arnaud.pele@LIGO.ORG - posted 17:16, Tuesday 11 February 2014 (9987)
SUS monitor channels added to frame/science frame #WP 4179

Today I finished up modifying the SUS monitor models as it has been deployed at LLO in december (alog 9964 and alog 10010), under ECR 1300261 in order to add the voltage and noise monitors in the frame + science frame for each individual osem. The following models were modified, recompiled, and commited to the svn :

h1susb123 (bs, itmx, itmy) rev 7070
h1susauxh2 (mc1, mc3, prm, pr3, im1, im2, im3, im4) rev 7087
h1susauxh34 (mc2, pr2, sr2) rev 7072
h1susauxh56 (srm, omc)  rev 7074
h1susauxex (etmx, tmsx)  rev 7083
h1susauxey (etmy, tmsy) rev 7085

Details of modifications are specified below :

For the quads :

M0/R0 Noisemon was changed from 1024 to 512Hz
M0/R0 Voltmon was changed from 1024Hz to 256Hz
L1/L2 Noisemon were created at a rate of 1024Hz
L1/L2 Voltmon were created at a rate of 256Hz

For the triple suspensions :

M1 M2 M3 Voltmon added at 256Hz
M1 Noisemon added at 512Hz, M2 at 1024Hz and M3 at 2048Hz

For the IMs and OMCs

M1 Voltmon added at 256Hz
M1 Noisemon added at 512Hz

For the TMS :

M1 Voltmon added at 256Hz
M1 Noisemon added at 256Hz

LHO General
patrick.thomas@LIGO.ORG - posted 17:10, Tuesday 11 February 2014 (9973)
Ops Summary
Fire department testing smoke detectors

09:15 Mike L. taking Kim on tour through LVEA, end Y
09:39 Jim B. upgrading GDS tools (WP 4432)
09:56 - 11:50 Mitchel R. cleaning up in LVEA west bay
10:31 Jaclyn S. to squeezer bay
10:35 Dave B. running OS updates on cdsssh, cdslogin, lhoepics
10:56 - 11:23 Stefan B. restarted h1odcmaster model, which crashed the h1oaf frontend and killed mxstreams. Dave B., Jim B. recovering
11:21 - 12:45 Joe D. working on PCAL assembly in H2 PSL enclosure
11:48 LN2 delivery ~10 min away
12:01 Jaclyn S. and Alexa S. to end Y to look at electronics
13:01 - 13:36 Justin B. to end Y to look at conditions for moving table in
13:03 Kris cleaning at end X
13:14 - 14:40 Karen at end Y
13:32 HVAC repair for chiller unit in MSR
? - 16:13 Mitchel R. cleaning up in LVEA west bay
X1 DTS
james.batch@LIGO.ORG - posted 16:48, Tuesday 11 February 2014 (9993)
Configuration changes for testing h1odcmaster
We are testing the h1odcmaster on the DAQ test stand, this has required that the H1.ipc file be replaced with a copy of the one from the h1 system (with host names changed from h1 to x1).  The previous H1.ipc has been saved as H1.ipc.11feb2014.  This will be restored when we are finished troubleshooting.
H1 ISC (ISC)
corey.gray@LIGO.ORG - posted 16:33, Tuesday 11 February 2014 (9981)
TMS-EY Alignment Work

(Corey, Doug, Keita, Yuta)

X-Translation Push

Started off the day securing the TMS with "hard stops" and installing (7) teflon-tipped dog clamps on the TMS Upper Structure & Pusher in preparation for our horizontal translation of ~6.5mm to the "left".  Doug watched us as we pushed the entire assembly this distance.

Yaw Alignment

After the big move, we then set up to fix the yaw of the structure.  This took a few iterations, because we noticed that after we Yawed and looked good, we would back off the Pusher and see that the structure springed back a little.  So we had to figure out how to overshoot and then have it spring back to nominal.  Once Doug was happy here, we torqued down all the Dog Clamps (25ft-lbs) and moved to pitch.

Pitch Alignment

The TMS was then freed up, balanced, and prepared for pitch work.  The back end of the TMS was seen to be pitched UP, so Yuta and I moved the TMS Table Counterweights toward the back, and we were able to get the Table pretty flat according to Doug.

