TITLE: 08/30 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Aligning
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 12mph Gusts, 9mph 5min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.10 μm/s
QUICK SUMMARY:
H1 unlocked about a half hour ago after a 16 hour lock stretch. Currently running an initial alignment and will start locking immediately after.
FAMIS 26446, last checked in alog79327
Both BRSs look good. BRS-X was drifting towards the upper limit, but has since started to turn around.
I was curious to see what files the two observatories actually share in their guardian user code. I've attached full lists, but here is a breakdown of the file comparisons.
LHO
Total number of nodes: 166
Total files used by nodes: 258
Files unique to LHO: 163
Files shared with LLO: 95
Files in common but only used by LHO: 30
A few key takeaways:
after_key = 'LHO_1409042566'
b4_key = 'LHO' # 1403879360 # Hot OM2, 2024/07/01 14:29UTC
(aligoNB) anthony.sanchez@cdsws29: python H1/darm_intergal_compare.py
Figures made by this script will be placed in:
/ligo/gitcommon/NoiseBudget/aligoNB/out/H1/darm_intergal_compare
Fetching data from nds.ligo-wa.caltech.edu:31200 with GPS times 1409042566 to 1409043166
Successfully retrieved data from nds.ligo-wa.caltech.edu:31200 with GPS times 1409042566 to 1409043166
data is 60.1 days old
Fetching data from nds.ligo-wa.caltech.edu:31200 with GPS times 1403879360 to 1403879960
Successfully retrieved data from nds.ligo-wa.caltech.edu:31200 with GPS times 1403879360 to 1403879960
running lho budget
Saving file as /ligo/gitcommon/NoiseBudget/aligoNB/out/H1/darm_intergal_compare/compare_darm_spectra_OM2_hot_vs_cold_no_sqz.svg
Saving file as /ligo/gitcommon/NoiseBudget/aligoNB/out/H1/darm_intergal_compare/compare_darm_range_integrand_OM2_hot_vs_cold_no_sqz.svg
Saving file as /ligo/gitcommon/NoiseBudget/aligoNB/out/H1/darm_intergal_compare/compare_cumulative_range_OM2_hot_vs_cold_no_sqz.svg
Saving file as /ligo/gitcommon/NoiseBudget/aligoNB/out/H1/darm_intergal_compare/cumulative_range_big_OM2_hot_vs_cold_no_sqz.svg
Script darm_intergal_compare.py done in 0.23667725324630737 minutes
H1's current coherence with jitter, LSC signals, ASC signals plot.
Bruco - Brute force coherence. Looks at the coherence between many channels. was the next thing I wanted to Try but I was greeted with a permissions denied when trying to ssh into the cluster with my A.E creds.
I ran a bruco on GDS CLEAN with data from the current lock after 11 hrs of lock time. My instructions. I have been running my brucos on the LHO cluster lately because it seems like every time I run on Caltech I get some error.
bruco command:
python -m bruco --ifo=H1 --channel=GDS-CALIB_STRAIN_CLEAN --gpsb=1409075057 --length=600 --outfs=4096 --fres=0.1 --dir=/home/elenna.capote/public_html/brucos/GDS_CLEAN_1409075057 --top=100 --webtop=20 --plot=html --nproc=20 --xlim=7:2000 --excluded=/home/elenna.capote/bruco-excluded/lho_DARM_excluded.txt
Results:
https://ldas-jobs.ligo-wa.caltech.edu/~elenna.capote/brucos/GDS_CLEAN_1409075057/
Greatest hits:
At Livingston, we have been using a separate DIAG_COH guardian to monitor the health of the feed forward. It is similar to Bruco in functionality, in sense that it computes coherence. It does it automatically every 3 hrs within an observing period and computes band limited coherence of channels according to a set config file. An additional feature is that it stores previous coherence values to a file as reference and compares current values with that. If it differs drastically say 2 sigma, it displays a message in DIAG_MAIN warning that certain FF is under performing compared to reference.
at 00:29 Fri 30 Aug 2024 PDT there was a short vacuum glitch in HAM6 detected by PT110_MOD1.
The pressure increased from 1.90e-07 to 2.06e-07 Torr.
Gitch was detected by VACSTAT.
VACSTAT is in testing mode, and the MEDMs still needs some polishing. The Overview and PT110 MEDM are attached.
Gitch time doesn't appear to be related to H1 locking or unlocking. 24 hour PT110 and H1 range trend attached.
The smaller, wider glitch on left was at 16:51 Thu and is known about (pump operations as part of noise hunting). The 00:29 glitch is the larger, sharp one to the right.
