Reports until 15:58, Thursday 05 April 2012
H1 FMP
jodi.fauver@LIGO.ORG - posted 15:58, Thursday 05 April 2012 (2548)
HAM2 ICC
The balance of the hand brushing, first vacuum and wipe down were completed. That leaves second vacuum and close-out for tomorrow.
*Note-It seems that the doors on HAM1 WERE NOT removed at the same time as the doors on HAM2 despite what the work permit says!The most likely scenario is that HAM1 doors will come off when the HAM2 doors go back on.
H2 AOS
douglas.cook@LIGO.ORG - posted 14:00, Thursday 05 April 2012 (2546)
H2 ITMy suspension initial alignment on the LSAT
First look at the repaired ITMy pitch shows the optic pitched down by 3.44 mrads and needs to be 12.5urads up.
Betsy and Travis will make a tweak and we will recheck.
Non standard measurement for reference use: UI mass top of top plat to TM centerline is 983.6mm

LSAT level left to right and front to back was perfect (beyond the optical level capability to see a difference)
H2 FMP
jodi.fauver@LIGO.ORG - posted 19:17, Wednesday 04 April 2012 (2545)
BSC6 Cartridge Install Clean-up
After the test stand was moved to the termination slab, the floor was clear for additional activities to take place. The Apollo crew (Mark D., Zack, Caleb) spent the morning dealing with the chores. 
-Betsy and I continue to be concerned about the flaking of the epoxy floor paint, so I thought it was worthwhile to try putting a layer of CPStat down over the paint and in front of the open chamber doors. ( We have used this tactic at West Coast Porcelain to provide a relatively clean floor during wrap-bag-tag activities after baffle part oxidation.)There is one layer of sheeting down now and we can apply additional layers as needed. I'll continue to assess the quantity of paint chips on the sticky mats and see if we're making any progress.
-The cleanrooms were restored to their original state and location and are available as clean workspaces.
H1 FMP
jodi.fauver@LIGO.ORG - posted 19:06, Wednesday 04 April 2012 (2544)
HAM2 ICC
Mechanical brushing started this morning and about half of the chamber was brushed by ~10am. The remainder of the chamber was finished shortly after lunch. Hand brushing the details was about 3/4 complete by the time the crew came out early for safety meeting. The drills did not do well today: 7+ for this chamber with 3-4 DOA (meaning they would not even turn).
LHO General
patrick.thomas@LIGO.ORG - posted 18:35, Wednesday 04 April 2012 (2543)
plots of dust counts
Attached are plots of dust counts > .5 microns.
Non-image files attached to this report
H2 AOS
keita.kawabe@LIGO.ORG - posted 17:01, Wednesday 04 April 2012 - last comment - 17:18, Wednesday 04 April 2012(2541)
TMSY is in the final configuration (Dan, Cheryl, Craig, Keita)

Uninstalled the transport safety restraint and installed the in-chamber restraint. 

TMS is now connected to the West wall of the chamber by the restraint bar. You can still go in and out of the chamber from the West door, but you need to duck.

We didn't care to balance things as the ISI is probably not level. Especially, the upper mass is still locked (but the telescope is not).

Comments related to this report
keita.kawabe@LIGO.ORG - 17:18, Wednesday 04 April 2012 (2542)

While Dan was shining the TMS primary mirror using Stinger, I noticed that when my eyes are behind the mirror but directly opposing Stinger, I can see a lot of small light dots. It look as if it's coming through many pin holes.

Now, we knew that there are thousands of point defects in the mirror surface, and we always suspected that they're coating defects. Quite disappointingly, that was true. We've already proven that the telescope's surface figure is good enough for our purpose (by injecting 1064nm beam with the same size as the IFO beam) on bench, but this is sad nevertheless.

