R. Crouch, J. Oberling, I. Abouelfettouh, O. Patane
As Ibrahim reported here, we have completed the first round of BBS position alignment (I say first round as we still have to do the pitch/yaw alignment, and that has the potential to change the position alignment so we may be doing this again). In the basis of our alignment equipment, which is set normal to the AR face of the BBS, the deviations from nominal are:
If we rotate these deviations to the XYZ axes using the BBS yaw we get deviations along those axes. I'm using the target BBS yaw for this (specifically, the AR surface yaw from the +X axis of 45.1056°), as we have yet to measure or align the actual BBS yaw, so this is more of an estimate at this point but will work for now (it takes a yaw change on the order of several degrees to change this calculation at the 0.1 mm level, so this is a pretty good estimate); this will be tightened up once we align the BBS pointing and revisit the positioning. The results (the tolerances rotate with the deviations, hence the change in X and Y):
The below table gives the target position of the center of the BBS's AR surface and the current position based on the above estimate of the XYZ deviations (all units are in mm):
| Axis | Target Position | Actual Position | Deviation | Tolerance |
| X | -160.4 | -160.5 | -0.1 | +/- 1.4 |
| Y | -226.3 | -225.8 | +0.5 | +/- 1.4 |
| Z | -83.1 | -83.2 | -0.1 | +/- 1.0 |
The next step in the alignment is to use the FARO to set up a total station/laser autocollimator combo looking along the target surface normal of the HR surface of the BBS. This will be used to align the BBS pitch and yaw. Once that is done we'll have to re-check the BBS position alignment (again, using the AR surface of the BBS) to ensure the pointing alignment did not change the optic's position (which may happen in this case as the BBS is currently, as Ibrahim reports, "quite yawed").
Ibrahim, Oli, Jason, Ryan C
Work done today:
Specifics of Optic XYZ Alignment
See pictures below.
Conclusion: T2000599-v6 (r13010) of bbssopt.m matches LHO parameters, T2000599-v5 (r12764) matches LLO's current configuration, but the difference in the models is negligible
The M3 mass listed in the bbssopt.m parameter set has been changed from 20.99kg to 20.909kg to match the mass of LHO's BBSS M3 optic.
This mass is given by the mass of the optic itself (20889g) + the mass of the primary prisms (7.56g and 7.46g) + the mass of the secondary prisms (5.57g together). This gives us 20.909kg. The change in the model from this is very minimal - the highest Pitch peak shifts down by a few hundreds of uHz, from 1.1656Hz to 1.1615Hz.
LLO's BBS02 is 21.065kg, and comparing this to the previous dummy's mass, the peaks are still in the same place, so the previous version of the model, v5 (r12764) matches with their current configuration, but since the shift is so minimal, using the same model that LHO is using would be fine.
I've updated T2000599 to -v6, and committed /ligo/svncommon/SusSVN/sus/trunk/Common/MatlabTools/TripleModel_Production/bbssopt.m to r13010.
The comparison for all DOFs for all three mass changes can be found at /ligo/svncommon/SusSVN/sus/trunk/BBSS/Common/Results/comparetripleparams/2026-05-20_M3_DummyvsLHOvsLLO/triplemodelcomp_2026-05-20_M3_DummyvsLHOvsLLO_M1toM1.pdf and committed to svn as r13011.
TITLE: 05/20 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
SEI_ENV state: MAINTENANCE
Wind: 2mph Gusts, 0mph 3min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.16 μm/s
QUICK SUMMARY:
We are currently in LASER SAFE for some BBSS and tomorrow's VAC work, but might go back to HAZARD in the afternoon for more HAM2 work.
Here's a look at some old data from september's OM2 heating test. I'm posting this incomplete alog, which has been sitting in my drafts for a long time, so that at least this information is available to people.
Background about the measurements:
Information about this dataset, bold numbers are used as assumptions in these models:
Here is a plot with all of this data, we took mean sqz data for all three data sets with the LO unlocked, we only took anti-squeezing data for the first OM2 hot data set without adjusting the SRCL offset. The low frequency non quantum noise is worse for the OM2 hot data because we didn't retune the feedforwards after heating up.
Some gwinc plots to help me understand the impact of mismatch phase:
In gwinc the mismatch is parameterized as IFO to OMC, and SQZ to OMC, since we have not adjusted the psams while we heated up OM2 we will need to increase both of these mismatches for the hot OM2 state, and both of them have a mistmatch phase which changes the frequency dependent rotation significantly. This rotation is most obvious for the mid- squeezing traces, where we adjusted the squeezing angle so that the high frequency noise matches what it is when there is no squeezing. To help myself understand, I made some plots that show a series of models with our measured mismatch magnitudes listed above and the squeezing angle fit for high frequency, varying ifo to omc mismatch phases, or sqz to OMC mismatch phases. Here is another plot showing only one type of mid sqz with both phases varying, which may also be helpful to look at to see the envelope of what varying the mismatch phase can do. Kevin has said that the only thing that matters is the relative mismatch phase, this plot illustrates that where most of the traces can't be seen because they are hidden by others with the same relative mismatch phase. The traces where two mismatch phases are identical are closest to the trace with no mismatch.
