Displaying reports 2661-2680 of 88961.Go to page Start 130 131 132 133 134 135 136 137 138 End
Reports until 12:09, Thursday 22 January 2026
H1 SUS (CDS)
oli.patane@LIGO.ORG - posted 12:09, Thursday 22 January 2026 (88812)
SUS Model Updates: h1susitmx

Jeff, Oli

We made several updates to multiple SUS simulink models as a continuation of our work converting SUSB123 and SUSH34 into SUSB13 and SUSB2H34 (88765). Our changes were based on our updated G2301306-v10_PARTII - for the conversion from SUSB123 into SUSB13 for ITMX, we referenced slides 5, 7, 12, 13, 18, and 19. These changes were svn'd as r34388 and were installed and started last week (88781).

These are the changes we had to make for h1susitmx:

CDS Parameter Block:
- Changed hostname from h1susb123 to h1susb13
- Removed specific_cpu

Input (before, after):
- Removed ADC1 since now all OSEM channels are on one ADC (called ADC1 in diagrams)
- Redid ADC to OSEM configuration
- Changed name of SHMEM channel that comes in for ITMX L3 BIO
    - Previously was H1:SUS-ITMX_BIO_L3_MON_SHMEM, now is H1:SUSITMY_2_ITMX_L3_BIOMON_SHMEM

Output (before, after):
- Changed DACs from several 20 bit DACs to two 28 bit DACs
- Lined up outputs to their corresponding DAC channels
- Lined up ESD bias and ESD sum outputs to their corresponding DAC channels
- Changed name of SHMEM channel that goes out for ITMX L3 BIO
    - Previously was H1:SUS-ITMX_BIO_L3_CTRL_SHMEM, now is H1:SUSITMY_2_ITMX_L3_BIOCTRL_SHMEM

Images attached to this report
Non-image files attached to this report
H1 SUS (CDS)
oli.patane@LIGO.ORG - posted 11:38, Thursday 22 January 2026 (88848)
SUS-C1, SUS-C2, SUS-C5, SUS-C6 rack photos and chassis serial lists updated

Jeff

The SUS rack photos for SUS-C1 (S1301868-v1), SUS-C2 (S1301869-v2 ), SUS-C5 (S1301872-v6), and SUS-C6 (S1301873-v3) have been updated with new photos taken after our upgrade last week. Rack photos prior to this upgrade are linked with the previous version of each rack.

We also have gone ahead in those same DCC SUS rack documents and updated the e-travelers and related documents for each rack, listing the currently installed electronics types and serial numbers.

LHO FMCS (PEM)
ryan.crouch@LIGO.ORG - posted 10:38, Thursday 22 January 2026 (88845)
HVAC Fan Vibrometers FAMIS Check (FAMIS 39851)

Closes FAMIS39851, last checked in alog88771

Everything looks as it did during the last check, no real changes over the past week that I can see.

Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 10:16, Thursday 22 January 2026 (88843)
Thu CP1 Fill

Thu Jan 22 10:05:43 2026 INFO: Fill completed in 5min 39secs

 

Images attached to this report
H1 IOO
jennifer.wright@LIGO.ORG - posted 09:58, Thursday 22 January 2026 (88839)
Checking JAC input alignment after body mode damper install

Jennie W, Rahul K, Jason O, Keita K

 

Rahul and Jason installed body mode dampers on the JAC, we did not expect it this to change the position of the JAC much.


Thinking of what could have cause this there are two possibilities:

  1. Installing the body mode dampers moved the JAC.
  2. The B & K hammering on Tuesday (see alog # 88822) of the input HAM1 periscope somehow shifted its mirrors.

After installation we tried to lock and could not. Checking the output of the JAC we could not see flashes. Lookign at the input coupler of JAC on a card the beam seems high coming into the JAC.

Jason and I recovered the alignment into the JAC somewhat with the PSL persicope PZT and the "JM1" fixed mirror which sits right after the HAM1 input periscope.

We finally managed to lock the JAC (altered the guardian timeout thredhold which checks the power level - I marked what I changed in the code so I can reverse it).

Below I include a comparison of the last time we were locked last Friday 16th, this is nominally in mW but we have not checked the calibration of this PD yet.

Yesterday afternoon we took this image of the locked level and it is worse by a factor of 2, see this image.

