Displaying report 1-1 of 1.
Reports until 08:46, Tuesday 16 June 2026
H1 SUS (SQZ)
oli.patane@LIGO.ORG - posted 08:46, Tuesday 16 June 2026 (90625)
OPO TFs after new lens+rebalancing

Transfer functions after the rebalancing of the OPO yesterday look great! There's no touching or new peaks. I compared them to the close out measurements we got back in February (89015) and they match very well.

Settings
- ISI Locked
- DAMP OFF

Data
/ligo/svncommon/SusSVN/sus/trunk/OPOS/H1/OPO/SAGM1/Data/2026-06-16_1500_H1SUSOPO_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz.xml
r13037

Results
/ligo/svncommon/SusSVN/sus/trunk/OPOS/H1/OPO/SAGM1/Results/2026-06-16_1500_H1SUSOPO_M1_ALL_TFs.pdf
r13037
/ligo/svncommon/SusSVN/sus/trunk/OPOS/Common/Data/alloposs_2026-06-16_H1SUSOPO_JunevsFeb2026_ALL_TFs.pdf
/ligo/svncommon/SusSVN/sus/trunk/OPOS/Common/Data/alloposs_2026-06-16_H1SUSOPO_JunevsFeb2026_ALL_ZOOMED_TFs.pdf
r13036