Reports until 17:16, Monday 01 December 2025
X1 DTS
joshua.freed@LIGO.ORG - posted 17:16, Monday 01 December 2025 - last comment - 13:57, Tuesday 02 December 2025(88293)
SPI Pathfinder, Phase noise measurements for Keysight 33600A waveform generator

J. Freed,

I took Phase noise measurements of the 2 channel Keysight 33600A waveform generator for its use in building SPI Pathfinder in the optics lab before install. Going only off of the phase noise graphs, it is sufficient as it shows comparable results to the SRS which had a phase noise considered to be good enough for SPI pathfinder.

Key.png Shows the phase noise results. C1, C2 are the phase noise results for Channel 1 and Channel 2 on the Keysight, respectively. (Set up shown below). Shown for comparison the SRS SG392, which was suggested as a possible frequency source for SPI. The last measurement shown is the direct measurement of phase noise between the 2 channels of the Keysight. This measurement reflects the intended use case of the Keysight for SPI. As we need 2 frequencies at slightly different frequencies locked to each other and SPI will be measuring the output phase difference. Note the 60Hz peak; most likely caused by unclean AC power. This is why we are not using an AC powered device in the final installation. 

Screenshot2025-12-01at50150 PM.png Shows the setup for C1, and C2. measurements. The SRS value was found with the same set up, just replacing the Keysight 33600A with a SRS. The C1-C2 is a direct measurement by plugging both channels into the BluePhase 1000. There is no 10MHz Ext back attachment in this measurment in order to best represent Keysight's theoretical performance in the optics lab.


 

Images attached to this report
Comments related to this report
joshua.freed@LIGO.ORG - 13:57, Tuesday 02 December 2025 (88316)

Edits to previous post. Graph: X axis label should be 'Frequency offset from 80MHz (Hz)' and y-axis label should be 'dbc'

Keyradwref.png and Keyradworef.png are the requirements for the phase noise of our oscilator with SPI having and not having a reference interferometer respectivly. In the final SPI pathfinder install, we will have a reference interferometer giving us much less stringent requirements on our oscialtors phase noise. But during the build, it may be nessisary to run tests without a reference interferometer, I plotted the without reference interferometer if that situation ever does come up. 

Images attached to this comment