thomas.shaffer@LIGO.ORG - posted 16:32, Tuesday 27 August 2024 - last comment - 16:56, Tuesday 27 August 2024(79747)
Observing 2330 and SQZ SDF diff

Back to Observing after fighting wind and then stepping through the CARM reduction slowly. There was a SQZ SDF diff that was changed when we lost lock this morning. I accepted it for now.

Images attached to this report
Comments related to this report
ryan.crouch@LIGO.ORG - 16:56, Tuesday 27 August 2024 (79750)

Lots of low frequency noise (high SRCL coherence), SQZing doesn't look great.

Images attached to this comment
LHO General
thomas.shaffer@LIGO.ORG - posted 16:30, Tuesday 27 August 2024 (79733)
Ops Day Shift End

TITLE: 08/27 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing
INCOMING OPERATOR: Ryan C
SHIFT SUMMARY: Maintenance day today. We started relocking around 12 but wind made recovery a bit more difficult. We were not making it past the CARM reduction, so Sheila and Jenne stepped us through it slowly. This worked and just got back to Observing.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
23:58 SAF H1 LHO YES LVEA is laser HAZARD 18:24
15:21 FAC Nellie FCES n Tech clean 16:39
15:21 FAC Karen EY n Tech clean 16:15
15:21 FAC Kim EX n Tech clean 16:25
15:22 GRD TJ CR n h1guardian1 reboot 15:42
15:32 PSL Jason, Ryan S PSL enc Yes PSL incursion, FSS 17:59
15:41 VAC Gerardo, Jordan LVEA Yes HAM6 RGA scans 16:45
15:55 FAC Tyler CS, MY n Forklift equipment around 17:50
16:01 CDS Jonathan MSR n Pulling two unused switches 16:30
16:32 PEM Robert LVEA Yes Install accel. near HAM6 18:40
16:44 PCAL Tony PCAL lab local Parts for LLO 16:50
16:59 CDS Fil LVEA - HAM6 Yes Test network cable 17:15
17:00 SEI Jim FCES n HAM8 testing 17:18
17:09 FAC Kim, Karen LVEA Yes Tech clean 18:22
17:21 PEM Sam, Genevieve FCES n PEM sensors 18:51
18:15 VAC Jordan LVEA Yes Turning off RGA by HAM6 18:22
18:52 PSL Ryan S CR n Rotation stage calibration 19:05
22:32 VAC Gerardo LVEA Yes Photography 22:52
H1 General
ryan.crouch@LIGO.ORG - posted 16:15, Tuesday 27 August 2024 (79744)
OPS Tuesday EVE shift start

TITLE: 08/27 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Wind
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: SEISMON_ALERT
    Wind: 23mph Gusts, 17mph 5min avg
    Primary useism: 0.13 μm/s
    Secondary useism: 0.09 μm/s
QUICK SUMMARY:

X1 SUS (SUS)
ibrahim.abouelfettouh@LIGO.ORG - posted 15:51, Tuesday 27 August 2024 (79743)
BBSS Bounce and Roll Modes Measured

Ibrahim A, Jeff K

On 8/22, Jeff K and I ran some measurements of our assembled BBSS and determined the Bounce and Roll modes to be:

I have linked this alog to the SWG Logbook in the case these results are needed for BRD development for the BBSS.
 
Screenshots of these peaks and the excitation spectra are attached and the Data can be found on the X1 (Test Stand) Network: at /ligo/svncommon/SusSVN/sus/trunk/BBSS/X1/BS/Common/Data/ under names:
Roll: 2024-08-22_1200_X1SUSBS_M1M2_Roll_V_0p001to40Hz.xml
Bounce: 2024-08-22_1200_X1SUSBS_M1M2_Bounce_V_0p01to40Hz.xml
 
Initial Measurement:
These values were found by taking a High BW Transfer Function using a noise injection into M1 Vertical in the 2Hz range in the neighborhood of the model values, reading at all BOSEMs in the M2 stage. (19-20Hz for Bounce and 31-33Hz for Roll). We then found the most coherent peak in that region to be our measured mode.
 
