Still no resolutions to the problems with HAM4 ISI transfer functions. I took more DTT tf's today, and they all look a little better than the Matlab tf's. The attached plot shows the 3 vertical pods. The 3 thinner lines are the Matlab tf's (which, though not shown, have been repeated 3 times, all 3 times look pretty similar) and the 3 thicker lines (red, green and blue) are the same sensors with a DTT whitenoise tf.
I should add that the H3 GS-13 looks bad, no matter what. It's just hard to see against the general awfulness of the vertical GS-13's. The first attached image shows the DTT tf in the .5 hz to 5 hz range on the H3 GS13. There should be a taller, pointier peak at ~1hz, which is now kind of squashed. Second image is the power spectra for all GS13s, H3 doesn't look exactly the same as the others at higher frequencies, but not bad. All the other sensors look okay.
T. Sadecki, B. Weaver
The now final (?) monolithic ETMy Quad has had both chains fully suspended. Electronics were hooked up and troubleshot for all stages. OLVs were taken for all stages and MEDM screens updated. We are satisfied with the current alignment enough that we are ready for the first IAS rough alignment look tomorrow. We have some rather gross yaw work to do on the main chain, but would like a preliminary look at the overall position on the table so we can take care of any gross, whole structure moves before continuing with finer alignments.
Monday 27 JAN 2014
-LVEA laser safe
-Diode room dust monitor needs investigating; false readings all day for 0.03um
920 -1100, 1120 -1200 Mitchell working on ACB in LVEA test stand area
1000- Betsy and Travis at EY
1120 Sheila, Alexa to EX
1155-1215 Richard to EY working with Betsy and Travis
1225 Kiwamu is transitioning the H1 PSL from commissioning to science mode.
1237 Kiwamu working ISCT1 inside the NHZ (LVEA remains laser safe)
1300-1600 Mitch and Scott (Apollo) working on ACB
1404 Hugh to LVEA for ITMX-CPS investigation
1454 - 1547 - Thomas and Aidan to LVEA for parts search
1457-1505 - Corey to LVEA ACB area for brief period
1533 - Sheila and Alexa to EX to toggle noise eater, etc
16:23 -- Alexa and Sheila returned from EX
St1 Before |
St1 After | St2 Before | St2 After |
X 6959nm | 41458 | -9643 | 2702 |
Y 8935nm | 9155 | 253 | 478 |
Z -22709nm | -22563 | -10660 | -10743 |
Rx 39467nrad | 39195 | 10198 | 10083 |
Ry -7904nrad | -5523 | 5845 | 7101 |
Rz 19481nrad | 20104 | 4123 | 4482 |
To help make the isolation of the ETMx less troublesome, this morning I reset the ETMX ISI position loop alignment targets to the tilted location. This way the ISI will not drive back to the untilted position putting larger outputs on the ISI and reliably running up the trilliums, tripping the WDs. If the local to cart matrices are somewhat correct, the position/alignment change is small & so far OK. Adding theStage1 & Stage2 cart position changes gives a 46um position shift of the ISI Optical Table with <4urads on the Ry. All the other dofs combine to 1um or 1 urad or less. Sheila did spend some time realigning but the ITM wasn't in nominal isolation either so that might have been hampering efforts there. Once the ITM was Isolated properly, the green locked much better.
So above is the before and afters.
From ~1400 to 1440--moving ladders, wiggling cables, general trouble making.
Betsy swapped all the BOSEMs and AOSEMs on ETMy and left them retracted for open light measurements. The resulting values with gains and offsets are:
prettyOSEMgains('H1','ETMY')
M0F1 30218 0.993 -15109
M0F2 29922 1.003 -14961
M0F3 29480 1.018 -14740
M0LF 31134 0.964 -15567
M0RT 30792 0.974 -15396
M0SD 31110 0.964 -15555
R0F1 31059 0.966 -15529
R0F2 30058 0.998 -15029
R0F3 29953 1.002 -14977
R0LF 30794 0.974 -15397
R0RT 31050 0.966 -15525
R0SD 30282 0.991 -15141
L1UL 29222 1.027 -14611
L1LL 30117 0.996 -15059
L1UR 25735 1.166 -12868
L1LR 22721 1.320 -11361
L2UL 21334 1.406 -10667
L2LL 24582 1.220 -12291
L2UR 21223 1.414 -10612
L2LR 20325 1.476 -10163
I entered the gains, checked that all the outputs were right (should be 15000*0.023333=350) and redid the safe.snap.
