Displaying reports 79781-79800 of 83355.Go to page Start 3986 3987 3988 3989 3990 3991 3992 3993 3994 End
Reports until 17:00, Tuesday 31 July 2012
LHO VE
kyle.ryan@LIGO.ORG - posted 17:00, Tuesday 31 July 2012 (3675)
Kyle, Gerardo -> replaced HAM6 annulus ion pump body
This pump finally shorted internally (typically expires @80,000 hours, lasted 105,000 hours)
H2 PEM
maggie.tse@LIGO.ORG - posted 16:59, Tuesday 31 July 2012 (3674)
Noise Sources in Optical Lever Spectra

Robert and I investigated noise sources contributing to the optical lever spectra for ETMY in this entry using recently installed PEM accelerometers. We were able to identify and remove three noise sources:

29.5 Hz
Annulus pump that was left on. The pump has since been turned off and will be removed from EY today.

90-91 Hz
Power supply for optical lever laser on the receiver pier. We moved this to the floor by the receiver pier and seismically isolated it from the floor using inflated clean room gloves.

56.5 Hz
Power supplies on floor near receiver pier. We placed these on a seismically isolated platform using springs.

The above changes removed the peaks in the optical lever spectra at 29.5 Hz and 90-91 Hz, and reduced the peak at 56.5 Hz by a factor of 2. We believe there are other sources contributing to the 56.5 Hz peak. Attached are the ETMY optical lever spectra before and after the above changes were made (including spectra for a recently installed PEM accelerometer mounted on the receiver pier), and pictures of the three noise sources.

The 29.5 Hz signal was also found in the OAT length error signal before the pump was shut down, as seen in an attached plot.

DTT template used is located here:
/ligo/home/robert.schofield/12sAug/OAToplev.xml

Images attached to this report
Non-image files attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 16:36, Tuesday 31 July 2012 (3673)
Decoupled pump from BSC6 annulus @ ~0900 this morning


			
			
H1 SEI
hugh.radkins@LIGO.ORG - posted 16:24, Tuesday 31 July 2012 (3672)
H1 HAM2 ISI Optical Table set to Elevation & Level--Ready for IAS
Corey, Mitchell, Hugh

We set the Optical Table in WHAM2 at -252.9 (E1000403) + 12.5 (T1100187-to correct for X-arm slope[local gravity plane vs LIGO Global Frame])= -240.4mm below the center of BSC2 where the local level and LIGO Global cross zero.  The Optical Table is level to +-0.2mm.

The HEPI remains floating and is ready for Initial Alignment horizontal adjustment.
LHO General
patrick.thomas@LIGO.ORG - posted 15:58, Tuesday 31 July 2012 (3671)
plots of dust counts
Attached are plots of dust counts > .5 microns in particles per cubic foot. The dust monitor in the H2 PSL laser room (H0:PEM-LVEA_DST13) has a calibration error and is not included.
Non-image files attached to this report
LHO General
patrick.thomas@LIGO.ORG - posted 15:54, Tuesday 31 July 2012 (3669)
Ops Day Summary
Sprague on site
Unifirst on site

8:45 Eric to work on HAM2
8:46 Cheryl to remove H1 ETMX
8:54 Volker to work in H1 PSL enclosure
8:55 - 9:35 Kyle to remove pump from BSC6 annulus
8:59 - 10:05 Corey to switch cable in HAM3
9:01 - 11:33 Filiberto to work near HAM2 on seismic cable
9:22 mopping, vacuuming in LVEA
~11:30 - 1:00? Robert to set up coils near BSC8
9:27 Michael R. to work in H1 PSL enclosure
9:42 - 9:55 Bubba, Hugh to remove tote from LVEA by hand
10:38 Robert, Maggie to work in optics lab
10:38 Corey to look for rigging in LVEA
11:28 Eric to work near BSC1 and BSC2
1:15 Richard to work in plenum?
1:30 - 3:50? Corey, Hugh, Mitchel surveying HAM2 ISI table
1:35 Jax, Elli to work at end Y
2:14 Apollo to move parts for optics extraction into mid Y
H2 ISC
eleanor.king@LIGO.ORG - posted 15:48, Tuesday 31 July 2012 (3668)
Doubler crystal temperature of the green laser