Looking at losses:
Looking at the plot with all the data, you can see that the mean squeezing is mostly the same for the three cases, above 200 Hz, suggesting that the losses are about the same for all three cases. Below 200 Hz, there is as usual some kind of extra noise that makes the mean sqz trace not useful for estimating loss. We only measured anti-squeezing for the hot OM2 case before we added the SRCL offset. This plot of a model of anti-squeezing shows that we don't expect much difference in anti-squeezing for the level of mode mismatch and SRCL offset that we have, except around 20Hz where technical noise would cover the quantum noise.
There are two ways to use these data to estimate unknown losses. I've used 2-2.3kHz since the technical noise is further from the shot noise than in other parts of the spectrum. Assuming an arm power of 330kW we can use no sqz data to estimate readout losses, doing that we need to add just a small amount of extra losses to explain our shot noise: 1.5% for cold OM2, 2.4% for hot OM2 no change in SRCL offset, 0.9% for hot OM2 with SRCL offset changed. This is encouraging since it means that we can account for most of our readout losses, however it doesn't agree with the estimate based on kappa C for the change in losses as OM2 was heated. We can also use the nonlinear gain and unlocked squeezing level to estimate unknown loss: cold OM2: 15.1%, hot OM2: 15.7%, hot OM2 SRCL offset: 19.8%.
J. Kissel Picking up from yesterday's work (2026-05-18, LHO:90270), today I: - Verified Bram's idea on how to resolve yesterday's UNSOLVED SIDEQUEST (1) regarding ~120 Hz noise seen on FBR_PWRIN_REF PD (LHO:90279) -- it's just the room lights. Check out this pic of o-scope timeseries of that PD readout with no lights, room lights, then clean room lights on. *sigh* Alright. RESOLVED: no big deal; when I want to use that PD, turn off the lights. - Resumed migration of optics: . Finish up REF beam path by migrating and installing M_B3 to steer REF beam into IFO_MEAS_A. Confirmed that Reflective surface is within the mount, and protruding AR surface faces M_M2 and IFO_MEAS_B. . Nothing to align here, it's a fixed optic, but checked alignment of REF beam into both IFO_MEAS_A and IFO_MEAS_B with irises in D2400143 breadboard holes 85 and 82. . Migrated D_IFO_MEAS_A and D_IFO_MEAS_B and aligned PD reflection into them with slop of PD housing moutning holes :: For D_IFO_MEAS_A Changed out smaller 1.0"x075" D1800140 Type 07 plate for bigger 1.16" x 1.16" D2000228-v1 DLC coated plate that we'd found to rectify the (lesser) stray beam from FBR_PWRIN_REF during D2400107-v4 build. As discussed, in the D2400107-v5 build, we're gunna need a whole different solution (LHO:90277). . Migrated and aligned M_B1 using holes 103 and 94. . Migrated M_M1, but needed to swap left-handed mount with right-handed mount due to interference of pico motor body with alignment pegs. More on this in the comments. . Aligned M_M1 using holes 92 and 87 (M_M4 and M_M5 are NOT installed) . Secured alignment adjustment screw set screws for M_F1 and M_B1 since these shall never be touched again. Tweaked M_M1 alignment to follow minor beam change when securing those screws. . Migrated and aligned M_M4 using holes 81 and 79 . Migrated and aligned M_M2, and . et voila! A little tweak of M_M2 and I had a heterodyne beat note to use for finishing alignment. Attached are End-of-day board status, and the MEAS IFO heterodyne victory lap. Crude estimate of the efficiency (without accounting for dark offsets): Max = 6.60 V Min = 1.08 V Amp = 5.52 V Mean = 3.85 V efficiency = Amp / (2*Mean) MEAS A 5.52 / (2*3.85) = 0.717 = ~72% MEAS B 5.52 / (2*3.82) = 0.723 = ~72% Excellent. Tomorrow we install the last two optics and align the REF IFO, and we're back in business! Estimated readiness for install May 26 2026!
As per WP 13257 we configured the network to create a dedicated subnet for IOCs and other services. This is in support of our work to improve our infrastructure and management of IOCs and services in CDS. This included network switch reconfiguration and extending the CDS environment. The edc will need a restart to be able to see the new subnet. This will be held off until we actually put something in the network that it needs to record.
