Reports until 00:02, Tuesday 24 July 2012
H2 ISC
bram.slagmolen@LIGO.ORG - posted 00:02, Tuesday 24 July 2012 - last comment - 00:15, Tuesday 24 July 2012(3555)
No PLL and No arm lock -> fiber beatnote out of range

Alberto and Bram.

With the strange increased power readings on the REFL_PD, which seemed to be linked to the Beckhoff PLL temperature feedback, and not being able to lock the PLL we headed to EY. After trying to find the beatnote (fiber output with the local laser), it turned out there was no light coming out of the fiber (even after Alberto relocked the RefCav). Keita re-relocked the RefCav.

With light coming out of the fiber, no beatnote was found until we tuned the crystal temperature from the original 31.27 deg C to 38 deg C (which seems to be it maximum). Even at that setting, it was at a few hundred MHz. All Beckhoff settings were turned 'off/cleared', as indicated on the screens.

To get the beatnote within range of the EY laser, I changed the RefCav to the next FSR (up in laser crystal temperature), with only a small change in beatnote frequency.

Adjusting the Lightwave (= RefCav) temperature down by ~0.5 deg C, requires the Innolight (EY) laser to adjust by 0.2 deg C (down), to follow the beatnote.

In the end I reduced the RefCav laser down by 5 deg C (to ~34.6 deg C), but tuning the laser temp of the Innolight laser did not to bring the beatnote down to ~40 MHz (still just above 100 MHz).

There is something fishy going on, but so far we didn't manage to find it.

Next we will unplug the temp feedback cable from the laser controller, just to be sure we don't have a secret offset.

Comments related to this report
bram.slagmolen@LIGO.ORG - 00:15, Tuesday 24 July 2012 (3557)

What is strange is that the RefCav FSR is 750 MHz (assuming 20 cm long RefCav), when locking the Lightwave laser to the 'next' FSR doesn't seem to move the beatnote frequency by a similar amount.

H2 SEI
fabrice.matichard@LIGO.ORG - posted 20:10, Monday 23 July 2012 - last comment - 23:43, Monday 23 July 2012(3553)
Rounded coefficents
Last week:
- we designed and installed new high blends in all 12 directions of the the BSC-ISI, in order to reduce low frequency gain peaking (both units)
- we designed and installed a .250Hz low blend in the longitudinal direction (Y) (both units)
this provided interesting improvement as reported in the last cavity measurement.

However,the measured isolation was not as good as expected (it did not follow the low blend as it should).

The problem seems to be in the blend filter conversion to the foton format: The low blend filters first coefficients are small and rounded.

The plot attached show the improvement after fixing the low blend in BSC8:
- Red curve is the reference (Stage 1 Damped)
- Blue Curve is Stage 1 Controlled, with rounded coefficients
- Green Curve is Stage 1 Controlled, with better filter coefficients

We'll work tomorrow on a solution to prevent this from happening again.

I leave both units up and running for Bram to possibly run cavity measurements. 







 
 
Non-image files attached to this report
Comments related to this report
bram.slagmolen@LIGO.ORG - 23:43, Monday 23 July 2012 (3556)

Thanks Fabrice, but no sigar ...

H2 AOS
thomas.vo@LIGO.ORG - posted 19:55, Monday 23 July 2012 - last comment - 17:17, Tuesday 24 July 2012(3551)
H2 ETMY/ITMY Optical Lever Quad Mappings via SUS Offsets
Jeff K, Jeff G, Thomas V


We thought it might be necessary to map out the quadrants of the ETMY and ITMY optical levers via offsets induced onto the suspension stages, namely M0.

After centering the beam on the optical lever QPD on ETMY, we offset the pitch on the M0 stage by a positive amount and expected the two lower quadrants of the QPD to have a spike in intensity (as deduced by the coordinate system set by SUS to determine which direction is positive and negative pitch and yaw).  Then offset yaw in a positive amount, which we expect to see the the two left quadrants to have a spike in intensity.  This was different than was what showing as pitch and yaw in the SUS medm and also is different than what I had used in my earlier calibration parameters.  This being said, the outcome of the mapping is as follows as viewed facing the QPD for ETMY:
13
42 

Repeating the test with the same process on ITMY OptLev, we found that the mapping didn't correspond once again to what was previously recognized as the mapping and is as follows:
42
13

