Displaying reports 4541-4560 of 89199.Go to page Start 224 225 226 227 228 229 230 231 232 End
Reports until 10:24, Thursday 25 September 2025
LHO VE
david.barker@LIGO.ORG - posted 10:24, Thursday 25 September 2025 (87142)
Thu CP1 Fill

Thu Sep 25 10:09:27 2025 INFO: Fill completed in 9min 23secs

 

Images attached to this report
H1 SQZ
camilla.compton@LIGO.ORG - posted 10:00, Thursday 25 September 2025 (87138)
Could not easily re-create low frequency SQZ noise with PSAMS

In 87071, we found that the low frequency SQZ noise could be removed by changing the ZM4 or ZM5 PSAMs and we left the ZM4 PSAMs at 6.2V strain (from 6.0V). 

Today I went back to the noisy area at 6.0V and lower on ZM4 and saw no increase in low frequency noise, plot attached. I did not try going higher than 6.2V yet, we could do that as 6.5V caused noise on Monday too. 

It is not surprising that this noise couldn't be re-created as it was coming and going in the past. 

Images attached to this report
H1 CAL
ryan.short@LIGO.ORG - posted 09:27, Thursday 25 September 2025 (87137)
Broadband and Simulines Calibration Sweeps

Following the procedure in the TakingCalibrationMeasurements wiki, at 15:30 UTC H1 dropped observing for the usual calibration sweeps.

Broadband - 15:30:52 to 15:36:02 UTC

Simulines - 15:36:37 to 16:00:00 UTC

Calibration monitor screenshot and calibration report attached.

File written out to: /ligo/groups/cal/H1/measurements/DARMOLG_SS/DARMOLG_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/PCALY2DARM_SS/PCALY2DARM_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L1_SS/SUSETMX_L1_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L2_SS/SUSETMX_L2_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L3_SS/SUSETMX_L3_SS_20250925T153638Z.hdf5

 

Images attached to this report
Non-image files attached to this report
LHO General
ryan.short@LIGO.ORG - posted 07:45, Thursday 25 September 2025 (87136)
Ops Day Shift Start

TITLE: 09/25 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 3mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.16 μm/s 
QUICK SUMMARY: H1 has been locked for 2.5 hours. Looks like the lockloss last night was from yet another earthquake, this time likely a M5.7 from Mexico. Calibration and commissioning time scheduled today from 15:30 to 20:00 UTC.

LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 22:00, Wednesday 24 September 2025 (87135)
OPS Eve Shift Summary

TITLE: 09/25 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Earthquake
INCOMING OPERATOR: Tony
SHIFT SUMMARY:

IFO is in LOCKING in LOCKING_ARMS_GREEN

Very calm shift with a fully auto Lock acquisition from the M6.1 Venezuela EQ this afternoon (alog 87128).

We had one Lockloss due to another earthquake from Venezuela, M6.3. Lockloss alog 87134. I turned on the ASC High Gain before it happened, but we understandably still lost lock.

I've just set IFO to auto-lock, which it will do once the EQ rings down (I expect it to take another 30 mins to get to a lockable state).

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:48 VAC Pump LVEA N AIP pumping on HAM6 23:43
22:10 ISC Jennie Opt Lab Local ISS array work 23:31
22:54 VAC Gerardo LVEA N Checking on AIP 23:16
H1 PEM (Lockloss)
ibrahim.abouelfettouh@LIGO.ORG - posted 21:20, Wednesday 24 September 2025 (87134)
Lockloss 04:01 UTC

Lockloss due to another EQ from Venezuela - 6.3 Magnitude

H1 ISC
elenna.capote@LIGO.ORG - posted 17:09, Wednesday 24 September 2025 (87133)
DARM offset step results

I previously reported results from DARM offset step tests in October and last week. The overall goal here is to figure out what overall effect the power outage has had on the IFO, especially since we have lost 1% optical gain. For example, see this alog about comparing modulation depth tests.

However, after looking closer at the DARM offset measurement results, and especially investigating the effect of whether leaving the OMC ASC on during the measurement matters, I would like to revise the previously reported measurement results.

Some background: this test aims to measure the contrast defect light by changing the DARM offset in mA while injecting strong PCAL lines to capture the DARM optical gain. We expect the relationship between optical gain (mW/pm) and DARM offset power (mW) to be quadratic. We assume by fitting the data to a parabola, the resulting y-intercept can tell us the contrast defect light on the DCPDs, if we had zero DARM offset.

