Richard, Jonathan, Fil, Marc, Daniel, EJ, Jeff, Oli, Ryan S, Dave:
h1susey was upgraded from Gen Std DACs (3x18bit, 2x20bit) to 2 LIGO-DACs.
Fil upgraded the three AI chassis to use the new SCSI interface board.
We removed the old Gen Std DACs and installed the two new LIGO-DACs.
The first LIGO-DAC drives the first standard AI, which in turn daisy-chain drives the second standard AI.
The second LIGO-DAC drives the special PI AI chassis.
After the hardware was changed the new h1iopsusey, h1susetmy, h1sustmsy and h1susetmypi models were rev-lock built and installed. A common model change for h1susetmy was accepted for its build.
EJ did a custom build for h1iopsusey due to rev-update issues found for watchdog source files. The user models were built with the production RCG-5.5.2
Following the model starts, the DAQ was restarted for new INI files due to removed DACs in h1iopsusey and h1susetmy.
With the daqd restarts we made a few changes.
H1:FEC-98_DAC_OVERFLOW_ACC_2_[0,1,2,3]
H1:FEC-98_DAC_OVERFLOW_ACC_3_[1,2,3,4]
We did two restart passes as h1susetmypi needed an updated to deal with DAC changes.
h1susey card information:
Removed:
* left at EY for h1iscey upgrade to 20bit-DAC