Displaying reports 73281-73300 of 84044.Go to page Start 3661 3662 3663 3664 3665 3666 3667 3668 3669 End
Reports until 11:43, Monday 31 March 2014
H1 AOS
thomas.vo@LIGO.ORG - posted 11:43, Monday 31 March 2014 (11095)
Beckhoff PLC3 for TCS
With the upcoming TCS CO2 work, we needed to work out any bugs in software land before trying to turn on the laser. A few additions (rotation stage) and deletions caused the old code to not work very well and so we re-mapped the configuration and re-linked the software to hardware.  This solved the problem of getting the hardware channels all the way through to EPICS.

A few more details:
- In order to not trip all the watchdogs in HAM2/3, Sheila turned off the connection between ISC to SUS temporarily as well as putting the MC Guardian in a down->paused state.
- We created a new boot project for PLC3.
- Committed the new system manager to SVN.
H1 ISC
alexan.staley@LIGO.ORG - posted 11:29, Monday 31 March 2014 (11093)
ISCTEX WFS

(Sheila, Alexa)

Since the beam on the WFS seemed a little large, we moved WFSA closer to the lens by 5 inches and WFSB closer to the lens by 9 inches. Hopefully, this helps keep the sensing matrix more consistent when we recenter the beams. TBD...

H1 SYS
daniel.sigg@LIGO.ORG - posted 09:26, Monday 31 March 2014 - last comment - 10:33, Monday 31 March 2014(11090)
Commissioning calendar for next 2 weeks

Here is the list of commissioning task for the next 7-14 days:

Green team:

Red team:

Blue team (ALS WFS):

Blue team (ISCTEY):

SEI/SUS team:

Comments related to this report
jeffrey.kissel@LIGO.ORG - 10:33, Monday 31 March 2014 (11091)SUS
Daniel means increasing the UIM driver range, as per ECR E1400164 and integration issue 762. Note, also, that after conversations with LHO staff and subsequent follow up with P. Fristchel, it has been agreed that we should modify all four UIM drivers (not just the ETM drivers as mentioned in the current versions of the ECR and II).

Also by ETMY Diagonalization, he means (once the optical lever installation is complete)
- Balancing the coils on the UIM and PUM stages
- Gathering measurements, designing, and installing filters for 
    - Length-to-Angle frequency dependent decoupling
    - P and Y plant compensation filters for WFS.
 (See LHO aLOG 11067 for list of measurements.)
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 08:58, Monday 31 March 2014 (11089)
CDS model and DAQ restart report, Monday 31st March 2014

Weekend report

model restarts logged for Sat 29/Mar/2014

no restarts reported for Sat

model restarts logged for Sun 30/Mar/2014

2014_03_30 00:42 h1fw0

Unexpected restart of h1fw0

H1 ISC
alexan.staley@LIGO.ORG - posted 18:53, Sunday 30 March 2014 (11088)
ALS WFS work

Sheila, logged in as Alexa. 

I spent some time working on the ALS WFS today.  The guardian can now be used to turn on and off the WFS, although to get started we still need to align the cavity pretty well by hand and then center the WFS by hand.

 I was able to lock one DOF (DOF_2_Y) with a 1Hz bandwidth (transfer function attached, this was using a gain of -0.05).  However, when I tried this after recentering the WFS it was no longer stable, so I've set it back to -0.001

One difficulty I had today is that the sensing matrix changes enough as the WFS centering changes to make the loops become unstable.  When I arrived I centered the WFS, locked all 4 DOFs, then the centering changed and the pit DOFs moved away from 0 until they saturated the DAC. 

The second screen shot attached is the measurent of the sensing matrix for ITMX pitch that Alexa and Daniel did on friday, next to one I did today.  You can see they are totally different.  After recentering I measured the sensing matrix again, and got back a value similar to Friday's. 

