james.batch@LIGO.ORG - posted 14:34, Friday 13 December 2013 (8950)
The default nds server is now h1nds1
I have changed the default nds server back to h1nds1.  This will be effective with new shells or processes.  
H1 ISC
alexan.staley@LIGO.ORG - posted 14:32, Friday 13 December 2013 (8949)
ALS Fiber Distribution Chassis S1202528

(Alexa, Daniel)

With the gain setting at 0dB for the external PD:

Transimpedance = 2kOhm

Ext PD Monitor = 1.28V

Beckhoff readback = 5.17V (there is a 4x gain between the monitor and controls output)

If the external PD is a silicon diode (responsitivty = .25A/W), this implies that we have 2.6mW on the diode. Alternatively, if we believe that we have ~1mW on the diode as referenced in alog, then the responsitivty of the diode is closer to .64A/W, which resembles that of an in-gas diode. The former is more likely.  **** Next time someone is in the PSL, can you please measure the power into the diode?

 

Meanwhile, I also confirmed that the gain setting on the ALS fiber distribution chassis is as expected. At the ext. PD input, I read -5.0V. Placing 100kOhm resistor at the input provides 50uA of current. Thus, one expects to read V=(50uA)(2kOhm) = 100mW at the ext PD monitor, which I verified. This is consistent with the test report for this chassis, which confirmed 3.13V at 30dB with 50uA. The factor for 30dB gives a factor of 31.26 in V.

 

Conclusion: ALS fiber distribution is okay. Diode is different than expected or we have more power ...

H1 ISC (ISC)
sheila.dwyer@LIGO.ORG - posted 14:16, Friday 13 December 2013 (8944)
End X PZT response to TMS motion

Starting at around noon, the QPD servos at end Y are running, we will leave this alone for about a day, then turn off a cleanroom to see if we have a repeat of the alingment drifts seen in end Y.

Before starting I moved the TMS  with the QPD servos locked, here are the responses in the PZT sensors and outputs:

Moving the TMS in yaw moves the PZT pit, and vice versa

  sensor mV/ urad TMS yaw output counts/urad TMS motion yaw  sensor mV/urad TMS pit output counts /urad TMS pit
PZT1 pit -2 -6.8 0.2 1.3
PZT2 pit 0.1 4.12 0.56 2.07
PZT1 yaw -22.9 7.26 0.594 1.39
PZT2 yaw -1.59 5.43 -.79 -2.1

for reference the times are:

nominal TMS positions PIT -3urad, Yaw -297urad UTC 18:45:35

PIT -3 urad YAW -197 urad  UTC 18:50

PIT -103 urad YAW -297 UTC 18:54:51

Images attached to this report
H1 AOS
alexan.staley@LIGO.ORG - posted 14:12, Friday 13 December 2013 (8947)
Script bug for HPI BS

transissioning to level 2 I get

   Isolating 4 dofs: X Y RZ HP
   Using filter module(s): H1:HPI-BS_ISO_X H1:HPI-BS_ISO_Y H1:HPI-BS_ISO_RZ H1:HPI-BS_ISO_HP
   Isolation level: 2
   Starting gain ramp...
Time::HiRes::sleep(-0.451209): negative time not invented yet at /opt/rtcds/userapps/release//hpi/common/scripts/HPItool2.pl line 516.
 

And then the script stops, without runing on X, Y, RZ and HP.

Additionally, I noticed that a lot of these HPI turn-on scripts don't actually move to the set target position, but instead just throw a warning, asking for a manual move.  This seems like an alignment time bomb that will get us sooner or later.

 

 

 

H1 CDS (DAQ, DCS)
david.barker@LIGO.ORG - posted 12:41, Friday 13 December 2013 (8946)
h1fw1 back after reboot of Solaris QFS file server

Initial restarts of daqd on h1fw1 did not fix the problem. The framewriter ran for about 10 mins and then its uptime counter stopped incrementing.

I power cycled h1ldasgw1, and had to manually remount the QFS file system and NFS export it. I then restarted h1fw1 and it has been running for  about 20 mins now.

