marc.pirello@LIGO.ORG - posted 09:31, Thursday 22 January 2026 (88842)
EOM Heliax Cables Pulled and Landed at PSL-R2

Per WP12989 we completed the EOM cable install from PLS-R2 to the HAM-1 feedthrough.  The old EOM cables from the PSL were disconnected at the patch panel, and the RF Combiner was powered down.  RF Cables were landed at the patch panel from left to right, 24MHz, 118MHz, 45MHz, 9MHz, with 2 empty spigots on the right hand side of the panel located at U2 in the rack near the floor.  We will shoot these with the TDR to once everything is installed to get total electrical length for calibration.

M. Pirello, F. Clara

H1 General
ryan.crouch@LIGO.ORG - posted 07:25, Thursday 22 January 2026 - last comment - 08:26, Thursday 22 January 2026(88840)
OPS Thursday day shift start

TITLE: 01/22 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: MAINTENANCE
    Wind: 5mph Gusts, 3mph 3min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.17 μm/s 
QUICK SUMMARY:

Comments related to this report
ryan.crouch@LIGO.ORG - 08:26, Thursday 22 January 2026 (88841)

There is a planned DAQ restart today as well.

H1 General
ryan.crouch@LIGO.ORG - posted 16:30, Wednesday 21 January 2026 (88835)
OPS Wednesday DAY shift summary

TITLE: 01/22 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: OPO and JAC work continued in HAMs 1 & 7 in LASER HAZARD.
LOG:                                                                                                                                                                                                       

Start Time System Name Location Lazer_Haz Task Time End
22:49 SAF LVEA IS LASER HAZARD LVEA YES LVEA IS LASER HAZARD \u0d26\u0d4d\u0d26\u0d3f(\u239a_\u239a) 16:49
15:31 FAC Randy EndY N Bring a forkflit down to the CS/High bay 17:24
15:53 FAC Kim, Nellie LVEA Y Tech clean 17:53
17:01 SUS RyanC CR N ZM4 Transfer function 17:24
17:07 JAC Matt, Sophie PREP lab N JOTE work 19:36
17:10 SQZ Sheila, RyanS, Kar Meng HAM7 Y Chamber work, Kar Meng in at 1900 21:02
17:17 PEM Robert LVEA HAM4 Y Laser vibro, in and out 18:17
17:57 ISC Jason LVEA Y HAM1 work 20:52
18:07 ISC Rahul LVEA Y HAM1 20:05
18:20 FAC Eric HAM Shaq N Changing filters 18:51
18:27 ISC Jennie LVEA Y HAM1 work 19:14
19:22 ISC Jennie, Keita LVEA Y HAM1 work 20:52
19:33 VAC Travis Optics lab N Put away part 19:43
19:36 FAC Kim H2 N Tech clean 19:54
19:53 CDS Oli LVEA, CER Y Take pictures of computer racks 20:05
21:21 VAC Travis MidX N Look for VAC equipment 21:39
21:21 VAC Jordan FCES N Look for VAC equipment 21:49
21:34 SUS Rahul Optics lab LOCAL Work on EOM 23:03
22:02 EE Marc, Fil LVEA Y Run cable for EOM from HAM1 to PSL racks Ongoing
22:33 VAC Gerardo VAC prep lab to LVEA Y Bring over glass piece to HAM1 23:38
23:38 VAC Gerardo VAP prep lab N Assembly work 00:19

19:24 OPO WD trip

H1 PSL
ryan.short@LIGO.ORG - posted 16:02, Wednesday 21 January 2026 (88838)
PSL 10-Day Trends

FAMIS 31121

Things are looking very stable over the past week; makes sense since very little has been asked of the PSL recently. The PMC and RefCav could still use an alignment touch-up, which I'll get to hopefully this week.

Images attached to this report
H1 SUS
oli.patane@LIGO.ORG - posted 15:26, Wednesday 21 January 2026 (88837)
Renamed SR3 OPLEV Whitening to AA cable

Jeff, Oli

We decided to rename the SR3 OPLEV Whitening to AA chassis cable from SUS_SR3-168 to SUS_OPLEV_SR3_AA. We decided to do this for a couple reasons: SUS_SR3-168 is not a very helpful name, it matches better with the other OPLEVs' cables, and anyway we had already been calling that cable SUS_OPLEV_SR3_AA in our wiring diagrams for a while. 
We relabeled both sides of the cable.

H1 SQZ (SYS)
karmeng.kwan@LIGO.ORG - posted 14:34, Wednesday 21 January 2026 (88836)
HAM7 progress update

[Sheila, Ryan, Karmeng]

We have returned ZM2 to the original position at the beginning of O4, we alleviated some of the saturation but the ZM3 iris still require adjustment.