Clean-Up & Tomorrow

So with this alignment, with a 0.0 Pitch & Yaw bias, we had a most excellent alignment of the TMS.  Doug and Jason were excused and we worked on finishing clean-up:  remove all Pusher hardware, re-install/torque removed Dog Clamps, re-adjust BOSEMs to new alignment, and then put TMS on EQ Stops.

NEXT UP:  Install tooling to secure the TMS for Cartridge insertion.

Images attached to this report
H1 SUS (CDS, SUS)
patrick.thomas@LIGO.ORG - posted 14:28, Tuesday 11 February 2014 (9989)
Arnaud restarting sus models
h1susauxb123
h1susauxh2
h1susauxh56
h1susauxh34
h1susauxex
h1susauxey
H1 SUS
jeffrey.bartlett@LIGO.ORG - posted 13:38, Tuesday 11 February 2014 (9986)
Noise in SR2 Power Spectra
   The in chamber power spectra (aLOG #9961) for H1-SR2 showed noise at around 10Hz and up which does not appear in the LLO spectra for the same type suspension and phase of testing. Per the testing group’s request, the spectra data for H1-SR2 was reprocessed to show the individual channel data for each DOF and with additional comparisons to other LHO HSTS suspensions. The plots are posted below. All scripts and data files have been committed to the SVN repository. 
Non-image files attached to this report
H1 ISC
patrick.thomas@LIGO.ORG - posted 13:20, Tuesday 11 February 2014 - last comment - 14:49, Tuesday 11 February 2014(9985)
Filiberto and Aaron ISC cabling at HAM6


			
			
Comments related to this report
patrick.thomas@LIGO.ORG - 14:49, Tuesday 11 February 2014 (9990)
Done.
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 11:40, Tuesday 11 February 2014 (9984)
Restart h1odcmaster caused DAQ data errors
Dave, Jim, Stefan

Stefan restarted the h1odcmaster model which has been running since Friday evening and it did two things: locked up the h1oaf0 machine and caused DAQ errors on both end station front end computers, plus LSC, ASC and SUSAUXB123. Running start-streamers on the affected front-ends did not fix the problem. We took h1oaf0 out of the dolphin fabric, its console was locked up so we power cycled the CPU. All the models started OK and the MX stream error resolved itself. A few minutes later the IOP IRIGB drifted upwards to 66 and then back down to 15.

I stopped the h1oafmaster model and then started it again. Once again it froze up the CPU but this time no MXSTREAM errors on other front ends.

We took h1oaf0 out of the fabric and power cycled, once again h1odcmaster started (with 0x2000 DAQ error due to INI file mismatch). 

We are now testing on the DTS.
H1 CDS (CDS)
jonathan.hanks@LIGO.ORG - posted 11:16, Tuesday 11 February 2014 (9983)
CDS Login server updated
The login machine has been updated today, including an updated authentication module.  This work required a reboot of the system.
H1 ISC
stefan.ballmer@LIGO.ORG - posted 10:33, Tuesday 11 February 2014 (9979)
added 3-way power combiner to 9MHz line - reduced IFR from +13dBm to +10dBm to avoid saturation
The first power amplifier after the IFR function generator was saturation at an input of +13dBm.

- I reduced IFR from +13dBm to +10dBm to avoid saturation.
- I backed out all 3dB attenuators in the various feed lines.
- I added a 3-way power combiner in the the 9MHz line to the EOM.

This increased the drive to the 9MHz EOM by 10*log10(3)=4.77dB (compared to the original setting), or 1.77dB compared to yesterday's setting.

We now should have a 9MHz modulation index Gamma of about 0.17.
X1 DTS
james.batch@LIGO.ORG - posted 10:22, Tuesday 11 February 2014 (9978)
New gds software installed on DAQ test stand
A new version of gds tools has been installed for the DAQ test stand computers running Ubuntu 12.04 or Mac OS X.  This will be effective with new logins or tools started from the command line of a new terminal window.  The changes include a fix for diaggui to allow it to display reference traces with the proper x-axis values (Bugzilla 254), and a fix for foton to increase the number of significant digits used in design strings.
H1 CDS
james.batch@LIGO.ORG - posted 10:07, Tuesday 11 February 2014 (9977)
New version of gds tools installed for control room machines
A new version of gds tools has been installed for the control room computers running Ubuntu 12.04 or Mac OS X.  This will be effective with new logins or tools started from the command line of a new terminal window.  The changes include a fix for diaggui to allow it to display reference traces with the proper x-axis values (Bugzilla 254), and a fix for foton to increase the number of significant digits used in design strings.
H1 General
patrick.thomas@LIGO.ORG - posted 09:31, Tuesday 11 February 2014 (9976)
Corey, Keita, Yuta to end Y
Translating TMS on test stand.

