Jeff, Betsy, Arnaud, Oli
Yesterday afternoon we went to the BBSS QOSEM satamp and checked the coil resistance. We did this by putting the BBSS in SAFE and then unplugging the Satamp to Duopus cables one by one. Each of these three cables have the pins for two QOSEMs, split as CH1 + CH2, CH3 + CH4, and CH5 + CH6. Each cable was unplugged and then the pins for each coil probed. Below are the results. These values are all pretty similar.
| Coil Resistance (Satamp to Duopus) | ||||||
| CH1-2 (SUS_BS_81) | CH3-4 (SUS_BS_82) | CH5-6 (SUS_BS_83) | ||||
| Channel | CH1 | CH2 | CH3 | CH4 | CH5 | CH6 |
| CH Coil Pins | 23 -> 10 | 18 -> 5 | 23 -> 10 | 18 -> 5 | 23 -> 10 | 18 -> 5 |
| OSEM Mapping | F1 | F2 | F3 | SD | LF | RT |
| Coil Resistance | 39.1 +/- 0.1 Ohm | 40.0 +/- 0.2 Ohm | 39.0 +/- 0.2 Ohm | 39.7 +/- 0.2 Ohm | 40.0 +/- 1.5 Ohm | 40.5 +/- 0.5 Ohm |
| Notes | Larger than normal variation | |||||
Indeed, the ~35ohm nominal coil resistance I quoted in LHO:90743 is measured at the QOSEM uDB9, so not including cable or feedthrough resistance.
After assembly I measured the coil reistance of these LHO BBSS QOSEMs to be:
34.6
35.6
35.1
35.7
36.4
35.6