Starting with yesterday's alignment, we tried to move the beam spots on IFI input baffle (using IM1), IFI output baffle (using IM2) and the baffle in front of IM4 (using IM3).
We had to make a huge change in all of the optics used: IM1 yaw: -500 urad, IM2 yaw: -900 urad, IM3 yaw: -720 urad.
Forgot to measure PIT/YAW position of IM4_TRANS (and we cannot trend it as we need fast data of segments) but the beam was on IM4_TRANS with almost full IM4_TRANS_SUM.
IMs didn't rail but this is clearly NOT the right way to move. Instead, what we need seems to be to shift the beam position on MC3 by a few mm in -X-Y direction.
We didn't bother to move PRM nor IM4 so the IFO REFL was again misaligned badly.
(One thing to note is that with this drastic change in the input alignment, the beam was still hitting the IM4 TRANS but not ISS QPD. However, just by turning IM3 we were able to regain the beam on ISS QPD. Now I'm fully confident that, whatever alignment we end up having after pumping down, as far as the beam is on IM4 TRANS, we can find the beam on ISS QPD by scanning IM3 (and IM2 if necessary) and slowly walk the beam using the picos to have a good alignment in the ISS path.)
We also measured the beam position (both height and Y position relative to the ideal beam path measured at specific X coordinate) to better understand the IMC beam. Numbers are to follow, but the gist is that the beam position is good (i.e. likely centered) on MC3 but is off in -Y direction by 3mm or so on MC2 even though MC TRANS is close to center.
So, for tomorrow:
My guess is that the beam has been (at least very close to) clipped by some of the IO baffles for a long time, and the above practice is to see if there is an alignment where the beam is at least not clipped.
Just want to add here that I undid all slider changes we made on Tuesday, specifically these: "IM1 yaw: -500 urad, IM2 yaw: -900 urad, IM3 yaw: -720 urad"
While these changes did not saturate the suspensions, IM2 and IM3 were very close to saturation, and we would not want to operate with them at those positions.
H1 ISI CPS Noise Spectra Check - Weekly Famis 39342
For some context, most of the Corner station is vented and the BS cartridge is not in the chamber which explains what is shown on the BS plots here..
TITLE: 05/19 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: More alignment work in HAM2 and BBSS alignment today. Now that the LVEA is back to Laser SAFE, more alignment steps for the BBSS and prep for the GV7 repair can be done starting tomorrow. The SPI NPRO continues to run overnight in the optics lab with proper curtains and signage in place.
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 19:43 | SAF | LASER HAZARD | LVEA | YES | LVEA IS LASER HAZARD | 23:11 |
| 14:43 | FAC | Kim | LVEA | - | Technical cleaning | 16:15 |
| 15:37 | SUS | Betsy | LVEA | - | Looking for parts | 15:51 |
| 15:49 | VAC | Jordan | LVEA | - | Pump checks | 15:56 |
| 15:49 | SUS | Ibrahim | LVEA | - | BBSS staging | 19:32 |
| 16:20 | SUS | Betsy | LVEA | - | BBSS staging | 17:47 |
| 16:30 | PEM | Ryan C | LVEA | - | Setting up DM at test stand | 16:46 |
| 16:34 | FAC | Kim | LVEA | - | Technical cleaning | 17:21 |
| 17:41 | VAC | Jordan +1 | EX, MX, EY, MY | - | N2 tank inspections | 20:22 |
| 17:52 | SUS | Betsy | LVEA | - | BBSS staging | 19:33 |
| 18:31 | SUS | Oli | LVEA/Opt Lab | - | Grabbing parts and delivering | 19:29 |
| 19:29 | SPI | Josh | LVEA | - | Checking cable | 19:37 |
| 20:26 | VAC | Travis | LVEA | - | HAM3 feedthrus | 21:23 |
| 20:28 | SUS | Betsy | LVEA | - | BBSS work | 22:48 |
| 20:29 | IAS | Jason, Ryan C | LVEA | - | BBSS surveying setup | 22:48 |
| 20:33 | SUS | Corey | Opt/Prep Labs | - | Looking for parts | 20:58 |
| 20:35 | SEI | Jim | LVEA | - | HAM3 feedthrus | 21:24 |
| 20:46 | SAF | Richard | LVEA | - | Safety checks | 21:15 |
| 20:48 | ISC | Keita | LVEA | Y | HAM2 beam alignment | 22:53 |
| 20:52 | ISC | Elenna | LVEA | Y | HAM2 beam alignment | 22:53 |
| 20:54 | SEI | Shoshana | LVEA | - | HAM3 feedthrus | 21:24 |
| 21:06 | SPI | Jeff | Opt Lab | Local | SPI alignment (working late) | Ongoing |
| 21:18 | EE | Fil | MER | - | HEPI electronics | 23:18 |
| 21:33 | CAL | Tony | PCal Lab | Local | Set up measurement | 22:32 |
| 21:33 | SUS | Ibrahim | LVEA | - | BBSS work | 22:35 |
| 22:21 | SEI | Jim | LVEA | - | Getting boxes from H2 PSL | 22:27 |
| 22:58 | SAF | Oli | LVEA | - | Transitioning to Laser SAFE | 23:10 |
A Norco tech came to the site to inspect the 8 LN2 dewars that feed the cryopumps. Inspection report will be posted to Q2000008 once received.