Keita suggested we do a scan with the on board PZT while the cavity was unlocked.

Taking the measured hom spacing which Masayuki and I measured in this unit at CIT, 1.75 radians and the FSR, caluclated from 1/time period between TM00 modes, we estimate the 01 mode is 0.75 of the FSR.

The scan of the transmitted beam in blue makes it look like we have two candidates for this 01 mode that are between 70 and 50% of the height of the TM00 mode, so we are most likely mis-aligned. The scan voltage is in red.


Last week we had aligned into HAM2 and out of the viewport that normally delivers the beam to IOT2L. We checked this late in the afternoon and the beam gets through the MC REFL periscope on the table and through one mirror, then clips on BS1.

This makes it likely that our output alignment has not changed badly and so we can go ahead with aligning the output of the JAC.

Our plan today is to spend half an hour checking the JAC is not locking to a HOM and then try and improve the alignment ontpo IOT2L with JM2 and JAC_M3 mirrors in HAM1.

We still have irises at the output of the HAM1 table, one of which was placed at the old output beam position before we installed the input periscope or the JAC cavity, so we will try and improve the line through this iris and onto IOT2L.


Yesterday the EOM arrived midday, Keita and Rahul concentrated on building this in the optics lab, this will continue to be worked on today.

Images attached to this report
H1 ISC (ISC)
marc.pirello@LIGO.ORG - posted 09:31, Thursday 22 January 2026 (88842)
EOM Heliax Cables Pulled and Landed at PSL-R2

Per WP12989 we completed the EOM cable install from PLS-R2 to the HAM-1 feedthrough.  The old EOM cables from the PSL were disconnected at the patch panel, and the RF Combiner was powered down.  RF Cables were landed at the patch panel from left to right, 24MHz, 118MHz, 45MHz, 9MHz, with 2 empty spigots on the right hand side of the panel located at U2 in the rack near the floor.  We will shoot these with the TDR to once everything is installed to get total electrical length for calibration.

M. Pirello, F. Clara

H1 General
ryan.crouch@LIGO.ORG - posted 07:25, Thursday 22 January 2026 - last comment - 08:26, Thursday 22 January 2026(88840)
OPS Thursday day shift start

TITLE: 01/22 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: MAINTENANCE
    Wind: 5mph Gusts, 3mph 3min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.17 μm/s 
QUICK SUMMARY:

Comments related to this report
ryan.crouch@LIGO.ORG - 08:26, Thursday 22 January 2026 (88841)

There is a planned DAQ restart today as well.

H1 General
ryan.crouch@LIGO.ORG - posted 16:30, Wednesday 21 January 2026 (88835)
OPS Wednesday DAY shift summary

TITLE: 01/22 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: OPO and JAC work continued in HAMs 1 & 7 in LASER HAZARD.
LOG:                                                                                                                                                                                                       

Start Time System Name Location Lazer_Haz Task Time End
22:49 SAF LVEA IS LASER HAZARD LVEA YES LVEA IS LASER HAZARD \u0d26\u0d4d\u0d26\u0d3f(\u239a_\u239a) 16:49
15:31 FAC Randy EndY N Bring a forkflit down to the CS/High bay 17:24
15:53 FAC Kim, Nellie LVEA Y Tech clean 17:53
17:01 SUS RyanC CR N ZM4 Transfer function 17:24
17:07 JAC Matt, Sophie PREP lab N JOTE work 19:36
17:10 SQZ Sheila, RyanS, Kar Meng HAM7 Y Chamber work, Kar Meng in at 1900 21:02
17:17 PEM Robert LVEA HAM4 Y Laser vibro, in and out 18:17
17:57 ISC Jason LVEA Y HAM1 work 20:52
18:07 ISC Rahul LVEA Y HAM1 20:05
18:20 FAC Eric HAM Shaq N Changing filters 18:51
18:27 ISC Jennie LVEA Y HAM1 work 19:14
19:22 ISC Jennie, Keita LVEA Y HAM1 work 20:52
19:33 VAC Travis Optics lab N Put away part 19:43
19:36 FAC Kim H2 N Tech clean 19:54
19:53 CDS Oli LVEA, CER Y Take pictures of computer racks 20:05
21:21 VAC Travis MidX N Look for VAC equipment 21:39
21:21 VAC Jordan FCES N Look for VAC equipment 21:49
21:34 SUS Rahul Optics lab LOCAL Work on EOM 23:03
22:02 EE Marc, Fil LVEA Y Run cable for EOM from HAM1 to PSL racks Ongoing
22:33 VAC Gerardo VAC prep lab to LVEA Y Bring over glass piece to HAM1 23:38
23:38 VAC Gerardo VAP prep lab N Assembly work 00:19

