PSL plots

For the 35W beam. The DBB plots were taken with the modecleaner locked and driving the FSS.

The amplifier diode power level has dramatically dropped but it is hard to know if that is an actual drop or a result of the decreasing humidity in the laser room, which is now at its lowest point yet.

Images attached to this report
Non-image files attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 22:19, Sunday 13 January 2013 (5091)
CP1 and CP2 level control still feezing when outside temps are below 23F (ongoing prob)

Efforts made months ago to fix the dryness of the instrument air supply to the Corner Station succeeded in drying out the instrument air but haven't fixed the freezing of the LLCV needle valves of CP1 and CP2 for sub 23F temperatures -> For tonight, I added low-wattage drop lights engulfed within foil tents to thaw stuck parts and then exercised/stroked the needle valves to free them up -> The next step is to stop flow through the transfer lines to allow them to warm-up to ambient and then to remove the needle mechanism to inspect/dry.  This will have to be done during the daytime.  

2225 hrs. local -> Leaving site now
H1 IOO
paul.fulda@LIGO.ORG - posted 19:49, Friday 11 January 2013 (5090)
IMC locking

[Cheryl V., Chris M., Kiwamu I., Keita K., Michael R., Rick S., Paul F.]

This afternoon the input modecleaner was locked!

Effectively the same feedback filters as in L1 were used.  The lock was achieved as soon as the VCO feedback path was enabled, at a visibility of roughly 70%. This is shown in the attached stime series; the first lock occurred at around GPS time 1041991500. Soon after, the alignment was adjusted to try to improve the visibility, probably causing the loss of lock. After relocking, the alignment was adjusted, bringing the visibility to around 94%.

One of the LVEA cameras is currently pointing at a TV monitor which shows the IMC_REFL beam (a very temporary solution until we can get the signal directly to the control room). The embedded youtube video shows the IMC_REFL beam at the best visibility so far achieved.

The second attached file shows the power spectrum of the PSL frequency control signal, calibrated in the same way as for L1. The character of the H1 IMC and L1 IMC frequency feedback signals are quite different below about 60Hz. Next week we can look at this, and many more features of the locked IMC, in more detail.

Images attached to this report
Non-image files attached to this report
LHO General
patrick.thomas@LIGO.ORG - posted 19:38, Friday 11 January 2013 (5089)
plots of dust counts
Attached are plots of dust counts > .3 microns and > .5 microns in particles per cubic foot from approximately 6 PM Jan. 10 to 6 PM Jan. 11. Also attached are plots of the modes to show when they were running/acquiring data. It appears from the plots of the status, that the dust monitor at location 15 alarmed once for a calibration failure (status 1).
Non-image files attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 18:10, Friday 11 January 2013 (5088)
Leak checked marked flanges on HAM3 and started pumping HAM4 annulus @ 1800 hrs. local
Hope to leak test HAM2 on Monday
LHO VE
kyle.ryan@LIGO.ORG - posted 18:09, Friday 11 January 2013 (5087)
HAM2-HAM3 pumpdown
1800 hrs. local 

1.5 x 10-6 torr @ turbo inlet
LHO VE
kyle.ryan@LIGO.ORG - posted 18:09, Friday 11 January 2013 (5086)
HAM2-HAM3 pumpdown
1800 hrs. local 

1.5 x 10-6 torr @ turbo inlet
LHO General
dale.ingram@LIGO.ORG - posted 16:28, Friday 11 January 2013 (5085)
Ops summary
* Apparently the tumbleweed crew worked throughout the day between MX and EX.
* The roll-up door crew at EY finished the door and departed.
* Apollo installed the blanks on the HAM 6 septum and put the doors back on HAM 6 (see WP 3645).  The HAM 6 doors are completely bolted and torqued.
* Currently the LVEA is in the modified laser safe condition.  It appears that this condition will remain in place throughout the weekend.  The IO table is shut and locked at present.

Dust monitor 15 alarmed periodically throughout the day.  H0PEM_DUST.adl shows 15 near BSC 4 (H2 BS); Patrick suspects that its actual location is near HAM 2 or HAM 3.  The monitor that the IO team unplugged might be #9.



H1 SUS
mark.barton@LIGO.ORG - posted 15:46, Friday 11 January 2013 - last comment - 15:05, Monday 14 January 2013(5083)
BS TFs
Mark B.

I did some TFs on the BS using DTT:

/ligo/svncommon/SusSVN/sus/trunk/BSFM/H1/BSTST/SAGM1/Data/2013-01-11_1230_H1SUSBSTST_M1_*_WhiteNoise.xml

It's pretty sick - L and Y are not to bad, but R, T, V and P are nothing like what they're supposed to be.
Non-image files attached to this report
Comments related to this report
betsy.weaver@LIGO.ORG - 15:05, Monday 14 January 2013 (5096)

Travis found and relieved a brushing barrel EQ stop on the top of the Penultimate mass.  Back to TFs.

H1 SEI
hugh.radkins@LIGO.ORG - posted 15:32, Friday 11 January 2013 - last comment - 13:54, Monday 14 January 2013(5082)
WBSC1 IAS/HEPI
IAS has given me numbers.  I am unable to move HEPI the requested amount (8mm -X) until the East side Caging Brace Front Faces are moved (removed).  Hopefully we'll be able to reinstall these which allows the HEPI to be locked.  Once this is done we can attempt the translation East.  I believe there is room otherwise for this move.  Caveat: The Actuators are not attached; they have interferences that are not directly determinable now.
For the record, contrary to many statements from SEI to date, BSC1 has only been marginally locked if it can be said it was locked at all.  Most lock bolts were not engaged and only a few of the 40 were doing anything.
Comments related to this report
betsy.weaver@LIGO.ORG - 13:54, Monday 14 January 2013 (5095)

Correction:  The error is in the -Y direction, not -X.