- No major issues found, some small leaks were repaired as they were found. The tank safeties and burst disks are due for replacement soon, since they should be replaced every 5 years.
The vacuum jacket pressures were also measured during inspection:
The vacuum jackets were all pumped down to <10 mtorr last May (excluding CP4 which is not in use), see alog 84290
| Dewar | Pressure (micron/mtorr) |
| CP1 | 22 |
| CP2 | 20 |
| CP3 | 110 |
| CP4 (not in service) | 350 |
| CP5 | 11 |
| CP6 | 10 |
| CP7 | 26 |
| CP8 | 24 |
While the corner station old HEPI pump controller (aka the Ben Box) is down and before the new Beckhoff replacement is rolled out, I've created a dummy ioc which serves the old channels. This has "greened up" the EDC, it no longer has 31 disconnected channels. To appease the alarms system, the PRESSn channels have non-zero values which lie in the good range. CDS now has zero active alarms.
I took the opportunity to update my docs on testing/installing containerized IOCs.
TITLE: 05/19 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
SEI_ENV state: MAINTENANCE
Wind: 4mph Gusts, 1mph 3min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.19 μm/s
QUICK SUMMARY: No alarms overnight, temperatures and dust counts looking normal. More alignment work in HAM2 planned for today along with continued prep on the BBSS in the LVEA west bay. CR workstations and wall FOMs are finishing their restarts.
J. Freed
Update from last time 90231:
1. The funny set up from 90231 (IMG_0627.jpg) Was removed! *Clap*. The Spitter was moved lower on the rack and attached on the side of slot 11 with zip ties Photoon5-18-26.jpeg Photoon5-18-26at317 PM.jpeg. Note: The metal of the splitter was not separated from the metal of the rack, unsure if nessisary. Attinuators were added on each side of the splitter.
2. The 4096 Hz CDS DAC signal was tested and connected to the Double Mixer, the double mixer output was quickly checked on an osciloscope and nothing looked different from testing.
3. The current expected RF power going into the SPI prep chassis is 31.6dBm for both the meas and ref paths (Using 2x 10dB and 2x 2dB attinuators) which is slightly underpower of the expected 32 dBm. Screenshot2026-05-18at52906 PM.png I figure a slight underpower is better than an overpower. But will add a comment here if that changes. Will also check to be sure once the 2W amp is installed
4. Found a box of SPI cables that we bought, (whoops forgot about those). All cables in the current set up were replaced with the bought ones. All cables were also labeled.
SPI_HAM3_Linear_ish_Layout.png: I realized now the Screenshot2026-05-18at52906 PM.png picture is impossible to read clearly. I made a more "linear" and hopefully clearer picture of RF chain.
[Keita, Rahul, Elenna]
Today I moved IM2 and IM3 to bring the beam back to our reference position on IM4 trans QPD and center it on ISS QPD after Keita's move of the mode cleaner mirrors in 90259. This requires some iteration back and forth on both suspensions.
Our desired positon on IM4 trans is P = 0.22 and Y = -0.06. On ISS QPD, it is centered, so P=0 and Y=0.
We are driving MC2 in length, so the mode cleaner is flashing, and there are bright flashes on the QPDs. I am pausing ndscope on a flash and measuring the height of the peak in the fast channel and calculating the pit and yaw position from each QDP segment.
| Start | End | ||
| IM2 P slider | 765 | IM2 P slider | 810 |
| IM2 Y slider | -187.7 | IM2 Y slider | -88.7 |
| IM3 P slider | -560.7 | IM3 P slider | -614.7 |
| IM3 Y slider | 320 | IM3 Y slider | 385 |
| IM4 trans PIT | 0.390 | IM4 trans P | 0.268 |
| IM4 trans YAW | 0.450 | IM4 trans Y | 0.010 |
| ISS QPD PIT | -0.379 | ISS QPD PIT | -0.059 |
| ISS QPD YAW | -0.455 | ISS QPD YAW | 0.08 |
Unfortunately, this still results in clipping on the baffles Keita notes above, so we will keep going.
At the nominal IM4 trans position, yaw is pretty well centered. I then moved IM2 to both edges of IM4 trans QPD.
