Displaying reports 74581-74600 of 83792.Go to page Start 3726 3727 3728 3729 3730 3731 3732 3733 3734 End
Reports until 12:32, Thursday 23 January 2014
H1 General
patrick.thomas@LIGO.ORG - posted 12:32, Thursday 23 January 2014 - last comment - 15:58, Thursday 23 January 2014(9495)
Mitchel to west bay to work on ACB


			
			
Comments related to this report
patrick.thomas@LIGO.ORG - 15:58, Thursday 23 January 2014 (9506)
Done
H1 SEI (CDS, ISC, SUS, SYS)
jeffrey.kissel@LIGO.ORG - posted 12:30, Thursday 23 January 2014 (9494)
BSC-ISI CPS Frequency Comb Strikes Again?
The Green Team has been been complaining of excess ~0.5 [Hz] pitch motion of the QUADs for the past few days. We had first chalked this up to the fact that the ISI were not yet at full performance. However, after Sebastien and company got both chambers up to a reasonable performance, and even going so far as to tailor the blend filters to get as much performance at 0.4-0.5 [Hz], the cavity pitch motion was deemed still too high. 

I finally just took a comparison ASD of both chambers and their optics' optical levers, and discovered a giant spike in ISI ITMX at 0.5 [Hz], and a subsequent comb cascading up to higher frequency. The 0.5 [Hz] (and its harmonics -- note that 0.5 [Hz] is *not* the fundamental) spike appears obviously in both the ISI ITMX ST2 GS-13s. Only the 0.5 [Hz] harmomic shows up in the SUS ITMX optical levers -- but it happens to perfectly coincide with the QUAD's first "pitch" mode at 0.5 [Hz]. Awesome.

Given that we've seen similar combs at LLO (and I believe DetChar has confirmed their presence at LHO), which were confirmed to be a result of the BSC-ISI's capacitive position sensor master oscillators beating against each other, I immediately suspect that this is our problem.

In order to confirm, we're going to run the corner BSCs in various configurations of OFF to assess whether the comb disappears, if not at least changes. 

Note: since this is systematic problem seen at both sites, a permanent fix to the problem is in the works (i.e. syncing all chambers to a single, GPS synced, master oscillator). It's just not ready yet.
Non-image files attached to this report
H1 ISC (ISC)
corey.gray@LIGO.ORG - posted 11:59, Thursday 23 January 2014 (9491)
EY TMS Cable Notes

With the recent cabling work at EY, wanted to record notes of where we're at and inventory. 

Problem Cable

Performed elec short tests on in-vac cables, and found an issue with one:  D1000223 S1104078's pin #1 & #23 (may be able to use a D1000225 in its place)

Screws On Cable Feedthru Connectors Confirmed To Be "Screwed In" All The Way

It has been brought up that a possible source of grounding are the sets of screws on the connector which attaches to a Chamber feedthru.  Through handling sometimes these screws can unscrew, and even a little bit will have them protrude out and short to the Chamber.  I went ahead and applied a torque to all of these connectors for TMS cables.  (attaching a photo of these screws on our 25-pin connetors....with head on shot and profile)

Missing Screws For Attaching To Feedthru

In the table below there is a note of cables which are missing screws which attach cables to the in-vac side of the feed-thru.

Cables Screwed Together & Unfortunate Cable Bracket Attachement

Since we've tested all these cables, I went ahead and screwed all cables together (noting we'll have to unscrew the bad D1000223 at some point).  In our Cable Layout drawing (D1200111), the D1000921 cables for the Picomotor & Beam Diverter are said to be OK with connectors "with or w/o metal ears".  If you do not have metal ears, you sort of make life hard for you in that if you have to look at a cable you can't simply unscrew it from the cable bracket.  You remove access to the Cable set screw, so you have to remove both cables and then disconnect cables---which is a pain.

ICS & Inventory Of Cables

I have made an Assembly Load for the TMS EY Cables (ASSY-D1200111).

