After taking the first tranfer function, we saw Length look weird, so we measured the wire lengths then replaced one of the wires and remeasured JM3. Before replacing the wire, and after replacing the wire. The wire replacement made everything look nicer, shifted the peaks into a more correct location. While it is still not perfect and looks a little different, we should be able to damp it, the magnitude differences are likely from the coil drivers.
The measurements are located in: /ligo/svncommon/SusSVN/sus/trunk/HTTS/X1/JM3/SAGM1/Data/
After replacing the wire:
2025-12-11_2300_X1SUSJM3_M1_WhiteNoise_L_0p02to50Hz.xml
2025-12-11_2300_X1SUSJM3_M1_WhiteNoise_P_0p02to50Hz.xml
2025-12-11_2300_X1SUSJM3_M1_WhiteNoise_Y_0p02to50Hz.xml
Before replacing the wire:
2025-12-11_2000_X1SUSJM3_M1_WhiteNoise_L_0p02to50Hz.xml
2025-12-11_2000_X1SUSJM3_M1_WhiteNoise_P_0p02to50Hz.xml
2025-12-11_2000_X1SUSJM3_M1_WhiteNoise_Y_0p02to50Hz.xml
Comparing JM3 to some previous TT measurements for OM3, PM1, and RM2.