Displaying reports 1-20 of 81709.Go to page 1 2 3 4 5 6 7 8 9 10 End
Reports until 13:18, Friday 25 April 2025
H1 PEM
ryan.crouch@LIGO.ORG - posted 13:18, Friday 25 April 2025 (84129)
Dust monitor alarm reset script

I keep forgetting to reset the dust monitor alarm levels after IOC restarts so I wrote a little script to check all of the dust monitor alarms levels and reset them if needed based on E1600132. It does not reset the LVEA ones, it just notifies, as these ones alarm levels are sometimes modified.

The scripts is called check_dm_alarms_lvls.py and lives at /ligo/cds/userscripts/. I've also updated the wiki with this info.

H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 12:52, Friday 25 April 2025 (84128)
HAM1 vent work - RM1 and RM2 (Tip Tilt suspension) re-installed into the chamber

Oli, Daniel, Fil, Betsy, Rahul

We have re-installed SUS RM1 and RM2 Tip Tilt suspensions into HAM1 chamber with the new ISI table - see several pictures attached below. RM1 is on the left (towards PSL side) hand side of this picture and RM2 is on the right (zoomed in). Both the suspensions are positioned roughly as per D1000313_V19, and will be fine tuned (pitch and yaw of the mirror as well) during beam alignment work. The blade springs of both the suspensions have been un-muted - see LHO alog 83984 for details.

Both RM1 and RM2 have been plugged-in to their electronics chain. We had to perform some troubleshooting with RM1 after observing grounding issues. At first Fil and Daniel ruled out any issues with the satellite box and in-air cables. After some investigation, the grounding issue was suspected to be coming from the in-vacuum quadrupus cable (connecting to the Bosem), which is an old cable and looks worn out too.

This morning I replaced the old Bosem cable (D1000228_v3 s/n S1105240) with a 55in long (D2100049 S/N 2200358) cable. Before connecting it to the Bosem, Daniel and I verified that the cable is fine (i.e not grounding) and later on we connected each pin (Bosem) one by one. We had to re-adjust the Kapton tape (installed for additional shielding) since the cable bosem connector was grounding to the chamber. After all these modifications, we can confirm that RM1 and RM2 are free from any grounding issues for now. We will check again before closing the chamber.

Next, I will start taking measurements on these suspensions and will post their health report soon.

Images attached to this report
H1 PEM
ryan.crouch@LIGO.ORG - posted 12:27, Friday 25 April 2025 (84127)
Pumpless dust monitor flow check

We have been running the dust monitor vacuum pumps with the assumption that they had to be set to 19 inHg according to T2100415, I went out to ENDY to test this out. When I first checked out the vacuum pump, it was at 20.5 inHg, the flow meter on the dust monitor was reading at 2.8 L/min which is what we want, I then lowered the pumps pressure down to 13 and rechecked the dust monitor and its flow was now at ~2.7. I then adjust the pump down to 10 which brought the dust monitors flow to 2.65. These 1/10s of a L flow are correctable by adjusting the internal flow rate on the dust monitor. Before I left I reset the pumps pressure to 19 inHg.

It seems like we do not need to have the vacuum dust monitor pumps set to 19 inHG, the dust monitors flow rate isn't as strongly dependent on it as we thought?

H1 PSL (PSL)
ibrahim.abouelfettouh@LIGO.ORG - posted 12:14, Friday 25 April 2025 (83720)
PSL Weekly Report - Weekly FAMIS 26397

Closes FAMIS 26397, last checked in alog 84018

Laser Status:
    NPRO output power is 1.841W
    AMP1 output power is 70.43W
    AMP2 output power is 140.4W
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 2 days, 23 hr 11 minutes
    Reflected power = 23.28W
    Transmitted power = 105.5W
    PowerSum = 128.8W

FSS:
    It has been locked for 1 days 20 hr and 35 min
    TPD[V] = 0.8003V

ISS:
    The diffracted power is around 3.7%
    Last saturation event was 1 days 20 hours and 35 minutes ago


Possible Issues:
    PMC reflected power is high

H1 SEI (SEI)
ibrahim.abouelfettouh@LIGO.ORG - posted 12:14, Friday 25 April 2025 (83717)
H1 ISI CPS Sensor Noise Spectra Check - Weekly FAMIS 26040

Closes FAMIS 26040. Last checked in alog 83968.

HAM4, HAM7, HAM8, EX ST1 all have elevated peaks, mostly around 1Hz line. This is expected due to high maintenance and vent activity.

