Below is a summary of the available controls on SEI platforms, for PRMI. Platforms should run under their preferred configuration at all time.
HAM1:
HAM2:
HAM3:
"Quick" circular AMP connectors had been added to this gauge pair wiring to facilitated disconnection/connection without the need to access the wiring terminations at the gauges themselves. I suspect a voltage drop across this additional connection accounts for the apparent signal error. Accessibility is not good to correct -> will live with this until opportunity to correct presents itself
During the last suspension model updates, many SUS models started reporting some additional status bits. I did a first round of cleanup of the overview screens to show all bits. - fixed the number of bits and the mask names in a series of model specific ODC screens. - ODC overview screen update: fixed the numer of bits and switched to the _LATCH version of the ODC channel ( this channel latches for 1 sec.) - set the bit mask for all SUS and SEI models. Still to do: - number of bits for the BS is different than ITMs - it needs it's own BSC screen - set the string bit label s - update IMC model - in collaboration with SEI: the ISI scripts currently incorrectly set he desired state channels.
A set of matlab transfer functions has been ran on the TMSX this afternoon after BSC9 was closed out, and while the ISI was damped. The pitch degree of freedom shows some excess noise between 0.7Hz and 3Hz, which is something I was not expecting to see since the measurement after the last work accomplished on TMSX was clean.
It might be related to some excess motion at the time of the measurement during the afternoon, which I will check by running undamped and damped TFs overnight in a quiter environment. We will see how repeatable the measurement is tomorrow morning.
It might also be the position of the cables that could have changed by themselves because of their weight. Or maybe some work that I'm not aware of that was done after the last measurement taken. In those cases we would probably need to physically check what is going on.
Results attached are described below :
(1) Comparison between the model and last TMSX TF
(2) Comparison between TMSX after the work done on Nov 20th (Orange trace), after the last time work has been done on Nov 22nd (black trace), and today (pink trace).
See the 5th page of (1) and (2) for the pitch dof
Looking at the new results from this morning, everything seems fine, and tf from overnight are not showing any problems. Plots pending.
In fact, after investigating a little bit more, and looking at the pitch motion at the suspension point of the TMSX, it looks like the ISI was moving in Pitch at low frequency during that measurement. The 3rd attached pdf shows in blue the pitch input motion during yesterday afternoon's TF (noisy measurement), and in green during the overnight TF.
The other two pdfs are showing the results with :
(1) a comparison between yesterday afternoon's TF and the one taken overnight
(2) a comparison between and LLO and LHO undamped TMSX tfs
Results are very clean and similar to model and LLO's data.
I am leaving the IMC locked for the night in order to see its stability. The ASC loops are all engaged.
It must be rock-solid because we fixed the EOM !!
An auto-locking script is locally running on opsws4. The script itself resides in /opt/rtcds/userapps/release/ioo/h1/scripts/imc/sballmer/ and it is called MClockwatch.
We obtained a 9 hours of overnight lock stretch. It dropped the lock at 7:55 am in this morning for some reason which I haven't investigated yet.
At 10 pm local last night, I found that the ref cav was having a trouble in relocking. It was due to a too low transmitted light, resulting in a non-ideal trigger for relocking. Probably some optics drifted and misaligned. I decreased the threshold from 0.9 to 0.7. Then FSS started locking fine.
The attached is trend of 12 hours from this past night.
Indeed, the alignment of the reference cavity somewhere got drifted. It seems that it is mainly in pitch. It is visible in the ref can reflected camera -- see the attached picture. The reflected light started showing a halo on the upper side.
Figure: Quad view of the PSL cameras, taken at around 9 am. this morning (6/December/2013)
RefCav Trans |
PMC Refl |
RefCav Refl | PMC Trans |
Table: Quad video's signal assignment.
BTW, you can now take a picture of the videos by running Chris's excellent script:
/ligo/home/christopher.wipf/Public/Data/2013-11-20_video/gst-camera-framecapture
Of course, this is a temporary place and we will move this script to somewhere appropriate at some point.