X1 SUS
jeffrey.bartlett@LIGO.ORG - posted 15:08, Friday 11 January 2013 (5081)
Phase 1b TFs on India HLTS PR3
   We run the Phase 1b transfer functions on the M1 level of the India HLTS PR3. The plots are posted in the attached files. The undamped results are in the files from 01/11/2013, the damped results are in the files from 01/10/2013. The results look good. If there are no issues with these data, we will run the power spectra tests.
Non-image files attached to this report
H1 SEI
vincent.lhuillier@LIGO.ORG - posted 15:02, Friday 11 January 2013 (5080)
BIO switches - ISI Teststand

Yesterday, while turning on the interface chassis of the ISI teststand, I was very surprised to hear the relays of the T240s switching at a very fast pace. That was the first time I intended to use the teststand since the RCG update (to 2.6.1). Since the other BSC-ISIs (ETMY - ITMY) are working, we doubted the problem could come from the new code but who knows?
Few elements were checked (Voltage, BIO interface chassis). Nothing was abnormal. The Binary card was swapped, but it didn't fix the problem.

We figured that h1isitst.mdl was broken (Who did a change?). The binary outputs were sent to the DAC and 2 of the GS13 broadcast signals were send to the Binary output. That's why turning on GS13s was switching relays. The model was fixed.

H1 SUS
jeffrey.bartlett@LIGO.ORG - posted 12:06, Friday 11 January 2013 (5079)
Install Beam Splitter Stay Legs
Andres and Jeff reinstalled the Beam Splitter Stay Legs this morning using the modified Stay Leg Brackets.  
H1 SUS
mark.barton@LIGO.ORG - posted 10:42, Friday 11 January 2013 - last comment - 11:16, Friday 11 January 2013(5077)
ITMy OL values
Mark B and Betsy.

The values being returned from the L2 AOSEMs were a bit odd, so Betsy and Travis investigated. It was decided to replace the LL OSEM which had been reading only some 14K counts. Old SN was 443, new is 491.

We also concluded that the gains and offsets in the L2 OSEMINF block weren't up to date, because even with the AOSEMs backed out to the OL condition, the UL, UR and LR _OUT values weren't +15000. So we remeasured the OLs and recalculated the gains and offsets:

prettyOSEMgains(logOLs('H1','ITMY'),'ITMY')

...
L2UL 25076 1.196 -12538
L2LL 22916 1.309 -11458
L2UR 17540 1.710  -8770
L2LR 18608 1.612  -9304

Betsy entered these and as hoped all four _OUT values were then very close to 15000.
Comments related to this report
mark.barton@LIGO.ORG - 11:16, Friday 11 January 2013 (5078)
For good measure we did the same again for L1 OSEMs:

L1UL 26053  1.151 -13027
L1LL 24627  1.218 -12313
L1UR 28574  1.050 -14287
L1LR 21056  1.425 -10528
H1 General
dale.ingram@LIGO.ORG - posted 09:15, Friday 11 January 2013 (5076)
Friday morning items
  • The NPRO turned off last night. Michael Rodruck will turn it back on today.
  • The IO team turned off the dust monitor at HAM 2.
  • Through 9:00 AM this morning we were in laser hazard. Keita is now transitioning the LVEA to laser safe.
  • Michael Rodruck anticipates that later today we will enter the dual-status condition in which the partitioned area around HAM 2 will be laser hazard and the rest of the LVEA will be laser safe. In this condition, any personnel working at height near the input arm will need to check with Michael Rodruck before starting the work.
H1 IOO
paul.fulda@LIGO.ORG - posted 18:32, Thursday 10 January 2013 (5075)
IOT2L progress

[Keita K., Cheryl V., Kiwamu I., Chris M., Paul F.,]

Today's progress on IOT2L:

LHO General
patrick.thomas@LIGO.ORG - posted 18:21, Thursday 10 January 2013 (5074)
plots of dust counts
Attached are plots of dust counts > .3 microns and > .5 microns in particles per cubic foot from approximately 6 PM Jan. 9 to 6 PM Jan. 10. Also attached are plots of the modes to show when they were running/acquiring data.
Non-image files attached to this report
H1 PEM
james.batch@LIGO.ORG - posted 17:28, Thursday 10 January 2013 (5073)
Installed new front end computer at Mid X
Installed network switches, h1pemmx computer, timing fanout, IRIG-B chassis, I/O chassis, and AA chassis in the Mid X VEA.  Tested switches and computer, they are operational.  The entire system is waiting for DC power and a connection from the vault seismometer to the AA chassis to start data collection.
H1 SUS
mark.barton@LIGO.ORG - posted 21:52, Wednesday 09 January 2013 - last comment - 16:20, Friday 11 January 2013(5062)
MC1 damped TFs
Mark B. and Cheryl V.

Cheryl adjusted the damping gains on MC1, MC2 and MC3 to fix the overdamping observed in recent TFs. A fresh round of MC1 damped TFs will commence shortly to confirm the effectiveness. The Measurement Status indicator will go on for the duration of the measurement and then go off automatically.
Comments related to this report
mark.barton@LIGO.ORG - 08:01, Thursday 10 January 2013 (5064)
Mark B.

Data taken, analysis pending: /ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/MC1/SAGM1/Data/2013-01-09-1041832585_H1SUSMC1_M1_0p01to50Hz_tf.mat
mark.barton@LIGO.ORG - 16:20, Friday 11 January 2013 (5084)
Mark B.

New damped TFs look rather more reasonable. However even better filters and gains based on LLO experience are being installed and there'll be another round of TFs in a few days.
Non-image files attached to this comment