Non-image files attached to this report
LHO FMCS (PEM)
ibrahim.abouelfettouh@LIGO.ORG - posted 12:03, Friday 25 April 2025 (83722)
Checking HVAC Fans - Weekly FAMIS 26375

Closes FAMIS 26375. Last checked in alog 84000,

Everything is under threshold. There are different fans on than last week so difficult to compare. MR FAN3 in CS has a slightly higher noise character in last 3 days

Images attached to this report
H1 SEI
jim.warner@LIGO.ORG - posted 11:53, Friday 25 April 2025 (84126)
HAM2 H3 HEPI L4C is ded

While prepping the HAM1 ISI for testing last night, I noticed that one of the HAM2 HEPI L4Cs seemed non-responsive. It seems this has been the case for quite some time, 206 days according to the attached ndscope. Something happened on Oct 1st last year that cause this L4C to suddenly have only ~1/50th the signal the H2 L4C on the HEPI  to sees. I have poked at it a bit, checked cabling on the pier and gently hit the foot to see if I could revive the sensor. I have a couple other tests I want to do before deciding to remove the actuator to replace the sensor, not exactly eager to have to do that.

Images attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 11:38, Friday 25 April 2025 (84124)
Check on Vacuum Components this Morning

Dry air skid checks, water pump, kobelco, drying towers all nominal.  Dew point measured at HAM1, reading -43.8°C. Main turbo pumps, XBM, YBM and OMT, all nominal, and temperature is good and stable.  HAM6 turbo pump and cart are nominal.

Images attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 11:32, Friday 25 April 2025 (84125)
h1cdsh8 IO Chassis power cycle

WP12488 HAM8 ISI binary readback broken following site outage FRS33798

Fil, Erik, Jim, Dave:

Yesterday we power cycled the binary input/output and interface chassis, which did not fix the issue. Today we stopped all the models on h1cdsh8, powered down the computer, then power cycled the IO Chassis.

This also did not fix the problem. Investigation continues.

H1 SUS
daniel.sigg@LIGO.ORG - posted 11:13, Friday 25 April 2025 (84123)
RM1/RM2/PM1

Some changes to the RM1/RM2/PM1:

All other things equal the local damping loops should have the correct sign now. Thic change will also require a sign change in the ASC centering servos.

H1 TCS
thomas.shaffer@LIGO.ORG - posted 11:05, Friday 25 April 2025 (84122)
TCS Chiller Water Level Top Off - Biweekly

FAMIS27814

Our cadence is a bit off from biweekly, but no matter since no water was added and the chillers look good (updated the T2200289 sheet). There is a small amount of blueish gunk starting to form in the wire mesh diffuser in the TCSX chiller. I didn't see any of these in our parts bin so I'll put a few on order. 

LHO VE
david.barker@LIGO.ORG - posted 10:55, Friday 25 April 2025 (84121)
Fri CP1 Fill

Fri Apr 25 10:08:57 2025 INFO: Fill completed in 8min 53secs

Gerardo confirmed a good fill curbside.

Images attached to this report
H1 CAL
francisco.llamas@LIGO.ORG - posted 10:46, Friday 25 April 2025 (84034)
Comparing sensing function mcmc model fit

FranciscoL, SheilaD, JoeB

In summary:

I regenerated the mcmc fit on report 20250327T160138Z at 5, 10 and 20 Hz, to compare how the mcmc model would change given different frequency ranges for the fit. We see that the residual of the measurement to the model is better as we increase the minimum frequency threshold. We think this is due to external effects, aside SRC, confusing the model.


We want to give more validity to our observations on alog 83592 -- that the sensing model at low frequencies is a good descriptor of the sensing function. On 83592, I claimed that the observed changes in the sensing function come from changes in SRCL detuning. This is backed up by a low residual on the measured/modeled sensing function. In today's alog, we evaluate how well the model describes the sensing function by (1) having a *low* residual and (2) decreasing the optical spring frequency.

The first three figures attached ('sensing_mcmc_compare_*Hz.png') are the sensing function plots generated by pydarm for measurement 20250327T160138Z. This measurement is not a representation of a calibrated interferometer, since we tuned the SRCL detuning (83585).

From the transfer function on the left, the green dots represent the measurement, the orange trace is the mcmc model from the measurement, and the blue trace is the mcmc model from the last validated calibration measurement (20250222T193656Z). The residual from the measurement to both models are plotted on the right, matching the colors of the respective models. We are interested in the change of the orange trace when we modify the frequency range of the fit.

Each figure has a pair of vertical lines denoting the limits on the frequency range used for the fit. I regenerated the measurement report for 05, 10, and 20 Hz for the minimum frequency. The maximum frequency at 1200 Hz was not changed.

