Closes FAMIS#27621, last checked 88194
Everything looking okay except PMC REFL being high. It does look like it's been higher since they recovered the PSL after yesterday's power outage, but we are also still recovering and more will still need to be done for the PSL anyway, so maybe this is expected/will be adjusted anyway.
Laser Status:
NPRO output power is 1.83W
AMP1 output power is 70.67W
AMP2 output power is 139.9W
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 0 days, 17 hr 9 minutes
Reflected power = 25.84W
Transmitted power = 105.6W
PowerSum = 131.5W
FSS:
It has been locked for 0 days 16 hr and 19 min
TPD[V] = 0.5027V
ISS:
The diffracted power is around 4.0%
Last saturation event was 0 days 0 hours and 0 minutes ago
Possible Issues:
PMC reflected power is high