I changed the IM2 yaw slider to -250.7, which brought the yaw position on IM4 trans to 0.85. This made the clipping worse.
I changed the IM2 yaw slider to 90.3, which brought the yaw position on IM4 trans to -0.88. This was still not good enough to fix the clipping on the baffle.
By making a very large move to IM2 yaw slider value of 710.3, this centered the beam in the IM4 baffle. This is a ~800 urad move according to the osems and slider. The IFO REFL beam is still clipped.
I undid the 800 urad move, so the IM2 yaw slider is back to -88.7 for now.
Keita and Rahul went out to measure the position of the beam on MCs 1,2 and 3. We think that we need to make a move of these three mirrors to see if we can unclip the beam on these baffles that way.
We want to note that the positive yaw move of IM2 corresponds to unclipping on the IM4 baffle, this is consistent with the beam motion observed in chamber in the -X direction. However, this is contradictory to the sign on IM4 trans QPD, which was moving to negative yaw when we did this move. We suspect that the segment defintion must be wrong somewhere.
Keita will say more later once we have a chance to analyze the positions in chamber.
To clarify, the slider values of IM2 and IM3 are left at the "end" positions on the table above.
I trended the power measured at IM4 trans compared to the IMC input power for the entirety of the run. Some notes:
Because of this recalibration, I decided to compare both the IM4_TRANS_INMON and IM4_TRANS_OUT16 to IMC-PWR_IN_OUT16
FM10 in the IM4 trans filter bank is a factor that Craig and Georgia determined in the alog linked above, 4.606. I trended the inmon channel and multipled it by this number, ignoring other calibration factors present in the filter bank.
Therefore, the plotted ratio of IM4_trans [IN, OUT] / IMC power IN will help us understand how the power at IM4 trans changed throughout the run, like perhaps if the amount of clipping on the way to IM4 has changed.
I am comparing the ratios of both IM4 trans IN and OUT just in case we get confused by the changing calibration of the diode. I took an hourly median of these channels so we are not confused by random variation and masked the times to only show when IMC lock was either in state 100 (locked) or 70 (ISS ON).
Overall, the amount of power arriving on IM4 trans has definitely changed throughout the run.
Notably, during the vent between O4a and O4b, the amount of power measured at IM4 trans dropped. We chose not to move the IMs, and instead Sheila picoed to recenter on IM4 trans, linked above. Another power drop occured again during O4b. This time, we moved IMs to fix it. At the start of O4c is when Sheila and I recalibrated the diode, hence the disagreement with IN and OUT channels.
The second plot attached shows how the pitch and yaw on IM4 trans has varied alongside the power.
Based on what Keita can see happening in the chamber with baffle clipping, it is possible that during these alignment shifts, the amount of clipping in HAM2 on the IO baffles was changing, so the amount of power making it to IM4 trans and the amount of power going into the IFO was changing.
Specifically, I want to emphasize that with Craig's integrating sphere measurements in HAM1, we assumed that all loss between HAM1 and the PRM in HAM2 was known, i.e. loss from the IFI, etc. However, if there was additional loss on that path that changed with input alignment shifts, that would explain the apparent IM4 trans power changes during O4. Notably, Sheila and I recalibrated IM4 trans in O4c because after we fixed the input alignment in late O4b, we got more power on IM4 trans than we had gotten all run (see the jump around day 600 on the attached plots). I thought this was not physically possible, so we adjusted the calibration. Perhaps instead, we changed the amount of clipping in HAM2, giving us more light on the PRM than we had seen all of O4.
This may also help explain some of our arm power measurement mysteries if we actually had less input power than assumed.
I just want to add a clarification that we have been trying to replicate the alignment onto IM4 trans QPD, so we are trying to align to the previous pitch and yaw position. However, for O4, the beam was nearly falling off the ISS QPD, so we don't want to replicate that alignment (the pitch and yaw values on ISS QPD were like +- 0.9). We have decided to go with centering the beam on ISS QPD, especially since we have adjusted the ISS pico mirrors to find that alignment.
