Telescope alignment and profile 0922 laser unit for CHETA X arm

The CHETA X arm table telescope was fully aligned as described in gitlab. The performance of the telescope was determined by profiling the beam after L2 and calculating the complex q factor. The profile was measured using the nanoscan beam profiler over a 55cm range between M4 and M5.

This unit was profiled twice the first with a slight angle on L1 to prevent back reflection and then with a reduced angle to determine if this was detrimentally effecting the beam propogation. A non-linear fit was used in both cases to find the q factor which was propogated to the ITM to determine ITM beamsize. Both results show the horizontal axis have a large deviation from the expected q factor and the beamsize at the ITM is reduced by ≈35% and reducing the angle had a minimal effect.

  w0 [um] z0 [mm] w @ ITM [mm]
Small angle on L1      
Horizontal fit 1486.4±180.5 -2448.4±269.1 36.919
Vertical fit 916.3±86.3 -1177.2±143.0 57.818
L1 normally incident      
Horizontal fit 1456.4+/-92.7  "-2428.4+/-149.4" 37.659
Vertical fit 1065.3+/-49.8  -1384.2+/-79.3 50.019

Unit 0922 is noticibly non gaussian, screenshot attached, which we believe is consistent with what was observed by thorlabs and is indicated in a paper datasheet shipped to us with the unit. We believe this may be contributing to the deviation from the model.

 