Power budget looks good so far. With 0.77mW exiting the OPO and reaches AM3 at the same power. We observed 0.75mW of OPO trans and 18.1mW of OPO refl on SQZT7 periscope.

H1 IOO (IOO)
rahul.kumar@LIGO.ORG - posted 13:04, Wednesday 21 January 2026 - last comment - 09:41, Wednesday 28 January 2026(88834)
HAM1 update - JAC damper

Jason. JennieW, Rahul

This morning we added the damper parts to the JAC in HAM1 chamber - see picture attached. The cables were re-routed slightly so that they don't interfere with the damper.

I have finger tightened the screws of the damper for now - since no torque spec was provided and we were worried that tightening them too much might affect its position/angle.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 08:27, Wednesday 28 January 2026 (88934)EPO

EPO-tagged.

jason.oberling@LIGO.ORG - 09:41, Wednesday 28 January 2026 (88941)

Masayuki and I fully tightened the JAC body mode dampers yesterday.  We were only able to get the screws ~1/2 turn past finger-tight before we could no longer tighten them.  No changes in JAC alignment were observed while we were doing this.

H1 SUS (SUS)
ryan.crouch@LIGO.ORG - posted 11:44, Wednesday 21 January 2026 - last comment - 08:25, Wednesday 28 January 2026(88828)
2nd HXDS received from LLO

Yesterday I picked up the shipment from LLO contain the 2nd HXDS (first HXDS shipment alog87998) and related ASSEYs and brought it to the Triples lab. 

Uwrapping and inspecting was uneventful, the zip ties were intact, the shock pads untripped and the humidity indicator was unchanged in color. The outer (front, back) and inner (front, back) bags were in good condition, unwrapping the suspension I found no play with the blades.  All the nuts and dowel pins were tight and well seated, the ICS looks good except for it seems to be missing the HDS Intermediate Mass Fixture (D2000230-V1, s/n 08). The part is scribed as V1, on the dcc there's a V2 with two added 8-32 helicoils taps that, based on D1900352 which uses V2, I should be able to see from the front and I don't, so this is a V1 plate and not an incorrectly scribed V2. I believe these two missing 8-32 holes are where we will attach the BOSEM Mounting plate (D2000257). The first HXDS we received in November had the V2 plate. We may have to do a small rework to update it to V2, or have LLO send the V2 part if possible.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 08:25, Wednesday 28 January 2026 (88933)EPO

EPO tagging

LHO General
eric.otterman@LIGO.ORG - posted 11:03, Wednesday 21 January 2026 (88832)
FCES air handler
The pre-filters were replaced in the FCES air handler. 
LHO VE
david.barker@LIGO.ORG - posted 10:16, Wednesday 21 January 2026 (88831)
Wed CP1 fill

Wed Jan 21 10:08:13 2026 INFO: Fill completed in 8min 10secs

 

Images attached to this report
H1 SEI
oli.patane@LIGO.ORG - posted 09:59, Wednesday 21 January 2026 (88829)
ISI CPS Noise Spectra Check Weekly FAMIS

Closes FAMIS#39334, last checked 88680

The elevated sensors listed are just from HAM1 and HAM7, which is expected
All chambers are looking normal relative to the current state that they are in (locked, vented, etc)

Non-image files attached to this report
H1 General
ryan.crouch@LIGO.ORG - posted 07:32, Wednesday 21 January 2026 (88826)
OPS Wednesday Day shift start

TITLE: 01/21 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: MAINTENANCE
    Wind: 5mph Gusts, 3mph 3min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.17 μm/s 
QUICK SUMMARY:

Some of the tasks on the board today:

H1 ISC (INS, ISC)
keita.kawabe@LIGO.ORG - posted 17:53, Tuesday 20 January 2026 - last comment - 18:40, Friday 23 January 2026(88822)
POP periscope stiffener was installed. B&K for POP as well as JAC input periscope (Rahul, Jim, Keita)

We have intalled the POP periscope stiffener.

Some dog clamps in the REFL path as well as the cable bracket for PM1 were relocated to accomodate. 1st and 2nd picture are "before" photo. 3rd one is after PM1 cable bracket relocation but before installing the stiffener.  There are also three "after" photos showing how things look on the table.

POP beam clearance:

I took the picture of the bottom periscope mirror through dichroic (HR for IR, transmission for green) to see the beam clearance. In the first such photo (POP_beam_clearance.jpg), the camera is close to the center of the dichroic and the short stiffener beam is close to the edge of the optic but not occulting the mirror, so we're OK. Just to make sure that we're absolutely safe, I moved the camera closer to the -Y edge (right on the picture) of the dichroic (POP_beam_clearance_extreme.jpg) and it still looks OK.

If it's hard to understand what was done, look at the annotated photo (the last attachment After_stiffener_installed_annotated.jpg), the cellphone camera was inserted to "Camera" position.

