I measured the MC cavity lock by adjusting the ifr frequency and measuring the transfer functions using an SR785. The source was injected into Exc A of MC common mode board and teed to ch1, meanwhile ch2 was connected to either IMON of REFL9 (1*FSR) and REFL45 (5*FSR). Data.txt is the data collected from REFL9 and Data_2.txt is that collected from REFL45. ArmCavityLength.m is the script that determines the length given the zero crossing of the projection. The attached plots show the results with a linear regression included.
The cavity length is determined to be
Implemented the new timing distribution system for the CPS's units (BSC's only) in the LVEA. Changes include using BSC2 CPS_1 unit as master. The timing signal is feed to a STS-2 Distribution chassis and distributed to all other CPS units. The STS-2 Distribution Chassis was placed in SUS-R6 floor rack.
Attached is a spectra of ITMX ST2 GS13 took after the timing system was changed (in blue), and compared with the reference from Seb's alog 9537 from last week (in yellow). Clearly, the combs do not appear any more.
Repeatedly in the past we had found the ISS PD's DC level jump to a state past 10Volts to somewhere arount 12V. The last time this occured last night (alog 9637). Staring at the PSL ODC screen, it told us that two things were wrong: BIT7 "ISS actuation saturation" and BIT11 "NPRO Relaxation Oscillation" were red.
Last night we fixed the ISS actuation saturation by readjusting the set point, but the PDs remaind in the odd 12V state.
Staring at the ODC screen this morning, BIT11 "NPRO Relaxation Oscillation" remained red. Taking that suggestion by ODC seriously, I went out to toggle the NPRO noise eater. Sure enough - the the ISS PD's switched back to the normal "<10V" state, and BIT11 went green.
Conclusions:
- Our long-standing ISS PD mystery is caused by noise eater oscillations.
- The PSL ODC system correctly announced this problem.
FYI, The ODC overview screen is available from the SYS tab on the site overview, and has links to all subsystem screens. The subsystem screens have a description string for each bit. Currently fully operational -including meaningful thresholds -are PSL, IMC, ISI and HPI for all currently commissioned systems, as well as SUS for MC1, MC2 and MC3. (All SUS's have a working ODC system installed, but the good settings are still a moving target due to commissioning.
I put the ODC master screen on video5. Due to the many not yet commissioned systems, there is still a lot of red there. In the attached plot, there is even more red due to the ongoing IMC length measurement. (stay tuned)
Related previous alogs:
Richard & company are moving this Dust Monitor to HAM4 (since it's an open chamber). Still working on bringing it online (it's been INVALID).
This was Dave & Jim doing work with the hardware watchdogs and they said it shouldn't have affected anybody.
I have been silently checking the signal chain of the REFLAIR and POPAIR RFPDs using the AM laser (a.k.a. PD calibrator) to make sure that they are functional expectedly.
Summary
The RF frequency of the AM modulation was adjusted in each measurement such that the demodulated IF signal was below 50 Hz.
Calibration of the amplitude modulation depth
We recalibrated the AM laser.
The current setting of the laser was changed recently because we opened up the current driver when we thought the laser diode had been dead in the early December. Then the laser head and its current driver were sent to Rich at Caltech for his extensive testing although the laser magically fixed itself and he didn't find anything wrong. So this was the first time for us to use the AM laser which had been fixed. Because of that mysterious event, I wanted to recalibrate the laser. First of all, Yuta and I measured the power to be 2 mW with an Ophir Vega without the attenuation filter. Then we measured the modulation depth for the amplitude modulation by using a Newfocus 1611 as a reference.
The new calibration for the amplitude modulation is:
P_am = 5.13 mW x (P_dc / 1 mW) * (1 V / V_drive)
where P_dc is the laser power at DC and V_drive is the drive voltage when it is driven by a 50 Ohm source. For example, if one puts this laser to a PD which then shows a DC laser power of say 2 mW, the AM coefficient is now 5.13 mW x ( 2 mW / 1 mW) /V_drive = 10.26 mW/V_drive.
REFLAIR_A_RF9 (S1203919)
Remarks:
The signal chain is OK. The PD response is smaller by 15% for some reason.
It seems as if the transimpedance is smaller by 15% than what had been measured at Caltech (LIGO-S1203919). The cable loss from the RFPD to the rack was measured to be 0.47 dB. Be aware that the demod gain is half of the quad I/Q demodulator because this is a dual channel demod (see E1100044). The demod conversion gain is assumed to be 10.9 according to LIGO-F1100004-v4.
REFLAIR_A_RF45 (S1203919)
Remarks:
The signal chain is healthy.
Found cable loss of about 1.5 dB. The measurements excellently agree with the loss-included expectation.
POPAIR_A_RF9 (S1300521)
Remarks:
The signal chain is healthy.
The measurement suggests that there is loss of 1 dB somewhere. I didn't measure the cable loss this time.
POPAIR_A_RF45 (S1300521)
Remarks:
The signal chain is OK. Though loss sounds a bit too high.
The measurement suggests a possible loss of 2.6 dB somewhere. I didn't measure the cable loss.