Jax, Bram, Elli.  We repeated the measurement we made last week on 26 July of measuring how the 532nm laser power output changes as you vary the doubler crystal temperature.  Measurements were taken over a greater range than last time.  The attached graph shows today's measurements (blue) and well as those taken last week (red).

The laser diode current was set to 1.725A and the 1064 crystal temperature was 42.14 degrees C.  As the graph shows, the maximum power output occurs at about 34.36 degrees C, althought the actual power output at this temperature varies between 40-45mW.

We also took a couple of measurements about the REFL_PWR_MON, GREEN_PWR_MON and IR_PWR_MON photodiodes.  First we measured the signal in counts of the output of these channels.  Then we covered each diode so no laser light hit it and measured the output signal again.  Then we measured the power incident on each diode in mW using the power detector.  The readings are as follows:

     Normal operation, beam incident on PD (# counts) No light incident on PD (# counts) Power (mW)
REFL_PWR_MON ~400 14276 1.2
GREEN_PWR_MON 8520 20 0.68
IR_PWR_MON 13350 170 0.73
Images attached to this report
H1 INS
jodi.fauver@LIGO.ORG - posted 15:38, Tuesday 31 July 2012 (3670)
BSC9 Optics Extract
Cheryl and I went into BSC9 to look at the particulate contamination in the viewports and at the drop-off by the gate valve. (Cheryl said that the contamination was the worst she has ever seen: my take is that this stuff looks very much like what we saw during the initial survey down the tube at BSC8.)We took some pix and a sample for analysis. Then, we removed OSEMs and preserved them clean. I came out: Scot and Ed went into the chamber with Cheryl. Mark D.and Bubba were also outside. The Lazy Susan, Slide Table and Lift Table were inserted. The SUS was removed, wrapped and set aside so that the optic can be put in a cake tin. The optics table was cleared and the telescope preserved clean. The VEA was cleared of SUS de-install tooling to make way for SEI stack extract tomorrow. 
H2 SUS
szymon.steplewski@LIGO.ORG - posted 13:32, Tuesday 31 July 2012 (3667)
Suspension Coil Driver grounding plugs

I have been working on assembling 10 grounding plugs for the TEST In connection on the Coil Driver boards, building on work described in https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?.submit=Number&docid=l1200193&version=  These grounding plugs make use of 100 Ohm resistors across pairs of pins and a 0.022 uF capacitor connected between the remaining pin and the DB9 connector metal (see page 16 of the document above for the schematic).  I will next measure the coil driver noise with these ground plugs in place across the TEST Input path.

Images attached to this report
H1 SUS
jeffrey.garcia@LIGO.ORG - posted 13:12, Tuesday 31 July 2012 (3665)
H1 SUS PR2 Chamber-side metal build TFs
Attached are the WhiteNoise transfer function measurements taken on 07/30/2012 on the HSTS metal build slated for becoming "PR2" in-chamber.  Coherence is excellent for all DoFs and resonances clearly visible.  A comparison of this data with data from the staging building build and the model will be posted.
Non-image files attached to this report
H2 ISC
bram.slagmolen@LIGO.ORG - posted 22:34, Monday 30 July 2012 (3660)
Arm Locking

From 17:30h to 18:36h (local time) the Cavity has been locking nice and steady, until it suddenly nose dived. Not sure why, the green power monitor increased maybe making the locking loop unstable (to much gain? haven't measured the UGF after the SHG temp mods).