The plot thickens:
As you can see, the two QPD's are not the same rotated about the center axis, which begs a confusing question of what might be going wrong. The most obvious answer is that the pitch signs are flipped, but where might this come from?  It could be a wiring issue in the hardware connecting the QPD to the ADC, a problem with a few negative signs somewhere in the readouts, or a difference in polarity of the output coil drivers F2 and F3 on the M0 stage of the ITMY.  The last option was found when we were trouble shooting and found that aLOG 2109 written by Jeff K. had addressed this issue but we couldn't find a log that resolved it.  It might be a combination of these things, still a work in progress.  It might be worth it to put in the glass viewport protectors and remove the enclosures to do a laser pointer test again on these QPDs.  Either way, the discrepancy between the two scenarios via the same process is perplexing.
Comments related to this report
thomas.vo@LIGO.ORG - 17:17, Tuesday 24 July 2012 (3573)
Opening up the Optical Levers on both ETMY and ITMY and used a laser pointer , I mapped out definitively the quadrants for both optical levers. This is as facing the front of the optical lever QPD and the channels correspond to the channels outputted by the SUS MEDM for ITMY and ETMY.

ETMY:
13
42

ITMY:
13
42

Now, the good news is that the OptLev electronics match between the ITM and ETM.  Also cross refrencing with LLO's HAM 3 optical lever QPD, it matches ours as well.  The interesting part is that when we give a positive pitch to the ITMY suspension, it shows exactly opposite of what we expect to see on the QPD.  A mystery still yet to be solved.  ETMY suspension and optical levers matched beautifully, a sign convention is the difference between them which come from a positive pitch misalignment on the suspension yields a higher intensity on the two lower quadrants of the optical lever QPD.  The same goes for positive yaw offsets on the suspension, both of these can be easily remedied.
LHO General
patrick.thomas@LIGO.ORG - posted 19:49, Monday 23 July 2012 (3552)
plots of dust counts
Attached are plots of dust counts > .5 microns in particles per cubic foot.
Non-image files attached to this report
H1 SEI
david.barker@LIGO.ORG - posted 17:57, Monday 23 July 2012 (3549)
h1isiham3 model recompiled and restarted

The h1isiham3 model running on h1seih23 was quite old, from June 8th. Before we start ISI testing we (Hugo and I) upgraded this model by bringing in the latest L1 HAM3 model and converting L1->H1. From now onwards we should maintain the h1isiham3 model as separate from the l1isiham3. The new model required new Dolphin IPC channels from the h1susmc2 model, so that model was rebuilt from L1  but was not restarted on h1sush34. I had some problems recompiling the h2suspr2 model, there are IPC errors between MC2 and PR2 which are currently unresolved.

There are still some errors with the H1 ISI HAM3 medm screens. There appears to be residual links to M1 and S1 systems. I'll work with Hugo to resolve the H1 screens.

H1 SUS
jeffrey.bartlett@LIGO.ORG - posted 17:36, Monday 23 July 2012 (3548)
Testing H1- MC3
   We have been working on resolving a post creek bake misalignment problem with the lower blade break off heights on H1-MC3. We have tuned out a lot of the difference, but the break off heights are still slightly out of spec. We are going to take transfer functions of the M1 level to see if MC3 performance is within tolerance despite the misalignment of the lower blade break off heights.  
  
   These are the open light values for BOSEMs on HSTS MC3. The offsets and gains, listed below' have been entered into MEDM. The BOSEMs have been adjusted to 50% light. Plan is to take a set of TFs tonight when the lab is quiet.    

OSEM    S/N    OLV       Offset     Gain

 T1     571    30360.8   -15180     0.988
 T2     565    28718.9   -14359     1.045
 T3     312    30896.5   -15448     0.971
 LF      74    27143.9   -13572     1.105
 RT     313    30112.7   -15056     0.996
 SD     694C   25573.4   -12787     1.173 
H1 SUS
betsy.weaver@LIGO.ORG - posted 16:38, Monday 23 July 2012 - last comment - 19:12, Tuesday 24 July 2012(3545)
More MC2 Phase 2b testing

This afternoon we obtained the spectra of the lower stages (M2 and M3) of MC2 in both the damped and undamped states (using plothsts_spectra.m).  The spectra were taken over the lunch hour so the LVEA activity should have been pretty quiet.  Data was committed to svn.  WIll post a PDF shortly.