Corresponding serial numbers as follows:
M0F1 094
M0F2 588
M0F3 266
M0LF 058
M0RT 041
M0SD 095
R0F1 050
R0F2 186
R0F3 422
R0LF 180
R0RT 056
R0SD 083
L1UL 477
L1LL 493
L1UR 485
L1LR 022
L2UL 473
L2LL 428
L2UR 332
L2LR 321
(Sheila, Alexa)
We made a few measurements at EX under the following configuration:
PLL Servo Board
PDH Servo Board
(Sheila, Alexa)
With the above servo board configurations and the HEPA fans off we measured the shot and dark noise again (in a previous alog I mentioned we didnt take the correct frequency ranges for the dark noise which is why we took it again). I have attached a plot of the power spectrum of the PDH error signal, the shot noise w. the fans off and on, and the dark noise with the fan off. All these spectra's were measured out of IMON. There is a considerable difference at low frequency of the shot noise with the fan off and on. The beam size might be too large on the BBPD? Another possibilty is fringe wraping. It is also suspicious that the shot noise is not flat and seems to drift if there is movement around the table...Are we really measuring pure shot noise??
Keita, Alexa, Sheila
Some of the excess noise on the ALS end refl PD is due to fringe wrapping. We looked at spectra of REFL B from the control room while the ITM was misalinged, ( so we should have been looking at the RIN on our green beam returning from the chamber). This is motivated by seeing that the niose at Imon with the ITM mislainged is accoustic, and seems to have a fringe wrapping shelf when the HEPA filters are on. By driving the ETM longitudnaly, we see a fringe wrapping peak on the DC PD. We tried misalinging the reaction chain, and don't see any difference. At some point we need to go bow on optics on the table and try to find the scattering path.
Kiwamu is transitioning the H1 PSL from commissioning to science mode.
The transition is done at 12:27. Now it is is the science mode.
Stefan logged yesterday that he was unable to get the ISI into operation state without tripping.
First time I did this today, I did Stage2 one dof at a time and had no problem. The ETMx blends at 750mHz on both Stages all dofs. Pretty sure this isn't as well as can be squeaked from the ISI. I took this opportunity to take spectra as I brought other blend filters to bare. Attached is the Stage2 GS-13 X In1 with various filter combos.
Ref3, red: As found 750mHz all dofs both Stages.
Ref7, blue: T(rilliums)750mHz all dofs Stage1, 750mHz Stage2---Maybe a factor of 2 improvement below 0.5Hz
Ref11, green: T250mHz all dofs Stage1, 750mHz Stage2---Much better above 0.5Hz but worse between ~.1 and 0.5Hz.
Ref15,thick brown: T100mHz.44NO Stage1 X Y & Rz else T250s, Stage2 750mHz---Big win below 1Hz to about 0.1Hz.
Ref18, pink:T100mHz.44NO Stage1 X Y & Rz else T250s, Stage2 250mHz---Better ~x2 or 4 above 0.8 but worse below especially 0.2Hz.
Ref23, aqua: T100mHz.44NO Stage1 X Y & Rz plus Notch X Y else T250s, Stage2 750mHz---Lost improvement above 0.8Hz but improves over brown between 0.8 and 0.2Hz.
Ref28, black: T100mHz.44NO Stage1 X Y & Rz plus Notch X Y else T750s, Stage2 750mHz---Some improvements below 0.2Hz and some above 1Hz.
I'm not sure this is exactly the best configuration such as maybe the T250 is better than the T750 for both Stages. Certainly incrementally there were tradeoffs. We need to combine the benefits in a better way.
So I have the ISI in this final arrangement. I took the ISI out of Isolation (it triggered then!) I then reset the ISI targets to the unisolated postions. The ISI was then able to go into Isolation with 'One Button'. Green team is evaluating if they can live with this minor alignment change.
Sheila to EX
Sheila and Alexa return 1215
Here is the list of commissioning task for the next 7-14 days:
Green team:
Red team:
TMS:
ISCTEY (starting next week):
Stefan , Kiwamu
The following TFs are now being measured.
All of them runs on opsws5. They started at 10:20 am local.
Summary: A camera with a flash right beside the lens was used to look for potential backscatter noise sites. Most baffles look good, and there was only one strong backscatter site within the arm cavities: the large gate valve seats by the ITMs. There were also a few retro-reflective sites outside the arm cavities: for the compensation plates, the HAM3 ISI, and the supports for the elliptical baffles and the TCS mirror 2 were quite retro-reflective. As an aside, I also noticed that the TCS alignment port was partially blocked by a nozzle. In the lower sensitivity input arm, the MCA1 and 2 baffles had bright reflections, as did the HAM3 ISI.
