Reports until 17:42, Wednesday 02 September 2026
H1 SPI (SPI)
sina.koehlenbeck@LIGO.ORG - posted 17:42, Wednesday 02 September 2026 - last comment - 12:10, Friday 02 October 2026(91794)
SPI Readout vs. ISI Sensors: A First Comparison

After the angle calibrations were applied to the SPI readout channels, I grabbed some data to check whether the SPI is roughly agreeing with the ISI sensors. I used the supersensor OUT channels for the ISI sensors (these are the signals behind the blending filters, before the summation for the blended supersensor). Screenshots from diaggui are attached.

For the differential X degree of freedom (SPI length), the visual agreement is remarkably good.

The optical lever (angle) measurements are harder to interpret. First of all, QPD A has an as-yet-unresolved noise issue, which Jennie is currently investigating. Aside from this, the results look very encouraging, since the CPS and GS13 at least somewhat agree with the SPI.

These spectra should be regarded as a first result, but not yet final. We will post a better analysis once the error has been resolved.

Note: If anyone has suggestions for which sensor channels would provide a better/more appropriate comparison than the supersensor OUT channels used here, please comment, your input is welcome.

Images attached to this report
Comments related to this report
sina.koehlenbeck@LIGO.ORG - 19:01, Wednesday 09 September 2026 (91861)

Brian pointed me to a set of new data channels that provide calibrated outputs directly in displacement units. Using these channels, I acquired spectra with DTT at UTC 2026-09-09 12:00:00 (BW = 0.01 Hz, 30 averages) and generated the plots that follow


Channels plotted — Angular (RY / Pitch):

Optical lever pitch signals (angular, rad/√Hz):

  • H1:SPI-H23_OL_ISI_J_PIT_OUT_DQ (ISI 2) — Optical lever ISI J pitch angle
  • H1:SPI-H23_OL_ISI_K_PIT_OUT_DQ (ISI 3) — Optical lever ISI K pitch angle

ISI blend / CPS RY signals (ISI RY witness channels):

  • H1:ISI-HAM2_BLND_CPSRY_IN1_DQ (ISI 2) — HAM2 blended CPS RY input
  • H1:ISI-HAM3_BLND_CPSRY_IN1_DQ (ISI 3) — HAM3 blended CPS RY input

ISI calibrated Cartesian RY signals (rad/√Hz):

  • H1:ISI-HAM2_CAL_CART_RY_OUT_DQ (ISI 2) — HAM2 calibrated Cartesian RY output
  • H1:ISI-HAM3_CAL_CART_RY_OUT_DQ (ISI 3) — HAM3 calibrated Cartesian RY output

Differential Z displacement channel (converted to rad/√Hz):

  • H1:ISI-DIFF_H23_SS_Z_OUT_DQ — Differential H23 SS_Z displacement. Given in nm/√Hz; converted to m/√Hz, then divided by the arm length (15.4 m) to yield rad/√Hz.

Channels plotted — Differential X motion (displacement):

  • H1:SPI-H23_DIFFDISP_MAIN_OUT_DQ — SPI H23 main differential displacement
  • H1:ISI-DIFF_H23_SS_X_OUT_DQ — Differential H23 SS_X (CPS + GS13 supersensor)
  • H1:ISI-HAM2_BLND_CPSX_IN1_DQ (ISI 2) — HAM2 blended CPS X input
  • H1:ISI-HAM3_BLND_CPSX_IN1_DQ (ISI 3) — HAM3 blended CPS X input
  • H1:ISI-HAM2_CAL_CART_X_OUT_DQ (ISI 2) — HAM2 calibrated Cartesian X output (GS13)
  • H1:ISI-HAM3_CAL_CART_X_OUT_DQ (ISI 3) — HAM3 calibrated Cartesian X output (GS13)

All scaled by nm_to_m (nm/√Hz → m/√Hz).


Unit conversions:

  • OL_ISI_J_PIT and OL_ISI_K_PIT plotted in rad/√Hz
  • SS_Z converted from nm/√Hz to rad/√Hz by dividing by a length of 15.4 m
  • Differential X-motion channels scaled from nm/√Hz to m/√Hz (nm_to_m factor)
  • Angular y-axis units in rad/√Hz; displacement y-axis units in m/√Hz
Images attached to this comment
sina.koehlenbeck@LIGO.ORG - 12:10, Friday 02 October 2026 (92168)

I have recompiled the data from Jeff’s earlier measurements and plotted the coherence of the SPI signals relative to CPS and GS13, as well as relative to the corresponding difference signals:

X-degree of freedom: (HAM3 X – HAM2 X)
RY-degree of freedom: (HAM3 Z – HAM2 Z) / 15.4 m
RZ degree of freedom: (HAM3 Y – HAM2 Y) / 15.4 m

Interpretation:

  • As expected, the SPI signals exhibit coherence with the CPS and GS13 signals.
  • For HAM2 and HAM3, the GS13 signals appear to be dominated by residual CPS noise.
  • A certain degree of coherence can be observed between the signals (HAM3 Z – HAM 2 Z) / 15.4 m and (HAM3 Y – HAM 2 Y) / 15.4 m and the SPI signals; however, this coherence is (for the most part) not dominant.
Images attached to this comment