I was trying to measure the TF from HEPI to the cavity error signal and the CPS. After about 30min I realised that there is no signal comming from the RFM into HEPI! HEPI looks at the same RFM locaiton as the ETM (H2:ISCEY-SUSETMY_LONG). I check in both models (h2susetmy and h2 hpietmy) which seem to be ok. I think Dave did some cleaning up on the weekend, will ask him tomorrow.

Attached plot shows a reference from last week, and a spectrum during the hour lock.

The 'ISI Low Blend (ITM+ETM)' is with a cross over frequency of 250 mHz on both stages (ITM/ETM). Currently the ISI runs with blend frequencies at 750 mHz for both stages, except in X and  Y (cavity axis) as indicated in the legend. The peaking at the 2.75 Hz is less, although an increase at the higer frequencies.

Images attached to this report
H1 SEI
hugo.paris@LIGO.ORG - posted 18:28, Monday 30 July 2012 - last comment - 11:55, Tuesday 31 July 2012(3659)
HAM3-ISI - In Chamber Testing Progress - Transfer Functions Running Overnight

Filiberto, Richard, Corey, Hugh, Hugo,

 

Today's effort brought us to the point where we are able to run transfer functions:

Transfer Functions are running overnight. They should be done by 6am.

 

Note: There is a mailing list to receive the alerts regarding these transfer functions. Please let me know if you would like your email address to be added to it.

Comments related to this report
hugo.paris@LIGO.ORG - 09:52, Tuesday 31 July 2012 (3662)

The computer overnight TFs were running on froze (black screen) around 11.30pm. I was able to retrieve the TF sections measured until then.

TFs were compared to the ones measured on LLO HAM3. Accordance regarding result on CPSs is good (plots attached).

Results on GS13s appeared filtered differently than LLO and need to be analyzed in more details.

GS13s were tested. Their spectra taken with when the ISI is tilted (most critical configuration for GS13s) show that they work properly (see attached plots).

Images attached to this comment
hugo.paris@LIGO.ORG - 11:55, Tuesday 31 July 2012 (3664)

LLO-HAM3 Transfer Functions used for the comparison plots (CPS-H, CPS-V) were recorded in-chamber, in-air, on January 18th 2012. These transfer functions are the one presented in LLO-HAM3 phase II testing report under the section 2.2.1 of the initial chamber testing chapter.

LHO FMCS
alberto.stochino@LIGO.ORG - posted 17:33, Monday 30 July 2012 (3658)
Temperature alarm H0:FMC-MY_CY_H20_SUP_DEGF

This afternoon at about 5.15pm the alarm handler went off for this channel: H0:FMC-MY_CY_H20_SUP_DEGF

I'm not sure what it refers to, but I plotted the trend for the past 2 days and it doesn't seem to be doing anything particularly unusual.

I aknowledged the alarm.

H1 SEI
hugh.radkins@LIGO.ORG - posted 10:50, Monday 30 July 2012 - last comment - 16:23, Monday 30 July 2012(3650)
H1 HAM3 HEPI Position Revisited
As I mentioned on Friday, I noticed a coherent shift of the HEPI platform on HAM3 as I was locking it up in preparation for ISI measurements.  Hugo gave me a window this morning so I unlocked the HEPI again and very thoroughly looked for interferences that might be pushing it out of our aligned location.  I could find no non-suspended elements touching suspended portions aside from a few of the temporarily run in-air cables.  I can't imagine that these contacts could pull the system around.
Once it was all released I pushed on the crossbeams to 'see' knocking.  I could not detect anything that looked like something running into something else but I did notice that the vertical moved fairly easily whereas horizontal motion was very stiff.  I'm trying to understand why this would be but nothing is so far revealed.  Thoughts?