Images attached to this comment
H2 SEI
vincent.lhuillier@LIGO.ORG - posted 16:46, Wednesday 04 April 2012 (2540)
DAQ Kill in ISI model - TEST

The new BSC-ISI model (isi2stagemaster.mdl) includes a DAQKill part. This part is used when one of the following condition is met:
- Rogue excitation (Voltage on the coil drivers readbacks when the masterswicth is turned off or the ISI watchdogs are tripped)
- Excessive motion on stage 1 CPS
- ISI ST1 watchdog tripped (If SUS watchdogs trip, ISI watchdogs trip too)

More details of the changes.

In the ISI models, there are 2 channels that cannot be turned off by the ISI watchdogs.

For the test I drove offsets on:
- HEPI outputs (valve drivers turned off)
- ISI outputs channels (cables unpplugged)
- The test channels of the ISI (not connected)

I tripped the stage 1 watchdog:
- The test channels of the ISI went to zero (DAQ_KILL part)
- The output of the ISI went to zero (Watchdogs + DAQKILL)
- The HEPI channels kept driving

The DAQKILL is working fine in the BSC-ISI model.

LHO VE
kyle.ryan@LIGO.ORG - posted 16:30, Wednesday 04 April 2012 (2539)
Vented Y-mid
Removed existing viewports and replaced with blank flanges with the exception of (1) each 10" to 2.75" zero-length reducer which was needed to re-mount the gauge pair that was left over from the BSC6 removal.  I am pumping the Y-mid overnight and will leak test the new flange joints tomorrow.  

Expect pressures to rise in both the Y1 and Y2 beam-tube modules as neither have a means to pump hydrogen while in the current VE configuration
H2 SEI
hugh.radkins@LIGO.ORG - posted 14:53, Wednesday 04 April 2012 (2538)
BSC6 ISI Optical Table Level Survey PartII
Got the Bench Mark Elevations from Scott(Apollo) and we can now put the numbers on the actual elevation.

Based on the end station zero datum being the middle of the B-nozzles on the BSC6 chamber, just as we've done in the LVEA (BSC2), the Optical Table is intended to be located at 1661.7mm (E1101071).  Our survey yesterday has the average elevation at 1661.8mm.  The reference bench marks we used were place at the end station last July when we rebuilt the External SEI, set the support tubes and installed the HEPI.

Very nice work Scott, Ed, Randy (Apollo); and GregG
X1 SEI
james.batch@LIGO.ORG - posted 13:56, Wednesday 04 April 2012 (2537)
Added EDCU channels to SEI test stand
Modified master file for daqd to include X1EDCU_ISIHAMX.ini and X1EDCU_IOPSEIBSC6.ini, restarted daqd process.
X1 SEI
james.batch@LIGO.ORG - posted 13:35, Wednesday 04 April 2012 (2536)
Restarted models on SEI test stand
For some reason not known, the IOP process on the SEI test stand was taking too much CPU time, running continuously at 17us with peak at 18us.  The x1isihamx and x1iopseibsc6 processes were restarted, the CPU time is now reasonable at 4us.
H1 FMP
jodi.fauver@LIGO.ORG - posted 07:31, Wednesday 04 April 2012 (2535)
HAM2 ICC
This report is for 03 April 2012. Viton rim was installed on the second dust barrier. The condition of the chamber was documented and then the support tubes were inspected. Then, the tubes were picked and vacuumed. Support tube/bellows protection and dust barriers were installed. The chamber is ready for brushing.
LHO General
patrick.thomas@LIGO.ORG - posted 18:32, Tuesday 03 April 2012 (2534)
plots of dust counts
Attached are plots of dust counts > .5 microns. I have included a plot of the mode of dust monitor 2 at end Y (H0:PEM-EY_DST2_MODE) to show when it lost communication.
Non-image files attached to this report
H2 SUS
betsy.weaver@LIGO.ORG - posted 17:57, Tuesday 03 April 2012 (2533)
ITMy HR Cleaning
Yesterday, Margot and Nicole worked on cleaning the ITMy.
While working with them I noticed divots under the metal prisms that were glued on last week by Gerardo and I, in 2 places.  We believe these are due to differential thermal expansion coefficients induced during the (what we believed to have been) low temp heating from a heat lamp.  Shown in the attached picture is one of the divots ~6mm in size under on of the 4 prism "legs".  The other divot is on the other prism (not show) and is much smaller at just under 1mm in diameter.  The prisms are not loose.