The PSL enclosure is now back in science mode. PMC/FSS/ISS are running fine.
Hugh noticed that some HEPI ITMY slow channels were not correct in the DAQ. I did some investigation and found:
Of the slow channels (datarate=16) in H1ISIITMY.ini :
channels associated with the FEC were correct
channels associated with Filter Modules were correct
other slow channels were incorrect, either showing zero or other data (like channel hopping)
Fast channels appear to be OK (within the limited sample I used)
The problem is not seen with HEPI BS or ETMX. Looking at the models they all look identical.
Reboot/Restarts:
I first restarted h1hpiitmy. Then tried restarting all models on h1seib1. Then tried a restart of the mx-streamers. Then recompiled and reinstalled the h1hpiitmy model and restarted it, followed by a DAQ restart (DAQ status was good, so this should have been unnecessary). All to no avail.
The only thing I haven't done is reboot the computer.
It appears this problem appeared when we upgraded to RCG2.8 on Tue 12 November.
We'll have to look at those channels here at LLO (once I figure out which ones they are...). A contributing factor may be that mx_stream driver update was not explicitly listed in the RCG 2.8 upgrade instructions. There were changes from 2.7.2 (but none appear linked to 2.8). This driver can be updated now, but would require mx_stream restarts (or even front-end restarts to load new kernel image).
Celine has checked some of those channels here at LLO and is not seeing the same issue, either on MEDM screens or values on the DAQ.
CP3 alarmed, appears to have been filled Cleaning at end Y Apollo torqueing bolts at end X Corner Station chiller 2 tripped, restarted by Ski 08:52 Joe and Gerardo swapping PSL light pipe shutter 09:05 Filiberto to look at end X transmon cabling 09:39 Joe and Gerardo done swapping PSL light pipe shutter 10:02 Sheila removing beam block from PSL table 10:34 Hugh reports that the corner station HEPI pumps tripped off early this morning 10:51 Dave, Kiwamu restarting H1 LSC model, DAQ restart 10:56 DAQ restart 11:16 Hugh modifying corner station HEPI pump controller (WP 4325) 11:19 Dave restarting all models on H1 ISC EY 13:02 Thomas V. working on ITMX optical lever enclosure/bellows 13:08 Safety inspection walking through the LVEA 13:28 Gerardo to LVEA to find Thomas V. 14:16 Arnaud running transfer functions on TMSX 14:17 Apollo removed dome and north door from BSC 10 (WP 4326) 15:22 Jeff K., Sheila and Arnaud purposely tripping chamber watchdogs 15:23 Hugh done modifying corner station HEPI pump controller (WP 4325) 15:36 Gerardo and Thomas V. done with optical lever enclosure/bellows
Last night the power watchdog on the PSL tripped again. I turned the last back on this morning after the shutter was installed.
Note: this morning I transitioned to science mode, then stefan and I reentered to make a measurement of the modulation depth with the OSA. When I transitioned back to commisioning mode I forgot to turn the AC on, so there was probably quite a temperature fluctuation. Now the psl is in commisioning mode.
Since the PSL is at least working reasonably now I updated safe.snaps for the iss, fss, and pmc.
(Joe, Gerardo)
PSL light pipe and shutter were removed from the chamber side (HAM01), the main PSL beam vieport was exposed to apply FC on it. FC application (Joe used a small bottle) was finished by 4:10 PM yesterday, to protect the viewport we attached the old vieport adaptor with 3 screws to the viewport, and covered the 3" center hole with aluminum foil.
Work will continue today.
(Sheila, Kiwamu, Joe, Justin, Gerardo)
New MB PSL shutter is now installed.
Joe removed the FC sheet from the viewport and made sure it was clean, then the viewport adapter was installed on the viewport without problems.
Next to be installed was the shutter, the 3" nozzle has a very tight fit into the viewport adapter, it went in but we had to use longer screws to be able to press it in, then those longer bolts were removed and shorter ones were used instead.
The light pipe tube slid out of the other light pipe and into the shutter without problems, this was secured with two screws.