For the record, based on the dial indicators, I see a CCW rotation of about 250urad caused by a shift to the east of the south side of about 0.5mm.
Comments related to this report
hugh.radkins@LIGO.ORG - 16:23, Monday 30 July 2012 (3656)
Maybe when I push up and down, I only need to move the local corner up & down or maybe the local and the opposite corner.  Whereas when I attempt to move it horizontally, I have to move maybe three or even four corners of the stiff system and hence the greater difficulty.
H1 SEI
hugo.paris@LIGO.ORG - posted 10:34, Monday 30 July 2012 - last comment - 16:33, Monday 30 July 2012(3648)
HAM3-ISI - In Chamber Testing Progress - WatchDogs Un-tripped

Issue with watchdog tripped - Richard, Filiberto, Hugo
Corrected, can now actuate

We managed un-tripping the coil driver chassis bio watchdog of HAM3 by:

 

In Vacuum cabling - Corey, Hugo
CPS H3 and CPS V3 were swapped - corrected 

I took spectra on our sensors this morning and noticed H3 and V3 were swapped. Corey went looking at it and located it on the in-vaccum side of the feedthrough.

Comments related to this report
hugo.paris@LIGO.ORG - 12:07, Monday 30 July 2012 (3653)

The binary word that should be used for HAM3, in ISI_CUST_CHAMBER_BIO.adl is:

65535

filiberto.clara@LIGO.ORG - 16:33, Monday 30 July 2012 (3657)
Replaced ISI Coil Driver SN S1103567 with SN S1103321 in Rack H1-SEI-C2 for HAM3 . Driver seems to have internal wiring issues with the outputs. When you drive CH1 (coarse1), it would drive CH4 (coarse 2). Same with CH2 (fine1) and CH3 (fine2). Unit also has a ±18V type connector for the input power of ±24V.

Filiberto Clara
H2 SUS
keith.riles@LIGO.ORG - posted 23:47, Sunday 29 July 2012 - last comment - 13:11, Tuesday 31 July 2012(3646)
Looking at OAT spectra with fine resolution
As part of gearing up for aLIGO CW and Stochastic analysis detector characterization,
a number of us are developing and starting to use new or enhanced spectral line finding
tools. We plan to apply these various tools to this summer's OAT data in the hope of
1) making the tools better via exposure to real data; 2) identifying spectral lines early
on that can be mitigated; and 3) cataloging lines that we have to live with in aLIGO.

A wiki clearinghouse for OAT spectral line studies can be found here. For now it has
mostly documentation on tools and channel selections, but over time it will be populated
with study results.

One important tool is an enhanced version of the familiar Fscan program. Greg Mendell
has started creating daily Fscan plots for locked-arm data (defined by the mean value of
H2:ISC-ALS_EY_REFL_PWR_MON_OUT16 over 1 minute lying in the range 4300-8000 -- see wiki
above for explanation). The Fscans generated so far cover only the days of July 16 and 20,
chosen because each had more than five hours of locked-arm data.

I have used the half-hour SFTs generated during the Fscanning to look at average spectra
for the calibrated OAT feedback channel H2:SUS-ETMY_M0_LOCK_L_IN1_DQ for July 16 (5 hours)
and July 20 (9 hours) with 0.56 mHz binning. Two different averages were computed, one
being a simple arithmetic average, the other a noise weighted average that is almost, but
not quite the same as a harmonic mean. (The same method was used to compute unweighted
and weighted averages for h(t) for all of S5 and S6, as discussed in more detail here.)
Generally speaking, one expects the weighted average to be lower than the unweighted
average and to have a smaller variance. Large differences between unweighted and
weighted averages indicate non-stationary bands.