I've promoted VACSTAT from a H3 test system to a H1 pre-production system. This allows me to add the channel H1:CDS-VAC_STAT_GLITCH_DETECTED to the alarms system (alarms was restarted 10:15). For testing it will send alarms to my cellphone.
Yesterday in the instrument air trend there was a brief period of pressure loss which was caused by the air dryer being turned off for several minutes while a line voltage switch was added to the unit. The bypass on the dryer was not opened, so additional air was not fed into the system. Once power was restored the dryer functioned correctly and allowed air back in.
Fri Aug 30 08:11:24 2024 INFO: Fill completed in 11min 20secs
TITLE: 08/30 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 146Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 9mph Gusts, 7mph 5min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.08 μm/s
QUICK SUMMARY:
IFO was locked and Observing for 8 Hours before I walked in to the Control room.
Looks like the IFO locked it's self last night with out calling Oli after losing lock at LASER_NOISE_Suppression.
H1's Current Status is LOCK at NOMINAL_LOW_NOISE and OBSERVING for 8 Hours and 27 Min.
H1EDC is reporting a NUC33 error :
Total Channels : 56980
Connected Channels : 56968
Disconnected Channels : 12
H1:CDS-MONITOR_NUC33_CPU_LOAD_PERCENT
H1:CDS-MONITOR_NUC33_CPU_COUNT
H1:CDS-MONITOR_NUC33_MEMORY_AVAIL_PERCENT
H1:CDS-MONITOR_NUC33_MEMORY_AVAIL_MB
H1:CDS-MONITOR_NUC33_PROCESSES
H1:CDS-MONITOR_NUC33_INET_CONNECTIONS
H1:CDS-MONITOR_NUC33_NET_TX_TOTAL_MBIT
H1:CDS-MONITOR_NUC33_NET_RX_TOTAL_MBIT
H1:CDS-MONITOR_NUC33_NET_TX_LO_MBIT
H1:CDS-MONITOR_NUC33_NET_RX_LO_MBIT
H1:CDS-MONITOR_NUC33_NET_RX_ENO1_MBIT
H1:CDS-MONITOR_NUC33_NET_TX_ENO1_MBIT
Did some troubleshooting on NUC33.
couldn't VNC or SSH into it.
Got a Keyboard and Mouse and couldn't get NUC33 to do anything, completely frozen.
I tried the REISUB linux rescue and no response. So I gave it a hard shutdown via the power button.
Once it powered back on everything came back up as normal. H1EDC went back to functioning as normal.
TITLE: 08/29 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Observing at 141Mpc
INCOMING OPERATOR: Oli
SHIFT SUMMARY:
H1 had a late-shift lockloss, and was pretty much automated (need INCREASE FLASHES & PRMI), but near the end there was a lockloss at LASER NOISE SUPPRESSION...most likely due to an EQ from Russia. H1's already giving it a go.
LOG:
Here is an investigation that might give us insight into the PRCL/CHARD/DARM coherences.
Today, we ran a PRCL injection for the feedforward where we injected directly into the PRCL loop. I took this injection time and looked at how the CHARD P and Y error signals changed, as well as their respective coherences to PRCL and DARM (figure). When injecting about 15x above ambient in PRCL from 10-100 Hz, there is a 2x increase in the CHARD P error signal and 4x increase in the CHARD Y error signal. The coherences of CHARD P and Y to PRCL increase as well, and the coherences of CHARD P and Y to DARM also increase. This injection allows us to measure a well-defined CHARD/PRCL transfer function. In the attached screenshot of this measurement, all the reference traces are from the injection time, and all the live traces are during a quiet time.
Meanwhile, I looked back at the last time we injected into CHARD P and Y for the noise budget, on June 20. In both cases, injecting close to 100x above ambient in CHARD P and Y did not change either the CHARD/PRCL coherence or PRCL/DARM coherence. There is some change in the PRCL error signal, but it is small. Again, in the attachments, reference traces are injection time and live traces are quiet time. CHARD P figure and CHARD Y figure.
I think that this is enough to say that the PRCL/DARM coupling is likely mostly through CHARD P and Y. This would also make sense with the failure of the PRCL feedforward today (79806). However, we may want to repeat the CHARD injections since there have been many IFO changes since June 20.
As a reminder, the previous work we did adding a PRCL offset did reduce the PRCL/CHARD coupling: read 76814 and the comments. We are currently not running with any PRCL offset.
I decided to compare the PRCL injection times back in March when I set the PRCL offset to reduce the coherence of DARM with LSC REFL RIN (76814, 76805). One conclusion of these tests was that a PRCL offset can reduce the REFL RIN/DARM coherence, but not necessarily improve the sensitivity. Also, the offset reduced the PRCL/CHARD Y coupling and increased the PRCL/CHARD P coupling.
