jeffrey.bartlett@LIGO.ORG - posted 10:09, Thursday 18 October 2012 (4475)
Phase 1b Testing of H1-SRM V2.0
   I reran the TFs last night on HSTS H1-SRM metal build. The results are much improved over the first set of transfer functions. All scripts, data, and results files have been added to SVN.

   The team is centering the AOSEMs on the M2 and M3 two stages so we can try to run power spectra during lunch.   
Non-image files attached to this report
H1 SUS
mark.barton@LIGO.ORG - posted 09:30, Thursday 18 October 2012 - last comment - 10:14, Thursday 18 October 2012(4474)
MC1 OL values, gains and offsets
Mark B. and Betsy B.

Yesterday afternoon, Betsy set up the MC1 for measuring OL values and I logged the values. Later, she set the OSEMs to half range (based on staging building gains and offsets).

This morning we computed the final offsets and gain and entered them into MEDM:

M1T1 25624.2 1.1708  12812.1
M1T2 28820.2 1.0409  14410.1
M1T3 27449.3 1.0929 13724.65
M1LF 27304.6 1.0987  13652.3
M1RT 23813.5 1.2598 11906.75
M1SD 25168.3  1.192 12584.15
M2UL 26288.6 1.1412  13144.3
M2LL 24343.8 1.2323  12171.9
M2UR   24540 1.2225    12270
M2LR 24557.9 1.2216 12278.95
M3UL 27280.5 1.0997 13640.25
M3LL 25514.3 1.1758 12757.15
M3UR 26364.8 1.1379  13182.4
M3LR 26645.6 1.1259  13322.8

I created two new scripts to make this easy:
^/trunk/Common/MatlabTools/logOSEMs_IN1_H1MC1.m logs the OL values
^/trunk/Common/MatlabTools/prettyOSEMgains.m creates a pretty table, as above.
Comments related to this report
betsy.weaver@LIGO.ORG - 10:14, Thursday 18 October 2012 (4476)

Images attached to this comment
LHO VE
kyle.ryan@LIGO.ORG - posted 19:20, Wednesday 17 October 2012 (4473)
PT343B discharged -> ignore alarms associated with this gauge until further notice


			
			
X1 SUS
jeffrey.bartlett@LIGO.ORG - posted 17:11, Wednesday 17 October 2012 (4472)
Phase 1b Testing of H1-SRM
   Ran TFs on H1-SRM last night, see attached plots. Results showed rubbing, suspected as being on the Top Mass side. A check of the suspension found both the left side EQ stops were touching the mass, which we backed off. We rechecked all other EQ stop and adjusted as necessary. We centered the Tablecloth, BOSEMs, and reset to 50% light. The TFs are setup to run tonight starting at 18:00.
Non-image files attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 16:43, Wednesday 17 October 2012 (4471)
Closed X1-9 beam tube pump port valve at X-mid
PT343B never "fired" following a previous power failure -> closing this valve isolates PT343B from the beam tube (pump) which may result in a large enough rise in pressure "seen" by the gauge to support discharge
H2 DAQ
james.batch@LIGO.ORG - posted 16:29, Wednesday 17 October 2012 (4468)
h2nds0 restarted, again.
daqd process died, final messages include 
"start net-writer 1034523398 17204 { long list of channels", 
"->4 kill net-writer 0", 
"connection on fd 15 closed", 
"read(seq_num) from NDS failed; errno=0", 
"connection closed on fd=16", and 
"Ask for retransmission of 20 packets; port 7097" which was repeated many times, then 
"Have to skip 22 packets (retry limit exceeded)" as the last entry.

Restarted daqd process.
H1 SUS
mark.barton@LIGO.ORG - posted 16:23, Wednesday 17 October 2012 (4467)
MC3 TFs
Mark B.

After Filiberto and Betsy had poked around on MC3, Betsy was able to start a yaw TF with DTT and see coherence, so the electronics is better off than it was.

To see where we're not at, I started a suite of Matlab TFs at about 16:20 and will check on it from home tonight.