The figures below show the results:
  1. Full spectrum (0-1000 Hz) for July 16 (unweighted and weighted averages)
  2. Full spectrum (0-1000 Hz) for July 20 (unweighted and weighted averages)
  3. Full spectrum (0-1000 Hz) for July 16 and 20 together (weighted only)
  4. Zoomed spectrum (0-2 Hz) for July 16 and 20 together (weighted only)
  5. Zoomed spectrum (2-10 Hz) for July 16 and 20 together (weighted only)
  6. Zoomed spectrum (10-100 Hz) for July 16 and 20 together (weighted only)
  7. Zoomed spectrum (100-200 Hz) for July 16 and 20 together (weighted only)
  8. Zoomed spectrum (200-800 Hz) for July 16 and 20 together (weighted only)
Some quick observations:
  • The full spectrum is "rich" with structure (how much of it is relevant to eventual aLIGO operations is not clear, however)
  • The July 20 noise is below that of July 16 over most of the full band (good to see!)
  • Here is a small sampling of line artifacts above 10 Hz (band of direct interest to the CW/Stochastic groups):
    • 120 Hz is easily seen, but 60 Hz is much weaker in July 16 data (stronger in July 20 data -- was there a 60-Hz notch on this channel on July 16?) There is a peak near 180 Hz (179.9 Hz), but I don't think it's a 60-Hz harmonic.
    • There is a persistent comb at 2.74 Hz seen out to at least the 25th harmonic in the July 16 data, but quite mitigated in the July 20 data
    • There is a very strong line at 29.375 Hz with rapidly weakening harmonics visible at 58.75 Hz, 88.125 Hz and 117.5 Hz
    • Other strong lines below 200 Hz include 77.30 Hz (and harmonic at 154.60 Hz), 78.15 Hz, 89.54 Hz (and harmonic at 179.08 Hz), 90.40 Hz (and harmonic at 180.80 Hz), and 105.92 Hz
Chores ahead of us include dealing with some auxiliary channel issues in Fscans, getting Virgo's NOEMI line-finding tool to work with only locked-arm data, resurrecting automated coherence measurements between the OAT feedback channel and ~100 auxiliary channels -- and digesting all of the results once they appear. Future reports will be posted both here in the alog and on the wiki page.
Images attached to this report
Comments related to this report
keith.riles@LIGO.ORG - 11:16, Monday 30 July 2012 (3652)
Here is a list of the SFT filenames from July 16. The start times are embedded in the names:
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026518956-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026520756-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026523546-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026525346-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026527146-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026528946-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026530746-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026532546-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026535786-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026538156-1800.sft

Here is a list of the SFT filenames from July 20.
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026785686-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026787486-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026789286-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026791086-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026792886-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026794686-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026796486-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026798286-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026803896-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026805696-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026807496-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026809296-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026811096-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026812896-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026814696-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026816496-1800.sft
H-1_H2_1800SFT_fscanH2_SUS-ETMY_M0_LOCK_L_IN1_DQ-1026818296-1800.sft
keith.riles@LIGO.ORG - 08:48, Tuesday 31 July 2012 (3661)
The initial entry included links but did not list the URLs. For future reference, here they are:

OAT spectral line wiki page: https://wiki.ligo.org/foswiki/bin/view/DetChar/OneArmTestLineInvestigations
Daily OAT Fscans (SUS channels): https://ldas-jobs.ligo-wa.caltech.edu/~pulsar/fscan/H2_OneArm/H2_OneArm_SUS/fscanNavigation.html
Averaged S5/S6 spectra: https://ldas-jobs.ligo.caltech.edu/~keithr/spectra/

keith.riles@LIGO.ORG - 10:53, Tuesday 31 July 2012 (3663)
Bram took a look at the transmitted-light photodiode for the reference cavity
in the optics lab (using a spectrum analyzer - no DAQ channel available).
He saw strong harmonics of 2.74 Hz and 4.0 Hz. So those lines showing up
in the ALS feedback signal are artifacts of the temporary light source being
used for the One Arm Test and should not be relevant to future aLIGO running.
bram.slagmolen@LIGO.ORG - 13:11, Tuesday 31 July 2012 (3666)

see post 3582 for the measurements of the 2.75 Hz and the 4 Hz peaks.

Displaying reports 79781-79800 of 83355.Go to page Start 3986 3987 3988 3989 3990 3991 3992 3993 3994 End