As well, I used awggui to try to verify that I could see drive on the coils of these lower stages (for example, exc H1:SUS-MC2_M2_COILOUTF_UL_EXC while watching the OSEMINF_OUT_DQ channel for UL).  This will need to continue later as at first pass, I couldn't see much signal on the M2 AOSEMs.

Comments related to this report
jeffrey.kissel@LIGO.ORG - 08:53, Tuesday 24 July 2012 (3559)
More details / results to come, but I figured we'd post the results for people to independently peruse, while we continue to discuss.

Conclusions at this point:
 - The top to top (undamped) transfer functions look mostly excellent
 - The only thing that has raised a yellowish flag is the first longitudinal mode at 0.68 Hz shows a sharp notch just above it in frequency. Given that the rest of the magnitude looks entirely clean (and independently looking at the DTT .xmls confirming good coherence), I'm confident it's a real feature. However -- after exploring the model in detail we (re)remembered that the first L mode is at the exact same frequency as the first T mode. So, perhaps, in this particular instance the two modes are just different enough to break the normally degenerate mode.
 - Spectra look as expected given the out-of-vacuum environment. The results are notably and encouragingly *less* noisy at high frequency, confirming that the surprisingly large excess motion at high frequency seen in this SUS's Phase 1b results was merely excessive ambient noise in the X1 Test Arena.

To come:
- Damped transfer functions (results taken, just not yet analyzed)
- TF comparisons with other, previous results
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 19:12, Tuesday 24 July 2012 (3576)
Status Update:

I attach new versions of the above plots, which now includes the 2012-07-20, damping-loops-closed transfer functions compared against MC2's phase 1b measurement (2012-06-15) and a token L1 HSTS (L1 SUS MC1, 2012-06-04). The conclusions are still the same; everything looks pretty darn good, except for the non-degeneracy of the first Longitudinal and first Transverse mode. Note that the non-degeneracy disappears with damping loops on, so the loops are modifying the dynamics enough that there is no further "cross-coupling" (if indeed it is a mechanical/physical cross-coupling and not just sensor cross-coupling from not-perfectly-aligned-flags or something).

HOWEVER, after looking into the assembly tolerance for Roll, we found that, as currently constructed, MC2 is out of spec. So, Betsy and Travis will need to adjust the Roll of the optic (on Thursday), and therefore all measurements will have to be re-done, since it's a major change that would affect the dynamics. (Booooo.)

Thinking out loud: Hrmm... if the Roll is out of spec ... Roll is fundamentally coupled with Transverse ... maybe the roll in the optic is modifying (increasing? decreasing? the frequency -- can't tell) the first Transverse / Roll mode of the SUS, breaking the above mentioned degeneracy ... maybe this is why it's obvious in this SUS and not any SUS's prior? But this still doesn't explain how a L / P drive could produce R / T, which should be completely orthogonal degrees of freedom.
Non-image files attached to this comment
H1 SEI
hugh.radkins@LIGO.ORG - posted 16:35, Monday 23 July 2012 (3547)
WHAM3 SEI/HEPI Status Update--Actuators/Initial Alignment
I really messed up but at least I remembered!  For the past couple weeks as we've been doing initial alignment and Actuator connection, I failed to correct the elevation numbers in HAM3 for the Xarm slope.  Oh boy...In case you care to know---The LIGO coordinate frame is not co-planar with the gravity field in most places and in fact there is a 2.4mm shift between them moving from BSC2 (where we set them equal) just to HAM3.  This shift is what I forgot to include in my elevation calcs when positioning HAM3 ISI Optical Table.
So today we disconnected and relocked the eight HEPI Actuators and cranked the HEPI DSCW Springs to pull the platform up to position.  Jason (IAS) was called in ~1400 and he gave us a bit of horizontal correction.  Tomorrow we'll reconnect the Actuators and ask for another IAS shot to put this back to bed.
At least we are getting better at Actuator connecting!

Thanks to Eric, Mitchell and Jason.
LHO General
gerardo.moreno@LIGO.ORG - posted 16:12, Monday 23 July 2012 (3546)
Ops Summary

Activities of the day of which I was informed:

- Elli K. to visit Y-END station, ALS table work, 8:57 - 11:18 am.