H1 ISC
christopher.wipf@LIGO.ORG - posted 11:58, Friday 13 December 2013 (8945)
ASC front end model updated

(Keiko, Chris)

The h1asc front end is now running with the latest version of the model. The previous version had been copied from L1 by Kiwamu back in May. Since then an Initial ALignment (IAL) system for the arm cavities (a dither scheme using the ALS green light) was added to h1asc by Stefan, but no other significant changes were made. Meanwhile the ASC model has been actively developed at LLO. Our goals were to update this model to track all the LLO changes, to merge in the IAL subsystem, and to start using library parts (which make it easier to verify that the LLO and LHO models are in fact the same).

Here's what's new:

H1 SYS (IOO, ISC, SUS)
jameson.rollins@LIGO.ORG - posted 11:33, Friday 13 December 2013 - last comment - 15:49, Friday 13 December 2013(8942)
Guardian IMC autolocker now running

The first new Guardian daemons are now running!

The IMC is now fully under guardian control.  The guardian components are (given by their full guardian identifier):

The paths listed are the fulls paths to the guardian system description modules that are being executed for each guardian process.  Given a system identifier FOO, guardian looks for a python module named FOO.py in the common/guardian subdirectories of the various relevant subsystems in the userapps path.

The IFO_IMC manager is the top level control.  It instructs SUS_MC2 and ISC_IMC to go to their ACQUIRE states if the IMC is not locked, and then transitions SUS_MC2 to turn on boost filters after ISC_IMC guardian reports that the IMC lock has been achieved.  If ISC_IMC looses lock, IFO_IMC sees this and moves SUS_MC2 and ISC_IMC back to their ACQUIRE states.

MEDM interface

The primary interface for controlling the processes is their medm screens, which are accessible from the "GRD" menu in the sitemap:

The REQUEST button selects the requested state of the system.  The MODE button selects the operational mode of the guardian process.  EXEC indicates that it is executing guardian code.  The STATUS reflects the guardian process status.  In this case, the IFO_IMC guardian is executing the RUN method of the LOCKED state.

The buttons at the upper right allow for displaying the log of the daemon, displaying the current system graph, and editing the underlying system module code.

guardian daemon controls

All three guardian processes are running under the new guardian control infrastructure on h1guardian0:

controls@operator1:~ 0$ guardctrl list
IFO_IMC * run: IFO_IMC: (pid 8739) 7805s, want down; run: log: (pid 13060) 78412s
ISC_IMC * run: ISC_IMC: (pid 6400) 9332s, want down; run: log: (pid 1346) 66758s
SUS_MC2 * run: SUS_MC2: (pid 2208) 66732s, want down; run: log: (pid 2050) 66737s
SUS_SR2 * run: SUS_SR2: (pid 32679) 96154s, want down; run: log: (pid 23722) 230173s
controls@operator1:~ 0$ 

The guardctrl command line interface can be used to start/stop/restart the daemon process, as well as create new ones as needed.

TODO

Some immediate todo task for the next week:

There are also an ever-growing list of guardian bugs (not a bad thing at this point, since it means the system is actually being used now) that I will start filling in in the new guardian bugzilla (thanks Jonathan).

Images attached to this report
Comments related to this report
jameson.rollins@LIGO.ORG - 15:49, Friday 13 December 2013 (8951)CDS

All the locking code guardian modules has been committed to the userapps repo at:

  • userapps:  r6650 
/opt/rtcds/userapps/release/ioo/common/guardian/ISC_IMC.py
/opt/rtcds/userapps/release/sus/common/guardian/SUS_MC2.py
/opt/rtcds/userapps/release/sus/common/guardian/SUS.py
/opt/rtcds/userapps/release/sys/common/guardian/IFO_IMC.py

The guardian and cdsutils versions are:

  • cdsutils: r127
  • guardian: r479

cdsutils and guardian are installed in their proper places in /ligo/apps/linux-x86_64/

H1 IOO (IOO)
cheryl.vorvick@LIGO.ORG - posted 11:06, Friday 13 December 2013 - last comment - 19:39, Sunday 26 March 2017(8943)
IMC beam spot centering measurement - all beam spots within 2.3mm of center.

I calculated the spot centering on the IMC mirrors, with help from Kiwamu and Jax (alog #6676).

All beam spots are with 2.3mm on center on all 3 mirrors, which is slightly better than what Jax measured in June.