In-Air Cable
 
Chamber
feed-thru
Seismically-Responsible Cables Cable Bracket In-Vac Cable Cable Bracket
on TMS
In-Vac Component
: Not sure of name : E6-7C1 D1000225 s/n S1104782
*no feedthru screws?
CB-5 , 1st Floor D1000234 s/n96-911 no CB OSEMS:  Face1, Face2, Face3, Left
: Not sure of name : E6-7C2 D1000225 s/n S1104778
*no feedthru screws?
CB-5 , 2nd Floor D1000234 s/n96-901 no CB OSEMS:  Right, Side, ---, ---
: Not sure of name : F2-2C1 D1000924 s/n S1104106
*no feedthru screws?
CB2 , 1st Floor (was 3rd Floor of BSC6 CB) D1000568 s/nS1104468 CB-primary, 1st floor Green QPD (D1000231 s/n C2Q001 aka S1400085)
: Not sure of name : F2-1C2 D1000924 s/n S1104469 CB2 , 2nd Floor (was 2nd Floor of BSC6 CB) D1000568 s/nS1104109 CB-primary, 2nd floor Red QPD (D1000231 s/n C4Q001 aka D1400087)
: Not sure of name : F2-1C1 D1000223 s/n S1104077
*no feedthru screws?
CB1 , 1st Floor (was 1st Floor of BSC6 CB) D1000921 s/nS1104116 CB-entry, 2nd floor Picomotors (D1000238 s/n1104586)
: Not sure of name : F2-2C2

D1000223 s/n S1104078--possible short w/ pin 1 & 23

CB1 , 2nd Floor (was 4th Floor of BSC6 CB) D1000921 s/nS1104114 CB-entry, 1st floor Beam Diverter (D1000237 s/nS1104289)
Images attached to this report
H1 ISC
sheila.dwyer@LIGO.ORG - posted 11:18, Thursday 23 January 2014 (9466)
X arm green modulation depth

Daniel, Alexa, Sheila

We have tried to measure the green modulation depth of the X arm by moving the modulation frequency (24.410610MHz) so that the sidebands are antiresonant in the cavity, and then locking the cavity on the carrier (+polarity on servo board) then the sidebands.  We used H1:ALS-C_COMM_A_LF_OUTPUT to measure the transmitted power.

  carrier sideband
no attenuator 491.3 184
6dB 620.34 81.67
12dB 664 hard to measure

I have changed the offset to -66 counts, so that with no light on the PD we get zero counts (-0.26 counts)  (someone had changed this from the -61 I originally set to -100 sometime last week, I am not sure why).  With the single shot beam we get 46.06 counts.

Below are plots ot the transmitted powers (normalized) when locked on the carrier and sideband, and the second plot is their ratio.  I used the ratio to do a simple fit for the modulation depth that we started with, and got 0.80  Now the modulation frequency is restored to the nominal 24.407079MHz and the 12dB attenuator is in place so our modulation depth should be 0.2.

Also, the ratio of the transmitted power on resonance/ single shot beam should be:

1/(1-sqrt(R1*R2))^2=22.6 for a perfectly mode matched carrier with R1=0.63, R2=0.99, taking into account the 0.2 modulation depth we would expect 22.17.  Since we curently measure a ratio of 14.4 between the single shot and locked trans power, which is 65% of what we would expect.  This means we have about 45% of our power in higher order modes. 

Daneil's alog about the RF powers is 9484

Images attached to this report
Non-image files attached to this report
H1 DAQ (CDS)
james.batch@LIGO.ORG - posted 11:08, Thursday 23 January 2014 (9488)
Writing longer frames on h1fw0
Writing of 64 second full frames has started on h1fw0.
H1 CDS
patrick.thomas@LIGO.ORG - posted 11:02, Thursday 23 January 2014 (9487)
Jim B. restarting fw0 and nds0


			
			
H1 General
patrick.thomas@LIGO.ORG - posted 10:49, Thursday 23 January 2014 (9485)
Jaclyn checking VCO signal into Beckhoff at end Y


			
			
H1 ISC
daniel.sigg@LIGO.ORG - posted 10:45, Thursday 23 January 2014 (9484)
ALS EX modulation depth

We were measuring the green modulation depth in EX by looking at the transmitted power when locking the cavity on carrier or sidebands. To make the interpretation easier we first used the ifr to move the sidebands to exactly mid-fringe. With a length of 3994.472 m and a multiple of 650.5 we get a frequency of 24.410610 MHz. The ifr was set to +11 dBm for the measurement.