- Karen R. clean floor in the LVEA, 9:00 - 11:58 am.

- Hugh R. work in and around HAM03, all day.

- Filiberto C. work in the corner station VEA, briefly around 10:38 am.

- Karen R. work @ Y-END, clean different areas, including the floor inside the vacuum support equipment room, 1:45 - 3:45 pm.

- Cheryl V. work at Y-END, TMS, 2:51 - 3:55 pm.

- Jason O. alignment work, 2:55 - 3:50 pm.

H2 ISC
alberto.stochino@LIGO.ORG - posted 15:28, Monday 23 July 2012 (3544)
ALS Reference cavity locked again

We found that the reference cavity had lost lock.

I relocked it by disconnecting the fast and slow inputs, tweaking a bit the temperature knob and re-connecting the input cables.

The cavity is now likely locked to a different FSR from before. This means that the beat note at the end station might now be falling out of the PD bandwidth.

We have to go to end Y and adjust the ALS laser temperature to get the beat note in the PD range again.

H1 AOS
thomas.vo@LIGO.ORG - posted 15:07, Monday 23 July 2012 - last comment - 08:50, Thursday 26 July 2012(3543)
H2 ETMY Optical Lever Calibration
I have taken the calibration data and analyzed it. The plot of the linear response curve is attached.

I used a polynomial fitting program in python to get these numbers.  This should give you the conversion from volts to micro-radians.

           Slope            Y-intercept
Pitch   [  1.01902201e+06   8.42886118e+00]
Yaw     [  9.83122943e+05  -1.19329010e+01]
Comments related to this report
thomas.vo@LIGO.ORG - 19:11, Monday 23 July 2012 (3550)
Please hold off on using these calibration parameters, it looks like the SUS readout channels that map out the quadrants on ETMY and ITMY were not correct and I will need to re-calculate if this is true. Updates soon to come.
thomas.vo@LIGO.ORG - 14:25, Tuesday 24 July 2012 (3564)
The QPD mapping did prove to be different than what was previously shown on the MEDM screens.

Here are the new values for the linear response curve. The units are micro-radians/volts. The lever arm is taken to be 6 meters still 

Pitch [ 1727.67131855    23.41350328]
Yaw   [ 1666.80728788   -33.14694715]


It is relevant to address that a positive pitch on the ETMY quad suspension MEDM on the m0 stage will yield a higher intensity on the two lower quadrants of the optical lever QPD, giving a negative pitch as viewed facing the QPD. 
Also, a positive yaw in the quad suspension MEDM will yield a higher intensity on the two left quadrants of the optical lever QPD, giving a negative yaw as viewed facing the QPD. For alignment purposes this will be important to note.  

The next step is to take measurements over the course of at least 2-6 hours and measure the drift as a function of time.
Images attached to this comment
thomas.vo@LIGO.ORG - 08:50, Thursday 26 July 2012 (3588)
I copy and pasted wrongly into the screen the pitch and yaw are reversed, this is signal(1/m) vs translation(micro-radians) which gives units of micro-radians*meters.
       Slope              Y-intercept
Pitch [1666.80728788   -33.14694715]
Yaw   1727.67131855    23.41350328]

Apologies
Images attached to this comment
LHO General
alexan.staley@LIGO.ORG - posted 14:19, Monday 23 July 2012 (3542)
Returned Broadband Photodetector S1200235
I have returned the broadband photodetector (S1200235) to its original box in MY. 
X1 SUS
james.batch@LIGO.ORG - posted 13:48, Monday 23 July 2012 (3541)
X1 Tripleteststand running intermittently
The tripleteststand is malfunctioning - Symptoms are GPS time for user model and IOP model freezes at some time (identical for both), killing the user model and restarting the IOP results in a GPS time of 0, and the IOP indicates no ADC/DAC cards exist.  There is an interesting dialog which can be brought up using the dmesg command:

