Displaying reports 8201-8220 of 84691.Go to page Start 407 408 409 410 411 412 413 414 415 End
Reports until 11:45, Monday 01 July 2024
LHO VE
david.barker@LIGO.ORG - posted 11:45, Monday 01 July 2024 (78780)
Mon CP1 Fill

Mon Jul 01 10:12:00 2024 INFO: Fill completed in 11min 56secs

Gerardo confirmed a good fill curbside.

Images attached to this report
H1 ISC
jennifer.wright@LIGO.ORG - posted 10:54, Monday 01 July 2024 - last comment - 14:33, Tuesday 02 July 2024(78776)
Improving f_CC with SRM moves

Sheila, Jennie W

We lost some range with the hot OM2 so we thought to see if we could move SRM and raise the coupled cavity pole to regain this range as we did before in this entry.

Took 5 minutes no squeeze time.

Opened POP beam diverter to monitor POPAIR

Following the procedure in this entry we opened the SRC1 ASC loops by turning off input and offsets simultaneously after changing ramp to 0.1s.

Set the ramp to 10s, turn gain to 0, clear history, gain back to 4.

Then we started to move SRM alignment sliders.

The starting alignment sliders are here.

started yaw steps down on SRM, wrong way as f_cc avg went down, changed to up in  yaw, then noticed POPAIR_RF18 was going down so switched to pitch steps down. Found a poitn at which it no longer chnaged fcc, went back to maximum found for f_cc. The trends can be seen here where the first vertical cursor is when we opened the beam diverter and the second is when we thought we had the highest coupled cavity pole (middle bottom trace). So we put the SRM sliders to their values for this time after finishing.

I then changed offsets in SRC1 ASC to -0.0417 for ASC-SRC1_P_OFFSET and 0.0977 for ASC-SRC1_Y_OFFSET but didn't turn them on while we grabbed 5 minutes no squeeze time.

17:05:30 UTC closed loops and switched on new offsets.

Sheila used the second period of no squeeze time to plot the how the range compares to June 24th when we had a cold OM2. This plot shows the DARM comparison on the top plot (Yellow is cold OM2, blue is hot OM2) and the range difference on the bottom. This plot shows the cumulative range comparison on the top (yellow is cold OM2, blue is hot OM2) with the same difference in range plotted on the bottom. From the bottom plot it looks like with hot OM2 we gain range between 25 and 40Hz, don't gain between 40 and 90 Hz and lose range for hot OM2 above this point.


So summary is, we didn't completely recover the range by raising the coupled cavity pole. Hot OM2 is good for low frequency range but not for high frequency. We still need to do some PSAMS tuning today so this might regain us range at higher frequencies.

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 11:23, Monday 01 July 2024 (78779)ISC

I changed the values (ASC_AS72_P and ASC_AS72_Y) in line 551 and 552 of lscparams.py to match the new offsets and reloaded ISC_LOCK guardian.

Then I accepted these offset values in sdf after changing the ramps back to their setpoint which was 5s.

Images attached to this comment
sheila.dwyer@LIGO.ORG - 14:33, Tuesday 02 July 2024 (78809)

The no sqz time for reference:

With OM2 hot, and SRM alignment adjusted as Jennie described above, we took no sqz time from 16:58- 17:05 UTC on July 1st.

H1 TCS
ryan.crouch@LIGO.ORG - posted 09:26, Monday 01 July 2024 (78775)
TCS Monthly Trend

Last checked in alog78246, closes FAMIS 28450.

For the CO2 lasers, ITMY and ITMX have dropped in power but only by a small amount (by ~1 and <1 [W]? I'm not sure on the units).

ITMY and ITMX HWS_SLEDPOWERMON dropped to zero last Tuesday as they were turned off due to alignment issues alog78654, and the SPHERICAL_POWER is still flat lined from the SR3 moves.

Images attached to this report
H1 ISC
camilla.compton@LIGO.ORG - posted 09:01, Monday 01 July 2024 (78774)
MICH FF tried but reverted

We tried the new hot OM2 MICH FF filter loaded in FM5 from 78688. The original FM6 FF was better, see attached, so we left everything as nominal. Is  MICH changing over time so that last weeks injections are not good today?

Images attached to this report
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 07:41, Monday 01 July 2024 - last comment - 07:51, Monday 01 July 2024(78772)
OPS Day Shift Start

TITLE: 07/01 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 155Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 3mph Gusts, 1mph 5min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.08 μm/s
QUICK SUMMARY:

IFO is in NLN and OBSERVING (3 hr 35 min lock)

Images attached to this report
Comments related to this report
ibrahim.abouelfettouh@LIGO.ORG - 07:51, Monday 01 July 2024 (78773)

Optics Lab (Lab 1) Dust Cts have settled since an excursion about 1 hour and 10 mins ago which saw elevated levels for 11 minutes at around 13:34 UTC (6:34 PT). They triggered the yellow alert, and almost got up to the red alert threshold but did not quite make it. Counts came back below threshold at around 13:45 UTC. Trend screenshot attached.

