Displaying reports 85441-85460 of 88618.Go to page Start 4269 4270 4271 4272 4273 4274 4275 4276 4277 End
Reports until 22:34, Tuesday 26 June 2012
H2 ISC
keita.kawabe@LIGO.ORG - posted 22:34, Tuesday 26 June 2012 (3272)
Starting over (Jax, Elli, Max, Keita)

Summary:

The beam was likely very close to the right edge of ETMY since Friday, and this was due to the fact that we were only using TMS angle for the input beam pointing. We fixed this problem by using QPD servo and started over from scratch.

Now we're back to the point where we see some interference, and the good news is that the demod signal is much bigger than before (1V pp). Also we're seeing some change in the reflected DC power due to interference, unlike yesterday.

We still seem to be getting ultra higher order modes all over the place, and we don't see anything on the camera, so we need to refine alignment tomorrow.

The ETM and ITM are wobbling too much at about 1Hz, though. Due to this, the reflected beam seems to be clipped by either the Faraday aperture or the PDH diode aperture, and as a result the DC level of the PDH diode is changing by 30 to 50% constantly. This is not an interference effect (it changes even if you misalign ITM).

Also today the CM board A was not working at all. No output, fast and slow, regardless of the Beckhoff setting.

Alignment details:

When we started playing around with the alignment today, the demod output was the biggest when some light blob was hitting the right cage  bar of ETMY.  We tried many things but we always got back to this position. The likely cause of this is the fact that we only used TMS for initial pointing.

Since we initially only used the TMS angle to point the injection to the center of ITMY without using the PZTs on the ALS table, and since we needed to rotate the TMS so much, actually the ALS beam was falling off of one of the green QPDs. This means that there was a very high chance that the beam is not going through the center of the TMS primary.

Elli and I finished the QPD servo (a servo that uses PZT-driven steering mirrors so the beam from the ALS table is centered on two in-vac QPDs on the TMS) quickly so we didn't have to worry about QPDs. See Elli's alog entry.

After this, of course the pointing of TMS changed, so we started over by first aligning the TMS (with QPD servo on) to the center of the ITMY using baffle diodes, then aligning ITMY to the center of the ETMY using cage bars, then aligning ETMY to center the second beam on ITMY using baffle diodes.

Now we're getting much larger demod signal than before, and we can see that something is going on in the DC level (that is different from the clipping). And we don't see any light blob on the cage bars, so we know that the thing is not close to the edge any more. That's as far as we got today.

H1 PSL
michael.rodruck@LIGO.ORG - posted 18:07, Tuesday 26 June 2012 (3271)
PSL in low power mode

PSL is currently in low power mode, with the HPO off and only the frontend running. After tweaking alignment into the PMC the transmitted power is about 10W. PMC, ISS, and FSS are locked. If IO needs high power we will turn the HPO back on, but for now we will run with the frontend only.

IO has installed their own power control stage if they need less than 10W.

DBB plots are attached, taken after transition. For the 35W beam.

Non-image files attached to this report
H2 SEI
hugh.radkins@LIGO.ORG - posted 16:17, Tuesday 26 June 2012 (3270)
BSC7 SE Corner HEPI Actuators Removed re WP 3302
EricA & HughR
We managed 1 corner before our crane moving window closed.  7 to go (BSC7 & 4)
X1 SEI
corey.gray@LIGO.ORG - posted 16:08, Tuesday 26 June 2012 (3268)
HAMISI#7 Optics Table Installation

(corey, greg, jim)

The granite table is now clear of the HAMISI#7 Optics Table.  This table was installed with minimal fanfare and no issues.  It has been torqued down. 