[13290.003398] x1ioptriple: Allocated daq shmem; set at 0xffffc9001a10d000
[13290.003399] x1ioptriple: configured to use 5 cards
[13290.003400] x1ioptriple: Initializing PCI Modules
[13290.003408] x1ioptriple: ADC card on bus a; device 4 prim a
[13290.003409] x1ioptriple: adc card on bus a; device 4 prim a
[13290.003415] x1ioptriple: pci0 = 0xffffffff
[13290.003423] resource map sanity check conflict: 0xffffffff 0x1000001fe 0xffc00000 0xffffffff reserved
[13290.003426] ------------[ cut here ]------------
[13290.003431] WARNING: at arch/x86/mm/ioremap.c:98 __ioremap_caller+0xd5/0x301()
[13290.003432] Hardware name: X8DTU
[13290.003433] Info: mapping multiple BARs. Your kernel is fine.
[13290.003434] Modules linked in: x1ioptriplefe(+) mbuf [last unloaded: x1ioptriplefe]
[13290.003437] Pid: 5760, comm: insmod Not tainted 2.6.34.1 #7
[13290.003438] Call Trace:
[13290.003442]  [] warn_slowpath_common+0x77/0x8f
[13290.003444]  [] warn_slowpath_fmt+0x3c/0x3e
[13290.003446]  [] __ioremap_caller+0xd5/0x301
[13290.003449]  [] ? T.484+0x13/0x15
[13290.003451]  [] ? pci_bus_read_config_dword+0x66/0x74
[13290.003452]  [] ioremap_nocache+0x12/0x14
[13290.003457]  [] mapAdc+0x65/0x251 [x1ioptriplefe]
[13290.003461]  [] mapPciModules+0x6c1/0x824 [x1ioptriplefe]
[13290.003465]  [] init_module+0x242/0x97d [x1ioptriplefe]
[13290.003468]  [] ? init_module+0x0/0x97d [x1ioptriplefe]
[13290.003471]  [] do_one_initcall+0x59/0x149
[13290.003475]  [] sys_init_module+0xd1/0x231
[13290.003477]  [] system_call_fastpath+0x16/0x1b
[13290.003478] ---[ end trace 31c35fdb3a0a9ba9 ]---
[13290.003481] ioremap reserve_memtype failed -22
[13290.003483] x1ioptriple: pci2 = 0xffffffff
[13290.003485] resource map sanity check conflict: 0xffffffff 0x1000001fe 0xffc00000 0xffffffff reserved
[13290.003487] ioremap reserve_memtype failed -22
[13290.003488] x1ioptriple: ADC I/O address=0xffffffff  0x0
[13290.003491] BUG: unable to handle kernel NULL pointer dereference at (null)
[13290.003919] IP: [] mapAdc+0xd8/0x251 [x1ioptriplefe]
[13290.004137] PGD 1b7cbc067 PUD 1b7c84067 PMD 0 
[13290.004352] Oops: 0000 [#1] SMP 
[13290.004562] last sysfs file: /sys/devices/pci0000:00/0000:00:1e.0/0000:27:01.0/class
[13290.004974] CPU 2 
[13290.004979] Modules linked in: x1ioptriplefe(+) mbuf [last unloaded: x1ioptriplefe]
[13290.005596] 
[13290.005801] Pid: 5760, comm: insmod Tainted: G        W  2.6.34.1 #7 X8DTU/X8DTU
[13290.006215] RIP: 0010:[]  [] mapAdc+0xd8/0x251 [x1ioptriplefe]
[13290.006637] RSP: 0018:ffff8801b7d49dc8  EFLAGS: 00010292
[13290.006846] RAX: 000000000000003f RBX: 0000000000000000 RCX: 000000000000003f
[13290.007059] RDX: 0000000000020cc5 RSI: ffffffff8179c135 RDI: 000000000000000a
[13290.007272] RBP: ffff8801b7d49df8 R08: 000000007ffffff2 R09: 000000000000000a
[13290.007487] R10: 0000000000000006 R11: 00000000ffffffff R12: ffffffffa000f130
[13290.007701] R13: ffff8801be8ad800 R14: 0000000000000000 R15: 0000000000000000
[13290.007914] FS:  00007f957e7836f0(0000) GS:ffff880001e40000(0000) knlGS:0000000000000000
[13290.008326] CS:  0010 DS: 0000 ES: 0000 CR0: 0000000080050033
[13290.008537] CR2: 0000000000000000 CR3: 00000001a165d000 CR4: 00000000000006e0
[13290.008746] DR0: 0000000000000000 DR1: 0000000000000000 DR2: 0000000000000000
[13290.008955] DR3: 0000000000000000 DR6: 00000000ffff0ff0 DR7: 0000000000000400
[13290.009165] Process insmod (pid: 5760, threadinfo ffff8801b7d48000, task ffff8801bdd0f200)
[13290.009571] Stack:
[13290.009771]  0000000000000000 ffffffffa000f130 0000000000000001 0000000000000000
[13290.009985] <0> 0000000000000000 0000000000000000 ffff8801b7d49e48 ffffffffa00087b7
[13290.010401] <0> ffff8801b7d49e18 00000000b7d49e58 ffff8801b7d49e48 ffff8801b7d49e58
[13290.011023] Call Trace:
[13290.011231]  [] mapPciModules+0x6c1/0x824 [x1ioptriplefe]
[13290.011445]  [] init_module+0x242/0x97d [x1ioptriplefe]
[13290.011659]  [] ? init_module+0x0/0x97d [x1ioptriplefe]
[13290.011872]  [] do_one_initcall+0x59/0x149
[13290.012081]  [] sys_init_module+0xd1/0x231
[13290.012291]  [] system_call_fastpath+0x16/0x1b
[13290.012501] Code: 00 02 00 00 e8 0c cc 01 e1 8b 35 91 e8 49 00 48 89 c2 49 89 c6 48 c7 c7 fc d3 00 a0 31 c0 e8 a9 57 00 00 4e 89 34 fd 20 5e 01 a0 <41> 8b 36 48 c7 c7 19 d4 00 a0 31 c0 e8 90 57 00 00 4a 8b 14 fd 
[13290.013247] RIP  [] mapAdc+0xd8/0x251 [x1ioptriplefe]
[13290.013468]  RSP 
[13290.018602] CR2: 0000000000000000
[13290.019122] ---[ end trace 31c35fdb3a0a9baa ]---