Doug or Jason watching alignment.
H1 ISC
sheila.dwyer@LIGO.ORG - posted 06:15, Tuesday 11 February 2014 - last comment - 11:03, Tuesday 11 February 2014(9970)
green team work yesterday

Yesterday we made some measurements for the noise model (end sation PDH spectra and open loop), and ALS COMM slow path, fast path and open loop). 

Then we moved on to trying to make some noise measurements.  We tuned the demod phase of RF A 9 to put the signal in the Q phase (we don't have enough range with the phase shifter to put it in the I phase).  The demod is now on local control (remote controls not working) with al switches down except for 8, 2, 1, and 1/2 (11.5 ns delay). 

We also tried to check the IR alingment into the arm cavity by locking at the transmitted PD.  We found the 00 mode with the COMM PLL offset at 1.23, and saw up to 90 counts in transmission.  We saw a misalingment mode (between pitch and yaw) at 6.48, with 17 counts in transmission.  We strugged to improve this, mostly because the transmitted power was fluctuating so much from the frequency noise. 

We engaged the boost on the PLL readback path, (adding more DC gain to ALS CARM).  This reduced the noise significantly, although we were still moving by mre than a linewidth, so we can't really make a linear measurement when locked with ALS COMM. 

Comments related to this report
daniel.sigg@LIGO.ORG - 11:03, Tuesday 11 February 2014 (9982)

Evan, Daniel

We found that the controls cable is missing from the second delay line, The cable we need is ISC-332 (DB25). We did not find any cable with this number, but we found two ISC-323. One is a DB25, the other a DB15. ISC-323 is for a shutter on ISCT6 (DB15). We think that ISC-332 (DB25) has been mislabeled. Unfortunately, it was pulled to the AS port rather than REFL.

H1 SUS
arnaud.pele@LIGO.ORG - posted 21:00, Monday 10 February 2014 - last comment - 09:23, Tuesday 11 February 2014(9965)
ETMY modes measurements

Brett S Arnaud P

This afternoon we measured the resonnance frequencies of the ETMY in two locking configurations in order for Brett to improve the parameters of the quad model. We used PUM/UIM osem signals as well as the "temporary" optical lever to measure them. We plugged the pd electronics to the three first channels of h1pemey ADC0 in order to read the op lev signal from dtt.

For the first measurement, the top mass was locked and PUM, UIM, test mass were free. At first we didn't have a really clear signal from the PD in Pitch and Yaw, and since the PUM flags seemed to be really low and touching the osems Brett had to adjust the earthquake stops of the reaction chain in order to get the lower masses freely suspended. After the few adjustements, measurement showed all the needed frequencies in the osems as well as in the op lev.

For the second measurement, the top mass and UIM were locked. This time, the reflected beam was moving too much due to excess pitch/yaw motion, and after waiting a long time for the suspension to settle down, we changed the plan, and pointed the beam to the right edge of the test mass (bottom edge of the ear) to measure only the bounce modes. The dtt spectra of the op lev sum looked somehow funny. Fortunately the PUM osem signals were sufficient and showed all the frequencies, even the ones of the bounce modes.

Brett will post the results when they will be processed

Comments related to this report
brett.shapiro@LIGO.ORG - 09:23, Tuesday 11 February 2014 (9975)
At first glance the mode frequencies look pretty good. The pitch modes of the double pendulum hang (UIM locked) were a few percent off and the bounce modes of the same case were few percent high. The bounce modes were similarly high in the single hang case, so that is consistent at least.

We left the ETMY with the main chain top mass locked and the reaction chain UIM and PUM locked, in case some measurements need to be repeated after I look at them in more detail.

