Reports until 10:58, Tuesday 18 November 2025
H1 ISC
matthewrichard.todd@LIGO.ORG - posted 10:58, Tuesday 18 November 2025 - last comment - 14:45, Wednesday 23 September 2026(88155)
Summary Table of OMC Scan and Gouy Phase measurements

M. Todd, J. Wright, S. Dwyer


Here is my attempt to summarize as many of the OMC scan measurements of the input beam overlap with the OMC mode, as well as PRC and SRC gouy phases -- all at different thermal states.

Measurement Time Test Masses CO2 [W] Ring Heater (per segment) [W] SR3 [W] OM2 [W] FOM aLOG
OMC Scan - Single Bounce off of ITMY 1443895154 Cold 0 0 0 0 Mismatch = 8.3% 87461
OMC Scan - Single Bounce off of ITMX 1443894875 Cold 0  0.45  0 0 Mismatch = 10.4% 87461
OMC Scan - Single Bounce off of ITMY 1443889943 Cold 1.7 0 0 0 Mismatch = 10.3% 87461
OMC Scan - Single Bounce off of ITMX 1443894875 Cold 1.7  0.45  0 0 Mismatch = 13.5% 87461
OMC Scan - Single Bounce off of ITMY 1431450536 Cold 0 0 5 0 Mismatch = 7.6% 85661
OMC Scan - Single Bounce off of ITMY 1403543046 Cold 0 0 0 4.6 Mismatch = 6.6% 78701
OMC Scan - Single Bounce off of ITMX 1431449762 Cold 0 0.45 5 0 Mismatch = 9.6% 85661
OMC Scan - Single Bounce off of ITMY 1431474471 Cold 0 0 5 4.6 Mismatch = 3.1% 85698
OMC Scan - Single Bounce off of ITMX 1431474101 Cold 0 0.45 5 4.6 Mismatch = 5.1% 85698
OMC Scan - Single Bounce off of ITMY 1444515634 Hot-ish 1.7 0 0 0 Mismatch = 7.1% 87461
OMC Scan - Single Bounce off of ITMX 1444515312 Hot-ish 1.7 0.45 0 0 Mismatch = 8.9% 87461
OMC Scan - SQZ Beam 1446952255 - - - - 4.6 Mismatch = 6.8% 88060
OMC Scan - SQZ Beam 1447088389 - - - - 0 Mismatch = 2.8% 88088
Gouy Phase - PRC 1255227492 Cold ITMY = 0.9, ITMX = 0.8 ITMY = 1.4, ITMX = 0.5 0 0 OneWay Gouy Phase = 23.2 [deg] 52504
Gouy Phase - PRC 1354415805 Cold 0 0 0 0 OneWay Gouy Phase = 20.7 [deg] 66215
Gouy Phase - SRC 1354410195 Cold 0 0 0 0 OneWay Gouy Phase = 19.9 [deg] 66211
Gouy Phase - SRC 1255907203 Cold ITMY = 0.9, ITMX = 0.8 ITMY = 1.4, ITMX = 0.5 0 0 OneWay Gouy Phase = 25.5 [deg] 52658
Gouy Phase - SRC 1255829128 Cold ITMY = 0.9, ITMX = 0.8 ITMY = 1.4, ITMX = 0.5 4 0 OneWay Gouy Phase = 29 [deg] 52641

 

Comments related to this report
sheila.dwyer@LIGO.ORG - 09:34, Wednesday 15 April 2026 (89898)

The measurements made with SR3 hot in May 2025 were done with SR3 heater requested power set to 2W, the readback of reported power was 1.9W.  The lines in the table that say 5W for SR3 power should say 2W.

sheila.dwyer@LIGO.ORG - 14:45, Wednesday 23 September 2026 (92035)

Some additions to this table are in 89056