19:24 OPO WD trip

H1 PSL
ryan.short@LIGO.ORG - posted 16:02, Wednesday 21 January 2026 (88838)
PSL 10-Day Trends

FAMIS 31121

Things are looking very stable over the past week; makes sense since very little has been asked of the PSL recently. The PMC and RefCav could still use an alignment touch-up, which I'll get to hopefully this week.

Images attached to this report
H1 SUS
oli.patane@LIGO.ORG - posted 15:26, Wednesday 21 January 2026 (88837)
Renamed SR3 OPLEV Whitening to AA cable

Jeff, Oli

We decided to rename the SR3 OPLEV Whitening to AA chassis cable from SUS_SR3-168 to SUS_OPLEV_SR3_AA. We decided to do this for a couple reasons: SUS_SR3-168 is not a very helpful name, it matches better with the other OPLEVs' cables, and anyway we had already been calling that cable SUS_OPLEV_SR3_AA in our wiring diagrams for a while. 
We relabeled both sides of the cable.

H1 SQZ (SYS)
karmeng.kwan@LIGO.ORG - posted 14:34, Wednesday 21 January 2026 (88836)
HAM7 progress update

[Sheila, Ryan, Karmeng]

We have returned ZM2 to the original position at the beginning of O4, we alleviated some of the saturation but the ZM3 iris still require adjustment.

Power budget looks good so far. With 0.77mW exiting the OPO and reaches AM3 at the same power. We observed 0.75mW of OPO trans and 18.1mW of OPO refl on SQZT7 periscope.

H1 IOO (IOO)
rahul.kumar@LIGO.ORG - posted 13:04, Wednesday 21 January 2026 - last comment - 09:41, Wednesday 28 January 2026(88834)
HAM1 update - JAC damper

Jason. JennieW, Rahul

This morning we added the damper parts to the JAC in HAM1 chamber - see picture attached. The cables were re-routed slightly so that they don't interfere with the damper.

I have finger tightened the screws of the damper for now - since no torque spec was provided and we were worried that tightening them too much might affect its position/angle.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 08:27, Wednesday 28 January 2026 (88934)EPO

EPO-tagged.

jason.oberling@LIGO.ORG - 09:41, Wednesday 28 January 2026 (88941)

Masayuki and I fully tightened the JAC body mode dampers yesterday.  We were only able to get the screws ~1/2 turn past finger-tight before we could no longer tighten them.  No changes in JAC alignment were observed while we were doing this.

H1 SUS (SUS)
ryan.crouch@LIGO.ORG - posted 11:44, Wednesday 21 January 2026 - last comment - 08:25, Wednesday 28 January 2026(88828)
2nd HXDS received from LLO

Yesterday I picked up the shipment from LLO contain the 2nd HXDS (first HXDS shipment alog87998) and related ASSEYs and brought it to the Triples lab. 

Uwrapping and inspecting was uneventful, the zip ties were intact, the shock pads untripped and the humidity indicator was unchanged in color. The outer (front, back) and inner (front, back) bags were in good condition, unwrapping the suspension I found no play with the blades.  All the nuts and dowel pins were tight and well seated, the ICS looks good except for it seems to be missing the HDS Intermediate Mass Fixture (D2000230-V1, s/n 08). The part is scribed as V1, on the dcc there's a V2 with two added 8-32 helicoils taps that, based on D1900352 which uses V2, I should be able to see from the front and I don't, so this is a V1 plate and not an incorrectly scribed V2. I believe these two missing 8-32 holes are where we will attach the BOSEM Mounting plate (D2000257). The first HXDS we received in November had the V2 plate. We may have to do a small rework to update it to V2, or have LLO send the V2 part if possible.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 08:25, Wednesday 28 January 2026 (88933)EPO