Images attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 20:58, Saturday 29 March 2014 (11087)
Pumping Y-end with MTP backed by QDP80
Assembled aLIGO RGA on BSC6 and rough pumped RGA volume -> replaced gauge controller and foreline pirani gauge on MTP (gauge exhibiting failure mode of losing "enhanced" or stable readings in turbulent flow conditions, controller losing power to B channel pirani (foreline) intermittently) -> Demonstrated MTP "Safety Valve" closing as a function of foreline pressure setpoint when in "Turbo" mode -> Pumping BSC10 annulus 

2100 hrs. local -> Kyle leaving site now
H1 SEI (SYS)
jameson.rollins@LIGO.ORG - posted 13:51, Saturday 29 March 2014 - last comment - 14:04, Saturday 29 March 2014(11077)
SEI guardian status update: BSC ISI guardians now running for BS, ITMX, ITMY, ETMX

BSC ISI guardians now running for BS, ITMX, ITMY, ETMX

After a lot struggle with the BSC ISIs this week, we finally have all commissioned BSCs under guardian control.

The BSC seismic stack is one of the more complicated systems we've deployed, as it currently consists of three nodes in a "managed" configuration: two separate nodes for each of the two ISI stages (e.g. ISI_ITMY_ST1, and ISI_ITMY_ST2), and a "chamber manager" (SEI_ITMY) that coordinates the activity of the two subordinate ISI stages.  USERAPPS location of the top level modules (and primary loaded library):

All single ISI stages share the exact same code, for each BSC stage as well the single HAM stage.  Here are the system graphs for the SEI manager and ISI stage 1:

The manager (left above) has three main requestable states: READY, DAMPED, ISOLATED.  The READY state puts both of the ISI stages in READY, the DAMPED state puts them both in DAMPED, and the ISOLATED state puts both into isolation levels that can be specified in the manager system description module (USERAPPS/isi/common/guardian/SEI_*.py).  At the moment all systems are using their default configuration, which is to operate the ISI stages in the HIGH_ISOLATED state, which corresponds to "lvl3" from the old "command" scripts.  This can't currently be changed on an individual BSC basis, due to a guardian core issue, but I'm working on it.

The degrees of freedom for which to restore cart bias offsets can also be specified on a per stage basis.  Note the ISI_ITMY_ST1 graph above has "RESTORE_ISO_CART_BIAS_*_RZ" states, and no corresponding states for the other degrees of freedom (X, Y, Z, RX, RY).  This indicates that we are restoring only RZ target cart bias offsets for this stage.

Notes about the BSC isolation procedure

There are a couple of important things to note about the current isolation procedure:

We had a lot of trouble dealing with the T240s in stage 1.  Their exterme sensitivity to pitch causes them to saturate if a large pitch cart bias offset is applied to the stage.  We there spend a lot of time trying to figure out how to gracefully bring them in and out of the loop, by reducing their gain and/or changing the blend filters during the isolation procedure.  All of theses things has problems, though:

This makes things very difficult if we need to hold pitch offsets on any of the ISIs.  It looks like we need to fix the synchronous switching of the T240 compensation filters before we can handle pitch cart bias offsets.

However, after speaking with the integration team, it seems that it's ok to operate temporarily in a mode where we only restore RZ cart biases.  Any needed pitch offsets can be held in the suspensions for now.  Therefore we do not need to change the T240 gains or blend filters during the isolation procedure.  This is good news, since it means we can run with the current guardian behavior.  Obviously this will have to be fixed down the line, as we will likely want to hold pitch offsets on the ETMs, but we have a workable solution for the moment.

Guardian BSC isolation configuration and procedure

With the above configuration we can successfully isolate the BSCs without touching the T240s or blend filters.  It seems to be fairly robust.

Important operator notes:

Notes about guardian SEI manager behavior

The SEI managers represent a new step for guardian, as their primary task is to control the state of other guardian nodes, which we haven't needed up to this point.  Getting their behavior robust has been quite tricky.  It necessitated a lot of work on the guardian built-in "manager" library, as well as a lot of experimentation with the SEI manager module itself.  The manager is designed to be fairly robust against commissioner noodling with it's subordinates, but that is also fairly tricky to get right in a robust and useful way.  Some things will need to be improved, but this is the behavior currently:

While things seem to working well at the moment, I certainly don't claim that all the bugs have excised.  If the manager does get hung in some state, resetting to the INIT state should clear everything, reset the ISIs, and get things back on track.