As of 10:35, the measurement has failed twice with the USER DACKILL (but not the WATCHDOG) tripping. Trying one more time…

USER DACKILL found tripped at 06:35, after end of third attempt - giving up for now.
LHO General
gerardo.moreno@LIGO.ORG - posted 16:10, Wednesday 17 October 2012 (4466)
Ops Summary
H1 SEI
hugh.radkins@LIGO.ORG - posted 15:38, Wednesday 17 October 2012 (4465)
WHAM2 HEPI Actuator attachment now complete
With my help, Ed & Zach of Apollo, installed and attached the HEPI Actuators on WHAM2.  We did have a few mechanical issues as the positioning of the chamber (and hence the Support Tubes and HEPI Housings) drifted from the Initial alignment baseline.  We had to increase the size of a few slots to allow travel of connecting components.  Still all attached and it appears we have no motion issues.   We'll see for sure when we make range of travel measurements after the HEPI plumbing has us back operating.
H1 SUS
jeffrey.kissel@LIGO.ORG - posted 12:39, Wednesday 17 October 2012 - last comment - 16:30, Wednesday 17 October 2012(4464)
Cursory check of MC3 Software -- Looks OK
B. Bland, F. Clara, J. Kissel

At the request of Betsy, given the confusion on the inability to drive MC3 (see LHO aLOG 4462), I took a quick look through the software side of things to be sure all was well, and indeed I see nothing out of the ordinary. 

Here's my assessment:
- There have been software changes involving the IOP model (modifying the IOP watchdog) on Oct 09 (see LHO aLOG 4402), and also hardware changes (swapping out MC3-only, T3LFRTSD quadrapuss cable) on Oct 11 (see LHO aLOG 4423), before Mark attempted to take transfer functions on Oct 12-13 (see LHO aLOG 4432) which show no coherence (again, according to LHO aLOG 4462).
- Filiberto reports that 
     - He can see drive going into (from the AI chassis) the coil driver chassis (though I didn't catch exactly to which cables he was referring), implying the that software checks out, i.e.
          - The BIO I/O is functional (arranged/routed properly in software, traveling well across shared memory)
          - The IOP watchdog is allowing signals to flow
     - He cannot see anything coming out of either the COIL or TEST outputs (going towards the SatAmp on the way to in vacuum).
- Betsy reports that 
     - MC1 is able to drive and transfer functions look normal, implying that
          - the shared cable between MC1 and MC3 remains functional
          - The BIO I/O is functional and allowing drive
          - The IOP watchdog is allowing signals to flow
     - She's cranked on the centering knobs for each OSEM involved with the T3LFRTSD and confirms correct sensor response implying that signal chains are connected to the right OSEM

From we conclude that something is awry with the drive signal chain between the input to the coil driver from the DAC / AI chassis to the tip of the in-vacuum quadrupuss cable for T3LFRTSD. This leaves the following possibilities:
- Coil driver board has malfunctioned
     - Can we see anything coming out the coil driver monitor boards in either TEST or COIL configuration? Looks like the noise monitor channels are white -- does this model not exist yet / is it not running?
- The Binary I/O switch chain had failed
     - (though the monitor bits indicate that the chain is functional, both in the TEST/COIL switch and LP switch)
- The rest of the ex-vac chain has failed somehow (Coil Driver to SatAmp cable, SatAmp pass-through, SatAmp to fake-feedthrough)
- The rest of the in-vac chain has failed (the quadrupuss has been swapped, but how about the D0900225?)

Good luck!
Comments related to this report
filiberto.clara@LIGO.ORG - 16:30, Wednesday 17 October 2012 (4470)
Found cabling issues inside Rack H1:SUS-C3. Cabling to Binary Ouput D1000725 (SUS-C3-12) and Binary Monitors for D1000725 (SUS-C3-11) were switched with those for PRM-PR3.
LHO VE
kyle.ryan@LIGO.ORG - posted 10:31, Wednesday 17 October 2012 (4463)
Opened unpumped iLIGO RGA volume into vacuum system at Y-end
(iLIGO RGA was valved out month(s) ago) 

Also, energized iLIGO RGA filament
H2 DAQ
james.batch@LIGO.ORG - posted 09:47, Wednesday 17 October 2012 (4461)
h2nds0 restarted
Medm overview showed white fields for h2nds0.

