The HAM7 suspensions are good to go. I am only presenting the results for the OPO (in 89015), ZM1, ZM3, and ZM4 since those are the four suspensions that were adjusted/in the way of the HAM7 work.
I accidentally remeasured ZM4 (RyanC had already measured it after it was moved 88827), so I processed my measurement along with the others I did.
I had remeasured ZM4 instead of ZM3 when the ISI was stil locked, so I had to measure ZM3 with the ISI unlocked and DAMPED, but it didn't seem to affect the measurements.
A before vs after comparison for these three ZMs(regular, zoomed) shows that they all look the same today as they did back in 2022.
ZM1 TF measurements
Settings
- ISI LOCKED
- SUS DAMP OFF
Data
/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM1/SAGM1/Data/2026-02-04_1700_H1SUSZM1_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
r12881
Results
/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM1/SAGM1/Results/2026-02-04_1700_H1SUSZM1_M1_ALL_TFs.pdf
r12881
ZM3 TF measurements
Settings
- ISI DAMPED
- SUS DAMP OFF
Data
/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM3/SAGM1/Data/2026-02-04_2000_H1SUSZM3_M1_WhiteNoise_{L,P,Y}_0p03to50Hz.xml
12883
Results
/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM3/SAGM1/Results/2026-02-04_2000_H1SUSZM3_M1_ALL_TFs.pdf
r12885
ZM4 TF measurements
Settings
- ISI LOCKED
- SUS DAMP OFF
Data
/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM4/SAGM1/Data/2026-02-04_1730_H1SUSZM4_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
r12880
Results
/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM4/SAGM1/Results/2026-02-04_1730_H1SUSZM4_M1_ALL_TFs.pdf
r12880
Comparison with last measurements
/ligo/svncommon/SusSVN/sus/trunk/HXDS/Common/Data/allhxdss_2026-02-04_ZM1ZM3ZM4CloseoutComparison_ALL_TFs.pdf
/ligo/svncommon/SusSVN/sus/trunk/HXDS/Common/Data/allhxdss_2026-02-04_ZM1ZM3ZM4CloseoutComparison_ALL_ZOOMED_TFs.pdf
r12886