Some considerations:

First question, does leaving the OMC ASC on matter? Jennie has been doing a lot of these measurements, and her experience has been, yes, because the OMC ASC offsets are set at one DARM offset and are not reset at each different offset. We do not expect the OMC alignment to drift significantly during the 15 minutes of this measurement, so the best practice has been to turn the OMC ASC OFF for the measurement. However, I didn't do it the first time I measured last week, but I got significantly better results (using Craig's code) with the ASC ON than I did with the ASC OFF.

Second question, what is the best way to fit this data? This apparently has been a question for a long time, see Gabriele's comment to 30573 in 2016. I believe the model I described above suggests that the vertex of the resulting parabola should be centered around x=0, that is zero optical gain. However, I discovered this week that Craig's code, which I used to report the results in October and last week (see first two links), fits a nonzero vertex (i.e. y = a(x-x0)^2 + b instead of y = ax^2 + b). Of course, that will change the answer significantly! I don't understand the mechanism that would cause us to move the center of this vertex away from x=0, so I refitting the data to follow y = ax^2 + b.

Results:

I am revising the fitted contrast defects to be:

Frequency October 2024 (OMC ASC on) [mW] September 2025 (OMC ASC on) [mW] September 2025 (OMC ASC off) [mW]
255.0 Hz 1.088 +- 0.033 1.219 +- 0.032 1.256 +- 0.005
410.3 Hz 1.086 +- 0.030 1.214 +- 0.042 1.240 +- 0.008

Conclusions:

This measurement shows that the contrast defect is higher now than it was in October. I don't know if we can attribute all of this to the power outage; we lost 1% optical gain comparing kappa c from before and after the vent, giving us at least 2% less optical gain between October and now. It also shows that there is perhaps a small effect related to the OMC ASC being off: slightly higher contrast and a better overall fit.

This also increases our upper limit on the possible homodyne angle- in October I stated that the homodyne angle upper limit would be about 7 degrees, but with this revised value it should have been more like 8.8 degrees. Now it is 9.3 degrees.

LHO General
ryan.short@LIGO.ORG - posted 16:30, Wednesday 24 September 2025 (87130)
Ops Day Shift Summary

TITLE: 09/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Earthquake
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Very quiet day of H1 observing until an earthquake caused a lockloss in the late afternoon. Once the ground motion rings down, Ibrahim will start relocking.

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:48 VAC Pump LVEA N AIP pumping on HAM6 23:43
16:10 FAC Kim Opt Lab N Technical cleaning 16:26
17:02 SEI Jim, Mitchell Opt Lab N Taking parts out of clean bags 17:30
17:58 CAL Tony PCal Lab N Looking for a cable 18:34
20:08 ISC Jennie Opt Lab Local ISS array work 21:00
20:41 CAL Tony PCAL Lab N Dropping off supplies 21:25
20:49 AOS Betsy Opt Lab N Checking bag contents 21:16
21:26 SPI Jeff, RyanS Opt Lab N Looking at SPI optics 21:41
22:10 ISC Jennie Opt Lab Local ISS array work 00:09
22:54 VAC Gerardo LVEA N Checking on AIP 23:16
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 16:22, Wednesday 24 September 2025 (87129)
OPS Eve Shift Start

TITLE: 09/24 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Earthquake
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
    SEI_ENV state: LARGE_EQ
    Wind: 6mph Gusts, 3mph 3min avg
    Primary useism: 0.69 μm/s
    Secondary useism: 0.18 μm/s 
QUICK SUMMARY:

IFO is DOWN due to EARTHQUAKE

Just waiting for this 6.2 EQ from Venezuela to pass through and then back to LOCKING

 

H1 General (Lockloss, SEI)
ryan.short@LIGO.ORG - posted 15:39, Wednesday 24 September 2025 (87128)
Lockloss @ 22:31 UTC - EQ

Lockloss @ 22:31 UTC after 11.5 hours locked - link to lockloss tool

Caused by S-waves from a M6.1 EQ out of Venezuela. Happened less than a minute after the start of the transition to EQ mode, so I didn't get a chance to try the ASC Hi-Gain feature to ride through it (although, in this case I don't suspect it would have helped). Holding H1 in DOWN as it's still 10 minutes before the R-waves are supposed to hit.

LHO General
ryan.short@LIGO.ORG - posted 07:39, Wednesday 24 September 2025 - last comment - 11:28, Wednesday 24 September 2025(87116)
Ops Day Shift Start

TITLE: 09/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 153Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 2mph Gusts, 0mph 3min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.14 μm/s 
QUICK SUMMARY: H1 has been locked for 3.5 hours. One lockloss overnight with an unknown cause, but relocking seems to have gone without issue. We will have a short planned commissioning window from 16:00 to 17:00 UTC.