REFL beam path:

I confirmed that the long stiffener beam doesn't interfere with the motion of the REFL beam diverter. Also, when the REFL beam diverter is open, I looked into the last steering mirror for the REFL air path from the viewport position to make sure that the short stiffener beam won't occult the REFL path. 

Some hiccups:

We used D2500433 -11 variant S/N 4 and -1 variant S/N1 even though page 11 of T2500339 suggests it should be -10 and -2 variant, respectively. We didn't have -10 variant, and -2 variant was absolutely too short.

B&K

We performed B&K hammer measurements before/after the stiffener installation for POP periscope. Before, there was a 70Hz-ish peak. After, it was pushed higher up in frequency. The transducer was attached to the ISI table and Jim hammered the top of the periscope.

Likewise we did B&K test for the input periscope of the JAC even though it was not absolutely necessary.

We haven't done B&K for the JAC output periscope because it's not even fully clamped down (we will move it).

Jim will post the data.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 17:56, Tuesday 20 January 2026 (88823)

Unused stiffener parts are in my office for now.

corey.gray@LIGO.ORG - 07:29, Wednesday 21 January 2026 (88825)EPO

Tagging EPO for photos.

jim.warner@LIGO.ORG - 10:13, Wednesday 21 January 2026 (88830)

These are the measurements we got with the B&K of the POP periscope before and after adding the stiffener. For the POP periscope, we mounted the accel to the table, right at the foot of the periscope, and did the impacts at the top of the post. The accelerometer was mounted with the sensors Z aligned vertically, the Y axis was roughly parallel to the edge of the ISI, so it was mostly pointing along the IFO X arm.  First and second images are impacts in the IFO X and Y dofs, you can pretty clearly see the 70hz post resonance. Third, fourth and fifth images are with the accelerometer in the same spot after adding the stiffener, it seems the mode has successfully been moved much higher to around 170hz. This should allow me to increase the ugf and loop gain quite a bit, to be more like the other ISI. I'll verify with tfs on the ISI after the vent. 

Last 3 images are B&K measurements of the JAC periscope, accel for these measurements was mounted on the edge of the table, with x,y,z sensor dofs aligned to the IFO x,y,z dofs. This also looks pretty good, first features are over 200hz.

Attached zip contains the csv data exports of each of the measurements. Names indicate the direction of the hit with the hammer, relative to IFO x,y,z conventions. 

Images attached to this comment
Non-image files attached to this comment
jennifer.wright@LIGO.ORG - 18:40, Friday 23 January 2026 (88881)

As per Jeff's request we took osme zoomed out photos of the periscope today.

Images attached to this comment
H1 General
oli.patane@LIGO.ORG - posted 16:29, Tuesday 20 January 2026 (88821)
Ops Day Shift End

TITLE: 01/21 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: Continued alignment for HAM7, and BNK measurements on HAM1 periscopes. Currently the LVEA is LASER HAZARD
LOG:                                                                                                                                                                                                                                                                                                                        

Start Time System Name Location Lazer_Haz Task Time End
15:37 FAC Randy LVEA n Craning rails (started at 15:15) 16:09
15:37 FAC Nellie LVEA n Tech clean 16:21
16:11 FAC Randy, Corey LVEA YES Craning 17:23
16:31 VAC Travis MX, EX n Turbo pump tests 21:10
16:35 SQZ Sheila LVEA YES HAM7 alignment 19:43
16:51 FAC Nellie LVEA YES Tech clean 17:30
16:57 TCS Matt JOAT Lab n Building up second table 19:21
17:17 SEI Jim, Keita LVEA YES BNK 19:52
17:23 PEM Robert LVEA YES Mounting accelerometers 18:53
17:37 SQZ KarMeng LVEA YES HAM7 alignment 19:43
17:37 SAF LVEA IS LASER HAZARD LVEA YES LVEA IS LASER HAZARD 20:30
17:46 FAC Randy, Corey LVEA YES Craning 19:00
18:09   Matt LVEA YES Grabbing laptop 18:26
18:13 VAC Jordan MY, EY n Turbo pump tests 22:04
19:04 SEI Rahul LVEA YES Helping Jim 19:24
19:17 PCAL Tony PCAL Lab y(local) Getting ready for a measurement 19:24
19:22 FIT Matt YARM n Running 20:30
19:43 SAF Corey LVEA YES Transitioning to Laser SAFE 20:03
20:09 EPO Mike, Rana, tour LVEA n Tour 21:09
20:33 EE Fil LVEA YES Cabling 00:17
20:41 FAC Randy LVEA n Securing railing 21:09
21:10 SAF Tony LVEA YES Transitioning to laser HAZARD 21:28
21:12 SQZ Sheila, KarMeng LVEA YES HAM7 alignment 22:34
21:40 FAC Tyler LVEA, MX, MY YES 3IFO checks 23:40
21:52 VAC Travis Optics Lab n Grabbing tool 21:58
22:08 JAC Jason LVEA YES Parts hunting and FARO spot hunting 22:31
22:46 PCAL Tony PCAL Lab y(local) Finding case for beamsplitter 23:17
22:49 PEM Robert LVEA YES Mounting accelerometers ongoing
22:49 SAF LVEA IS LASER HAZARD LVEA YES LVEA IS LASER HAZARD ongoing
22:50 JAC Jason, Jennie Optics Lab n Hunting for parts 23:02
00:18 SEI Jim, Keita LVEA YES BNK measurements 00:20
H1 SQZ (SQZ)
karmeng.kwan@LIGO.ORG - posted 16:06, Tuesday 20 January 2026 - last comment - 22:28, Tuesday 20 January 2026(88806)
HAM7 progress