All items were cleaned prior to installation, including the PSL pipe, Sheila and Kiwamu pushed a wad of presoaked wipes thru it to clean it.
Stefan, Sheila and Kiwamu with remote assistance from Rich A. and Volker
The mysterious jump in the RF phase (see alog 7941) is now understood and fixed. It was due to a loose connection at DB15 connectors in the EOM box and not due to the SMA connector (see alog 8811 and 8813 for our early detective story). We applied two small in-situ modifications on the EOM box. As a result, now it doesn't show the mysterious RF jump any more.
The box:
I briefly explain the EOM box for those who are not familiar with our custom-made EOM box. The EOM box consists of two boxes -- one contains LC resonant circuits and the other contains the EOM crystal. This two-boxes-design allows one to tune the resonant frequencies by tweaking the LC circuits without messing up the alignment of the EOM crystal because one can simply take out the electronics box and leave the crystal box for solder or tuning some parts in the circuits. To apply voltage across the EOM crystal for normal operation, the LC circuits need to be connected to the EOM crystal. This is done by a DB15 connector attached on each box -- female DB15 on the crystal box and a male DB15 on the electronics box (see pictures shown below). In this way, the two boxes are electrically connected.
A picture of the actual EOM box. The gloved hand is me pressing the SMA downward in order to reproduce the RF jump.
When I pressed the SMA connectors downward in this morning, the RF characteristic of the EOM box changed as if something jumped. This was repeatable, although it seemed that the condition to make it jump was random -- occasionally, pressing the SMA didn't make it jump and sometime pushing the SMA toward the box made it jump. Anyways, at this point, it was clear that the EOM box was the culprit and not the RF cables.
The causes:
At the beginning, we thought the culprit was the SMA connectors (see alog 8811 and 8813). However, this turned out to be wrong as we investigated it further. With a remote assistance from Volker, Stefan and I took the electronics box apart from the crystal box while keeping the crystal alignment. In the process of removal, we found that the electronics box was attached to the crystal box merely by friction of the DB15 connectors and three pieces of adhesive tape. So the orientation of the electronics box was not so solid with respect the EOM box. We then checked the return loss of the LC circuits without connecting the EOM and confirmed that wiggling the SMA connector didn't change its impedance. Instead, we discovered that the DB15 connector can easily change the amount of its stray capacitance -- the frequency of the resonant notch could shift by approximately 1 MHz by very gently touching the DB15 connector with our latex-gloved-hand. We put the electronics box back on the EOM crystal box and wiggled the orientation of them. Indeed, it changed the resonant frequency by about 1 MHz in a discontinuous way. So we determined that the loose connection in the DB15 connectors was the culprit and the mysterious RF jump was induced by some change in the orientation of the two boxes.
It seemed that the DB15 connectors were not all the way in because the two aluminum boxes contacted first.
The repair/modification fixed the issue:
We did the following two repair/modification:
The DB15 connector have two screws to support it and we put a washer for each screw. They raise the height of the DB15 connector like shims. Also, we newly installed four screws to make the connection of the two boxes more solid. There were already four screw holes on each box to accommodate them. So we just installed them. After these modifications, we checked the return loss of all three RF ports. We didn't observe the mysterious jump at all, even when the electronics box was wiggled hard. Of course, strongly pressing the electronics box downward shifts the notch position by an order of 10 kHz due to the change in the stray capacitance at the DB15 connectors, but the shift is smooth and not in a discontinuous way any more. So the RF jump issue is now solved.
A picture of the electronics box when apart from the EOM crystal box. This DB15 connector was shimmed by washers.
A picture of the EOM crystal box when the electronics box is taken away. There is a threaded screw hole on each corner and these are the ones we used for installing the new screws.
A top view of the EOM box with the lid off. The green circles indicate the screws that we newly installed.
A top view of the EOM box with the lid off. The green circles indicate the screws that we newly installed.
Here are the return loss of all three ports after the modification:
Also raw ASCII data of these plots in csv format are attached too.