TODO for SEI guardian

Images attached to this report
Comments related to this report
jameson.rollins@LIGO.ORG - 14:04, Saturday 29 March 2014 (11086)

Note, the following code version number are applicable to the above configuration:

  • guardian: svn r878
  • cdsutils: svn r211
  • USERAPPS/isi/common/guardian: svn r7619
LHO VE
kyle.ryan@LIGO.ORG - posted 10:15, Saturday 29 March 2014 (11085)
Kyle on site for rough pumping of Y-end -> Should be here several hours -> Will make aLOG entry when leaving


			
			
H1 ISC (CDS, SYS)
jeffrey.kissel@LIGO.ORG - posted 19:21, Friday 28 March 2014 - last comment - 11:56, Monday 31 March 2014(11083)
ALS Electro-optical Control Schematic
A. Staley, S. Dwyer, J. Kissel 

After speaking with Sheila and Alexa, trying to understand the spectrum in LHO aLOG 11026, I realized that -- because of so much unexpected excess noise -- the current ALS setup has diverged significantly from the baseline plan. As such, we've worked together to create a diagram that is to-the-best-of-(my)-knowledge accurate, and makes sure to show how every control signal gets around to its actuators, including passes through digital land. The only things I did not include were the demodulator boards between the REFL diodes (ARM REFL, MC REFL, and the two green XARM and YARM REFLs) and lame things like whitening, AA, and AI chassis that facilitate getting in-and-out of digital land.

In particular, I draw your attention to the unorthodox and new features that were not part of the original design plan:
(1) The PRM is misaligned. This way we can use IFO REFL red diode signal (after normalizing / linearizing it with the red arm cavity trans) to measure the arm and offload at low frequency to the CARM board in the second input. This is unorthodox because normally we would feed REFL straight to the CARM board in all analog and retain the large bandwidth, but because the noise is so large, it must be linearized in the front-end (the only place the Arm TRANS signal exists), and fed out one of the corner station LSC DACs. Note that this configuration is just a special technique for these particular series of measurements and is not designed to become a normal part of the lock acquisition scheme.

(2) The Voltage Controlled Oscillator (VCO) actuated, Phase-Locking Loop (PLL) that's wrapped around the corner-station COMM and DIFF Phase-Frequency Discriminators (PFD) are used to reduce the noise enough to keep the PFD in the linear, phase detecting mode (check out pgs 6-8 of the AD9901 datasheet -- it's a really nice explanation of how it works).

(3) Before the hand-off, there's another temporary step which is to use the Beckhoff system ("Slow ADC" in the diagram) to feed directly to the PSL VCO (again at low frequency) to reduce the noise enough for the COMM hand-off to work. 

This is most certainly an impressively complicated system!

Note that this diagram has been committed into the noisebudget repo here:
${NBSVN}/trunk/HIFO/Common/
and I imagine will become an "as built" DCC document in the glorious future when everything is working beautifully.
Images attached to this report
Non-image files attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 11:56, Monday 31 March 2014 (11096)
Comments and my replies on the diagram from Bram, to be added/fixed later:

You are missing the Fiber Distribution Box, which shifts the laser frequency by -160 MHz before it gets shipped to the end-stations to compensate for the PSL AOM. The lasers at the end-station are then locked at the + (X-end?) and - (Y-end?) 40 MHz around the ‘0’.
Ah, yes. I had debated putting the distribution box in there, and decided against it for clarity because I forgot the important frequency shift feature.
See D1200136 for details or Figure for in T1000555 for schematical diagram.

Also, the F and S in the ‘servo’ block diagrams, is the S beckhoff? 
No -- the "F" and "S" are the "Fast" and "Slow" paths of what has been traditionally called a "common mode" board everywhere -- I was trying to start a clarifying revolution with this diagram...