A bruco shows that there is once again significant coherence with REFL RIN and DARM. I compared the PRCL injection time from yesterday with the PRCL injections with different offsets. The PRCL/ CHARD couplings have increased for both pitch and yaw: plot. I also included the coherences of CHARD to DARM for these times but then realized that data might actually be confusing to compare. However, the PRCL offset has an effect on the PRCL/CHARD coupling, so it could be one strategy to combat this coupling. Unfortunately, it has opposite effects for pitch and yaw.
There was a lot of work done to move the alignment of the PRC around in May; here are some alogs I found to remind myself of what happened: 77736, 77855, 77988. Seems like the goal was to reduce clipping/center on PR2. I wonder if this alignment shift caused the increase in the PRCL/CHARD coupling from March to now.
We should consider checking the PRCL and CHARD coupling while adjusting the PRC alignment. The yaw coupling is stronger, maybe because the beam is much further offcenter of PR2 in yaw than in pitch?
Overall, I think the benefit of this investigation would be a reduction of the noise in CHARD, which would help improve the sensitivity.
We did a 10 minute on/off HAM1 FF test today. I determined from that test that improvement could be made to the feedforward to CHARD P and INP1 P, so I used that off time to train new filters (time here: 79792).
I took a screenshot of the DTT results. The golden traces represent the various loop spectra with the feedforward OFF. Blue represents the previous feedforward, and red represents the new feedforward I fit today. First, look at the top and bottom plots on the left, showing CHARD P and INP1 P. You can see that the old feedforward was having minimal benefit (gold to blue), and the new feedforward is performing much better (gold to red).
Next, look at the middle plot on the left showing CHARD Y. This plot shows that the feedforward is making CHARD Y WORSE (blue worse than gold). Therefore, I just turned it off. I am not yet happy with the fitting results for CHARD Y, so I will continue to work on them.
You can also see in the middle right plot, PRC2 P current feedforward is performing well (gold to blue), so I made no change, red still matches blue.
Note! This means the significant change here must be related to RF45- PRC2 is only sensed on RF9, while INP1 and CHARD use RF45.
Finally, DARM on the right hand side shows improvement from blue to red. It looks like HAM1 FF off is better below 15 Hz, maybe due to CHARD Y.
The new CHARD P and INP1 P filters are all saved in FM9 of their respective banks. I SDFed the new filters, and the CHARD Y gains to zero, second screenshot.
I made one mistake which is that I did not SDF these changes in SAFE, only in OBSERVE, which means that needs to be done before an SDF revert, since they are not guardian controlled!
I ran A2L while we were still thermalizing, might run again later. No change for ETMY but the ITMs had large changes. I've accepted these in SDF, I reverted the tramps that the picture shows I accepted. I didn't notice much of a change in DARM or on the DARM blrms.
ETMX P
Initial: 3.12
Final: 3.13
Diff: 0.01
ETMX Y
Initial: 4.79
Final: 4.87
Diff: 0.08
ETMY P
Initial: 4.48
Final: 4.48
Diff: 0.0
ETMY Y
Initial: 1.13
Final: 1.13
Diff: 0.0
ITMX P
Initial: -1.07
Final: -0.98
Diff: 0.09
ITMX Y
Initial: 2.72
Final: 2.87
Diff: 0.15
ITMY P
Initial: -0.47
Final: -0.37
Diff: 0.1
ITMY Y
Initial: -2.3
Final: -2.48
Diff: -0.18
I am not sure how the A2L is run these days, but there is some DARM coherence with DHARD Y that makes me think we should recheck the Y2L gains. See attached screenshot from today's lock.
As a reminder, the work that Gabriele and I did last April found that the DHARD Y coupling had two distinct frequency regimes: a steep low frequency coupling that depended heavily on the AS A WFS yaw offset, and a much flatter coupling about ~30 Hz that depended much more strongly on the Y2L gain of ITMY (this was I think before we started adjusting all the A2L gains on the test masses). Relevant alogs: 76407 and 76363
Based on this coherence, the Y2L gains at least deserve another look. Is it possible to track a DHARD Y injection during the test?
Since I converted this script to run on all TMs and dofs simultaneously, its performance hasn't been stellar. We've only run it a handful of times, but we definitely need to change something. One difference between the old version and the new is the frequencies the injected lines are at. As of right now, they range from 23-31.5Hz, but perhaps these needs to be moved. In June, Sheila and I ran it, then swapped the highest frequency and the lowest frequency to see if it made a difference (alog78495) and in that one test it didn't seem to matter.
Sheila and I are talking about the AS WFS offset and DHARD injection testing to try to understand this coupling a bit better. Planning in progess.