plots of dust counts
Attached are plots of dust counts > .3 microns and > .5 microns in particles per cubic foot requested from March 15 01:00 UTC to March 16 00:00 UTC. Also attached are plots of the modes to show when they were running/acquiring data.

Data was taken from h1nds1.

mode:
120 seconds worth of data was unavailable on this server
1380.0 minutes of trend displayed
read(); errno=9
read(); errno=9
T0=13-03-15-01-00-00; Length=82800 (s)
No data output.

status:
120 seconds worth of data was unavailable on this server
1380.0 minutes of trend displayed
read(); errno=9
read(); errno=9
T0=13-03-15-01-00-00; Length=82800 (s)
No data output.

> .3 microns
120 seconds worth of data was unavailable on this server
1380.0 minutes of trend displayed
read(); errno=9
read(); errno=9
T0=13-03-15-01-00-00; Length=82800 (s)
No data output.

> .5 microns
T0=13-03-15-01-00-00; Length=82800 (s)
120 seconds worth of data was unavailable on this server
1380.0 minutes of trend displayed
Non-image files attached to this report
H1 SEI
hugo.paris@LIGO.ORG - posted 18:08, Friday 15 March 2013 (5826)
HAM2 & HAM3 ISIs - Safe.snap Updated

Safe.snap files were updated for HAM2 and HAM3 ISIs, so they comply with the latest model/screen updates.

The new safe.snap files are commited under the SVN:
HAM2-ISI - r3992
HAM3-ISI - r3993

Pictures of the state of the ISIs, at the snap, are attached.

Non-image files attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 17:04, Friday 15 March 2013 (5824)
IOP model safe.snap files updates

I updated the safe.snaps for the IOP models in the corner station (the non-test stand ones). Main changes are in handling the alarm levels of channels associated with the software watchogs.

When I tried to burt restore the h1iopsusquadtst safe.snap settings, the epics process for this front end crashed. This is reproducible, so I'll stop the upgrade for today and fix this on monday.

H1 SUS
jeffrey.kissel@LIGO.ORG - posted 16:29, Friday 15 March 2013 (5823)
H1 SUS ITMY
Was poking around H1 SUS ITMY to try and assess SUS sensor noise, and found that ITMY needed some infrastructure TLC similar the HAM SUS. 
- CART2EUL and EUL2CART matrices were still using H2 values, so I've updated them
- OSEMINF to_um calibration in FM5 was old, stale, and off. It's now correct, fresh, and ON.
- Stuck from_um anti-calibration filters is the DAMP filter banks such that we can use the same loops.
 
I've since captured a new safe.snap, and committed both the new filter file and h1susitmy_safe.snap to the userapps repo.
H1 SUS
betsy.weaver@LIGO.ORG - posted 16:18, Friday 15 March 2013 (5822)
Other MC optic blade tip EQ stops swapped

Today, I swapped the sketchy top mass blade tip EQ stops on MC1 (or, better known as the one near the center of the table).  IO will readjust this suspension to continue alignment.

H1 SUS
betsy.weaver@LIGO.ORG - posted 16:15, Friday 15 March 2013 (5821)
PR3 IAS alignment

Betsy, Travis, Jason

Indeed, things went too well earlier in the week.  This morning we spent a few hours diagnosing why the beam was low on the PR3 after transiting through the PLX (although the PLX assy was level and the incoming IAS beam was verified to be at the correct height).  We finally resolved to pitch the PLX in order to center the beam on the PR3 center and move on.  We then spent many hours this afternoon walking the PR3 pitch everywhere but onto the nominal alignment.  After pouring through some documentation on the HLTS, we discovered the correct knob to turn (not called out well in documentation, and was not obvious to us QUAD-centric folk).  Just as we were back on track, we *ehem* had a bump of the IAS equipment, so we decided to quit and cut our losses.  Long story short, now that Jason will be with us at LHO next week, we will finish up in ~a few hours Monday morning, bright-eyed-and-bushy-tailed.  Sorry, IO, we need one more day!