Mechanical Mode Confirmation:
After this, we ran an ASD and injected a sine wave at the found frequencies one-by-one, ringing up our modes for 5 averages before stopping the injection. We then watched the ASD and confirmed that given the number of averages it took to percieve a gradual ring-down at those frequencies (over 20 averages of for each at 0.02BW assuming a Q-Factor of ~1000, which is conservative), that the readout was indeed mechanical, and that these were indeed our Bounce and Roll modes.
 
Note: While the screenshots may say M3, this is simply because of our set-up being hooked up to the M3 stage electronically. These results were read from the M2 BOSEMs on the PUM of the BBSS. This cabling discrepancy has since been fixed.
Images attached to this report
H1 ISC
sheila.dwyer@LIGO.ORG - posted 12:51, Tuesday 27 August 2024 (79740)
added more DOFs back to DRMI ASC

TJ, Sheila

TJ is having difficulty locking in the high winds this afternoon.  We've been running without DRMI ASC other than the BS, so that operators have to manually adjust PRM and SRM sometimes.  This afternoon we added back in INP1, PR1 and PRC2 loops, but left SRC ASC off.  This seems to be helping with POP18, although the operators may still need to adjust SRM by hand.

LHO General
thomas.shaffer@LIGO.ORG - posted 12:27, Tuesday 27 August 2024 (79739)
Ops Day Mid-Shift Report

Maintenance activies completed around 30 min ago. I ran an initial alignment at 1100 PST, then started locking at 1206 PST. The wind is really starting to pick up and we just lost lock at the state Resonance. We will keep trying but this might be a tough relock.

H1 PSL
ryan.short@LIGO.ORG - posted 12:18, Tuesday 27 August 2024 (79738)
PSL Rotation Stage Calibration

Since it hasn't been done since before the vent, and the output of the PMC has continued to slowly drop, before locking today I calibrated the PSL rotation stage following the steps in alog79596.

The measurement file, new calibration fit, and screenshot of accepting SDFs are all attached.

  Power in (W) D B (Minimum power angle) C (Minimum power)
Old Values 97.882 1.990 -24.818 0.000
New Values 94.642 1.990 -24.794

0.000

Images attached to this report
Non-image files attached to this report
H1 AOS (INJ)
keith.riles@LIGO.ORG - posted 12:12, Tuesday 27 August 2024 - last comment - 16:26, Tuesday 27 August 2024(79737)
CW hardware injections seem to be disabled
Now that CIT's ldas-pcdev2 machine is back up following last week's data center troubles, my CW hardware injection monitoring shows no apparent signals since the return to observing mode on Saturday. Is that by design, or did the injections restart fall through a crack? Thanks.
Comments related to this report
david.barker@LIGO.ORG - 14:39, Tuesday 27 August 2024 (79741)

I just checked the CW hardware injection signal is running with no problems. The h1hwinj1 server had crashed on 25jul2024 and I rebooted it on 29jul2024 and verified it started ok. Attachment shows a month trend of the signal, the restart can been seen on the left.

Images attached to this comment
keith.riles@LIGO.ORG - 16:26, Tuesday 27 August 2024 (79746)
My bad. There was indeed another disruption to the monitoring that was affected by the pcdev2 shutdown, which I hadn't noticed. Thanks for the quick follow up and sorry for the noise.
H1 PSL
jason.oberling@LIGO.ORG - posted 11:35, Tuesday 27 August 2024 (79736)
PSL FSS RefCav Tune-up (WP 12057)

J. Oberling, R. Short

The FSS RefCav TPD had been trending down again, and Ryan did not get enough improvement from a remote alignment tweak last week.  So today we went into the enclosure to do an on-table alignment of the FSS RefCav beam path.  As usual, we began with a power budget:

No, we're not getting magic amplification between the FSS In and AOM In power measurements; that kind of small discrepancy is normal with the small stick power meter head we use for this alignment (they are also very AOI dependent, so it's entirely possible we had the head better aligned to the beam for the AOM In measurement).  Immediately we see the single-pass diffraction efficiency is lower than it normally is (generally hangs out in the lower 70% range these days), so we began by adjusting the AOM alignment to improve it.  This also means we have to then adjust M21 to improve the double-pass diffraction efficiency.  After the adjustments we had:

This is the lowest we've seen the diffraction efficiencies of this AOM, and suggests maybe it's time to swap the AOM for a new one (pretty sure this is the same AOM installed with the PSL in 2012, would have to go back through the alog to see if it's been swapped since install).  As usual, we had to do some small tweaks to the FSS EOM (provides the 21.5 MHz PDH sidebands for RefCav locking), and had to tweak the beam alignment into the RefCav using the input iris.  Once done, the RefCav locked without issue and our picomotor mounts were used to maximize the RefCav TPD; we began with a TPD of 0.57 V and ended with a TPD of 0.82 V.  We've had it higher in the past, but we have less power out of the PMC than usual (due to our slowly increasing PMC Refl that we have yet to figure out the cause of), and less power out of the PMC means less power available for the FSS RefCav which means lower maximum RefCav TPD.  To end we realigned the beam onto the RefCav RFPD and took a visibility measurement:

This finished the FSS RefCav tune-up, so we left the enclosure.  We left the ISS OFF while the enclosure returns to thermal equilibrium; it will be turned back ON once that equilibrium is reached, and Ryan will do a rotation stage calibration.  This closes LHO WP 12057.

H1 CDS
david.barker@LIGO.ORG - posted 11:03, Tuesday 27 August 2024 - last comment - 09:00, Wednesday 28 August 2024(79735)
CDS Maintenance Summary: Tuesday 27th August 2024

WP12061 Upgrade RCG h1susex, remove LIGO-DAC delays

EJ, Erik, Jonathan, Dave, Daniel, Marc:

h1susex RCG was upgraded to a custom 5.3.0 specifically for the IOP to remove a delay in the new LIGO 28AO32 DAC. We compliled all of the user models with this RCG as well.

Our first restart was a simple model restart, but we got Dolphin errors. So our second restart was to fence h1susex from the Dolphin fabric and power cycle h1susex via IPMI. After this the models started with no issues.

No DAQ restart was required.

Code path is /opt/rtcds/rtscore/advLigoRTS-5.3.0_dacfix

WP12063 Alert System

Dave:

The locklossalert system code was modified to permit an alert window which spans over midnight, needed for the new owl shift hours.

Note that the business/every day filter is applied after the minute-in-day filter, so a window starting Friday evening and extending into Saturday morning will cut off at midnight if business days only is selected.

One other change:

For everyone subscribed to Guardian alerts, if for any reason the Guardian EPICS record cannot be read (node down, h1guardian1 down) the alert will now default to SEND.

Guardian Reboot

TJ:

h1guardian1 was rebooted at 08:21 PDT. All nodes except TEST came back automatically. TJ worked on TEST and got it going.

MSR Rack Cleanup

Jonathan:

Jonathan removed two test switches from the MSR racks which are no longer needed.

Comments related to this report
david.barker@LIGO.ORG - 14:43, Tuesday 27 August 2024 (79742)

I updated the o4 script which reports where we are in O4 and reminds us of important dates (break start/end, vent start/end).

 

Images attached to this comment
david.barker@LIGO.ORG - 09:00, Wednesday 28 August 2024 (79760)

Tue27Aug2024
LOC TIME HOSTNAME     MODEL/REBOOT
08:50:45 h1susex      h1iopsusex  <<< 1st try, restart model
08:50:59 h1susex      h1susetmx   
08:51:13 h1susex      h1sustmsx   
08:51:27 h1susex      h1susetmxpi 
08:57:11 h1susex      h1iopsusex  <<< 2nd try, reboot
08:57:24 h1susex      h1susetmx   
08:57:37 h1susex      h1sustmsx   
08:57:50 h1susex      h1susetmxpi 
 

H1 CDS
jonathan.hanks@LIGO.ORG - posted 09:58, Tuesday 27 August 2024 (79734)
WP 12058 Removed two temporarily installed switches from the h1-daq-1 rack.
I removed two switches from the h1-daq-1 rack that had been put in for testing.  They are no longer needed in racks.

I also took a moment to sort and clean up a number of SFP transceivers that had been left out on the MSR work area.
LHO VE
david.barker@LIGO.ORG - posted 08:12, Tuesday 27 August 2024 (79732)
Tue CP1 Fill

Tue Aug 27 08:08:21 2024 INFO: Fill completed in 8min 17secs

Jordan confirmed a good fill curbside.