(The springs were pulled down, Flexures installed, the Springs are now pulling up on the Flexures, Pull-Down hardware is still in places [it will be removed once we have the Optics Table torqued down])

LHO VE
kyle.ryan@LIGO.ORG - posted 11:40, Tuesday 26 June 2012 - last comment - 11:48, Tuesday 26 June 2012(3266)
Leak from HAM3-BSC2 septum annulus redemonstrated
Both the inner and outer annulus O-rings which are integral to the HAM3-BSC2 septum itself (BSC2 side) were shown to leak previously.  These leaks were re-checked following the most recent venting of the input MC volume and found to have been unchanged by the change in load across the septum.  This 2nd test was not documented and some doubt as to whether or not the bellows tie-rods had been relaxed prior to this 2nd test arose as they were discovered to be in tension when Apollo began to remove this septum plate for inspection yesterday -> So, today I rechecked the outer O-ring and demonstrated that it still leaks and can confirm that the bellows tie-rods are now loose.  As such, the septum will need to be removed for inspection/repair.
Comments related to this report
john.worden@LIGO.ORG - 11:48, Tuesday 26 June 2012 (3267)

The outer leak appears to be located on the BSC side of the septum plate and near 7:00 (bottom left)  looking towards HAM1.

LHO VE
john.worden@LIGO.ORG - posted 11:21, Tuesday 26 June 2012 (3265)
BSC 6 Pumpdown Progress

Current status as the Y END pressure continues to come down.

Images attached to this report
H2 AOS
keita.kawabe@LIGO.ORG - posted 09:05, Tuesday 26 June 2012 (3264)
BSC8 baffle diode medm was "fixed"

As reported before, PD1/2/3/4 on the ITMY baffle correspond to H2:PEM-CS_CHAN_30/29/28/27 and the MEDM screen didn't make any sense because of this.

The real issue should be some cabling somewhere (in-vac or out-of-vac), but instead of a proper fix I changed the medm screen.

PD1/2/3/4 still correspond to CHAN_30/29/28/27, but now MEDM  shows correct number at correct location.

LHO General
patrick.thomas@LIGO.ORG - posted 18:30, Monday 25 June 2012 (3262)
plots of dust counts
Attached are plots of dust counts > .5 microns.
Non-image files attached to this report
H2 ISC
keita.kawabe@LIGO.ORG - posted 17:56, Monday 25 June 2012 - last comment - 19:02, Monday 25 June 2012(3261)
First flavor of interference (Max, Keita)

We're still aligning the arm, it's difficult because we cannot see the beam at all at this stage on the mirror.

Anyway, the beam goes to the ITM, and then to the ETM, then to the ITM and comes back to the ETM. We're seeing some signal from the demod board, which goes zero when the ITM is misaligned. So this is the real interference signal!

Unfortunately it's not like the arm is flashing or anything, I suspect that our alignment is off mainly in yaw because I can tilt the ETM in yaw and see the beam on both righ and left side of the ETM at about the same height. 

Tomorrow we'll look at the interference pattern in the reflection from the arm, DC level of the reflection, and the demod signal amplitude to help our alignment effort. We'll also optimize the demod phase from Beckhoff world.

Comments related to this report
keita.kawabe@LIGO.ORG - 19:02, Monday 25 June 2012 (3263)

The alignment jumped twice by a large amount and it took us some time to recover.

Once it was supposedly Kyle hitting something, and ITM didn't come back exactly to the same position.

Another time I don't know what happened.

LHO VE
kyle.ryan@LIGO.ORG - posted 16:26, Monday 25 June 2012 (3260)
Moved 500 l/s ion pump from HAM5 to HAM6
Kyle, Mark L. (Apollo)
LHO General
gerardo.moreno@LIGO.ORG - posted 16:24, Monday 25 June 2012 (3259)
Ops Summary

- Installation of IO components on H1 laser table started today.  Currently laser beam is dumped and we get no video signal, also some MEDM screens are red, for more information see work permit #3303.
- Apollo, work around HAM03 area to remove septum plate.
- Cable tray removal from around BSC02 area, Filiberto.