LHO General
corey.gray@LIGO.ORG - posted 16:06, Friday 15 March 2013 (5820)
Ops DAY Summary

Day's Activities:

Code change for HAM-ISI 2 & 3.  Required re-starting model and DAQ around 10am (Barker, Warner)

Photon Calibrator "dirty" Alignment (Daveloza, Rodruck, Savage)

ISCT1 craned into place by Apollo (west side of HAM1)

Betsy/Travis worked mainly in HAM2

Cyrus doing rackwork in out-buildings

Tour in Control Room around 12:30pm

Crew working on baffle in West Bay cleanroom

RO Alarm toward the end of the day

Dust Alarms:

H1 CDS
david.barker@LIGO.ORG - posted 13:50, Friday 15 March 2013 (5816)
h1 isi ham2,3 restarted

We noticed a strange 0x2000 daq status of the h1isiham3 which should not be the case (models were restarted, then the DAQ). So to be safe I have just restarted the h1isiham2,3 models again. I cleared the Dolphin IPC errors this raises on the suspension systems in ham2,3.

BTW, the safe.snap files for h1isiham2,3 are out of date and generate many errors when ran manually.

H1 CDS
corey.gray@LIGO.ORG - posted 10:56, Friday 15 March 2013 (5813)
Continuing To Modify CDS Software SUS Watchdog Screens

This morning I asked the question: 

"Who is OSEM[*] in the Alarm Handlder??"     (  *  can be 1-6 in the Alarm Handler)

Dr. Kissel handed over a crumpled Post-It note which decrypted the OSEM mystery.  Currently, to understand what a SUS alarm means, one needs this Post It note because the Alarm handler names SUS Watchdog channels as OSEM1, OSEM2, ...OSEM6 (which isn't really useful since the SUS Expert screens don't use the OSEM names; they use F1, F2, F3, Left, Right, Side).  I posted the Post It notes to the CDS SUS wiki.  [As for the actual alarm in question, it was related to OSEM5&6 (Left & Right) "Centering"; and this is most likely due to not being pumped down].

I did actually do something here.  Jeff showed me the IOP Watchdog screen, and how some of the custom medm screens on here have been changed from using "OSEM" to "useful" names.  But we noticed some sub-screens weren't updated (namely EY BSC6).  So, I updated this screen from the OPS Station.  Here are steps taken:

(I also changed names on the H1CDS_SOFTWARE_SUS_WATCHDOGS_TST.adl screen as well)

LHO General
patrick.thomas@LIGO.ORG - posted 10:30, Friday 15 March 2013 - last comment - 21:11, Monday 18 March 2013(5815)
optics labs dust monitors temp & rh sensors
I added temperature/relative humidity sensors to the dust monitors in the optics lab, vacuum prep lab and bake oven room.
Comments related to this report
patrick.thomas@LIGO.ORG - 21:11, Monday 18 March 2013 (5838)
It seems that the dust monitors in the optics lab have been repeatedly giving errors since this was done. Strange.
H1 CDS
david.barker@LIGO.ORG - posted 10:30, Friday 15 March 2013 (5814)
new h1isi ham models installed

We restarted the h1isiham2 and h1isiham3 models. The new models have ODC parts added. The DAQ was restarted.

This closes out WP3768

H1 SEI
hugh.radkins@LIGO.ORG - posted 08:02, Friday 15 March 2013 (5812)
LVEA HEPI Fluid Flush Status

Re 5807 I did another hands on inspection of all the welded and threaded joints at HAM3 and find no indication of fluid.  The servo is still running at 26psi and the reservoir fluid level has not dropped.  We will run at substantially higher pressure once the Actuators are introduced to the system but I do believe we are OK here.  I'll continue to monitor this closely.

H1 IOO
deepakmanoj.nandakumar@LIGO.ORG - posted 20:52, Thursday 14 March 2013 (5811)
Input Optics Work

 

[Deepak N., Rodica M., Chris M.]

* After Besty's realignment of the MC3 BOSEMs we found that the mode cleaner was not aligned.  Deepak checked the top stage earthquake stops, and they all seemed to be fine.  After noticing that the RT BOSEM was mis-centered by ~4000 counts we decided to recenter it.  This brought the mode cleaner much closer to flashing, but we still had to make some small tweaks afterwards.  The moral of the story is that the meaning of the alignment numbers is rather sensitive to the BOSEM centering.  I guess this isn't terribly surprising because, as Jeff K. pointed out, the magnetic field of the BOSEMs is designed to roll off quickly outside of the opening.  