For the residual plots, we see that the magnitude is largest when fitting to 5 Hz, and that the change between the 10 and the 20 Hz fits is negligible. For illustration purposes, I plotted the different sensing models, along with the measurement, together in the last figure (20250327T160205Z_compare_models). The phase plot is not the same as the plots given by pydarm due to missing factors. We see that the three models match above 30 Hz in magnitude.

For the optical spring frequency, we see a decrease from 2.19 Hz to 0.43 Hz when fitting to 10 Hz, and 20 Hz, respectively. This decrease on the SRC spring suggests that the model does not fit the sensing function well, below 20 Hz.

So, regenerating the sensing model shows that the model is good when we see anything above 20 Hz. But we see from the measurement that the SRCL detuning does change the sensing function at low frequencies. The model, however, cannot fit the data from the measurements at these frequencies. We have seen before (80267) that angle to length cross coupling influences the sensing function in the low frequencies. A scenario of cross coupling affecting the sensing function could explain why the model cannot fit the measurement.

Repeating the analysis on cross-coupling, as done in 80267, could improve the sensing function.


TIL; pydarm edition:

The approach that enabled me to regenerate pydarm reports, albeit messy, was to

  1.  Modify the pydarm_cmd_H1.yaml that is in the report to be regenerated, in this case 20250327T160138Z. In particular, the mcmc_fmin value.

  2. Once the YAML file is configured, run

pydarm report 20250327T160138Z --skip-gds --regen

This command will re-run the calibration report, and, by using the configuration that lives in that directory, make a new model to the sensing function measurement.

Images attached to this report
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 07:47, Friday 25 April 2025 (84118)
OPS Day Shift Start

TITLE: 04/25 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: MAINTENANCE
    Wind: 1mph Gusts, 0mph 3min avg
    Primary useism: 0.06 μm/s
    Secondary useism: 0.09 μm/s
QUICK SUMMARY:

IFO is in PLANNED ENGINEERING for VENT

Some planned tasks for today:

Work safe!

LHO VE
janos.csizmazia@LIGO.ORG - posted 00:48, Friday 25 April 2025 - last comment - 08:20, Friday 25 April 2025(84117)
Wire seals finally tested with a positive result!
Jordan, Janos, 04-24-2025, ~15:00

The wire gaskets for VBO-C and VBO-D have been leaking continuously for a very long time. Lots of vendors have been contacted, and after a lot of unsuccessful tests, it seems that we finally found the good one manufactured by Torr Scientific (UK). This gasket is extremely annealed, butter-soft, needs super careful handling, but even after some suboptimal installation steps (mea culpa), it finally sealed a new chamber successfully.

We tested it right after receiving (had a thriller with customs) with a new C+B chamber, pumped it down and tested on the same day. After the test, we let the pump running during the weekend just to see the ultimate pressure it can reach.

They are still very pricey though, 193 GBP (= 257 USD)/pc., even in the case of purchasing 100 pcs., it's 147 GBP (= 196 USD)/pc. So, certainly not an optimal, long-term solution, but a good bridge between having nothing and manufacturing our own.
Comments related to this report
jon.feicht@LIGO.ORG - 08:20, Friday 25 April 2025 (84119)
Good news! Let's take some detailed measurements (as possible), also how close they fit over the flange lip, etc. Will help us when we start fabrication in house. 
H1 ISC
camilla.compton@LIGO.ORG - posted 17:01, Thursday 24 April 2025 - last comment - 22:22, Thursday 24 April 2025(84115)
Starting to repopulate HAM1 with ISC Components

Oli, Matt, Camilla 

We started to repopulate HAM1 with ISC components following the locations show in D1000313-v19
 
On the +X side, we placed: M14, HW1 (facing wrong direction), M17, M1, M7, M13, M10, M11, M12, BD4, BD5, BD6a, BD6b, BD7, BD17 (in different location due to masses), BD18, HiPWR BD (needs to be moved after ASC REFL beam profiling). Photos here and here
We could still place: L2 and M15 and flip the direction of HW1 (see Corey's original photo)
We want to wait to place: M16, the periscope, Beamdiverters, LSC POP diodes, BD1, BD2, as it will make alignment or beam profiling of the diodes tricky.  
 
On the -X side, we placed: M2, M6, L1, M18. Photo here
We could still place: M5 (once we are happy with RM1), the majority of the beamdumps, the SLED.
We want to wait to place: the ASC and LSC diodes until some beam profiling/ beam size measurements have been done.
 