Images attached to this report
LHO General
thomas.shaffer@LIGO.ORG - posted 07:35, Tuesday 27 August 2024 (79731)
Ops Day Shift Start

TITLE: 08/27 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 6mph Gusts, 3mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.09 μm/s
QUICK SUMMARY: Locked for 2.5 hours, seems like an automated relocked before that. PEM magnetic injections are running now. Planned 4 hours of maintenance today.

 

H1 CDS
erik.vonreis@LIGO.ORG - posted 07:17, Tuesday 27 August 2024 (79730)
Workstations updated

Workstations were updated and rebooted.  This was an OS packages update.  Conda packages were not updated.

H1 General (SUS)
oli.patane@LIGO.ORG - posted 22:02, Monday 26 August 2024 (79729)
Ops Eve Shift End

TITLE: 08/27 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Commissioning
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Observing and have been Locked for 5.5 hours. Quiet evening with nothing to report. Feedforward measurements were run, and I also just tuned squeezing up.

Little comparison on the damping for new PI over the past 5 days (ndscope)
Since starting back up last week, we have been seeing a new PI (79665), and it had caused a few locklosses. On Friday afternoon (08/24 ~00:00UTC - crosshairs on ndscope) it was decided that the easiest way to try and control it for now would be to change our input power from 60W to 61W. This, along with some changes to the damping gain, has helped keep it under control. Comparing the PI's behavior to the length of our lock stretches, we only see the ringups during the first 1.5-2 hours of each lock, so once we're thermalized past a certain point, they stop happening. The changes to damping (I'm not sure what was done besides raising the damping gain) also means that they get caught and damped quickly.
 

LOG:

23:30 Observing and Locked for 2 minutes
04:27 Dropped Observing to take FF measurements
04:46 Feedforward measurements done, running SQZ alignment
04:55 Back to Observing

Images attached to this report
H1 General
oli.patane@LIGO.ORG - posted 16:38, Monday 26 August 2024 - last comment - 17:25, Monday 26 August 2024(79726)
Ops Eve Shift Start

TITLE: 08/26 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 9mph Gusts, 6mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.07 μm/s
QUICK SUMMARY: Observing and just got to NLN a few minutes ago. Once we're thermalized I'll be taking LSC FF measurements.

 

Comments related to this report
oli.patane@LIGO.ORG - 17:25, Monday 26 August 2024 (79728)

SDF diff turning the PRCLFF gain off ws accepted to get into Observing

Images attached to this comment
H1 ISC
ryan.short@LIGO.ORG - posted 14:31, Monday 26 August 2024 - last comment - 16:24, Tuesday 27 August 2024(79721)
OMC Locking with Higher DARM Offset

Before the vent, we had lowered the DARM offset used at the end of the DARM_OFFSET state for locking the OMC since we had seen the PRG fall off and cause a lockloss with the nominal offset of 9e-05 (see alog79082 for details). When locking this afternoon, I raised the offset from 6e-05 back to 9e-05 after running through DARM_OFFSET, and seeing that the PRG didn't plummet and cause a lockloss, we continued locking. The OMC locked on the first try, something that hasn't been the case recently, so having more carrier light there seems to help. I compared OMC scans from this lock against the last lock, which used the lower DARM offset; attachment 1 shows the scan with the higher offset and attachment 2 with the lower offset. According to the OMC-DCPD_SUM channel, we get ~10mW more carrier light on the OMC when locking with this higher DARM offset.

I've put this DARM offset of 9e-05 back into ISC_LOCK and loaded it. We can watch over the next couple of lock acquisitions to see if the problem wth the PRG dropping off resurfaces.

Images attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 14:37, Monday 26 August 2024 (79722)OpsInfo

Tagging OpsInfo

If you see the power recycling gain start falling soon after DARM_OFFSET, you can turn off the LSC-DARM1 filter module offset, lower it, and turn it back on until the PRG stays steady, then proceed with OMC locking.

ryan.short@LIGO.ORG - 16:24, Tuesday 27 August 2024 (79745)ISC, OpsInfo

Now that we've locked several times successfully since yesterday with this higher DARM offset, I've also rearranged the state order in ISC_LOCK so that the DARM offset is applied before any ASC so that the OMC can work on locking while ASC converges (this is how the order used to be before the DARM offset issues started).

See attached for the new state progression around this point in locking.

Images attached to this comment
Displaying reports 7121-7140 of 84553.Go to page Start 353 354 355 356 357 358 359 360 361 End