The lasers in the end-station get the slow temp control directly from the beckhoff from the demos error signal monitor.

The beckhoff links are a little confusing, maybe make a little color legend?
H1 SUS
arnaud.pele@LIGO.ORG - posted 18:58, Friday 28 March 2014 (11082)
ETMY M0 and R0 TF running overnight

started at ~18:50 will run for 8 hours.

H1 ISC
sheila.dwyer@LIGO.ORG - posted 18:43, Friday 28 March 2014 (11072)
ALS WFS better beam quality

Yesterday I went to end X to investigate the beam shape for the ALS WFS.  (This is to follow up the measurements we made in alog 10774)

First I tried changing the alingment of the reaction chain, this didn't make a difference.  Then I tried changing the TMS QPD offsets.  This moves the lock point of the input alingment loops, changing the alingment of the green beam as it goes through TMS.  This certaintly changes the beam quality on the WFS, screenshots of several different offsets are attached.  In the end I left the offsets at ALS-X_IP_ANG_PIT at -50, ALS-X_IP_POS_PIT at 150. 

I am not sure why changing these offsets makes such a difference in the beam quality. 

This moved the QPD A by 0.07 counts and QPD B by 0.04 counts.  Pablo and I redid the pointing of TMS to the baffle PDs, but it hadn't changed more than the normal change between ISI trips.  However, switching these offsets does move the beamspot on ISCT1, I moved PR3 by 0.9urad in pitch to bring the spot back.  

After realinging the arm, we re checked the phasing of the WFS. We did this by putting an exictation on LSC-X_EXTRA_AO_1_EXC at 50Hz and looking at the signals of the individual segments, which seems fine.  This is saved in /home/sheila.dwyer/ALS?HIFOX/WFS/WFS_PhaseDTT_March282014.xml

Then we started the sensing matrix measurement in Alexa's alog 11073

Non-image files attached to this report
H1 SUS
arnaud.pele@LIGO.ORG - posted 18:39, Friday 28 March 2014 - last comment - 18:41, Friday 28 March 2014(11067)
ETMX UIM filter tested

This week, I tested the ETMX UIM P2P and Y2Y filters (described in alog 10912). The results are showing that filters are working as expected giving a single pendulum transfer function falling at 1/f^2 after the resonnance (f=0.45Hz for pitch, f=0.6Hz for yaw). Also the coupling between pitch and yaw is small (~5%) so unless this is too much, there's no need for diagonalization for UIM.

Attached pdf are showing P2P P2Y (first pdf) Y2Y Y2P (second pdf) transfer functions with/without the filter.

Left to do for next week :

ETMX : Long decoupling on UIM/PUM

ETMY

ITMX/ITMY

Non-image files attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 18:41, Friday 28 March 2014 (11080)

sorry for the 90degrees rotated pdfs ! couldn't find a way to rotate and save them.

H1 SEI
hugh.radkins@LIGO.ORG - posted 18:03, Friday 28 March 2014 (11078)
WHAM4 SEI--HEPI MEDM Populated & Safe snapped--Measurement Underway

I got the medms populated so the cartesian location values are now valid.  These initial values today are the 'Aligned' numbers.  They will probably be a little different after these range of motion/linearity tests complete and the stress of the system are further released/shaken out.

H1 SEI
sheila.dwyer@LIGO.ORG - posted 16:55, Friday 28 March 2014 - last comment - 11:45, Monday 31 March 2014(11070)
HAM3 ISI trip

This trip happened around the time of a beckhoff restart.  A beckhoff restart causes a crash of the IMC guardian, which causes it to stop.  Its not clear to me why this should cause MC2 to get a large signal, but it seems to.  This causes a cascade of trips.  Even though I don't know why this happends, creating a safe state in the IMC guardian so that it handels missing channels better could help with the problem. 

Another solution is to have the SUS WDs not trip HEPI.  Can we get that fix soon?

HAM2 also tripped at the same time, Jeff and Hugh brought that back. 

