[Oli, Jenne]
Oli brought all of the IMC optics back to where they had been yesterday before the power outage. We squiggled MC2 and the IMC PZT around until we could lock on the TEM00 mode, and let the WFS (with 10x usual gain) finish the alignment. We offloaded the IMC WFS using the IMC guardian. We then took the IMC to OFFLINE, and moved MC1 and the PZT such that the DC position on the IMC WFS matched a time from yesterday when the IMC was offline before the power outage. We relocked, let the alignment run again, and again offloaded using the guardian. Now the IMC is fine, and ready for initial alignement soon.