 

* We attempted to set the polarizaiton of the recycled MC Refl beam using the HWP as mentioned in yesterday's post, but ran into a confusing inability to do so. Hmm...

H1 SUS
jeffrey.kissel@LIGO.ORG - posted 19:21, Thursday 14 March 2013 (5810)
Another Round of TMSY De-rubbing
J. Kissel, B. Bland

Betsy took another crack at de-rubbing the TMSY this morning. Her words which I've paraphrased: "I definitely readjusted the stiffness of the cables between the optical table and top mass, and around the top mass. But that aside, it really does look totally free -- I softly bounced, rocked and rolled it and it's definitely has a good range of motion before hitting its stops."

Since, I've taken another set of transfer functions (see attached). It looks looks significantly better than previous measurements:
- Vertical Modes are no longer double-peaked
- The most prominent Roll feature between 0.65 and 0.8 Hz has changed shape to be less complicated, lower in frequency, and dare-I-say fewer modes
- Whatever high frequency noise that appeared in yesterday's Tranverse TF has gone away again
- The second L/P mode at 0.8 Hz has moved down to 0.75 Hz
- TFs in general are a little bit cleaner
But until we get a similarly detailed measurement under vacuum, or a model we can trust, I can't definitively say "we're good." But I certainly can be convinced that what noise that's left that we see is from air-currents and ambient noise. I'll let team-TMS make the final call.

Non-image files attached to this report
H1 General
ryan.fisher@LIGO.ORG - posted 17:48, Thursday 14 March 2013 - last comment - 18:32, Wednesday 19 June 2013(5809)
ODC Summary Bits Enabled for all LHO Channels and Remaining ODC Work:
Each ODC channel has a Summary Bit that is used to indicate whether each system is in a nominally good operating state.  This bit is generated by checking that the necessary bits in the ODC channel are ON, as specified by the subsystem commissioners.  With the input from Jeff Kissel and Sebastien Biscans, we have set all of the EPICS variables needed to correctly populate the Summary bits for all active ODC channels in H1.  The EPICS variable Bitmasks are decimal numbers that correspond to a bitmask, where for each bit in the ODC channel that should be on when the system it is characterizing is in a good working state.  Here is an example:
H1:SUS-BS_ODC_COMBINE_ODC_BITMASK = 126
In binary, this is 0111 1110, which indicates that bits 6, 5, 4, 3, 2, and 1 should all be ON when the SUS-BS is operating nominally.  If you are curious what those bits mean, you can do the following:

controls@opsws1:~ 0$ caget H1:SUS-BS_ODC_BIT1
H1:SUS-BS_ODC_BIT1             Master Switch ON/OFF Status
controls@opsws1:~ 0$ caget H1:SUS-BS_ODC_BIT2
H1:SUS-BS_ODC_BIT2             DACKILL Status

These BITMASK EPICS records were set using scripts located in /ligo/home/ryan.fisher, and attached to this entry (with modified file extensions so the alog will post them).

With this completed, the EPICS description strings for the bits set and the MEDM screens reworked, we believe we have most of the active SUS and SEI subsystem ODC channels operational.  

The remaining systems for which ODCs need to be implemented under SUS and SEI are: the HAUX (HAM-Auxilliary) front ends and the HAM-ISI front ends.  The latter should be implemented with the next update from the cds_user_app svn (anticipated Mar 15).

We will be generating new safe.snap files specifically for the updated ODC EPICS settings for all models and sending them to CDS-announce for incorporation into the SVN and live systems.
Non-image files attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 14:11, Friday 15 March 2013 (5817)
I added scripts for setting the ODC bit labels and masks in /opt/rtcds/userapps/release/cds/h1/scripts

h1setODCbitmask:    Set ODC bit masks for HPI, ISI and SUS
h1setODCbitstrings: Set ODC bit labels for HPI, ISI and SUS
ryan.fisher@LIGO.ORG - 15:27, Wednesday 24 April 2013 (6193)
Missed one file for ISI-HAM bitmasks, it should be attached.