X1 SEI
corey.gray@LIGO.ORG - posted 16:06, Monday 25 June 2012 - last comment - 13:18, Wednesday 27 June 2012(3258)
HAMISI#7 Work and Serial Numbers

(corey, greg, jim)

HAM7ISI Springs have been installed, pulled-down, and Spring Safety hardware is in place.  Did a first run through on the Optics Table to pull out any broken/loose tangs.  The Optics Table can be installed tomorrow.  At this point, I was able to note quite a few serial numbers (s/n) while they were easily viewable.  Here are some of them:

Springs

(where the Tip is associated with each Corner;  i.e. a Spring which is bolted down to a Spring Post of Corner-3 actually has its Tip in the Corner-1 area)

Spring Posts

Flexures

Stage0:  014

Stage1:  010

Optics Table:  009

Comments related to this report
corey.gray@LIGO.ORG - 13:18, Wednesday 27 June 2012 (3276)

More Serial Numbers:

Actuators (installed 6/27)

  • H1:  179
  • V1:  097
  • H2:  178
  • V2:  118
  • H3:  176
  • V3:  007

CPS Probes (as slotted in satelite boxes)

  • H1:  12009
  • V1:  12025
  • H2:  12012
  • V2:  12016
  • H3:  12036
  • V3:  12030
H2 ISC
daniel.sigg@LIGO.ORG - posted 12:50, Monday 25 June 2012 (3257)
ALS cavity locking
The common mode board for the ALS cavity locking, S/N S1102637, has been modified to account for the fact that the green cavity pole is expected to be around 600Hz for the one arm mirrors we have (up from 200 Hz). The common compensation is now a 40Hz/600Hz pole/zero pair. We also modified the first boost filter, so it can be engaged. It is a 100Hz/1kHz pole/zero pair. In detail:
C121 -> 220u (from 33n)
R69 -> 18.0k (from 120k)
C118 -> 1u (from 100n)
R77 -> 165 (from 82)
We left R70 at 1.2k which reduces the HF gain by 7%.
H2 ISC
rana.adhikari@LIGO.ORG - posted 21:31, Saturday 23 June 2012 - last comment - 10:36, Monday 25 June 2012(3247)
ITMY Pitch gets the green beam back to ETMY

Using ~1/2 of the ITMY pitch bias range, the green beam gets back to the ETMY.

After enabling the ITMY HEPI + ISI, I scanned the ITMY bias sliders and looked at the ETMY face camera. The ITMY EUL2OSEM matrix has gains of ~5-10 (unlike the TMSY ?) and so the +/-15000 cts of the slider screen is enough to go full scale...and consequently, I tripped the watchdogs a couple times.

The yaw alignment was really good - the return beam to the ETM was nearly centered horizontally, but needed ~1/2 of the ful pitch range of the ITMY TOP stage.

With the illuminator off, it was easy to see the beam flashing around after the watchdog tripped. With the illuminator ON, but turned down to 0.5 Amps, I could see the stuff in the chamber as well as the green beam. Then I used the usual method of putting the beam on the 4 compass points of the ETMY cage and centering it on the mirror by averaging.

Moving the yaw slider to the right moves the beam on the cage to the right. Moving the pitch slider to the right moves the beam down. It takes +1000 slider units of torque to move the beam from top to bottom of ETMY.

The values of ITMY bias that center the beam on ETMY are P = +6300 & Y = +900.

So the ITMY alignment was off by ~6 mirror diameters =>  (0.210 m / 4000 m) / 2 = 25 micro radians.

Now that the beam goes both ways, the next thing to do is to align the ETMY to center the beam on ITMY using the baffle PDs as before. Then perhaps we'll be able to hear the cavity flashes if we can route the audio from the baffle PDs or the ALS green PDs into the control room.

Images attached to this report
Comments related to this report
rana.adhikari@LIGO.ORG - 23:05, Saturday 23 June 2012 (3248)

And...lots of YAW bias gets the ETMY aligned to get the beam back on to the ITMY baffle PDs. The ETMY bias for rough alignment is:

P = -2700

Y = +14200

I haven't seen any flashing yet that looks like a Fabry-Perot resonance, but at least all of the suspensions are close enough in alignment to proceed. Since the TMSY and ETMY both have large yaw biases, we can still hope that they can be mostly relieved by yawing the HEPI (assuming they need the same direction of yaw). Because of the double-loop nature of the quads, we should also get some ITMY pitch relief from pitching its HEPI.