 

Beam Centering Measurements: 12/12 P2L/Y2L gain smallest peak amplitude EL/EP alpha alpha in % distance from center of mirror, in mm
      0.25/5.2382 (EL/EP)*P2L/Y2L gain   % * 0.375
             
MC1            
P -0.90 0.05 0.048 -0.043 -4.3 -1.6
Y 0.60 0.30 0.048 0.029 2.9 1.1
             
MC2            
P -1.30 0.05 0.048 -0.062 -6.2 -2.3
Y -1.00 0.50 0.048 -0.048 -4.8 -1.8
             
MC3            
P 0.85 0.12 0.048 0.041 4.1 1.5
Y -1.10 0.05 0.048 -0.052 -5.2 -2.0
Comments related to this report
cheryl.vorvick@LIGO.ORG - 19:39, Sunday 26 March 2017 (35095)

I'm revisiting beam centering measurements, and have recalculated the beam centering logged here, based on my enhanced knowledge, and with both the procedure used in 2013 and in 2017.

My sign conventions in this 2013 alog are incorrect, and I'm now correcting them, and posting values using both the 2013 and 2017 procedure.

In 2013, though it's not stated, I used the Eul2OSEM values that match the UR OSEM, which matches the procedure I used in 2017, alog 34973

Updated Results (details in attached pdf):

Non-image files attached to this comment
H1 DAQ (CDS)
james.batch@LIGO.ORG - posted 08:07, Friday 13 December 2013 (8940)
Change of NDS server to h1nds0 (temporary)
I have changed the default nds server to h1nds0 until we can sort out what is wrong with h1fw1.  Note that this only affects newly opened shells and newly started programs.

Also note that h1nds0 does not offer the same set of channels, it is set up to write science frames instead of commissioning frames.

I will change the default back to h1nds1 as soon as h1fw1 is functional.
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 21:28, Thursday 12 December 2013 - last comment - 11:06, Friday 13 December 2013(8937)
vertex interferometer alignment: PR2 steered, a spot found on BS camera, ongoing

[Filiberto, Keita, ChrisW, Arnaud, Kiwamu]

We steered PR2 to get the beam on the center of PR3. Now we can see a beam hitting somewhere in the BS chamber. This beam spot appears to be the dark port beam.

Then one time, we were able to make the PR-Y cavity flash, but after some time of damping works, we lost it and never found it again. Tomorrow we are going to try a more systematic alignment. The attached is the alignment from the time when we saw fringes.

 

HAM3 spool Camera:

We first tried GigE cameras to look at the front surface of PR3 from the HAM3 spool position. This didn't work. The cameras were not so sensitive enough to resolve weak light. Then we removed a GigE and installed an analog camera on the 0 o'clock position on the spool. This then allowed us to see a spot on the PR3 cage as well as the swiss cheese baffle. We locally hooked up a monitor on the floor.

Later, Keita told me that it could be much more sensitive if we changed a gain in medm.

PR2 steering:

In the process of steering PR2, I found that PR2 kept tripping its watchdogs. It was due to the OSEMS hitting the open light threshold. I had to offload the alignment biases to IM3 and IM4 to mitigate this issue. The offloading worked well -- I was able to steer the beam onto PR3 while keeping the beam on both IM4 trans and POP_A. Unfortunately POP_B lost the beam but the recovery should be easy as we still have the beam on POP_A. The centering on PR3 was good in yaw and OK in pitch. I used scattered light from the lower part of the cage for aligning yaw and used the baffle to align the beam height. Also, PRM was aligned to obtain the retro-refletion back to PSL.

We got PR-Y flashing, but the spot seemed moving a lot:

Then we started steering PR3 by looking at the BS camera. The idea is to align the PR-Y cavity without wildly touching BS and ITMY which are thought to be a reference. Since I couldn't find a power cable for the illuminator on BS, the BS camera view has been simply dark. At some point, we found a moving spot out of this dark monitor. Because this spot responds to the angle of ITMY and BS, we believe this was the dark port beam which is now hitting somewhere in the BS chamber. By a yaw scan in PR3, we started seeing fringes in POP_A_SUM and REDL_A_LF. POP_A_SUM could go up to 10 counts when flashing while it is 5 counts nominally. However, the fringes didn't look stable -- sometime it didn't flash at all and sometime it flashed a lot. We suspected that something was moving and indeed PR3 was moving the spot position a lot. This was visible in the BS camera when we introduced an intentional offset in PR3 in yaw. The spot was moving vertically at about 1 Hz with an amplitude of approximately three beam sizes. Chris investigated why PR3 wobbled so much and eventually found that turning the HAM2 ISI on at level 3 made it quitter. After the investigating, we restored PR3 back. However we became unable to find the fringe any more. It is unclear what happened.