Images attached to this report
Comments related to this report
jameson.rollins@LIGO.ORG - 17:54, Friday 28 March 2014 (11075)

To be clear, IMC guardian did not "crash" in this particular situation.  The guardian responded exactly as it's currently programmed to respond, which is to go into ERROR when it looses communication with any of the channels it's monitoring.  I want to distinguish and ERROR condition, which is something that guardian handles, to a "crash", which means that the guardian process died unexpectedly.

Here's my guess for the sequence of events:

  1. Beckhoff action somehow caused one of the PSL channels (H1:PSL-PERISCOPE_A_DC_ERROR_FLAG) to momentarily disappear.
  2. IMC guardian, which is monitoring the above channel, goes into ERROR when said channel disappears
  3. Around this time, the IMC looses lock, presumably for the same reason the PSL channels disappeared.  Since the IMC guardian is in ERROR and no longer monitoring the IMC situation, it doesn't shut off the feedback to MC2.
  4. Large control signals into MC2 cause it to trip.  A typical watchdog cascade causes everthing from here to Walla Walla to trip.

It's possible guardian could be made slightly more robust against loss of some of it's channels, but that only helps up to a point.  Eventually guardian has to drop into some error condition if it can't talk to whatever it's trying to control.  It could try to move everything to some sort of safe state, but that only works if it can talk to the front-ends to actually change their state.

daniel.sigg@LIGO.ORG - 09:30, Saturday 29 March 2014 (11084)

A Beckhoff restart also causes the IMC servo board to be reset, as well as all whitening for photodiodes, wavefront sensors and QPDs. I assume that the resulting transient caused the MC to trip. It would be interesting to know, if this is due the length or alignment system. Is it the initial transient or a run-away integrator? In either case this should not result in a trip. A better action would be to simply turn off the ISC inputs.

arnaud.pele@LIGO.ORG - 11:45, Monday 31 March 2014 (11094)

After taking a look at the time of WD trips, it seems like HAM3 ISI trips before MC2, see green plot vs red plot (the X axis is the number of seconds after gps=1080083200)

Images attached to this comment
H1 SUS
betsy.weaver@LIGO.ORG - posted 14:31, Friday 28 March 2014 - last comment - 13:08, Monday 31 March 2014(11064)
EY closeout

Margot and I performed all of BSC10 closeout tasks listed in my previous alog regarding this closeout.  We pulled the FC from the ETMy optic a little after 11am, and the door was going on at ~1:30.  It took "so long" (ha) because we ended up having to spend some extra time calling in the troops about one of the ISC viewports having what appeared to be a scratch on it's inside surface.  More details to come as needed.

Comments related to this report
betsy.weaver@LIGO.ORG - 18:44, Friday 28 March 2014 (11081)

Here are some further pictures of the ACB swing back and the FirstContact spray cone attachment to the HR side of the ETMy QUAD structure from yesterday.  Yes, the cone protrudes into the baffle.  Yes, Margot still was able to fit her head and an arm with the spray bottle in the cone.

The last picture shows the placement of the horizontal wafer and verticle 1" witness optic in the center of the chamber, placed today just before the door went on the chamber.

Images attached to this comment
margot.phelps@LIGO.ORG - 13:08, Monday 31 March 2014 (11097)
Notes and dcc numbers for Contamination Control Samples (ones that came out and ones that went in) to WBSC10 before doors went on:
1. Vertical 1’’ optics on QUAD: T1400246 (SN1195) out and T1400247 (SN 262) in. 
2. Vertical 1’’ optic under quad: T1400248 (SN261) placed at end of closeout.
3. Vertical wafer on quad: T1400249 attached to quad under HR of ETM.
4. Horizontal wafer on floor under quad:T1400250 placed at end of closeout.
5. Horizontal 1’’ optic: T1400251 was left on beam tube floor between BSC10 and BSC6. Looking up SN/history. Didn’t take it out, because we had no clean optic containers handy.
6. 24’’ PET swipe sample taken in tube near purge today close to end of closeout. Labeled, and taken to PET microscope area. 
H1 SUS
betsy.weaver@LIGO.ORG - posted 19:27, Thursday 27 March 2014 - last comment - 18:38, Friday 28 March 2014(11049)
ETMy First Contact Applied

This morning, SEI, TMS, and SUS gave a cursory "approval" of their systems in order to proceed with chamber close-out tasks.  So, after lunch, Margot and I applied First Contact to the ETMy-HR surface which will be pulled just before closing the door tomorrow.  The order of events this afternoon (and late morning in order to catch the alog up):

Jim locked the ISI and put final touches on chamber feedthru cable routing.