Does this mean that someone would like to modify the 3IFO version of the EOM box?
A DB15 connector to pass RF signals, not to mention they're the RF signals that produce all the interferometer error signals ... speechless.
Steps taken to commission the ITMY ring heater pair: - Updated to the recently committed Beckhoff code that Joe B and Adam M worked on yesterday. - Added the calibration values that Aidan B had recommended for running the ring heaters. - Opened up a strip tool to monitor the vacuum pressure with the closest gauge (HVE-LY:Y1_120BTORR) - Turned the ring heaters on to low power (~.01 Watts) to make sure that the channels were working properly - Increased the power to 1 Watt for a 4 hour period starting at 12:00PM Pacific Time, keeping an eye on the pressure gauge previously mentioned. Upper: Requested Power = 1.00 Watts Measured Voltage = 5.94 Volts Measured Current = 0.17 Amps Calculated Power = .99 Watts Calculated Resistance = 34.94 Ohms Lower: Requested Power = 1.00 Watts Measured Voltage = 5.92 Volts Measured Current = 0.16 Amps Calculated Power = .98 Watts Calculated Resistance = 37.0 Ohms A four hour trend of the pressure gauge and ring heater settings is attached. A quick, back-of-the-envelope calculations shows that the ring heaters are working properly. The next step for measuring the thermal lensing would be to measure the beam diameter of a single reflection off of the back of ITMX independently and then changing the radius of curvature of ITMY using the ring heater to match. This requires looking at the REFL beam either in the HAM1 chamber or on the ISCT1 table, depending on how much power we are able to get out of REFL. The Thorlabs beam profile we have on site requires a minimum of ~1e-5 Watts of incident beam power to function.
NOTE: This pair of upper and lower ring heater segments is due to be replaced, per "Bug255" https://services.ligo-wa.caltech.edu/integrationissues/show_bug.cgi?id=255 See, also, T1300463
J. Kissel, T. Vo Because of the recent resonant features seen in ISI-ITMX that seemed to have been fixed by a simple lock and unlock of the ITMX Arm Cavity Baffle (see LHO aLOG 8632), I asked "has anyone ever taken B&K hammer transfer functions of the ACB?" The answer was a resoundingly loud "no" (I am Jack's complete lack of surprise). Me and my big mouth volunteered to do it, and there was a tiny window of opportunity before Thomas cleaned up the chamber, so Thomas and I pioneered the first B&K measurements of the ITM ACB. Details of the measurement setup and execution below. Again, I don't know how to properly export the data (as indicated in LHO aLOG 8654), so you'll have to be patient regarding the results. Note that obtaining official plots of the results should not in any way be considered as a hold-up for chamber close out. Of course, after taking the measurements and cleaning up the chamber -- because we *touched* the baffle -- we had Sebastien run a quick set of ISI-ITMX transfer functions, and it informed us we're not-at-all done battling this bumbling oaf of a baffle (see LHO aLOG 8653). This should be considered a hold-up for chamber close out. Round three, first thing tomorrow! ----------- Details: We took two distinct measurements: (1) With the accelerometer on the suspended baffle itself, using a unused slotted bolt hole in the inside middle, closest to the ITM HR surface. This test was just a shot in the dark, to see if we could get a nice driven transfer function of the suspended stage, since there have been no prior attempts with this generation of the ACB baffle. At first glance from the B&K software plots, it looks like a complete mess, so it will most likely either be a completely confusing / useless or completely depressing result. DSCN0206.jpg (or pg 3 of the .pdf) shows a picture of the accelerometer from inside the baffle looking out back toward ITMX. As is (hopefully) resolvable in the picture, ACC +X = ITMX -L, ACC +Y = ITMX -T, ACC +Z = ITMX +V. The Y impact was on the (ITMX) +T face, bottom corner, closest to the ITM, in the (ITMX) -T direction. The X impact was along the bottom edge of the (ITMX) -L face, in the +L direction. (2) With the accelerometer on the bottom of the support structure's tube (shown in DSCN0209.jpg). One can't see it in the picture, but the accelerometers axes were aligned with the global IFO's axes, such that ACC +X = Points down X arm towards ETMX, ACC +Y = Points down the Y Arm towards ETMY, and ACC +Z = Points up with local gravity (the same as ITMX's +V). The X & Y impacts were made towards the bottom of the outer, support structure, "eddy current damping 8 dia," tube (D1002564, of the assembly D1200275). Man, that thing rung like a bell when Thomas whacked it... The files live on the B&K laptop only, in C:Users\ligo\Desktop\SUS Hammer Test\ITMX\BandK\ and are called (1x) SimpleHammerDisplay3-ArmCavityBaffleBaffle-ISIfloating-SuspendedElements_Ximpact.pls (1y) SimpleHammerDisplay3-ArmCavityBaffleBaffle-ISIfloating-SuspendedElements_Yimpact.pls (2x) SimpleHammerDisplay3-ArmCavityBaffleBaffle-ISIfloating-StructuralElements_Ximpact.pls (2y) SimpleHammerDisplay3-ArmCavityBaffleBaffle-ISIfloating-StructuralElements_Yimpact.pls
Results are attached below.