ITMX. In Figure 1, the top image shows approximately what the beam spot on ITMX “sees”. I use these photos to look for potential scattering problems; the camera is placed very close to the center of the optic where the beam spot will be. The flash mimics the wide-angle scattering of interferometer light from the beam spot and lights up regions that will retro-reflect interferometer light back to the beam spot, where it can recombine with the main beam to produce scattering noise. Of course the technique is limited by different angular distributions of scattered IFO light and light from the flash, as well as by color differences, but it gives a rough idea and also shows glints, which are hard to predict. The camera is about 10 cm in front of ITMX, so it sees a little more through the hole in the baffle than the beam spot on the optic will. Nevertheless I expect that the baffle will not block the view of the gate valve seat, the shiny reflecting arc at the right of the view through the baffle aperture. This is the only site found in the high-intensity arm cavities. I think the first step is to double check that the valve seats are actually visible from the ITM beam spots.
The lower image in Figure 1 is the view from the X-manifold spool of ITMX, with the camera located in the IFO beam path (the camera is in the beam path when the flash retro-reflects off of the optic). The arm cavity baffle looks good: the small reflecting rectangle at the bottom center of the baffle is the connection for the safety lanyard and is actually under the bottom side of the baffle and thus should be outside the clear aperture from ETMX. It looks like it is inside the baffle because the far part of the baffle box sags down a little and we are seeing a little of the underside of the baffle.
ITMX RM (compensation plate).Figure 2 shows the view from the beam spot on the compensation plate (the reaction mass). There are strong retro-reflections from the HAM3 ISI. The glints are off of stage 1 of the ISI so there is some isolation of the reflecting surface. There are also glints off of the elliptical baffle, which has one stage of passive isolation like the ISI, and also off of the mount for TCS mirror 2, which is not seismically isolated.
The view in TCS mirror 2 (the copper-colored circle), shows that the TCS alignment port is partially blocked by the nozzle of the port. Thus there is not a complete view of the compensation plate from all parts of the port, which may make alignment more difficult. We also need to watch for backscatter sources on the TCS table.
BS.The first page of views from the beam splitter, Figure 3, show bright reflections from the frames of the ITMs. I have reported this before (Link) and Mike Smith’s subsequent calculations suggested this would not be a problem, though I don’t think that specular reflection was considered.
Another potential source of retro-reflections to the spot on the BS are the reduction flanges for the tubes holding the gate valves between BSC1 & 8 and BSC3 & 7.
The second page of BS views shows the view back out the input arm. As with the view from the ITMX compensation plate, there are particularly bright glints from the HAM3 ISI (notice the beams of light shining on the floor of the tube), including one from the table edge, just a few centimeters below the path to/from PR3.
PR3. Figure 4; looking toward the BS, there are fairly strong reflections from the electro-polished aluminum baffle, MCA1, as well as from masses on the HAM3 tabletop.
PR2.The worst reflection in Figure 5 seems to be from the edge of the MCA2 baffle.
PRM. Figure 6 shows a fairly bright reflection from the MCA1 baffle, a wide-angle reflection from the circular baffle, and reflections from objects on the HAM3 table.
MC1 & 3. Figure 7 shows that the MCA1 baffle again seems quite reflective and that there are also reflections from objects on the HAM3 table.
MC2. The reflection from the MCA1 electro-polished surface in Figure 8 does not appear as bad from this side, possibly because of the greater distance.
When I came in today, the arm was unlocked. For some reason the tidal relieve to HPI ETMX was still integrating. When I cleared its history, all watchdogs on ETMX, its ISI and HPI tripped. HPI and SUS were easy to recover. When trying to re-engage state 1 for ISI stage 1 though, the ISI consistently tripped when trying to move it to the final alignment biases.
Daniel, Stefan With the PLL running reliably, I successfully handed off the mode cleaner control to the ALS COMM signal. - I copied all old HIFOY scripts in /als/h1/scripts/ into /als/h1/scripts/HIFOY_safe - I updated the CARM_down script and started working on the CARM_handoff script. - The initial part of the CARM_handoff script ran fine, and I was able to turn off the MC feed-back to MC2. - The CARM loop before engaging the AO path has a UGF of 20Hz - rather low, and the phase margin is questionable (0deg margin at 100Hz) - Then I noticed that CARM FM6 was an outdated filter, which needed to be updated for the new PLL. I added FM10 "shape", which is a z1.4:p1000 with the gain matching the previous filters FM6 and FM7. This effectively got rid of a p40:z7k - a filter that we now have in analog in the PLL. With that I increased the UGF to 90Hz. - I tried ramping up the AO path, but always lost it before I was able to measure the OLG. More work to be done there. Updates to the CARM_down and CARM_handoff scripts: - The MC2_M3_ISCINF_L FM6 used to be a z150:p500 lead filter for extending the actuation lead to 500Hz. I restored it. - We are now using H1:SUS-MC2_M3_LOCK_L FM9 instead of FM8. I adapted the scripts - Use CARM FM10 instead of FM6 and FM7. - I took all ARM HEPI feed-back engaging out of the script - that is handled by the ALS at the end station. - I took any reference to RFAIR_9 out of the down script.
Great. Now you can try the power-recycled X-arm.