J. Kissel Picking up from last Friday (2026-05-15, LHO:90253), today I: - Secured D_R_P1 and D_M_P1 and confirmed that TFP REFL beams went into the dumps - Started aligning REF path and migrating optics over from the D2400107-v4 breadbord's CVM100 mounts into / onto D2400107-v5 IXM100 mounts: . R_M1 migrated. . Re-set up the power meter gantry assembly (since its only me, and I need the power meter stable for maximizing throughput thru irises) . Placed irises in holes 83 and 84, adjusted R_F1 and R_M1 to walk and align the beam thru irises, confirming with IR camera view of iris holes and power meter thru-put. . Secured R_F1 and R_M1 IXM mount 8-100 alignment screw shaft set screws, since these optics shall not move again. . R_B1 migrated. . Aligned beam through holes 86 and 95, repeating process above, but only adjusting R_B1. . Uncapped IFO_REF A and B PDs, PWRIN_REF PD and hooked them up to the o-scope (discovering a minor re-cabling issue; see LHO:90268) UNSOLVED SIDEQUEST (1) :: PWRIN_REF PD has a ~120 Hz oscillation on it. . R_B2 migrated. . Aligned beam through holes 98 and 99, repeating process above. UNSOLVED SIDEQUEST (2) :: Now, with D_FBR_PWRIN_REF so far away, the known issue with its PD seating within the generic enclosure means the PD reflected beam is ~2 inches above the board. Tried a huge 2"x3" dump but that was too big. . R_B3 migrated. . With nothing to align, just checked that beam went through irises in holes 100 and 97, and that beam looked well centered on PD. They did. . Migrated D_IFO_REF_B and D_IFO_REF_A, confirmed PD reflection landed on there and it did. Done for the day! The MEAS PATH alignment tomorrow!
Regarding UNSOLVED SIDEQUEST (2) :: Now, with D_FBR_PWRIN_REF further away [...] the PD reflected beam is ~2 inches above the board. Here're some picture to aide the discussion. I say "this is a known issue," because we'd encountered this particular PD's pitch of the diode when assembling the CVM100 version of the ISIK transceiver, D2400107-v4, but in that layout, the PD was angled away from the incoming beam in the opposite direction and the the dump was closer. We didn't have room behind R_B2 to have the same dump location with the upgrade to IXM100 mounts. None of the other PDs have this issue, the beam reflects back at the same 1 [inch] beam height that's the beam height for the board. We were in too much of a rush and had too little experience/expertise with these to consider cracking open the Generic PD enclosure (D1600083) to try to fix it. We suspect, in this Type 3 assembly for the FFD200 PDs, that either - the PD (line item 12 T1000573) is not seated well in its retaining ring (line item 4 D1600082), - there's a defect or something stuck between the retaining and its seat in the enclosure housing (line item 1 D1600079) - the retaining ring is not *exactly* the same size as the Q3000 for which it was designed and there's some slop in the assembly, or - the PD's pins were soldered into the PCB at an angle (line item 11 D1700116) But we're just guessing. This is S2401094, whose assembly technique is discussed in CIT:837 and record of completion is mentioned in CIT:898. Starting up the conversation with CIT.
Regarding UNSOLVED SIDEQUEST (1) :: PWRIN_REF PD has a ~120 Hz oscillation on it. See attached o-scope picture -- channel 3 (purple). 2V peak-to-peak wobbly 120 Hz oscillation on a 4.5 [V] mean. We've not tested this PD since we updated its transimpedance circuitry (see LHO:90105, and TIA Variant 3 D1002481 instantiation S2500713), so there may be a grounding issue there. Hard to believe it's the TIA op-amp itself, given that the RC feedback impedance we installed yields a 13.497 [kHz] pole. I've only tested the IFO MEAS and IFO REF channels of the Variant 2 chassis with light so far, but they don't show any issue. Laser power level on this PD during yesterday's test is the expected ~0.8 [mW]. I checked (but didn't re- or triple-check) that the electrical ground situation is well-managed. As in March 2026, the electrical ground of power supply feeding the TIAs banana'd-to-BNC'd-to-clip-doodled to the shield the o-scope's clip-doodle dongle whose signal is reading out PD ADC output voltage from a breakout board. Nowhere along this adpater/doodle/breakout/cable chain is metal touching metal. I'll look through the chain again, try reading it out differentially, and consult with local experts. It isn't a high priority right now, but we need to understand this before the transceiver leaves the optics lab.
J. Oberling, R. Crouch, J. Warner, B. Weaver, I. Abouelfettouh
This week we surveyed the position of the components that reside in WBSC2: The BS SUS cage (BSS), the ISI optics table (ISI Stage 2), and the 2 ITM Elliptical Baffles.
BS and the SUS Cage
The first picture shows our FARO survey of points on the BS SUS cage, chiefly along the bottom of the main support structure. These were surveyed by holding the FARO SMR against the hole being measured; the PolyWorks software handles the compensation from the center of the SMR to the point being measured. As can be seen, each point is very close in both X and Y axis position, being less than 0.1mm from its nominal location. The Z axis deviations are larger, but the largest of them is just over 0.25 mm, so every point is well within the positioning specifications used during installation and alignment in 2013.