First page is comparing X-X and Y-Y of the "structural element"
* Main resonnances are around 80Hz, 120Hz, 180Hz
Second page is comparing X-X and Y-Y of the "suspended element"
*Data is very noisy
J. Kissel, C. Vorvick Calum and Betsy installed the new PR3 baffles (D1300957) yesterday (see LHO aLOG 8619), but only roughed in their alignment with respect to the optic / prisms / wires. Today Cheryl and I went into HAM2, and I aligned the baffles to-the-best-of-my-ability, by-eye, and then tool tightened them to PR3 HLTS structure. My left-right metric was "just covering the edge of the bevel on the optic with the straight portion of the baffles" and my up-down metric was having the "bottom curves of the baffles following the curvature / bevel of the lower half of the optic." It was certainly a ball-park activity, given that the goodness of alignment depends heavily on one's angle, orientation, and height of viewing. From most face-on views, however, the baffle looks to be baffling the wires and not clipping the optic. I took many pictures, but did not find out until later that the memory card on the camera was full, so the pictures were not being stored and lost. FFFFUUUUDDDGGEEEE. Cheryl, Calum, Betsy, and Kate have graciously volunteered to retake pictures tomorrow before / after they pull of the first contact on all the optics. We also took B&K hammer transfer functions after securing the baffles (with PR3 freely suspended, and the ISI Locked, and probably with me leaning on the table). I placed the accelerometer in the exact same location and configuration as in LHO aLOG 6014, (4th picture attached) with ACC +X = PR3 -L, ACC +Y = PR3 -T, and ACC +Z = PR3 +V, in the upper left corner of the cage on the HR face, "front" of the cage. The X impacts with the hammer were just below the accelerometer, on the HR face. The Y impacts were on that same strut / corner, at the same height as the accelerometer, on the -T face of the cage, in the +T direction. Comparing these results with the previous results taken before the baffles were installed -- by-eye they look roughly equivalent. The resident LHO expert of post-processing the data is off to LLO for a few days, so we'll post a comparison next week (or I can ask Calum / Stuart tomorrow, we'll see.) For now, the saved templates live only on the laptop, and live in C:\Users\ligo\Desktop\SUS Hammer Test\PR3 and are called SimpleHammerDisplay3-PR3Baffle-ISIlocked-Ximpact.pls SimpleHammerDisplay3-PR3Baffle-ISIlocked-Yimpact.pls
Attached are the plots comparing the dynamical behavior of PR3 structure with and without the baffles.
The first page shows the X-X response, and second page Y-Y
Green curve is the old measurement without the baffles, and blue curve is the new measurement with the baffles.
As we were expecting, and since the baffle is really light compared to the structure, no main difference is observed.
Note: The HEPI-BS position loops are in the second set of filter banks. One should use the isolate BS lv2 command button, instead of lv1, for this chamber only.