Images attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 22:02, Thursday 12 December 2013 (8938)

We also fully turned back on ISI and HEPI of ITMY and BS, which notably reduced the beam motion seen on the BS camera . The configuration of both systems was set to the one described by Hugo in his alog 8860. We increased the threshold of the T240 watchdogs for ITMY ISI to avoid the system to trip on the T240 when starting the loops. For the BeamSplitter ISI, we had to exclude the T240 out of the loop to turn it on. We then lowered the blends and included the T240. The isolation was turned on independentely for both stages of each ISI.

kiwamu.izumi@LIGO.ORG - 11:06, Friday 13 December 2013 (8941)

This is a picture of the HAM3 spool camera that looks at the front surface of PR3 through the swiss cheese baffle.

The green circle in the picture indicates the scattered light off of the upper side of the PR3 cage. This is the original position before we started aligning anything.

Images attached to this comment
LHO General
cheryl.vorvick@LIGO.ORG - posted 16:40, Thursday 12 December 2013 - last comment - 07:36, Friday 13 December 2013(8933)
OPS Summary:
- Corey to EY
- Gerardo to BSC2 for camera work
- Keita/Filiberto input spool camera 
- Justin, card access changed - doors now lock
- Richard to EX for fiber - Dale tour at 11AM
- praxair delivery
- Mounts Lock here to work on doors
- Townsend Elect. on site
- EY - alignment then welding
- Jamie - guardian work on IMC locking
Comments related to this report
justin.bergman@LIGO.ORG - 07:36, Friday 13 December 2013 (8939)

Card access not changed; door closing mechanisms adjusted.

H1 ISC
keita.kawabe@LIGO.ORG - posted 16:31, Thursday 12 December 2013 - last comment - 16:01, Sunday 12 January 2014(8932)
HAM1 WFS: awesome

After an initial assessment that stated "not too bad, not too good" (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=8919), which was based on just two data points (beam radius on WFSs), I added the third point further downstream of the WFS2 and it turns out that actually it's excellent.

In the attached plot, blue circles represent the measured beam width, blue crosses on both sides of blue circle represent the potential position error (Jamie claims +-1cm though probably it's too generous), red line is a fit over the three points, green is the curve generated by the design parameters.

As you can see, in both horizontal and vertical direction, the waist is very close to the middle of the WFSs, and the waist radius is very close to the designed 250um.  Gouy separation for X (horizontal) and Y (vertical) are 82.6 and 87.1 degrees, respectively.

This was obtained from just two iterations of measurming at WFS1 and WFS2 and pushing RM2 toward RM1 based on the measurement. Downstream was measured just once after we were satisfied with WFS1 and WFS2.

  relative position (mm) position error (mm) X diameter mean (um) X diameter std X goodness of fit X goodness of fit std Y diameter mean Y diameter std Y goodness of fit mean Y goodness of fit std
WFS1 0 +-10 746.79 4.6 0.01 0.000 688.9 4.2 0.00 0.000
WFS2 369 +-10 746.25 10.1 0.00 0.000 758.18 4.4 0.00 0.000
downstream 763.2 +-10 1574.92 4.95 0.02 0.000 1666.38 1.79 0.01 0.000

Distance from RM3 to the first lens on the sled is, according to Jamie, between 48.0 and 48.25 inches.

Also, as noted earlier, the above data was obtained after having moved RM2 toward RM1 by 22.5mm. Everything else is the same as what Sheila reported much earlier.

We also measured at one point between RM3 and the sled (14.5" downstream of RM3).  Together with upstream number measured yesterday (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=8893), these are:

  position position error X diameter (um) X std X goodness of fit mean X goodness of fit std Y diameter Y std Y goodness of fit Y goodness of fit std
After RM3 14.5" downstream of RM3 +-10mm 3750.18 4.885 0.01 0.000 3923.63 2.137 0.00 0.000
Upstream of telescope 44.7" downstream of 50:50BS right after the HWP +-1" 3851.01 1.982 0.01 0.000 3959.23 1.232 0.01 0.000
Upstream of telescope, head rotated 45 degrees same as above same as above 4117.74 1.475 0.02 0.000 3845.62 0.763 0.00 0.000

In all of the above measurements, "Profile averages" was 10, "Rolling profile Averages" was 3, the actual number of measuremets (i.e. the number of scans performed before I stopped the measurement) were larger than 10 but I don't know if the software was taking more than 10 points into the statistics or not.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 17:04, Thursday 12 December 2013 (8934)

Several things to note.