Apollo+Joe+Gerardo had a viewport party at the BSC10/6 area.

Margot and I locked the lowest masses of the ETMy QUAD.  Then we,

Secured the ACB locking bracket and used the wedge to swing it back and prop it into place.  See notes on this in attached alog.

Attached the FC QUAD cone to the structure and sprayed FC (standard 3 coats) on the HR surface.

Removed the cone from the suspension.  Finished FC layer reinforcement.

Removed the wedge from the ACB and restored it's nominal hanging position. Removed the ACB locking bracket screws.

Started stowing all left over bags of "junk" left outside of the chamber door on various table surfaces by SUS and every other team that visited this last month.  If you're missing something (and you noticed), I took ownership of it.

Jim arrived and unlocked the ISI.  The ETMy QUAD will remain locked during dome replacement.

Apollo started gearing up to remove walking plates from upstairs and crane the dome on the chamber.

 

Tomorrow's to-do list which Margot and I will do until the last step:

Inspect insides of viewports.

Adjust ACB to check magnet gap which may or may not have shifted slightly due to the swing back from today.

Clean the chamber, vacuum, etc.

Take final PCL swipes.

Blow various other optic surfaces.  Or not.

Swap in final 1" witness optic on QUAD structure.

Check/Remove all tools.

Remove ETMy-HR with N2 blow.

Set and lock all 64 QUAD EQ stops and nuts, accounting for buoyancy.

Set QUAD BOSEMs to account for buoyancy.

Quick QUAD health check with DTT TFs.

Lay final witness plates.

Exit chamber.

Finish stowing everything around door and move it out of the way.

Put door on.

Comments related to this report
betsy.weaver@LIGO.ORG - 20:10, Thursday 27 March 2014 (11050)

So, when we went to swing back the ACB and stuff the wedge into place, the wedge could not be inserted in the gap all of the way due to interference with the round balance weight attached on top of the baffle.  We pushed the wedge in ~2/3 of the way which gave us ~2/3 more working area than we had with out swinging it back at all.  Good enough.  (Removing the round weight would have cause the balance of the ACB to be out of alignment so that wasn't an option.)  Also, even though the locking bracket was engaded which provided some support to the down tube of the ACB, the ACB was still a bit "rickety" during this maneuver.

betsy.weaver@LIGO.ORG - 18:38, Friday 28 March 2014 (11079)

The first picture below is of the ACB wedge not quite installed as far in as it could.  This meant that the screw holes to secure the wedge in place could not be engaged.  ACB engineers must have anticipated this so they provided us with a "safety" chain which one could tie up "somewhere" such that the wedge doesn't come flying out at you if you bump the baffle while it's up.  The chain was of limited length, so they best Margot and I could do was wind it around the post loosely a few times.  Note, the wedge seemed wedged in well enough that it didn't budge when we did happen to bump the baffle a bit here and there with the cone work.

 

Also note, the ACB did not hang the same way after we restored it.  Mitchell had to make some minor tweeks to the balance today in order to recover the correct gap setting for the magnets (specifically 0.09 inches! - Really?  Not 0.08" or 0.1", but 0.09", huh!?  Seems a bit tight given the grossness of both the swing back application and the shear size of the baffle... but what do I in SUS know.)  In the end, as usual the ACB comes with in a few mm of the QUAD structure at the head of one of it's screws (see 2nd picture with red circle around it).

Images attached to this comment
Displaying reports 73281-73300 of 84044.Go to page Start 3661 3662 3663 3664 3665 3666 3667 3668 3669 End