The 4 optics that are still in pans (L2, M15, M16, M5) have all been moved to pan 2 (lower shelf) and the periscope is on the bottom shelf.
Images attached to this report
Comments related to this report
betsy.weaver@LIGO.ORG - 22:22, Thursday 24 April 2025 (84116)
Late entry,
Jim and I worked the last few half days to lace up all of the cabling for SUS, SEI, and IO/ISC in prep for the main reinstall of table components. Fil, some Richard, and I did preliminary ground loop checks on all SUS and IO/ISC cables so we have a handle on the situation before complicating with components.

I routed cables on the table top in an attempt to be near where things plug in but also not in the way of landing objects. Further dressing and coiling will be needed after troubleshooting plugins with components as we go.
Pics of the pre round of cabling.
Images attached to this comment
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 16:42, Thursday 24 April 2025 (84114)
OPS Day Shift Summary

TITLE: 04/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY:

IFO is in PLANNED ENGINEERING for VENT

Productive day where most of the work was spent on HAM1 cabling, followed by RM1 install.

Of note:

LOG:

Start Time System Name Location Lazer_Haz Task Time End
14:39 FAC Kim LVEA N Tech clean 15:52
14:39 FAC Nellie LVEA N Tech clean 15:52
15:45 VAC Jordan LVEA N Purge air checks 16:03
15:45 SUS Betsy LVEA N Cabling HAM1 18:59
15:52 FAC Nellie H2 N Tech clean 16:07
16:07 SUS Randy YARM N QUAD Moving 23:24
16:33 SUS Richard LVEA N Cabling/H.O.R.S.E. 17:03
16:50 SUS Mitchell EX, EY n FAMIS tasks 22:50
16:56 SUS Travis LVEA N Cabling 17:06
17:07 FAC Kim, Nellie MY, MX N Tech clean 18:51
17:53 VAC Jordan MX, MY N Emergency Pump Survey 18:38
17:53 VAC JaNos MX,MY N Emergency Pump Survey 18:37
18:23 ISC Camilla, Matt LVEA N Measuring viewport heights 18:54
18:39 PCAL Tony PCAL Lab N TSA to France + stickers 20:07
19:01 SUS Jim LVEA N Cabling 23:00
19:31 FIT Ryan C LVEA N Respiratory and Perspiratory Regulation Test (Walk) 20:04
19:56 ISC Matt, Rahul, Camilla, Oli LVEA n HAM1 ISC installation 23:23
19:59 VAC Jordan MY, MX N Turbo Pump Checks 22:50
20:00 SUS Betsy LVEA, Opt Lab N Walkabout + Feed Through Cable Work 21:56
20:03 VAC Travis Opt Lab N Feedthru Cable Work after turning compressor off 22:48
20:19 VAC Janos LVEA N VAC checks 20:28
20:41 CDS Fil, Dave FCES N HAM8 Post Outage Glitch 22:59
21:07 VAC Vacuum Convoy YARM N Beam Tube Experiment Walkabout 21:56
21:49 PCAL Tony PCAL N Measurements and general lab affairs 22:22
22:59 EE Fil HAM1 N helping Rahul 23:24
23:24 SUS Betsy. Fil, Rahul, Daniel LVEA N RM1 Things 23:40
H1 PSL (ISC)
ryan.short@LIGO.ORG - posted 16:45, Thursday 17 April 2025 - last comment - 10:57, Friday 25 April 2025(83989)
SPI Pick-off Path Install Day Three: Complete!

S. Koehlenbeck, J. Freed, J. Kissel, J. Oberling, R. Short

The SPI pick-off path installation on the H1 PSL table is now complete. The beam in the new SPI path has been reduced to 20mW and is currently being dumped with a razor dump between SPI-L1 and SPI-L2. Pictures attached reflect the final installation and layout, which will be be reflected in the updated as-built layout at a later date.

Associated entries: 83925, 83933, 83956, 83961, 83978, 83983 (and more to come)

Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:28, Friday 18 April 2025 (83998)ISC, SQZ
ECR E2400083
IIET 30642
WP 12453

Here's Ryan's birdseye view labeled with all the components.
For details of the components, see the SPI BOM, T2300363, exported from its google sheets to -v4 as of this entry.
Images attached to this comment
corey.gray@LIGO.ORG - 09:53, Monday 21 April 2025 (84019)EPO

Tagging EPO for photos.

joshua.freed@LIGO.ORG - 10:57, Friday 25 April 2025 (84120)

83996 Power In ALS / SQZ / SPI Paths Post SPI Pick-off Install

Displaying reports 1-20 of 81709.Go to page 1 2 3 4 5 6 7 8 9 10 End