Line 1 in the picture was created from the first and last survey points and represents the pointing of the BS SUS cage; all angles are reported in degrees. Some things to note here: I'm using the Acute Angle datum in PolyWorks, which is the angle measured from the closest axis. For the HR surface normal of the BS, the X Acute Angle is measured from the -X axis, the Y Acute Angle is measured from the +Y axis, and the Z Acute Angle is measured from the +Z axis. Since Line 1 is roughly perpendicular to the surface normal of the BS HR face, the axes the angles measure from are changed: The X Acute angle is now measured from the +X axis, the Y Acute Angle is still from the +Y axis, and the Z Acute Angle is now from the -Z axis. In addition, since Line 1 is nominally perpendicular to the BS HR surface normal I would expect the X and Y Acute angles to be swapped (BS X Acute = Line 1 Y Acute; BS Y Acute = Line 1 X Acute), but they aren't exactly. This appears to be a small error in the CAD model, if we make the assumption that the BS HR surface and the HR side of the BS SUS cage are nominally pointing in the same direction. This does, however, change the deviations for the X and Y Acute angles for Line 1. The table below shows what the data for Line 1 should be:
| Nominal | Measured | Deviation | |
| X Acute Angle | 44.9699 | 44.9073 | -0.0626 |
| Y Acute Angle | 45.0301 | 45.0927 | 0.0626 |
| Z Acute Angle | 90.0000 | 89.9661 | -0.0339 |
This means the BS SUS cage is yawed 0.0626°, or ~1.09 mrad, in the clockwise (CW) direction when looking from the top down (since Line 1 is closer to the +X axis than it should be). The Z Acute Angle represents a slight counterclockwise (CCW) roll of the SUS cage, when looking directly at the HR surface of the BS.
To attempt to better locate the BS in the IFO coordinate system, several measurements were taken with a ruler from points on the "Figure 8" section of the BS SUS cage to the BS optic itself. All measurments except one were done using a scale with 0.5 mm tic marks (so accurate to +/- 0.25 mm). The 10:00 "Figure 8 face to BS HR face" measurement had to be done using the side of the scale in inches, with 1/32" tic marks (so accurate to +/- 1/64") and then converted to mm (so accurate to +/- 0.4 mm). The measurements positions are listed like the BS HR surface is a clock, and assumes you are looking directly at the HR surface. The below table gives those results:
| Outside Edge of Figure 8 to BS optic edge (mm) | Front face of Figure 8 to BS HR face (mm) | |||||
| 3:00 (-X/-Y) | 6:00 (-Z) | 9:00 (+X/+Y) | 2:00 | 6:00 | 10:00 | |
| Measurement | 49.0 | 48.75 | 48.5 | 24.75 | 27.0 | 26.2 |
| CAD Nominal | 49.0 | 49.0 | 49.0 | 26.2 | 26.2 | 26.2 |
| Deviation | 0.0 | -0.25 | -0.5 | -1.45 | +0.8 | 0.0 |
The BS sits decently centered in the Figure 8 portion of the SUS cage, a little bit low and to the +X/+Y side. I would say not as much horizontally as it looks from the table, given the inherent error with reading the scale (the BS is not wider than its 370.0 mm specification, it's actually 0.15 mm narrower at 369.85 mm). The pointing implied by this measurement, however, is more than a little alarming. The 2:00 and 10:00 measurements show a significant yaw of the BS optic w.r.t. the SUS cage, and in the same direction as the yaw of the SUS cage as measured by the FARO. There is ~320.0 mm between the 2:00 and 10:00 positions on the BS, so that 1.45 mm difference in depth is a 4.53 mrad CW yaw. When added to the CW yaw of the SUS cage, this measurement shows that the BS optic is yawed 5.62 mrad CW from its nominal yaw. Even assuming the errors fall in our favor (so the 2:00 at 25.0 mm and the 10:00 at 25.8 mm), that's still a 3.59 mrad CW yaw (2.5 mrad BS and 1.09 mrad SUS cage). In addition, the 6:00 measurement implies a significant downward pitch of potentially several mrad, although with no way to measure the top of the optic we can't actually measure it. I have to be honest, I'm having a very hard time believing this measurement; we will revisit this once the BS cartridge has been moved to the test stand, where we have a better field of view for the FARO, more room to work and much better lighting around the BS, and can take direct measurements of the BS position and pointing using a total station and laser autocollimator (although there is no guarantee that the optic will be pointing in exactly the same direction after being craned across the LVEA). More to come on this.
ISI Optics Table
The second attachment shows the ISI positions as measured by the FARO. I've corrected the Z axis positions for the length of the rod we use to hang the SMR from the ISI so they give a better idea of the Z axis position. Not much can be said here, as LLO discovered that while these rods are good for measuring the Z axis position, they are not at all good at measuring X and Y. This makes sense as they were designed to be accurate in length and only length, so there's no guarantee that X and Y are repeatable. We plan on measuring the X and Y errors of this particular set of rods in the coming days (align to a table with a known hole pattern, attach the rod and measure with the FARO, repeat multiple times to see how the X and Y positions change). For now, we can say that the ISI is lower on the -X side vs the +X side, and lower on the +Y side vs the -Y side. I'm not alarmed by the deviations in Z axis position, as this ISI was supposed to be lower by ~2.5 mm (to place the BS in proper Z axis position, since it's lower in the IFO coordinate system but the SUS is the same length as the QUADs), but this was never captured in the CAD files.