REFLAIR_B_RF27 (S1200234)
Remarks:
The signal gain is bigger than the expectation by a factor of 2.3.
REFLAIR_B_RF135 (S1200234)
Remarks:
The signal gain is bigger than the expectation by a factor of 1.5
POPAIR_B_RF18 (S1200236)
Remarks:
The signal gain is bigger than the expectation by a factor of 2.3
POPAIR_B_RF90 (S1200236)
Remarks:
The signal gain matches with the expected value, but I don't believe this.
There was a typo:
P_am = 5.13 mW x (P_dc / 1 mW) * (1 V / V_drive)
P_am = 5.13 mW x (P_dc / 1 mW) x (V_drive / 1 V)
For 27MHz and 136.5MHz, the RF gains are +19.8dB and +50.7dB, respectively. S1400079
The response of the BBPD isn't really flat over all frequencies. See D1002969.
The description in D1002969 is for the initial version. (The schematics seems up-to-date.)
The latest version has the rf performance as attached.
This is a follow up of the calibration measurements for REFLAIR_B and POPAIR_B.
I have updated the expected signal gain for these photo detector chains using more realistic gains which Koji gave (see his comments above). Now all the values make sense. Note I did not perform any new measurements.
In the following calculations, the quantity in red represent the updated parameters.
REFLAIR_B_RF27(S1200234)
Remarks:
The signal chain is healthy. There is loss of 0.92 dB somewhere.
REFLAIR_B_RF135(S1200234)
Remarks:
The signal chain is OK. There is loss of 3.9 dB somewhere.
POPAIR_B_RF18 (S1200236)
Remarks:
The signal chain is healthy. The signal was bigger by 9% than the expected.
POPAIR_B_RF90 (S1200236)
Remarks:
The signal chain is healthy. There is loss of 1.2 dB somewhere.
From these measurements, we can use POPAIR to infer the calibration for POP.
I looked at a recent lock acquisition while the interferometer was trying to engage the outer ISS loop. The LSC is relatively stable during this time, and the POP beam diverter is still open.
After undoing whitening gain and digital gain (2 ct/ct for POPAIR9/45, and 32 ct/ct for POP9/45), we find the following TFs:
This implies calibrations of 1.7×106 ct/W for POP9 and 1.8×106 ct/W for POP45.
There's a factor of 4 difference in power between POP and POPAIR (17 mW versus 68 mW with a PSL power of 23 W), so the values I gave above are off by a factor of 4. The demod gains should be 6.4×106 ct/W for POP9 and 7.2×106 ct/W for POP45.
I've attached a plot showing that the IMC has 'not been in a good state' (i.e. red) for the last 21 hours due to the ASC control switch not being on (summary bit 1). I can't see anything obvious in the alogs to determine why this is, anyone know? Thanks
The IMC WFS was off for approximately 8 hrs from Jan-30 01:50 to Jan-30 10:03 in UTC and it have been up in the rest of the time in the past two days. Probably it is because of the master gain which was lowered to 0.25 for some reason yesterday.
I've had a look at the last four hours and the IMC ODC is still red for the same reason (see first attached plot), but the PSL is looking green (2nd plot). I'm currently trying to make the PSL ODC plot for the last day but it's taking some time, so as soon as I have them I'll post it. But if this is to do with the NPRO relaxation oscillation (alog 9674) which Stefan reported turned the PSL ODC green it has not affected the IMC ODC.
(Correction Jan/31: Sheila's made a similar plot for a period (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=9691) where the cavity was locked all the time, and it seems like the plot presented here was sampling only a small edge of the 00 lock range.)
As of now, it seems like both ETM and ITM should be within 0.3 urad pk-pk from the optimal alignment to stay locked on 00 mode.
Plot 1: I took OL and green transmission data for 5 minutes when the cavity was going back and forth between 00 and 01 mostly without losing lock.
In the top panel, the transmission time series is actually the sum of the SHG and the transmission, but it's on 00 mode when it was 1400 counts, 01 when it was between 1000 and 1200 counts.
Plot 2: 2D color map of the above data.
X axis is ITM OL PIT, Y is ETM. I arbitrary set the threshold of 1300 counts for green transmission and subtracted that from the transmission data, and set the color map such that orange-red color is above 0 (i.e. locked to 00 mode), yellow-green is negative (01 mode).
Each cell is normalized by the number of data points in that cell. Blue means there's no data point in that cell.
Sorry that X and Y axis scaling is different. To aid your eyes, I put white and pink diagonal lines showing the direction of common (soft) and differential (hard) misalignment.
Plot 3: 2D plot of where the OLs were staying.
You cannot tell from plot 2 where the OL was staying the most, so here is the number of data points in each cell divided by the total number of data points. Adding everything on this map together you'll get 1.
On average EX=-28.12, IX=-6.28.
Conclusion:
Note that this is dependent on RF sideband frequency.
Note the bogus or inconsistent sign convention of OLs.
For ETM, when the optic tilts down (increase in PIT slider), OL PIT increases, while for ITM OL PIT decreases.
That's the reason why the hard (differential) mode is parallel to the line ETMX=ITMX.
Done
Done
Starting work.