From the flashing that can be seen from the ITMY spool cam, it seems like the return green beam is hitting the ITMY baffle at the edge around the mirror, at the 2 o'clock position look at the ITM from the spool.

richard.mccarthy@LIGO.ORG - 07:22, Sunday 24 June 2012 (3249)
While trying to get the video rerouted to the control room as well as EY I will look at hooking up a FiBox to the green light signal. Will talk to Keita about this.
peter.fritschel@LIGO.ORG - 10:50, Sunday 24 June 2012 (3250)

There is something funny about the numbers here. The claim is that the ITM alignment was off by about 25 micro-radians, and that correcting this uses about 1/2 of the pitch control range (I assume this is at the TOP stage). However, according to T1100595-v1, the control range of the quad, from the TOP stage, is 1.14 mrad in pitch; it is a little unclear as to whether this is +/- 1.14 mrad or a 1.14 mrad total swing; either way, it is much larger than 25 urad -- it should take less than 1/10 of the range to correct it.

rana.adhikari@LIGO.ORG - 17:50, Sunday 24 June 2012 (3251)

I think its funny too. Using ~50000 cts (as measured at the DAC outputs for the face OSEMs on TOP) seems like its close to the 217 limit for the DAC.

Should have checked the coil driver readbacks to find out how much current/voltage was being used to make sure there is no attenuation happening downstream.

peter.fritschel@LIGO.ORG - 09:32, Monday 25 June 2012 (3253)

Jeff K and I found the basic problem with the ITMY alignment numbers. There is a factor of 10 error in the observed pitch misalignment number. The 0.21 m above should be 2.1 m (6 * 0.34 m). The pitch offset is thus 250 micro-radians. We also noted that there seems to be a factor of 2 discrepancy between the designed input range for the TOP driver and the DAC output range, so that we can only access half the current range for each BOSEM: the TOP driver was designed to give +/- 200 ma for each coil, but we can only go to +/- 100 ma. So, the 250 micro-rad of offset is about half the pitch bias range we have available, which is about half of what we should have.

jeffrey.kissel@LIGO.ORG - 10:36, Monday 25 June 2012 (3254)
P. Fritschel, J. Kissel

The confusion lies in Rana's calculation of the total distance for beam travel:
1000 cts of torque = beam moves top to bottom of ETMY
the ETMY is 34cm in diameter.
So, a 6300 cts offset, is
6300 [ct]* 34/1000 [cm/ct] = 214 [cm] = 2.14 [m]
not ~0.210 [m] as Rana uses in his calculation (i.e. he's off by a factor of 10). So, correcting the number in his calculation, we get

pitch offset = torque [ct] / (2 L) % factor of two from "optical lever" effect see G1200698
             = (2.14 m / 4000 m) / 2 
             = 250 micro radians. 


This jives with the predicted range of the QUAD TOP driver and the current monitors:

Predicted Range
T0900232 says, in section 6.1: 
"As the input voltage is specified as 10V* and the voltage gain of the circuit is 1.2** the voltage swing of each output terminal is +/-12[V]. As each output resistor is 40 Ohms, and the [B]OSEM coil resistance is 40 Ohms, the total output resistance is 
     (40 [Ohms] + 40 [Ohms] + 40 [Ohms]) = 120 [Ohms]. 
The voltage across this resistance will be 24[V] maximim, so the [maxmimum] current will be
     24 [V] / 120 [V/A] = 200 [mA] "  

However,
* I've confirmed via direct measurement that the 18-bit DACs can only do +/- 10 [V] differentially NOT per leg. So the max possible voltage is
     DAC_MaxV = 10 [V] (differential output)
** The DC gain of the circuit is 
     CD_VoltageGain = (1+R3/R10) = 1.22 [Vin/Vout] 
(OK, 2% error, but just wanted to make sure we understood from where Ron's numbers came)
And just in case, to remove all suspicion:
From the QUAD Top Driver board schematic, D0902747:
-      Rout = (R1+R5)*(R90+R91)/(R1+R5+R90+R91) 
          = (22+22)*(220+220) / (22+22+220+220) [Ohms]
          = 40 [Ohms],
so that's correct, and Stuart just measured the BOSEMs to have a resistance of 
-      Rcoil = 42.7 [Ohms] , 
so that's correct as well.