ITM Elliptical Baffles
The final four attachments show our survey of both ITM elliptical baffles. Our view of the baffles and available fiducials to take measurements from were both limited, but we can say a few things.
ITMx Elliptical Baffle
We were able to get two points along the +Y bottom edge of the baffle, a single point along the +Y top edge, and single point near the center of the -X bottom edge of the baffle. From this I made a couple of planes that represent the +Y and bottom sides of the baffle and are shown in the third and fourth attachments; I, J, and K are the direction cosines of the surface normal of the plane, while the listed angles are the angle from the surface normal to the +X, +Y, and +Z axes. Interestingly, the point on the top edge looks very well aligned, within 1.0 mm all around, while the points along the bottom of the baffle are all low by several mm. In addition, there appears to be a significant upward pitch to the baffle. Jim did note that when attaching the transport bracket he had to push the baffle in the +X direction to clear ~0.5 mm at the point where the bracket attaches to the suspended portion of the baffle. This point is roughly 476 mm away from the baffle's suspension blade, so this is an ~1.05 mrad angle. Applying this same angle along the bottom of the baffle box gives an ~ -0.33 mm Z axis move of that bottom -X edge of the baffle, so this does not account for the measured deviation. In addition to the pitch, the bottom plane also shows a large roll (CCW when looking at the ITMx in WBSC3), while the side plane shows a large yaw (CCW when looking from the top down). We know these baffle panels aren't exactly straight, so it's hard to say if this significant pointing is also present on the elliptical hole of the baffle (we couldn't see it, so we couldn't measure it directly).
ITMy Elliptical Baffle
Similar to the ITMx baffle, we were only able to get a handful of points along the -X side and the bottom of the baffle. I made planes from these points representing the -X side and the bottom of the baffle (fifth and sixth attachments). As seen with the ITMx baffle, the points along the top of the baffle all look good while the points on the bottom are too low by several mm. There is a significant upward pitch to this baffle as well, as well as a large roll (CCW when looking at ITMy in WBSC1) and yaw (CCW when looking from the top down), although none are as large those as seen on the ITMx elliptical baffle. Again, we could not see the elliptical hole in the baffle to measure it, so we can't say if this pointing is an artifact of the panels or also present on the actual baffle portion of the baffle.
This completes our in-chamber measurements of the WBSC2 cartridge assembly, and closes LHO WP 13171.
I also want to note, Ryan and I also preformed some in-chamber FARO measurements in WHAM3 (ISI, MC2 SUS cage, PR2 SUS cage, MC2 and PR2 baffles) on April 10th; I will post those as soon as I get a chance to process the data in PolyWorks.
Some action photos while Jason & RyanC were chamberside at HAM3.
Ignore the very first table in the above alog, concerning the yaw of the BS SUS cage, as I apparently forgot how to do math when creating that. In writing a master alog trying to tie together the series of measurements we took on the BS (still a WIP) I went back to PolyWorks and, in a separate file so to not compromise the existing data, rotated the CAD model so the HR face of the BS SUS matched the target yaw of the BS HR surface. The results of that are shown in the attachment here. The angles listed for Line 3 are measured from the positive axis; i.e. X Ang is measured from the +X axis. Takeaways:
Honestly, ignore everything I wrote above about BS yaw. Not only was my math bad on the rotation correction (see previous comment), the error bar on the scale measurements used to calculate the yaw is so large that the only thing we can definitively say about the BS in-chamber yaw is that it is yawed CW w.r.t. the SUS cage.
As explained above, we did 2 measurements of the BS optic distance from the front face of the HR side Figure 8, one at roughly the 10 o'clock position and one at roughly the 2 o'clock position. The only Class B ruler we had at the time had mm on one side and inches on the other; the mm side had 0.5 mm tic marks for a +/- 0.25 mm error bar, while the inches side had 1/64" tic marks for an error bar of +/- 0.4 mm. 2 measurements were required at each position, one from the HR Figure 8 to the gold wire baffle and then one from wire baffle to the BS itself. Carrying the error bars from the multiple ruler measurements all the way through the yaw calculation yields a yaw of 4.53 +/- 4.06 mrad CW. With the error bar almost as large as the result, the only definitive thing we can say is that the BS optic was yawed w.r.t. the SUS cage, but we can't put an accurate number on it with the measurement method used.