Images attached to this comment
H1 DetChar (DetChar)
sukanta.bose@LIGO.ORG - posted 06:55, Monday 01 July 2024 (78771)
Data Quality Shift Report 2024-06-10 to 2024-06-16

Link to report.

LHO General
ryan.short@LIGO.ORG - posted 01:00, Monday 01 July 2024 (78769)
Ops Eve Shift Summary

TITLE: 07/01 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Observing at 157Mpc
INCOMING OPERATOR: Oli
SHIFT SUMMARY: One lockloss early in the shift, but recovery was simple and have been observing since. H1 has now been locked for 7 hours.

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:08 SAF LVEA LVEA YES LVEA IS LASER HAZARD Ongoing
18:23 SQZ Terry Opt Lab local SHG work 21:47
21:05 SQZ Kar Meng Opt Lab local SHG work 00:18
22:40 PEM Robert LVEA y Looking for noise source 23:00
01:01 - Kar Meng EY n Driving to EY 01:25
H1 AOS
robert.schofield@LIGO.ORG - posted 18:02, Sunday 30 June 2024 (78767)
More ITM bias adjustment during observation

This is a continuation of the work discussed in 78734

 

Start of change (GPS)

End of change (GPS)

ITMX bias at start (V)

ITMX bias at end (V)

ITMY bias at start (V)

ITMY bias at end (V)

1403804070

1403804192

 

0

0

0

77

1403804365

1403804489

 

0

0

77

-80

1403807629

1403807751

 

0

0

-80

-20

1403809555

1403809816

0

-40

-20

-40

1403812204

1403812324

 

-40

36

 

-40

-40

1403816907

1403817127

 

36

36

-40

0

H1 PSL
ryan.short@LIGO.ORG - posted 17:11, Sunday 30 June 2024 (78766)
PSL FSS RefCav Remote Alignment Tweak

After H1 lost lock this afternoon, I took the opportunity to do a quick RefCav alignment tweak since the TPD was showing the transmission was low at around 680mV. Using the two picomotor-controlled mirrors in the FSS path, mostly adjusting in pitch, I was able to improve the signal from about 680mV to 870mV. More of an increase than I was expecting, but this should be good to last until the PMC is swapped out on Tuesday (where this alignment may need to be done again). I suspect I could've spent more time to improve this further, but I stopped here so as not to delay IFO locking any longer.

Images attached to this report
H1 General (Lockloss)
ryan.short@LIGO.ORG - posted 16:49, Sunday 30 June 2024 - last comment - 18:04, Sunday 30 June 2024(78765)
Lockloss @ 23:39 UTC

Lockloss @ 23:39 UTC - link to lockloss tool

Locked for 26 minutes. No obvious cause, but I see the ETMX motion about 100ms before the lockloss this time.

I'm going to take this opportunity to do a quick FSS RefCav alignment adjustment so that it can hopefully make it until Tuesday when we swap out the PMC.

Images attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 18:04, Sunday 30 June 2024 (78768)

H1 back to observing at 00:57 UTC. Pretty much a fully automated relock; I just slightly adjusted PRM to make buildups during PRMI better, but it might've caught on its own.

LHO General
thomas.shaffer@LIGO.ORG - posted 16:25, Sunday 30 June 2024 (78764)
Ops Day Shift End

TITLE: 06/30 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 151Mpc
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY: Just got back to low noise after we had a lock loss that ended a 15 hour lock. Relocking was fully auto and it didn't even run an initial alignment.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:08 SAF LVEA LVEA YES LVEA IS LASER HAZARD 10:08
18:23 SQZ Terry Opt Lab local SHG work 21:47
21:05 SQZ Kar Meng Opt Lab local SHG work 01:05
22:40 PEM Robert LVEA y Looking for noise source 23:00

 

LHO General
ryan.short@LIGO.ORG - posted 16:09, Sunday 30 June 2024 - last comment - 16:22, Sunday 30 June 2024(78761)
Ops Eve Shift Start

TITLE: 06/30 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 7mph Gusts, 4mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.09 μm/s
QUICK SUMMARY: H1 lost lock about an hour ago but is relocking well; so far up to TRANSITION_FROM_ETMX.

Comments related to this report
ryan.short@LIGO.ORG - 16:22, Sunday 30 June 2024 (78762)PSL

The RefCav TPD is reading down to 0.69V, showing a warning on DIAG_MAIN, and has been falling over the past couple weeks. Since this is likely due to the increased PMC loss lowering the output of the cavity, I don't suspect fixing the RefCav alignment will get too much out of it, but I can try an alignment this evening if the IFO loses lock.

Images attached to this comment
H1 General
thomas.shaffer@LIGO.ORG - posted 15:21, Sunday 30 June 2024 (78760)
Lock loss 2216UTC

Lockloss1403820932

Ending a 15 hour lock. This lock had the ETMX wiggles as well, but they happened almost a second prior to lock loss and were larger than I have seen.