So, the maximum output current of the circuit is actually

     DAC_MaxV * CD_VoltageGain / (2*Rout+Rcoil) = 10 [V] * 1.22 [V/V] / (40+40+42.7) [V/A] = 0.09943 [A] = 99 [mA]

which if all other numbers from T1100595 are correct, then this would produce a maximum displacement of

Pitch: 560 micro rad
Yaw: 580 micro rad


Current Monitors
The current monitors have a calibration of 

Amps across the BOSEM coil = ADC [V/ct] * Fast_I Mon Circuit Gain [V/V] / BOSEM Coil [V/A]
                           = 40/2^16 [V/ct] * 3/2 [(single ended Vout) / (differential Vin)] * / 42.7 [V/A]
                           = 2.1441e-05 [A/ct]

The current monitors for the primary Pitch OSEM (F1) on the ITM, with a 6400 [ct] torque offset from the sliders, show -2736 [ct] of current. (I use F1 and not either F2 or F3, because F2 and F3 also contain the Yaw offset). That means the current across the coil is

Icoil = 2736 [ct] * 2.1441e-05 [A/ct] = 0.059 [A] = 60 [mA] 

which is indeed about half the range of the coil driver, which jives with both the [corrected] calculation of 250 urads and the observation that Rana saw ~50000 cts (or half the range of the 2^17 ct DAC max).



Note that the original design intent was to have ~1 mrad of range from the TOP stage, so if we want to restore that intent, we should reduce the output resistance of each leg by a factor of 4, such that the range is

DAC_Max * CD_VoltageGain / (2*Rout+Rcoil) = 10 [V] * 1.22 [V/V] / (10+10+42.7) [V/A] = 0.19458 [A] = 194 [mA]

which could be done by changing the values of R1, R5, R90, and R91 to 10 Ohm resistors.


H2 General
rana.adhikari@LIGO.ORG - posted 21:04, Friday 22 June 2012 - last comment - 11:05, Monday 25 June 2012(3244)
Green Beam Found at ITMY!

Daniel, Keita, Rana, Aidan, Dale, Dick

Using the ITMY baffle PDs, we were able to get the green beam from EY to ITMY today! The aLIGO beam tubes are able to transmit light.

We spent hours today trying out various ideas about how the beam might be missing, but in the end we found it by moving the EY Transmon Suspension.

The MEDM screen for the TMS had a range of +/- 15000 cts. In fact, the TMS has an 18-bit DAC and so its full range is +/- 2^17 = +/- 131072 cts.

 

With the current ALS table alignment, the TMS biases which give the maximum voltage on each baffle PD channel are:

				chan name
				P
				Y
				CS_CHAN_27
				-10000
				+2900
				CS_CHAN_28
				-69000
				+1200
				CS_CHAN_29
				-48000
				+1000
				CS_CHAN_30
				-46700
				+17800

We used these numbers to "triangulate". Then we used the ITMY spool cam to further center the beam on the ITM. It is not at all visible on the ITM surface, but can barely be seen on the gate valve behind the ITM as well as a little glint on the edge of the hole in the baffle where the cavity axis is. Our random search pattern initially just gave flashes of ~50 cts, so the new digital filters helped after all. Eventually, the signal on each PD can separately be made to saturate at 32768 cts as we align the beam onto it.

The final TMS settings for "good alignment" to the ITM are P = -16000 and Y = +9250.  Correction: there were some digital offsets in the 'test' filter bank. After zeroing those out, the new numbers are P = -42800 & Y = +59200.yes

Vincent has turned on the ETMY HEPI with low frequency boosts (although no integrators yet) so the pointing should be more immune to the HEPI hydraulic pressure fluctuations. Still need to fix up the ITMY HEPI. There's some seismic measurements going on overnight, but we can next align the ITMY to get the beam back to EY and then lock the green laser to the arm.