EPO tagging

LHO General
eric.otterman@LIGO.ORG - posted 11:03, Wednesday 21 January 2026 (88832)
FCES air handler
The pre-filters were replaced in the FCES air handler. 
LHO VE
david.barker@LIGO.ORG - posted 10:16, Wednesday 21 January 2026 (88831)
Wed CP1 fill

Wed Jan 21 10:08:13 2026 INFO: Fill completed in 8min 10secs

 

Images attached to this report
H1 SEI
oli.patane@LIGO.ORG - posted 09:59, Wednesday 21 January 2026 (88829)
ISI CPS Noise Spectra Check Weekly FAMIS

Closes FAMIS#39334, last checked 88680

The elevated sensors listed are just from HAM1 and HAM7, which is expected
All chambers are looking normal relative to the current state that they are in (locked, vented, etc)

Non-image files attached to this report
H1 ISC (INS, ISC)
keita.kawabe@LIGO.ORG - posted 17:53, Tuesday 20 January 2026 - last comment - 18:40, Friday 23 January 2026(88822)
POP periscope stiffener was installed. B&K for POP as well as JAC input periscope (Rahul, Jim, Keita)

We have intalled the POP periscope stiffener.

Some dog clamps in the REFL path as well as the cable bracket for PM1 were relocated to accomodate. 1st and 2nd picture are "before" photo. 3rd one is after PM1 cable bracket relocation but before installing the stiffener.  There are also three "after" photos showing how things look on the table.

POP beam clearance:

I took the picture of the bottom periscope mirror through dichroic (HR for IR, transmission for green) to see the beam clearance. In the first such photo (POP_beam_clearance.jpg), the camera is close to the center of the dichroic and the short stiffener beam is close to the edge of the optic but not occulting the mirror, so we're OK. Just to make sure that we're absolutely safe, I moved the camera closer to the -Y edge (right on the picture) of the dichroic (POP_beam_clearance_extreme.jpg) and it still looks OK.

If it's hard to understand what was done, look at the annotated photo (the last attachment After_stiffener_installed_annotated.jpg), the cellphone camera was inserted to "Camera" position.

REFL beam path:

I confirmed that the long stiffener beam doesn't interfere with the motion of the REFL beam diverter. Also, when the REFL beam diverter is open, I looked into the last steering mirror for the REFL air path from the viewport position to make sure that the short stiffener beam won't occult the REFL path. 

Some hiccups:

We used D2500433 -11 variant S/N 4 and -1 variant S/N1 even though page 11 of T2500339 suggests it should be -10 and -2 variant, respectively. We didn't have -10 variant, and -2 variant was absolutely too short.

B&K

We performed B&K hammer measurements before/after the stiffener installation for POP periscope. Before, there was a 70Hz-ish peak. After, it was pushed higher up in frequency. The transducer was attached to the ISI table and Jim hammered the top of the periscope.

Likewise we did B&K test for the input periscope of the JAC even though it was not absolutely necessary.

We haven't done B&K for the JAC output periscope because it's not even fully clamped down (we will move it).

Jim will post the data.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 17:56, Tuesday 20 January 2026 (88823)

Unused stiffener parts are in my office for now.

corey.gray@LIGO.ORG - 07:29, Wednesday 21 January 2026 (88825)EPO

Tagging EPO for photos.

jim.warner@LIGO.ORG - 10:13, Wednesday 21 January 2026 (88830)

These are the measurements we got with the B&K of the POP periscope before and after adding the stiffener. For the POP periscope, we mounted the accel to the table, right at the foot of the periscope, and did the impacts at the top of the post. The accelerometer was mounted with the sensors Z aligned vertically, the Y axis was roughly parallel to the edge of the ISI, so it was mostly pointing along the IFO X arm.  First and second images are impacts in the IFO X and Y dofs, you can pretty clearly see the 70hz post resonance. Third, fourth and fifth images are with the accelerometer in the same spot after adding the stiffener, it seems the mode has successfully been moved much higher to around 170hz. This should allow me to increase the ugf and loop gain quite a bit, to be more like the other ISI. I'll verify with tfs on the ISI after the vent. 

Last 3 images are B&K measurements of the JAC periscope, accel for these measurements was mounted on the edge of the table, with x,y,z sensor dofs aligned to the IFO x,y,z dofs. This also looks pretty good, first features are over 200hz.