Comments related to this report
ryan.short@LIGO.ORG - 11:28, Wednesday 24 September 2025 (87122)

The hour of commissioning time did not happen today due to calibration report issues at LLO. This has been added onto tomorrow's regularly scheduled commissioning time, so we will have a total of 4.5 hours (with the first 30 minutes dedicated to calibration sweeps, as usual).

H1 ISC
elenna.capote@LIGO.ORG - posted 18:17, Tuesday 23 September 2025 - last comment - 16:36, Wednesday 24 September 2025(87114)
Modulation Depth test results - March vs September

Back in March, I ran a modulation depth test, with several goals in mind related to providing useful calibration information for modeling. That alog is still unfortunately sitting in my drafts. However, I was able to use the results to make a side-by-side comparison with a modulation depth test Sheila ran last Thursday, after the power outage. We are still trying to understand what the overall effect of the power outage was on the IFO. Namely, we have lost 1% of optical gain, 86964.

Some background: the modulation depth test aims to measure the fraction of carrier, 9 MHz and 45 MHz power at each port. This is done by measuring the powers at the nominal settings, and then individiually stepping the 9 and 45 MHz modulation up or down by a known value. Using a measured calibration of V/dBm and rad/V (see alog 62883), and using the bessel functions, the power fraction of each field can be measured at each diode based on how much the total diode power changes at each step. (note: I still have a to do list item to better calibrate the modulation depth in radians using OMC scan data).

Procedure: I stepped both down and up in modulation depth, resulting in 5 different measurements (nominal, 9 down, 9 up, 45 down, 45 up). In March, I stepped down by 3 dBm and up by 2 dBm, but in September we were able to step both down and up by 3 dBm. I measured for 3 minutes at each step in March, and 1 minute at each step in September.

EDIT: I realized I made an error, and the first results I report below are actually from Feb 2025. The significant difference here is that in my February measurement I only stepped down by 3 dBm for 9 and 45 MHz each, so there is less data to fit. In my March measurement, I stepped up 2 dBm and down 3 dBm, getting 5 total different measurement times. There should be very little difference in the interferometer between February and March 2025, so the differences in the results I believe are due to the fact that more points (5 versus 3) gives you a much better fit to the data. I would compare March and now for a more accurate understanding of the differences.

February results, well before power outage, only fit from 3 data points:

Field Input POP REFL AS
carrier 0.9771 0.9842 0.9419 0.3386
9 MHz 0.01278 0.01527 0.02950 0.1787
45 MHz 0.01003 0.000474 0.02860 0.4827

March results, before power outage, fit from 5 data points:

Field Input POP REFL AS
carrier 0.9779 0.9831 0.9164 0.2134
9 MHz 0.01212 0.01332 0.04059 0.2761
45 MHz 0.009687 0.002302 0.04431 0.5413

September results, after power outage (and reduction of PSL power, attenuation at IMC REFL), fit from 5 data points:

Field Input POP REFL AS
carrier 0.9783 0.9833 0.9333 0.3556
9 MHz 0.01195 0.01325 0.02533 0.1852
45 MHz 0.009486 0.001909 0.04199 0.4641

EDIT: Including the March results (and trusting them more than the February results) changes the conclusion. The input ratios are very similar between all three measurements. This is also true for POP where carrier and 9 ratios are concerned. 45 MHz is the hardest field to measure because the 9 and carrier are so strong at POP. There may be less 45 MHz at POP, or this is just the measurement uncertainty. At REFL, there may be an increase in carrier light now after the power outage, and there may be half as much 9 MHz as in March.

The most dramatic differences are at the AS port. Just comparing the February and March measurements, there may be considerable uncertainty in how much of each field is at AS in general. However, if we choose to believe the March results, this would suggest a significant increase in carrier at AS, and a significant decrease in 9 and 45 MHz. However, comparing February and now, the 9 MHz and carrier are nearly the same, and the 45 seems to have decreased.

Comments related to this report
elenna.capote@LIGO.ORG - 11:30, Wednesday 24 September 2025 (87123)

Rereading this alog, I see that I should say specifically which diodes measure these powers:

Input == IM4 trans

POP == POP A LF

REFL == REFL A LF

AS = AS_C DC NSUM

elenna.capote@LIGO.ORG - 16:36, Wednesday 24 September 2025 (87131)

Here are plots of various channels during the March and September mod depth tests. The shaded region indicates which step was being taken at the time, and the dotted line matching each shading color indicates the median of the channel at that time.