After consult with the larger SUS team, 
1) Gluing of all SUS optic/metal components is being evaluated to maintain the bonding strength and outgassing requirements of the glue joints, while minimizing the differential thermal expansion issue seen here.

2) We decided to reinforce the existing joints under the remaining legs with more glue and cure at room temp.  The VRB is being consulted vacuum compatibility of this since the outgassing step will not be complete.

These bonds were cured at room temp for ~24 hours, then the heat lamp was turned on for an additional ~24 hours.

Note that the other 2 metal/optic bonds here at LHO were performed on the TCP and the ERM and were successful.  However, we followed a slightly different procedure which called for immediate heat lamp heat for 16-24 hours.  As well, the configuration of these bonds was also different in that the prism "feet" ran the full length of the prism so the bond was much larger.  We are investigation what parameters factor into these failed bonds, and assessing the actual temperatures seen from heat lamp heating.
Images attached to this report
H2 SEI
hugh.radkins@LIGO.ORG - posted 17:14, Tuesday 03 April 2012 (2532)
BSC6 ISI Optical Table Level Survey
Scott(Apollo) & Jason

I don't have the absolute elevation worked out yet but the five shots we took on the Optical Table puts the level of the Optical Table to ±0.2mm.  If the In-Chamber leveling requirement is the same as that on the Test Stand (±100urad per E1101071); we are so very close.  The Optical Table is ~1955mm in diameter.
LHO General
jonathan.berliner@LIGO.ORG - posted 16:05, Tuesday 03 April 2012 (2531)
Tuesday Ops Log
- SUS cartridge installed in BSC6 yesterday. SEI still has work to do (undressing cartridge, plumbing, cabling, HEPI dial indicator, assess level of support tubes)
- Installation of OMC MC tube
- CP1 and CP5 filled by PraxAir
- Sno Valley visits
- Stoneway Electric visits
- Dust monitors: #2 (EndY?) disconnected, unsure why
- Air handler at MY alarmed, possibly due to seasonal changes (no chilled water flowing at MY)
- CDS: Many reboots at BSC8 and BSC6
- OSB doors locked at 1600
H2 CDS
david.barker@LIGO.ORG - posted 12:46, Tuesday 03 April 2012 (2530)
PSL and BSC8 ISI front end work
The h2psl0 front end models were restarted by Michael, Jim and Vern.

The latest ISI model (and associated C code) were built using the latest trunk 2.5 by Dave and Vincent. This code was installed on h2seib8, and the original HEPI code was also started (test models from yesterday were removed).
H2 INS
dale.ingram@LIGO.ORG - posted 12:02, Tuesday 03 April 2012 (2529)
Links to BSC6 Cartridge Install Photos
Here are ResourceSpace links to four collections from yesterday's cartridge installation at EY.

** Part One -- Preparations
** Part Two -- Locking the Quad
** Part Three -- Flying the cartridge
** Part Four -- Landing the cartridge

You might need to sort the collections by date once or twice to see the contents in the proper order. Elsewhere in RS you might find better shots of the uncovered transmon than the video screen shots you'll see in the Part I of the collection above.
Images attached to this report
H1 AOS
patrick.thomas@LIGO.ORG - posted 10:53, Tuesday 03 April 2012 (2528)
plots of dust counts
Attached are plots of dust counts > .5 microns for April 2, 2012. Note that the dust monitor at location 2 at end Y (H0:PEM-EY_DST2_5) has lost communication. I have included a plot of its mode (H0:PEM-EY_DST2_MODE) to show when.
Non-image files attached to this report
Displaying reports 85901-85920 of 88367.Go to page Start 4292 4293 4294 4295 4296 4297 4298 4299 4300 End