Here we report the measured RF levels. We used an Agilent N1914A with probe E4412A (S/N MY50270005). A zero check and calibration was performed before the measurement. An attenuator was added at the RF patch panel in the field rack between the balun and the cable driving the Pockels cell. As a comparison we list the attenuation values measured with the RF source at a frequency of 24.407079 MHz as well. We left the 12 dB attenuator installed and switched back to the RF source.

Attenuator ifr +11dBm
24.410610 MHz
RF source
24.407079 MHz
none 13.69 dBm 13.49 dBm
6 dB 7.74 dBm 7.52 dBm
12 dB 1.72 dBm 1.50 dBm
LHO General
patrick.thomas@LIGO.ORG - posted 10:29, Thursday 23 January 2014 - last comment - 11:36, Thursday 23 January 2014(9483)
Gerardo soft closing GV7, Gerardo and Stefan installing beam block on top of HAM2


			
			
Comments related to this report
patrick.thomas@LIGO.ORG - 11:17, Thursday 23 January 2014 (9489)
Done with install, Gerardo opening GV7
patrick.thomas@LIGO.ORG - 11:36, Thursday 23 January 2014 (9490)
GV7 open
LHO General
patrick.thomas@LIGO.ORG - posted 10:19, Thursday 23 January 2014 - last comment - 15:30, Thursday 23 January 2014(9482)
Keita transitioning end Y to laser HAZARD


			
			
Comments related to this report
patrick.thomas@LIGO.ORG - 11:47, Thursday 23 January 2014 (9492)
For WP 4409
patrick.thomas@LIGO.ORG - 15:30, Thursday 23 January 2014 (9504)
Work done, VEA transitioned back to laser SAFE.
H1 CDS (DAQ)
james.batch@LIGO.ORG - posted 10:17, Thursday 23 January 2014 - last comment - 11:54, Thursday 23 January 2014(9480)
Implement long frame writing
The h1nds0 daqd will be shut down for up to 10 minutes to build RCG 2.8.2 versions of daqd to support longer frames.  Longer frame collection should be implemented about 11:00 PST for h1fw0.  Following about an hour of testing, and assuming testing is successful, longer frame writing will be implemented on h1fw1 about 12:00 PST.  
Comments related to this report
james.batch@LIGO.ORG - 11:54, Thursday 23 January 2014 (9493)
Due to frame writer 0 crashing 35 minutes after starting while writing longer frames, we are delaying the change to framewriter 1 until 13:00 PST at the earliest.
H1 SEI
sebastien.biscans@LIGO.ORG - posted 10:17, Thursday 23 January 2014 (9479)
Model modification on ISI-HAM4 & ISI-HAM5 top-level

HAM4 and HAM5 are different compared to HAM236, because of the ST0-L4Cs. Consequently, we need to use the proper blocks for the BIO cards.

The changes have been made (see pic attached), the models recompiled and restarted. Everything went fine.

Images attached to this report
LHO General
patrick.thomas@LIGO.ORG - posted 10:15, Thursday 23 January 2014 - last comment - 10:54, Thursday 23 January 2014(9478)
Justin to end X


			
			
Comments related to this report
patrick.thomas@LIGO.ORG - 10:54, Thursday 23 January 2014 (9486)
Done
LHO General
patrick.thomas@LIGO.ORG - posted 10:10, Thursday 23 January 2014 - last comment - 10:18, Thursday 23 January 2014(9477)
Gerardo opening rollup door to LVEA


			
			
Comments related to this report
patrick.thomas@LIGO.ORG - 10:18, Thursday 23 January 2014 (9481)
Done
Displaying reports 74581-74600 of 83792.Go to page Start 3726 3727 3728 3729 3730 3731 3732 3733 3734 End