Non-image files attached to this comment
H1 SUS
filiberto.clara@LIGO.ORG - posted 13:27, Tuesday 23 September 2025 - last comment - 11:48, Wednesday 24 September 2025(87103)
Sat Amps Modified: PRM, SRM, ITMY, ITMX, BS, and BS

WP 12802
ECR E2400330
Modified List T2500232

The following SUS SAT Amps were upgraded per ECR E2400330. Modification improves the whitening stage to reduce ADC noise from 0.05 to 10 Hz.

Suspension Old New OSEM
PRM M2 S1100080 S1100064 ULLLURLR
PRM M3 S1100106 S1000285 ULLLURLR
SRM M2 S1100078 S1100095 ULLLURLR
SRM M3 S1000274 S1100091 ULLLURLR
ITMY L1 UIM S1100070 S1100129 ULLLURLR
ITMX L1 UIM S1100141 S1000287 ULLLURLR
BS M2 S1000295 S1100117 ULLLURLR
ETMY UIM S1100159 S1100088 ULLLURLR

 

F. Clara, J. Kissel, O. Patane

Comments related to this report
jeffrey.kissel@LIGO.ORG - 10:56, Wednesday 24 September 2025 (87118)ISC
Here's the characterization data and fit results for S1100064, assigned to PRM M2's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100064_PRM_M2_ULLLURLR_20250917.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PRM      M2       S1100064         CH1                UL           0.0961:5.24     120.25        zpk([5.24],[0.0961],1,"n")
                                   CH2                LL           0.0950:5.18     120.25        zpk([5.18],[0.0950],1,"n")
                                   CH3                UR           0.0961:5.24     120.25	 zpk([5.24],[0.0961],1,"n")
                                   CH4                LR           0.0962:5.25     120.00	 zpk([5.25],[0.0962],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:00, Wednesday 24 September 2025 (87119)
Here's the characterization data and fit results for  S1000285 , assigned to PRM M3's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1000285_PRM_M3_ULLLURLR_20250917.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PRM      M3       S1000285         CH1                UL           0.0956:5.22     120.250       zpk([5.22],[0.0956],1,"n")
                                   CH2                LL           0.0967:5.27     120.250       zpk([5.27],[0.0967],1,"n")
                                   CH3                UR           0.0955:5.21     120.375       zpk([5.21],[0.0955],1,"n")
                                   CH4                LR           0.0950:5.18     120.375       zpk([5.18],[0.0950],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:03, Wednesday 24 September 2025 (87120)
Here's the characterization data and fit results for  S1100095 , assigned to SRM M2's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100095_SRM_M2_ULLLURLR_20250917.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SRM      M2       S1100095         CH1                UL           0.0955:5.22     120           zpk([5.22],[0.0955],1,"n")
                                   CH2                LL           0.0975:5.33     120           zpk([5.33],[0.0975],1,"n")
                                   CH3                UR           0.0951:5.19     120.25        zpk([5.19],[0.0951],1,"n")
                                   CH4                LR           0.0955:5.20     120.25        zpk([5.20],[0.0955],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:08, Wednesday 24 September 2025 (87121)
Here's the characterization data and fit results for  S1100091 , assigned to SRM M3's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100091_SRM_M3_ULLLURLR_20250917.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SRM      M3       S1100091         CH1                UL           0.0983:5.37     120.25        zpk([5.37],[0.0983],1,"n")
                                   CH2                LL           0.0959:5.23     120.25        zpk([5.23],[0.0959],1,"n")
                                   CH3                UR           0.0955:5.23     120.00        zpk([5.23],[0.0955],1,"n")
                                   CH4                LR           0.0957:5.24     120.00        zpk([5.24],[0.0957],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:35, Wednesday 24 September 2025 (87124)
Here's the characterization data and fit results for  S1100129 , assigned to ITMY L1 (UIM) ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100129_ITMY_L1_ULLLURLR_20250916.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
ITMY     L1       S1100129         CH1                UL           0.0956:5.23     121.75        zpk([5.23],[0.0956],1,"n")
                                   CH2                LL           0.0966:5.28     120.00        zpk([5.28],[0.0966],1,"n")
                                   CH3                UR           0.0978:5.34     120.00        zpk([5.34],[0.0978],1,"n")
                                   CH4                LR           0.0966:5.27     120.00        zpk([5.27],[0.0966],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:41, Wednesday 24 September 2025 (87125)
Here's the characterization data and fit results for  S1100287 , assigned to ITMX L1 (UIM) ULLLURLR OSEMs.
	(Note the typo in Fil's main entry -- he quotes S1000287, but it's S1100287.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100287_ITMX_L1_ULLLURLR_20250916.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
ITMX     L1       S1100287         CH1                UL           0.0958:5.23     120           zpk([5.23],[0.0958],1,"n")
                                   CH2                LL           0.0966:5.28     120           zpk([5.28],[0.0966],1,"n")
                                   CH3                UR           0.0941:5.14     120           zpk([5.14],[0.0941],1,"n")
                                   CH4                LR           0.0955:5.23     120           zpk([5.23],[0.0955],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:45, Wednesday 24 September 2025 (87126)
Here's the characterization data and fit results for  S1100117 , assigned to BS M2 ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100117_BS_M2_ULLLURLR_20250917.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
BS       M2       S1100117         CH1                UL           0.0970:5.31     120           zpk([5.31],[0.0970],1,"n")
                                   CH2                LL           0.0975:5.33     120           zpk([5.33],[0.0975],1,"n")
                                   CH3                UR           0.0967:5.30     120           zpk([5.30],[0.0967],1,"n")
                                   CH4                LR           0.0955:5.22     120.375       zpk([5.22],[0.0955],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 11:48, Wednesday 24 September 2025 (87127)
Here's the characterization data and fit results for  S1100088 , assigned to ETMY L1 (UIM) ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100088_ETMY_L1_ULLLURLR_20250916.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
ETMY     L1       S1100088         CH1                UL           0.0968:5.29     120           zpk([5.29],[0.0968],1,"n")
                                   CH2                LL           0.0956:5.23     120           zpk([5.23],[0.0956],1,"n")
                                   CH3                UR           0.0955:5.22     120           zpk([5.22],[0.0955],1,"n")
                                   CH4                LR           0.0959:5.24     120           zpk([5.24],[0.0959],1,"n")
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
LHO VE
jordan.vanosky@LIGO.ORG - posted 12:45, Tuesday 23 September 2025 - last comment - 10:43, Wednesday 01 October 2025(87101)
HAM6 Annulus Ion Pump (AIP) Replacement