Images attached to this report
H1 PSL
thomas.shaffer@LIGO.ORG - posted 10:45, Sunday 30 June 2024 (78759)
PSL Status Report - Weekly

FAMS26260


Laser Status:
    NPRO output power is 1.821W (nominal ~2W)
    AMP1 output power is 67.15W (nominal ~70W)
    AMP2 output power is 137.3W (nominal 135-140W)
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 32 days, 23 hr 34 minutes
    Reflected power = 22.75W
    Transmitted power = 105.2W
    PowerSum = 127.9W

FSS:
    It has been locked for 0 days 11 hr and 42 min
    TPD[V] = 0.6936V

ISS:
    The diffracted power is around 2.0%
    Last saturation event was 0 days 11 hours and 43 minutes ago


Possible Issues:
    PMC reflected power is high
    FSS TPD is low

LHO VE
david.barker@LIGO.ORG - posted 10:15, Sunday 30 June 2024 (78758)
Sun CP1 Fill

Sun Jun 30 10:09:55 2024 INFO: Fill completed in 9min 52secs

Images attached to this report
H1 CAL
thomas.shaffer@LIGO.ORG - posted 14:34, Saturday 29 June 2024 - last comment - 18:27, Wednesday 03 July 2024(78746)
Calibration Sweep 2106 UTC

Calibration sweep taken today at 2106UTC in coordination with LLO and Virgo. This was delayed today since we were'nt thermalized at 1130PT.

Simulines start:

PDT: 2024-06-29 14:11:45.566107 PDT
UTC: 2024-06-29 21:11:45.566107 UTC
GPS: 1403730723.566107

End:

PDT: 2024-06-29 14:33:08.154689 PDT
UTC: 2024-06-29 21:33:08.154689 UTC
GPS: 1403732006.154689
 

I ran into the error below when I first started the simulines script, but it seemed to move on. I'm not sure if this frequently pops up and this is the first time I caught it.

Traceback (most recent call last):
 File "/var/opt/conda/base/envs/cds/lib/python3.10/multiprocessing/process.py", line 314, in _
bootstrap
   self.run()
 File "/var/opt/conda/base/envs/cds/lib/python3.10/multiprocessing/process.py", line 108, in r
un
   self._target(*self._args, **self._kwargs)
 File "/ligo/groups/cal/src/simulines/simulines/simuLines.py", line 427, in generateSignalInje
ction
   SignalInjection(tempObj, [frequency, Amp])
 File "/ligo/groups/cal/src/simulines/simulines/simuLines.py", line 484, in SignalInjection
   drive.start(ramptime=rampUp) #this is blocking, and starts on a GPS second.
 File "/var/opt/conda/base/envs/cds/lib/python3.10/site-packages/awg.py", line 122, in start
   self._get_slot()
 File "/var/opt/conda/base/envs/cds/lib/python3.10/site-packages/awg.py", line 106, in _get_sl
ot
   raise AWGError("can't set channel for " + self.chan)
awg.AWGError: can't set channel for H1:SUS-ETMX_L1_CAL_EXC
 

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 09:48, Sunday 30 June 2024 (78757)

One item to note is that h1susex is running a different version of awgtpman since last Tuesday.

erik.vonreis@LIGO.ORG - 10:01, Monday 01 July 2024 (78777)

This almost certainly failed to start the excitation.

I tested a 0-amplitude excitation on the same channel using awggui with no issue.

There may be something wrong the environment the script is running in.

 

 

louis.dartez@LIGO.ORG - 13:48, Monday 01 July 2024 (78784)
We haven't made any changes to the environment that is used to run simulines. The only thing that seems to have changed is that a different version of awgtpman is running now on h1susex as Dave pointed out. 

Having said that, this failure has been seen before but rarely reappears when re-running simulines. So maybe this is not that big of an issue...unless it happens again.
louis.dartez@LIGO.ORG - 15:17, Wednesday 03 July 2024 (78842)
turns out i was wrong about the environment not changing. according to step 7 of the ops calib measurement instructions, simulines has been getting run in the base cds environment...which the calibration group does not control. That's probably worth changing. In the meantime, I'm unsure if that's the cause of last week's issues.
erik.vonreis@LIGO.ORG - 16:51, Wednesday 03 July 2024 (78846)

The CDS environment was stable between June 22 (last good run) and Jun 29.

 

There may have been another failure on June 27, which would make two failures and no successes since the upgrade.

 

The attached graph for June 27 shows an excitation at EY, but no associated excitation at EX during the same period.  Compare with the graph from June 22.

Images attached to this comment
erik.vonreis@LIGO.ORG - 18:27, Wednesday 03 July 2024 (78851)

On Jun 27 and Jun 28, H1:SUS-ETMX_L2_CAL_EXCMON was excited during the test.

Displaying reports 8201-8220 of 84691.Go to page Start 407 408 409 410 411 412 413 414 415 End