I also placed corrected versions of Stefan's scripts in:
/ligo/home/ryan.fisher/EPICS_Scripts

The equivalent files are attached.
Non-image files attached to this comment
ryan.fisher@LIGO.ORG - 17:38, Wednesday 19 June 2013 (6812)
I have added EPICS bit label and bit mask settings for the ODC channels for the seismic systems listed below to the scripts:

/ligo/home/ryan.fisher/EPICS_Scripts/h1setODCbitstrings 
and 
/ligo/home/ryan.fisher/EPICS_Scripts/h1setODCbitmask :

Seismic systems added:

HPI-HAM4
HPI-HAM5
HPI-HAM6
HPI-ITMX
ISI-HAM4
ISI-HAM5
ISI-HAM6
ISI-ITMX

I have also added HPI-ETMX and ISI-ETMX to the script for when/if these are added.

I have run these scripts to re-apply all settings as of GPS: 1055723826
H2 SUS
keith.riles@LIGO.ORG - posted 17:29, Thursday 14 March 2013 (5806)
A belated look at spectral lines in quiet H2 OAT channels
To check for possible spectral lines in the H2 One Arm Test due purely to the aLIGO DAQ / electronics, I picked a couple of channels that appeared very quiet in Greg Mendell's Fscan plots for SUS channels and applied the same spectral averaging procedure as used in previous studies of the  OAT feedback channel.

Many detailed spectra are posted here, with a sampling attached to this report. As can be seen in the first column of plots at the above link, the first channel examined, H2_SUS-ETMY_L1_OSEMINF_LL_IN1_DQ, has an unsurprising comb of 60-Hz harmonics (indicated with 'A' in the plots), with odd harmonics much stronger than even harmonics. In addition, though, it has an even stronger comb of 56.8407-Hz harmonics (indicated by 'B' in the plots, with two aliased reflections in the 900-1000 Hz band indicated by 'R' in the plots). This latter comb was previously seen in the OAT feedback channel and showed correlation with EY seismometer, accelerometer, magnetometer and microphone channels.

I tried looking at the corresponding ITMY chanel, H2_SUS-ITMY_L1_OSEMINF_LL_IN1_DQ, and found the 60-Hz harmonic comb to be lower and the other comb to be absent. The 2nd column of plots at the above link shows the spectra for the ITMY channel using the same scale as for the ETMY channel. The 3rd column of plots shows the ITMY channel with a zoomed-in vertical scale to look for evidence of other spectral lines. With the exception of a few lines that appear only on August 25, no other sharp structure above 10 Hz is apparent (although there are broad bumps here and there, for example, near 400 Hz).

In the table of plots, each spectrum for each channel is shown for three cases: 1) individually for the days of July 16, 20; August 4, 22, 25; 2) individually for the days of August 28, 31; September 7, 13; and 3) averaged together over all the days of July 16, 20; August 4, 22, 25, 28, 31; and September 7, 13. These are all the days for which enough "science mode" OAT data was produced to warrant Fscan SFT production. Click on thumbnails to get pdfs.

Jeff Kissel confirmed that both of these channels were indeed connected to OSEM sensors during the One Arm Test, but nominally the full band above 0.5 Hz should have been dominated by sensor noise, as shown in this plot he provided.

The spectra attached to this report show some examples of the above for the full 0-1000 Hz band, where approximately 200 half-hour SFTs (0.5 mHz binning) have been averaged together with inverse noise weighting.

Non-image files attached to this report
H1 SEI
stefan.ballmer@LIGO.ORG - posted 16:48, Thursday 14 March 2013 - last comment - 17:31, Thursday 14 March 2013(5804)
ISI ODC screens updated
Finally, I also updated the HAMISI ODC screens to include the bit label read-back.

Note that the HAMISIs don't have the ODC code in the front end - this still needs to be one.