The only way to recover is to power the computer off, then power the I/O Chassis off at the power supply, wait, then power up the I/O Chassis, then the computer.  The IOP model can be started normally at that point, followed by the user model.  So far, the system has died three times since Wednesday July 18.
H2 ISC
alberto.stochino@LIGO.ORG - posted 13:44, Monday 23 July 2012 - last comment - 23:39, Monday 23 July 2012(3540)
ALS laser input power increased?

[Bram, Alberto]

Both the ALS REFL PD and the TMS green QPDs are showing larger counts than before the weekend. It looks like the trend started sometime between Friday and Saturday night.

See attached plots. The REFL PD is set with a negative gain so that it reads 0 when the arm was unlocked, and it beceomes postive when the arm is locked.

Now we read around -4000 cts when the arm is unlocked.

Also the green QPD segments, both A and B read around 2200 cts, vs about 2000 as they did before.

We're not yet sure if this means that there's actually more physical power in the beam going. If that was the case, then it would be quite strange.

Could it be a DAQ artefact?

Images attached to this report
Comments related to this report
bram.slagmolen@LIGO.ORG - 23:39, Monday 23 July 2012 (3554)

This maybe have to do someting with the PLL being out of lock, with the intergrator railing both the temp and the PZT feedback. Although I don't know the mechanism, but it may seem that there is more green being generated???

After we disabled the PLL servo and cleared the integrator (in Beckhoff), the pwoer seemed to be fine. When we engaged the temp controller (slow servo and integrator in the Beckhoff), after a little while (2-5 min), the power on the green REFL_PD_DC seemded to 'increase'. 

We haven't measured the power levels in the end-station, as we had bigger fish to fry... see a later post.

H2 AOS
thomas.vo@LIGO.ORG - posted 07:38, Monday 23 July 2012 (3539)
H2 ETMY Optical Lever Install and Calibration
The H2 ETMY optical lever is fully installed with the beam reflecting off the front surface of the optic, picture of the beam is attached to show beam size.

Calibration data has been taken and being processed, the linear response curve needs to be implemented in the C-code written by Jeff Kissel.  The hysteresis within the gears of the translation stage on the receiver side could be a factor in the accuracy of the calibration.
Non-image files attached to this report
H2 SEI
vincent.lhuillier@LIGO.ORG - posted 23:17, Saturday 21 July 2012 (3538)
Status of the seismic subsystems used for the single arm

This alog gives a quick status of the seismic systems (HEPIs and ISIs) used for the single arm test (BSC6 and BSC8).

