Displaying report 1-1 of 1.
Reports until 17:07, Wednesday 04 February 2026
H1 TCS
matthewrichard.todd@LIGO.ORG - posted 17:07, Wednesday 04 February 2026 (89038)
Turning on the Ring heaters again!

M. Todd, S. Muusse, C. Compton, S. Dwyer


I wanted to get another measurement of what the HWS think the coupling factor is for thermal lens from ring heater power.

The HWS were not on, so after filling out a work permit Camilla and I went out and turned on the HWS SLEDs. Then we restarted the HWS codes in the individual computers. We also asked Jim to take ITMY ISI to fully isolated.

After waiting about 25 minutes for the HWS to get a baseline reading we turned up both ITMY and ITMX ring heaters by 4W (2W/segment). The HWS will track the defocus and I will compare with my models tomorrow morning. I wanted to do both ITM ring heaters to get a self-consistent measurement.

We also plan on doing single bounce OMC scans tomorrow morning with the ITMs being sufficiently thermalized after the RH turn on. This should give us another lens (punny pun here) to look at the thermalization business.