This concludes my round of medm screen updates. 
Comments related to this report
ryan.fisher@LIGO.ORG - 17:31, Thursday 14 March 2013 (5808)
Note that the HAM-ISI models will be restarted tomorrow and should include the ODC channels.
H1 General
ryan.fisher@LIGO.ORG - posted 21:50, Wednesday 13 March 2013 - last comment - 12:17, Friday 03 May 2013(5788)
Setting ODC EPICS Descriptive Bit Strings
Ryan Fisher, Stefan Ballmer

We have set the EPICS records for the labels that describe the ODC bits for the SUS and ISI models currently running in H1.*  The labels provide a short description of what every bit in each ODC channel represents, for each ODC channel.  The higher order bits in the ODC channels that do not have labels are currently unused, but may be used in the future.

These records were set using scripts located in /ligo/home/ryan.fisher, and attached to this entry.  

* Excluding H1:SUS-TMSY_ODC_BIT* for now.
Non-image files attached to this report
Comments related to this report
ryan.fisher@LIGO.ORG - 12:00, Thursday 14 March 2013 (5791)
I have now also set all of the EPICS strings that describe the bits in the ODC channels for the HPI models.  The script is attached.
Non-image files attached to this comment
ryan.fisher@LIGO.ORG - 13:56, Thursday 14 March 2013 (5792)
I have now also set all of the EPICS strings that describe the bits in the ODC channels for the TMSY model.  The script is attached.

Note that there is a small bug in the models for these channels that currently mislabels bit4 as bit6: 
H1:SUS-TMSY_ODC_BIT6
(should be H1:SUS-TMSY_ODC_BIT4)

We will go through the process of changing this small bug in the corresponding Simulink model library part, after emailing cds_announce and sus team members.  
Non-image files attached to this comment
stefan.ballmer@LIGO.ORG - 14:12, Friday 15 March 2013 (5818)
I added scripts for setting the ODC bit labels and masks in /opt/rtcds/userapps/release/cds/h1/scripts

h1setODCbitmask:    Set ODC bit masks for HPI, ISI and SUS
h1setODCbitstrings: Set ODC bit labels for HPI, ISI and SUS
ryan.fisher@LIGO.ORG - 15:19, Wednesday 24 April 2013 (6194)
Correction to scripts included above: 

The ISI-HAM ODC channels should only have 5 bits, and should not be part of the other ISI channel script.  The corrected scripts are attached, and the correct scripts are in the following location on llocds.ligo-wa.caltech.edu:

/ligo/home/ryan.fisher/EPICS_Scripts

Non-image files attached to this comment
ryan.fisher@LIGO.ORG - 12:17, Friday 03 May 2013 (6257)
I corrected a small typo (M2 where it should have been M1) in SUS_OMC_Labels.txt for this log book entry only.  The scripts in /ligo/home/ryan.fisher were corrected approximately a month ago.  The corrected txt file is attached.
Non-image files attached to this comment
H1 ISC
keita.kawabe@LIGO.ORG - posted 22:31, Tuesday 12 March 2013 - last comment - 14:15, Friday 15 March 2013(5774)
50-50 IR beam splitter is a good AR for green (Corey, Keita)

Seems like the problem we had when using a green beam reflected off of a 50-50 IR beam splitter was because of the fact that 50-50 surface for 1064nm works as a very good AR for 532nm.

The BS in question is this: https://dcc.ligo.org/LIGO-E1000671-v1

Corey used a fiber coupled green laser as the source, placed a PBS cube for 532nm to ensure that the P-pol light is transmitted, and used a HWP for 532nm to change the polarization.

There are two beams in reflection, one from 50-50 IR BS surface and one from the back surface. There is one beam spot in transmission, because the 50-50 IR BS surface is very small. Anyway:

  P-pol S-pol
50-50 IR surface reflection 0.92% 0.044%
AR surface reflection 6.3% 20.7%
Transmission 92% 79%

On TMS ISC table (and in the arms) the green light has S-polarization, and according to Corey's measurement we're totally screwed.

Now that we know about this, we could either align the red QPD path assuming that there is an invisible beam next to a visible beam at a known distance, or we could place a HWP on the ALS table and inject a wrong polarized beam to see if we get anything.

Corey will write a one-page document and add it to E1000671.

Comments related to this report
corey.gray@LIGO.ORG - 14:15, Friday 15 March 2013 (5819)

Note added to E1000671v1 is here.