ones from 87363 didn't have good enough coherence.
Settings:
- ISI in ISOLATED
- SUS in ALIGNED
- Estimators off (using regular damping)
- L DAMP gain to 20% (-0.1), P DAMP gain at 20% (-0.1), Y DAMP gain to 100% (-0.5)
Results:
M1 to M1
/ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/SAGM1/Data/2025-10-21_1600_H1SUSSR3_M1_WhiteNoise_EstL_{L,P}_0p02to50Hz.xml r12738
These have a lot better coherence now