Attached zip contains the csv data exports of each of the measurements. Names indicate the direction of the hit with the hammer, relative to IFO x,y,z conventions. 

Images attached to this comment
Non-image files attached to this comment
jennifer.wright@LIGO.ORG - 18:40, Friday 23 January 2026 (88881)

As per Jeff's request we took osme zoomed out photos of the periscope today.

Images attached to this comment
H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 13:59, Friday 16 January 2026 - last comment - 09:49, Wednesday 21 January 2026(88788)
HAM7 - ZM4 physically moved in Yaw

Sheila, Karmeng, Rahul

This morning we physically moved ZM4 (since the sliders were saturating the DAQ output) in YAW to better align it with ZM5 and ZM6. More details (including health check results) will be posted later.

ZM4 is now dogged down to its new position. 

Comments related to this report
ryan.crouch@LIGO.ORG - 09:49, Wednesday 21 January 2026 (88827)SUS

I ran a quick transfer functions on ZM4, a few things to note. The suspension is in air so I changed the Coil Driver state from 2 to 1 (reverted once I was done) as the wiki suggested, the ISI is also tripped but the damping was engaged on the suspension. We were also slightly shaking due to an 6.1 earthquake from near Japan.

The suspension is looking healthy.

/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM4/SAGM1/Data/

2026-01-21_1700_H1SUSZM4_M1_WhiteNoise_L_0p01to50Hz.xml
2026-01-21_1700_H1SUSZM4_M1_WhiteNoise_P_0p01to50Hz.xml
2026-01-21_1700_H1SUSZM4_M1_WhiteNoise_Y_0p01to50Hz.xml

Images attached to this comment
H1 SUS (CDS, SEI)
jeffrey.kissel@LIGO.ORG - posted 08:45, Tuesday 13 January 2026 - last comment - 11:59, Thursday 22 January 2026(88745)
H1SUSB123 and H1SUSH34 SUS and SEI Systems prepped in SAFE for Conversion to SUSB13 and SUSH34
J. Kissel
ECR E2500296, E2400409
WP 12962

D2300401 (for susb13) and D2300383 (for susb2h34)

We begin the major upgrade of the H1SUSB123 and H1SUSH34 SUS and SEI Systems converting them to SUSB13 and SUSH34 a la G2301306 today. We're focusing on upgrading the DACs in the IO chassis () and all the downstream surrounding impact of that  analog electronics This will take down the following computers, and they will be resurrected with new names as follows:
    FORMER NAME           FORMER SUS              NEW NAME           NEW SUS
    h1susb123             ITMX, ITMY, BS          h1susb13           ITMX, ITMY
    h1sush34              MC2, PR2, SR2           h1susb2h34         BS, MC2, PR2, SR2, LO1, LO2
    h1susauxb123          ITMX, ITMY, BS          h1susauxb13        ITMX, ITMY
    h1susauxh34           MC2, PR2, SR2           h1susauxb2h34      BS, MC2, PR2, SR2, LO1, LO2

As such, I've 
    - brought the ITMX, ITMY, BS, HAM3, and HAM4 SEI systems to ISI_DAMPED_HEPI_OFFLINE (so we don't risk any "hard" trips of HEPI during all this).
    - brought all impacted SUS gaurdians to AUTO mode, then to the SAFE state
    - Increased the bypass time on all impacted software watchdogs to bypass time to a large number (90000000 secs), and hit BYPASS
Comments related to this report
marc.pirello@LIGO.ORG - 16:20, Tuesday 13 January 2026 (88752)

Updated timing cards on newly named SUSB13, SUSB13 AUX, SUSB2H34, SUSB2H34 AUX, formerly known as SUSB123, SUSB123 AUX, SUSH34 and SUSH34 AUX.

Timing FPGA Version 1589

marc.pirello@LIGO.ORG - 11:59, Thursday 22 January 2026 (88849)

While we were modifying the power systems in the SUS racks, the AA chassis S1202818 lost its negative voltage rail.  We pulled the chassis and replaced the power board.  This chassis was placed directly back into service.

M. Pirello, O. Patane, J. Kissel

Displaying reports 2661-2680 of 88961.Go to page Start 130 131 132 133 134 135 136 137 138 End