Comments related to this report
keita.kawabe@LIGO.ORG - 10:44, Monday 25 June 2012 (3255)

Angle calibration of TMS OSEM:

When tilting TMS without tilting the ALS beam on the table, the beam leaving the TMS telescope would tilt by almost the same amount (19/20 or 21/20) as the TMS itself. Here I just ignore 5% error and assume that the beam tilts as much as the TMS.

Using the baffle diode (https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?.submit=Number&docid=d0901376) we can calibrate the TMS OSEM. Note that the PD locations are not the same for H1 and H2. We're looking at H2 ITMY (left hole on page 3 of the above document).

There are three PDs horizontally aligned. The top one (PD3) is about 6.25 in away from the middle (PD2). The bottom one (PD4) is 11in lower than the middle. We know (from experience) that increasing P offset moves the beam down. And there's a fourth diode (PD1) that is at the same height as the middle one and horizontally off to the left by 11 in. PD1-PD4 naming convention is explained in E1100867, page 15.

 

PD3=CHAN_28

(P.Y)=(-69000, +1200)

PD1=CHAN_30,

(P,Y)=(-46700, +17800)

PD2=CHAN_29,

(P,Y)=(-48000, +1000)

 

PD4=CHAN_27,

(P,Y)=(-10000, +2900)

For PIT, PD3-PD2 is about 6.25 in for 21000 counts, so it's 6.25*25.4e-3/21000/4000 = 1.9 nrad/count.

Also for PIT, PD2-PD4 is about 11 in for 38000 counts, 11*25.4e-3/38000/4000 = 1.8 nrad/cout.

For YAW, PD1-PD2 is about 11 inches  for 16800 counts, so 11*25.4e-3/16800/400 = 4.2 nrad/count.

There's no funny thing in the output matrix, PIT to coil matrix elements are just +-1, same for YAW, so you can multiply 131072 to get the range.

We can also get the initial alignment number using the offsets as of now (-42800 for PIT and +59200 for YAW)

  PIT YAW
Calibration

1.9E-9 rad/count

Positive offset moves the beam down.

4.2E-9 rad/count

Positive offset moves the beam towards the inside of L.

Range +-0.25E-3 rad +-0.55E-3 rad
Initial alignment

Off by +81E-6 rad

The beam was pointing down.

Off by -250E-6 rad

The beam was pointing outside the L.

keita.kawabe@LIGO.ORG - 11:05, Monday 25 June 2012 (3256)

Note that the order of the channel numbers (27, 28, 29, 30) is reversed. Right now PD1, 2, 3 and 4 = Chan30, 29, 28 and 17.

X1 SUS
stuart.aston@LIGO.ORG - posted 11:37, Friday 22 June 2012 - last comment - 07:20, Monday 25 June 2012(3229)
PR2 (HSTS) Phase 1b testing M1-M1 transfer funtions plus M1, M2 and M3 power spectra
[Stuart A, Betsy B, Deepak K, Andres R, G2]

After taking an initial set of M1-M1 transfer functions earlier in the week on PR2 (HSTS), it was observed by Jeff B that somehow a magnet had become detached from the M2 mass (see LHO aLog entry 3157). 

Thanks to the efforts of the assembly team, this magnet was rapidly re-attached and left to cure for a period of 24 hrs. PR2 was re-suspended yesterday, and a small pitch offset corrected. AOSEMs were then adjusted to their final operating positions and aligned. The canopy/cover was fitted over PR2. Following this re-work, it was prudent to take another M1-M1 transfer function with damping loops OFF for all degrees of freedom, which can be found below (see 2012-06-21_1700_X1SUSPR2_M1_ALL_TFs.pdf). 