[Sheila, Karmeng]

FC1 alignment was adjusted to account for the buoyancy, and we coaligned the FC1 green/IR beam. We managed to recover the FC refl green on SQZT7 periscope, but not on the photodiode.

The OPO IR transmission is misaligned, and we managed to recover the IR on SQZT7 PD. As we recovered the IR refl (top) on SQZT7, we saw an improvement from ~600 to ~1680 (we may have been clipping on the refl PD previously).

By changing the setting on ZM2, we managed to improve the IR trans (bottom).

Old setting ZM2 PIT: 424 ; YAW: -235

New setting ZM2 PIT: 444 ; YAW: -315

Will need to check on the iris tomorrow, to make sure we did not deviate from the original path.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 22:28, Tuesday 20 January 2026 (88824)

Some additional information:

In december we checked that the alignment sliders on FC1 had not moved since the observing run.  Noting that the red and green co-alignment were off, this morning I trended osems as well and saw that FC1 had a shift of +200urad in pitch at the time of the vent.  The ZMs all had similar shifts.  When I moved FC1 to make sure the osem values matched the observing run pitch, the beam was then clipped on B:P1.  We adjusted B:M1 to clear the VIP, then adjusted B:M4 to recenter the beam on the ZM4 iris.  Then Kar Meng adjusted ZM4 alignment to match the irises on SQZT7 in the homodyne path.  After this we saw that the beams were still on the QPDs in HAM6, but not well centered.  We will have to return to that tomorow.  

After that we turned our attention to the co-alignment of the FC green sensing with the squeezed beam.  The shift in the FC1 pitch wasn't enough to make a difference in the large misalingment of these two.  We became confused for a while looking at the FC green beam in the SK path.  The old HAM7 layout shows the SK path on the -X side of the three beams leaving the chamber toward the green periscope, this is correct but the two beams cross as they travel down the periscope so that the FC green refl path is on the -X side by the time it reaches the bottom, and looking at the SQZT7 layout shows the SK path located on the table in the area where the FC green refl path is. 

Once we understood this path, we were able to walk H:M1 to coalign red and green right after A:DC2, and using H:M2 to co-align them at the iris after ZM3.  Then we adjusted the pointing to the periscope using H:M4.  

While we were in laser safe, I unlocked the OPO suspension, since we hope that we are finished with the alignment work on the VIP.  This caused enough of a shift that the reflected seed beam was clipping on its way to the diode, perhaps on the lens in HAM7.  We moved this beam so that it is still mis centered on the lens but not close to the edge.  

H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 13:59, Friday 16 January 2026 - last comment - 09:49, Wednesday 21 January 2026(88788)
HAM7 - ZM4 physically moved in Yaw

Sheila, Karmeng, Rahul

This morning we physically moved ZM4 (since the sliders were saturating the DAQ output) in YAW to better align it with ZM5 and ZM6. More details (including health check results) will be posted later.

ZM4 is now dogged down to its new position. 

Comments related to this report
ryan.crouch@LIGO.ORG - 09:49, Wednesday 21 January 2026 (88827)SUS

I ran a quick transfer functions on ZM4, a few things to note. The suspension is in air so I changed the Coil Driver state from 2 to 1 (reverted once I was done) as the wiki suggested, the ISI is also tripped but the damping was engaged on the suspension. We were also slightly shaking due to an 6.1 earthquake from near Japan.

The suspension is looking healthy.

/ligo/svncommon/SusSVN/sus/trunk/HXDS/H1/ZM4/SAGM1/Data/

2026-01-21_1700_H1SUSZM4_M1_WhiteNoise_L_0p01to50Hz.xml
2026-01-21_1700_H1SUSZM4_M1_WhiteNoise_P_0p01to50Hz.xml
2026-01-21_1700_H1SUSZM4_M1_WhiteNoise_Y_0p01to50Hz.xml

Images attached to this comment
Displaying reports 2681-2700 of 88975.Go to page Start 131 132 133 134 135 136 137 138 139 End