I plan to post the measurements to the log as I look at them in more detail.
LHO General (CDS, PEM, PSL)
patrick.thomas@LIGO.ORG - posted 17:40, Monday 10 February 2014 - last comment - 14:15, Tuesday 11 February 2014(9964)
Lighthouse LMS Express 7 crashed
This is the commercial software controlling the Lighthouse particle counters in the H1 PSL enclosure.

Logging into h0dust (Windows 7 Pro SP1 64 bit virtual machine) through VNC this afternoon I found the error message in the attached pictures.

I found two related events under Event Viewer (Local) -> Windows Logs -> Application on 2/7/2014 11:02:01 AM. The text of the errors is attached.

Searching for related information I found the following: http://support.microsoft.com/kb/2640103
Images attached to this report
Non-image files attached to this report
Comments related to this report
cyrus.reed@LIGO.ORG - 14:15, Tuesday 11 February 2014 (9988)

It looks like the LMS software install includes the .NET 4 runtime, but it was probably out of date at install.  4.5.1 is available through Windows Update, I would assume it includes the fix mentioned in the KB article (since it's over a year old now), but I haven't been able to confirm what's included in the update.  It's probably the first thing to try though, but it means downtime for the updates to install.

H1 SEI (SEI)
hugo.paris@LIGO.ORG - posted 15:38, Monday 10 February 2014 - last comment - 15:23, Tuesday 11 February 2014(9955)
HAM ISI Command Scripts Work Again

HAM3-ISI tripped recently and needed to be turned back on. Commissioners noticed that the command buttons were not working anymore.

It turns out that someone ran an SVN update on the following file: /opt/rtcds/userapps/release/isi/common/scripts/HAMISItool, which was to be updated along with new models, screens and scripts, after testing at MIT. Having a model/script mismatch, the command script would be looking for inexistent channels, and fail.

I reverted HAMISItool to -r5830. This script should not be updated from the SVN again, until the latest HAM-ISI update gets propagated.

HAM3-ISI was turned back on with Isolation Lv3 and 01_28 blend filters. Target values were not changed to ensure returning to previous alignment.

Comments related to this report
hugo.paris@LIGO.ORG - 10:53, Tuesday 11 February 2014 (9980)

The WD plotting software, which lives in the same folder, had stoped working too. It turns out that the related scripts were updated too, while they should not have been yet.

I reverted opt/rtcds/userapps/release/isi/common/scripts/wd_plots/ to -r6477, and the WD plotting software works again.

arnaud.pele@LIGO.ORG - 15:23, Tuesday 11 February 2014 (9991)

My bad, I updated the directory before modifying BSCISItool !

H1 SYS
kaitlin.gushwa@LIGO.ORG - posted 09:20, Thursday 06 February 2014 - last comment - 10:29, Tuesday 11 February 2014(9850)
HAM4 witness plates & optics
Jeff B. & Kate G.

A horizontal 4" witness plate (wafer) and horizontal 1" optic (S/N 1247) were placed near the center of the SEI table on Feb 5th. A vertical 1" optic (S/N 1245) was mounted on the HR side of SR2. A vertical wafer still needs to be set near SR2. 

Important notes: Witness samples were not placed immediately after the chamber was opened. Jeff did some cleaning before SR2 was installed and the wafers and optics were placed. Also, the 1" optics were removed from their PET G containers, put in PEEK holders, double bagged, and stuck in a bag with other tools before going in chamber. I didn't see obvious scratches in the First Contact, so hopefully the samples weren't damaged. 
Images attached to this report
Comments related to this report
kaitlin.gushwa@LIGO.ORG - 10:29, Tuesday 11 February 2014 (9974)

On Feb 7th, I used a handheld particle counter in the cleanroom around HAM4 and in chamber before starting work. The counts seemed really high. The particle counter (used for trending data) on the +X side of the chamber was not working.

Size (um) Cleanroom Count Chamber Count
0.3 10725 17177
0.5 2518 3895
0.7 1075 1562
1.0 586 798
2.0 169 213
5.0 1 0

I placed a vertical wafer near the HR side of SR2, and inspected the cleanliness of the chamber with a flashlight array. There were a few particles on the witness samples (which had not seen work since they were placed two days earlier), and a lot of contamination on the TCS HWS optics and SEI table. The dark iPhone photos don't do it justice.There was only time for a quick cleaning using the TigerVac and IPA soaked wipes. It would be helpful to get a PET swipe sample the next time work is done in HAM4. 