During Tuesday maintenance, we swapped the HAM6 AIP (Starcell). Note this annulus system is connected to HAM5 via the septum plate. We vented the lines with dry nitrogen and left a continuous nitrogen purge(~.3 psi) of the line during the pump swap. Nitrogen attached to HAM5 pump out port while HAM6 pump out port was left open to atmosphere.

No issues during the swap, annulus system is now pumping at both HAM5 & 6 ports with an aux cart and turbo pair. As of end of maintenance, the HAM6 cart was at ~3E-5 Torr, HAM5 cart at ~1E-4 Torr. These pumps will continue running until pressure is <1E-5 Torr at which point the ion pumps will be powered on.

Carts are placed on foam for isolation, and a piece of foam between the flex hose running up to HAM6 pump out port and HAM6 chamber. See attached pictures. 

Work permit will be closed once pumps are disconnected from chambers.

Images attached to this report
Comments related to this report
gerardo.moreno@LIGO.ORG - 17:10, Wednesday 24 September 2025 (87132)DetChar, VE

Update.

IFO was out of lock due to an earthquake, I went in to the LVEA to check on the aux-carts pumping down on the annuli for HAM5 and HAM6.  HAM5 aux-cart was good and pumping down on the annulus, however HAM6 aux-cart safety valve somehow managed to trip between yesterday and today, time is unknown as of now, I restored aux cart, and opened the valve.  Aux-cart for HAM6 was reporting a dubious pressure number of 1.26 x 10-07 Torr.

After restoring pumping to HAM6 annulus, both aux carts are reporting more believable numbers.

Images attached to this comment
gerardo.moreno@LIGO.ORG - 10:43, Wednesday 01 October 2025 (87245)VE

(Jordan, TJ, Gerardo)

Late entry.

TJ powered ON the ion pumps over the weekend, that allowed for the pumps to reach very good vacuum pressure on the shared annuli system, then on Tuesday morning, Jordan isolated the annuli system for HAM5 and HAM6 from the mechanical pumps and turned off the aux carts.

A couple of hours later we removed the small can turbos, flex hoses and aux carts from the HAM5/6 area, to conclude the replacement of the HAM5 annulus ion pump body.

 

 

Images attached to this comment
Displaying reports 4541-4560 of 89199.Go to page Start 224 225 226 227 228 229 230 231 232 End