Found daqd process on h2nds0 had problems - last log message indicated "couldn't create interpreter thread; pthread_create() err=11" and "connection dropped on port 38943 from ; fd=17"

Two nds processes were running, one with an uninterruptible sleep state.  

Then the processes disappeared.

Restarted nds and daqd processes.
LHO General
patrick.thomas@LIGO.ORG - posted 19:32, Tuesday 16 October 2012 (4460)
testing dust monitor code
I am now running the modified dust monitor code at end Y.
LHO General
patrick.thomas@LIGO.ORG - posted 19:32, Tuesday 16 October 2012 (4459)
testing dust monitor code
I am now running the modified dust monitor code at end Y.
LHO General
patrick.thomas@LIGO.ORG - posted 18:50, Tuesday 16 October 2012 (4458)
plots of dust counts
Attached are plots of dust counts > .3 microns and > .5 microns in particles per cubic foot. The data was taken from approximately 6 AM to 6 PM.
Non-image files attached to this report
H1 SUS
betsy.weaver@LIGO.ORG - posted 18:19, Tuesday 16 October 2012 (4457)
H1 MC1 optic hung in suspension

Betsy, Travis

This afternoon, we swapped the metal MC1 dummy mass out and replaced it with the IMCF09 mode cleaner flat optic.  Using the auto optical lever to look at the prisms on each side, we set the roll of the mass to within ~0.1mm.  We then noticed a small pitch error by eye which we corrected for using the top mass sliding pitch mass.  This brought the 6 top BOSEMs back into range, as set during the build phase across the street.  We'll continue to tally and align all 6 DOF errors tomorrow.

H1 SEI
hugo.paris@LIGO.ORG - posted 16:33, Tuesday 16 October 2012 (4455)
HAM3-ISI - New medm screens for payload (SUS) monitoring

JimB, DaveB, HugoP,

HAM3-ISI used to have temporary medm screens for payload state monitoring. We rebuilt the HAM3-ISI model with a call for the create_hamisi_payload post-build script. This script automatically creates unit-specific medm screens for payload state monitoring. The call for create_hamisi_payload was set in the block properties of the HAM3_PAYLOAD block, in the simulink model of HAM3-ISI (see attachement #1).

We did the same for HAM2-ISI.

make install would then return an error due to issues with CDS. These issues were corrected.

I updated the HAM-ISI overview screen so it automatically points to the new payload screens. The unit-specific payload monitoring screens now work for both HAM2-ISI and HAM3-ISI (see attachement #2)

Images attached to this report
H1 PSL
michael.rodruck@LIGO.ORG - posted 16:29, Tuesday 16 October 2012 (4456)
Alarms changed

I removed the shutter alarms from the PSL alarm configuration file and added alarms for the power watchdogs on the amplifier and oscillator.

H1 SUS
filiberto.clara@LIGO.ORG - posted 16:24, Friday 12 October 2012 - last comment - 10:04, Wednesday 17 October 2012(4436)
MC1 Bottom Stage
Looked at MC1 (Bottom stage) UL channel, which showed low counts ~9000 compared to the expected 24,000 counts. Traced problem to IO Chassis (h1sush2a). Replaced ADC card SN110201-04 with SN110203-20. Problem still persisted, replaced ADC Adapter Board (D0902006 SN S1102390) and cabling inside IO chassis with a spare (no SN number). Have requested a replacement from CIT. Will replace sometime next week, but SUS is able to continue transfer functions on MC3.

Filiberto Clara
Comments related to this report
betsy.weaver@LIGO.ORG - 10:04, Wednesday 17 October 2012 (4462)

Mark reports that the TFs taken Friday, looked noisy.  The situation with MC3 deteriorates.  Picking up where Filiberto/Jeff left off last week, I find that indeed no EXC is getting through to the MC3 tops, hence the lack of coherence in Friday's measurements.  Because this is possibly related to what Fil did on Friday with respect to switching out some boards (alog above), he is now looking again at the boards.

 

While I was in the cleanroom, I chased a moth and finally caught it as it was perched on the MC1/MC3 shared in-vacuum quadra-puss.  Ugh - now I have to switch that cable out as well.

Displaying reports 81681-81700 of 86031.Go to page Start 4081 4082 4083 4084 4085 4086 4087 4088 4089 End