BSC-ISI:
ISI-BSC8 has been running with damping loops and Isolation loops since the end of May with decent isolation performance. https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=2976

ISI-BSC6 commissioning started more recently. An overall amplification of the motion was immediately noticed around 1Hz (QUAD mode). By the time, writing the commissioning scripts and controlling HEPI was the priority and the no time was spent on ISI-BSC6 to fix the “1Hz” issue.

This week, some modifications were made to the complementary filters that are used to build the blend filters. The initial filters were replaced by less aggressive ones (smaller bump around the blend frequency (750mHz, 250mHz) and less roll off on the low pass. I have attached a link to bode plots of the 4 complementary filters currently used.
LHO_ISI_BSC6_BSC8_OAT_Complementary_Filters_2012_07_20.fig

The blend and isolation filters currently implemented on the two ISIs are designed using the following parameters:
- Blend at 750mHz on all degrees of freedom except on stage 1 in the Y direction where a 250mHz Blend is used instead.
- UGF: 15Hz on all DOFs
- Phase margin > 45deg
- Gain margin>20dB
- Gain peaking <2

The isolation filters used on the two ISIs are pretty closed. Spectra were measured on the ISIs in the undamped and the isolated state. The performances are OK for now. Optimization will be done later.

Stage 2 GS13 spectra are presented below. The scrip that is used to measure the spectra in the different configuration is currently being reworked. The tails of the spectra in the two uncontrolled and the controlled configurations should match better.

BSC-ISI – ITMY (BSC8):
- LHO_ISI_BSC8_ASD_m_CART_ST2_GS13_Undamped_vs_Isolated_2012_07_18_162209.fig

BSC-ISI – ITMY (BSC6):
- LHO_ISI_BSC6_ASD_m_CART_ST2_GS13_Undamped_vs_Isolated_2012_07_19_151614.fig

The HEPI was commissioned fairly quickly to lock stage 0 of the ISI to the ground. Recently, the TMS-QUAD alignment offsets were offloaded on HEPI-BSC6 (250*10^-6 rad in the +RZ direction).

The HEPI are tuned using the following parameters:
- Blend at 800mHz
- UGF: 10Hz
- Phase margin > 45deg
- Gain margin>20dB
- Gain peaking <2
- Pringle modes are controlled (AC)
- No sensor corrections but the STS-2 are in place, connected, leveled and working.

I have attached spectra of the super sensor signals when the HEPI is uncontrolled and controlled.

HEPI – ITMY (BSC8):
- LHO_HPI_BSC8_ASD_m_Super_Sensors_H_Undamped_vs_Isolated_2012_06_20_170452.fig
- LHO_HPI_BSC8_ASD_m_Super_Sensors_V_Undamped_vs_Isolated_2012_06_20_170452.fig

HEPI – ETMY (BSC6):
- LHO_HPI_BSC6_ASD_m_Super_Sensors_H_Undamped_vs_Isolated_2012_07_02_155724.fig
- LHO_HPI_BSC6_ASD_m_Super_Sensors_V_Undamped_vs_Isolated_2012_07_02_155724.fig

Calibration check:
Some measurements were made to check the calibrationand the basis change matrices used on HEPIs and ISIs. Spectra All sensors when no damping on the ISI. Some glitches still need to be fixed (factor of two on HEPI). Couple of glitches still needs to be fixed.
- LHO_ISI_BSC8_ASD_HEPI_BSC_ISI_Undamped_2012_07_18_020214.fig
- LHO_ISI_BSC6_ASD_HEPI_BSC_ISI_Undamped_2012_07_18_020214.fig


During the “1Hz chase” on ISI-BSC6, the reconstruction scripts that are used to build the hundred of filters, required to control the ISI, from a unique set of measurements (ISI undamped) were fully verified for the first time. Excellent agreements between measurements and simulations were seen even on the cross couplings between stages. These scripts allow a huge gain of time during commissioning.
A nice step forward was done during the last few weeks.

Displaying reports 85181-85200 of 88640.Go to page Start 4256 4257 4258 4259 4260 4261 4262 4263 4264 End