Reports until 16:24, Wednesday 01 July 2026
H1 SUS
elenna.capote@LIGO.ORG - posted 16:24, Wednesday 01 July 2026 - last comment - 09:27, Thursday 02 July 2026(90852)
Status of BBSS cross coupling

[Betsy, Oli, Ibrahim, Elenna, with online help from Arnaud, Marie and Gabriele]

Today we began by attempting to follow the same steps that LLO did to decouple pitch and yaw, as described in 81817. To summarize, they applied a length offset and adjusted coil driver gains to ensure the same amount of motion is sensed on M0 F2 and F3 oseminfs.

Betsy and I immediately found that following that same process was not going to work. As a reminder, Ibrahim and Oli swapped the F3 osem yesterday due to ground loops, 90838. To get straight to the point, here are some results:

Arnaud notes that overall this is an improvement, because last week the P2Y cross coupling was much worse, roughly 1x pitch to 2x yaw. The overall numbers are different depending on which sensor you trust. However, Betsy and I think that the top mass drive versus sensed motion (master outs versus oseminfs) should be relatively straightforward. It seems like we have a mechanical imbalance on the suspension. However, the improvement that has occured was only due to swapping an osem, which seems electrical.

Oli will revisit the derivation of the osem2eul/eul2osem matrices. We also plan to recheck the suspension transfer functions to see if the P2Y coupling is reduced.

Images attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 09:27, Thursday 02 July 2026 (90875)

Last week's measurement showing worse P2Y coupling than in this alog: 90739