We decided that we did NOT want to propagate the upstream measurements to downstream, because it was difficult to obtain more than one data points. (ModeMaster is supposed to overcome this, but in reality there are many caveats and you should know how to tell when which caveat applies.)

With NanoScan and limiting ourselves to the measurements with small beam, it was still difficult to obtain good data because of some kind of glitches. It's not clear if it was due to NanoScan or the beam, the beam was well damped and was not moving on the viewer card, there was no noticable intensity glitch either. But the symptom was that the statistics window shows nice steady data for anywhere from one second to 30 seconds, then there's some kind of glitch and the scan/fit image looked noticably different (not necessarily ugly), the diameter mean becomes larger and the stddev jumps to a big number (like 10% or more of the mean, VS up to a couple % when it's behaving nice), and the goodness of fit also becomes large. Somehow no glitch made the beam diameter number smaller. I just kept waiting for a good period and cherry-picked.

When the beam was moving it was impossible to obtain good data.

Another kind of glitch was "saturating" glitch where the software says there is a saturation. We disabled AGC of NanoScan and lowered the gain by 3dB in an attempt to eliminate saturation, it seemed to help but we couldn't kill that error completely.

keita.kawabe@LIGO.ORG - 17:08, Thursday 12 December 2013 (8935)

We (I and Jamie) will go in HAM1 tomorrow to set the eddy current damper spacing (now that Bram wrote a procedure to do that, plus it turned out that Jamie didn't check the ECD spacing on the back plates).

keita.kawabe@LIGO.ORG - 17:24, Thursday 12 December 2013 (8936)

Measurement apparatus. We flipped one steering mirror on the sled to direct the beam to Nanoscan that is placed at the same distance from the steering mirror as the WFS. Made measurement, flipped the mirror back, and moved to the next one.

Images attached to this comment
lisa.barsotti@LIGO.ORG - 04:26, Tuesday 17 December 2013 (8980)
Very nice!

This will become version 12 in D1000313.

It would be good to add the corresponding alamode file with the final distances here: https://dcc.ligo.org/T1300960-v1, I added a note to remind us that this is the relevant log entry.

kiwamu.izumi@LIGO.ORG - 16:01, Sunday 12 January 2014 (9227)

A update on 2014.Jan.12:

Alexa, Koji and I changed the position of both WFSs to dump the reflected light off of the diode (see alog 9226).

  • The position of WFS_A (or equivalently WFS_1) was shifted by 9.4 mm toward the west.
  • WFS_B (or equivallently WFS_B) was shifted by 10 mm toward the west.
H1 SUS
angus.bell@LIGO.ORG - posted 16:30, Thursday 12 December 2013 (8931)
End Y monolithic welding preparation
end-Y spent day moving suspension onto stand, setting up metrology and aligning test masses. End of play we have masses aligned and set in height. We have one fibre in position ready to start welding (S1301759). Tomorrow we will double check mass separation and pitch before commencing welding.
Angus, Alastair, Doug and Travis
H1 ISC
alexan.staley@LIGO.ORG - posted 18:52, Tuesday 10 December 2013 - last comment - 14:22, Friday 13 December 2013(8898)
ALS Fiber Distribution Chassis S1202528

As noted in alog6341, the gain on the Dual PD Amp board was set to 30dB (photo 1). I have reduced the gain to 10dB (photo 2). However, I still see saturation at 10V in the external PD monitor. Tomorrow I will try reducing the gain to 0dB.  

Images attached to this report
Comments related to this report
alexan.staley@LIGO.ORG - 11:14, Thursday 12 December 2013 (8923)

I have reset the internal PD gain to 30dB (since we were never saturating there), and set the external PD gain to 0dB.

For the external PD:

Transimpedance = 20kOhm

Responsivity = .268A/W

Power @ ext PD = 0.93mW (w/ a drift of 7% of the ref cav alignment)

Gain = 0dB

Volts = (resp) x (power) x (trans) x (gain) = 5V

alexan.staley@LIGO.ORG - 14:22, Friday 13 December 2013 (8948)

...Ignore above calculation for external PD...