Damping loops were then turned ON and a further complete set of M1-M1 transfer functions taken overnight. All the transfer functions obtained have now been plotted against all other HSTS suspensions previously measured on both LLO and LHO test-stands (see allhstss_2012-06-22_AllHSTS_ALL_ZOOMED_TFs.pdf).

n.b. 
Yellow trace = X1 PR2 M1 (2012−06−21_1700) with damping loops OFF
Purple trace = X1 PR2 M1 (2012−06−21_2120) with damping loops ON

The transfers functions obtained again demonstrate good agreement with the model and the spread of all HSTS measurements obtained thus far. 

Power spectra have been taken with damping loops both ON and OFF for each stage (012-06-22_0800_X1SUSPR2_M*_ALL_Spectra.pdf).

Power spectra plots, with both damping ON and OFF have been produced, which compare LHO PR2 and LHO MC2 measurements (allhstss_2012-06-22_ALL_Spectra_Don.pdf and allhstss_2012-06-22_ALL_Spectra_Doff.pdf).

In addition, power spectra for specific degrees of freedom (L, P and Y) can be more conveniently compared across multiple stages (M1, M2 and M3) of the same suspension in the final plots found below (allhstss_2012-06-22_X1SUSPR2_M1M2M3_Spectra_ALL_Don.pdf).

A BURT snapshot has been taken of the current functioning environment "20120622_x1sushxts27_PR2.snap", which has been stored in the following directory:-
opt/rtcds3/tst/x1/cds_user_apps/trunk/sus/x1/burtfiles. This BURT snapshot directory has also been tidied to remove old or incomplete snapshots.

All of the above data, plots, scripts, and snapshots have been committed to the SUS svn as of this entry.

This should now provide sufficient measurements to complete Phase 1b testing of the PR2 suspension and allow it to progress to Phase 2 (chamber-side) testing.
Non-image files attached to this report
Comments related to this report
stuart.aston@LIGO.ORG - 15:01, Friday 22 June 2012 (3232)
The M3 stage watchdog was observed to be perpetually tripping and could not be reset. This is due to watchdogs being triggered for the OPELV_RMS and OPLEV_SUM, which are not visible/settable in the medm screens. 

To rectify this I have manually set the following:- 
caput X1:SUS-HXTS_M3_WD_OPLEV_RMS_MAX 5
caput X1:SUS-HXTS_M3_WD_OPLEV_SUM_MIN -10

For now, this prevents the watchdog from tripping. 
jeffrey.kissel@LIGO.ORG - 07:20, Monday 25 June 2012 (3252)
These results look excellent. 

The only thing that concerns me (where the level of concern (from 1 = "it's awesome. no worries" to 10 = "OMG take it apart and rebuild it") is a 5.5) is the cross-coupling I see in the individual comparison with the model, i.e. in
2012-06-21_1700_X1SUSPR2_M1_ALL_TFs.pdf.
In there, I see cross-coupling between degrees of freedom which we don't normally expect to be there:

1st Yaw Mode (@ 1.09 Hz) into T and L
2nd Yaw Mode (@ 2.04 Hz) into T and L
2nd Roll/Pitch Mode (@ 1.51 Hz) into Y
2nd Vert or 3rd Long Mode (@ 2.80 Hz) into Y

I'm not terribly concerned, because -- as usual -- the cross coupling is reduced to barely visible with damping loops ON. But, it's something to watch out for with this guy as he progresses through the testing (i.e. we'll look with more scrutiny after the optic is swapped). To refresh one's memory -- the reason why we care:  
(1) When the model doesn't match measurements, we can't trust the model to accurately predict other transfer functions which we can't normally measure
(2) If the transfer functions (which are representative of the "plant" upon which we design control loops) are *different* between suspensions by some (as-yet-to-be-quantified) significant amount, then it will make copying and pasting control systems more difficult and time consuming to design.
I think with this suspension, and all others, we're doing fine in both these departments -- but again, we won't really know until we start locking some cavities and really try to push the SUS's to their limit.
Displaying reports 85441-85460 of 88618.Go to page Start 4269 4270 4271 4272 4273 4274 4275 4276 4277 End