(Mark & Slim from Apollo)
Just to give an update on the Staging Bldg Cleanroom issues noted last month. Last week, the flimsy castors (actually, it was just poorly designed cleanrooms---weak castors not installed securely in thin boxbeam legs) were replaced with much beefier replacements (after seeing cross-section of legs, still wish they were thicker/stronger like the VEA cleanrooms). Photos are attached. These were replaced on the two (of four) cleanrooms we move the most down there.
(obviously both of the attached photos should be rotated clockwise 90deg, seems like the alog doesn't like my photos!)
(Corey, Eric, Hugo, Jim)
Investigations continue for the big mechanical resonance peak around 120Hz. Hugo has given us a rough location of where to look (Somewhere around the H3 GS-13). On Monday, we pulled out H3 (replaced with mass), but there was no change. Also on Mon we pulled some walls in hopes of finding anything obvious. One issue noticed were a pair of 90degree Wall Brackets which were loose (at V3); see attached "photo(3)". These were torqued down, but the 120Hz peak was still there---Hugo did notice broadband noise reduction in the high frequencies by torquing these brackets down.
On Tuesday, Jim started banging on things to see if he could hear any rattling/ringing. He noticed a ping when hitting the big Leg Element weight on the Optics Table. We decided to put some Viton pads underneath it. Hugo went with a checkerboard pattern ([48] 0.5"x0.5" 0.0625" thick pads); see attached photos....grrrr, the last photo should be flipped, looks fine on my computer before I uploaded. Waiting to hear how the measurements turned out on these guys.
UPDATE!
Jim just informed me that his banging-on-things technique totally vanquished the ugly 120Hz peak! (I'm sure Hugo will post something in the alog)
Attached are plots of dust counts > .5 microns. I have included a plot of H0:PEM-LVEA_DST15_MODE to show when the dust monitor at location 15 in the LVEA (in the clean room over BSC8) was disconnected (when the mode stops changing).
Roof insulation and drywall installed today. Crew will most likely need the crane for the next few days for fall protection
We have re-applied FirstContact to the ITMy HR and the TCP AR. When we peeled the "old" sheet from the ITMy HR we observed ~10-20 particles per square inch. (We did have continuous N2 filter flow over the surface during th epeel and during the inspection.) I used the alpha swabs in an attempt to free some of the larger particles from the surface - I was successful on most of the larger which were more like ~1 per few square inches. We also observed alot of the streaking observed and recorded before. We hope the new sheet which we plan to peel tomorrow will remove the streaks and most of the particulate. The TCP AR surface appeared much cleaner, likely because we never pull this sheet for IAS purposes. After re-suspending the masses, we adjusted one of the PUM OSEMs and one of the UIM magnet flags as Garcia needed another round of tweeking on those. The ITMy has been left suspended. The FMy is also suspended. As of this writing the ISI is still locked.
Also, we removed the UIM tray that had been on the QUAD since early January (to facilitate ACB work). The tray does not bolt to the structure but sits loosely on the lower stucture - likely this was causing some ~50 Hz peaking Kissel observed in the January spectra.
J. Garcia, J. Kissel I've reprocessed J. Garcia's results from the latest X1 SUS QUAD03 (an ETM-type QUAD) to include a few extra measurements and the appropriate model. Results on both chains look identical to QUAD04 (now H2 SUS ETMY), and match the model(s) quite well. For both suspensions, all measurements are with lacing cables down the reaction chain. The earlier measurements of each (2011-11-19 for QUAD04 and 2012-01-24 for QUAD03), are the first attempt at lacing the cables and the later measurements are after some adjustment. As one can see, the main chains (M0) are for-all-intensive-purposes identical in all four measurements. On the reaction chains (R0), the only the pitch transfer functions appear different between the cabling adjustments, and most notably the second mode at ~1.7 Hz. As for the over all shape difference from the model (e.g. much lower magnitude at DC, over shift up in mode frequencies), we understand this is due to cables stiffening up the pitch degree of freedom in general. Because we've seen this now in 3 suspensions that have had their reaction chains cabled: X1 SUS QUAD 04, X1 SUS QUAD 03, and H2 SUS ITMY (ITMY is not shown in these plots because it's a different reaction chain type, but you can see the change referenced in aLOGs 1831, 1769, etc); AND the change in stiffness appears consistent and reproducible, we are not concerned. Pending no further mechanical additions / adjustments need to be made (i.e. he's as assembled as he can be before moving on to get have his lower have exchanged for glass), this suspension is good to go. However, we should leave him on the mechanical test stand for exploratory measurements (e.g. OSEM to OSEM basis transfer functions, long-term noise and stability studies). Details ------- The script that produces these plots is ${SusSVN}/sus/trunk/QUAD/Common/MatlabTools/plotallquad_dtttfs.m to which I've made several improvements to while making these plots (because of the bug in how the script adds models to the plots): - Now recalculates appropriate model on the fly according to quadType of last measurement, as opposed to using loaded model precalculated in the last measurement analysis. - Cleaned up messages to command window regarding loaded files. - Now only loads files that are in useMeasts. - No longer addpath genpaths, because it was unnecessary to begin with. - Moved useMeasts and figFileTag to top, so it's more visible. - Corrected 2012-01-06 measType. - Other small changes that I've now forgotten. So please svn up before using this script again!
I often find myself bouncing between the LHO, LLO, and now SEI aLOGs. It would be nice to have links to other aLOGs be added to each of these respective (and any future) aLOGs. Maybe in the second row of header link buttons (Home, Search, Add Report, etc.) but justified to the right-hand-side, and have the button be the color of the other aLOG (LHO is Blue, LLO is Green, and SEI is Red).
The H2 ETMY QUAD has been cabled from the clean cables routed through the ISI to the mounted teststand feedthrough. The "dirty" cables outside of the clean room have also been connected, and readout of the ETMY OSEM signals was confirmed via MEDM.
The aLOG update is complete. The quick search area on the top of the screen is now a little more intelligent. There is more that can (and will be done). Links in the entry headers now have a white background so that they do not disappear if a low contrast background is set for that section. As a reminder when you log in to the aLOG that log in will eventually expire. If you are working on a entry and your log in expires your data is not lost. Simply log back in and look in the drafts area (4th tab when you are logged in).
Yesterday Travis and I finished lacing the OSME cables and suspending the ETMy reaction chain. The chain is hanging side-shifted 1-2mm relative to the main chain, so we will need to make some adjustments to fix that. We made one round of coarse pitch adjustment and now Jason is looking at the DOFs closer with the theodolite/autolevel.
Set up an optical level to look at the ERM and while I currently have no way of quantitatively measuring the ERM height, I can at least confirm that it is in fact sitting slightly high and little to the left (when viewing from the HR side of the ETM). If I had to guess, I'd say the ERM is sitting roughly 1mm high.
Earlier this month, we had made a complete assessment of all aLIGO SUS signs and flips across suspension types (see T1200015) and currently functional builds of those types (see SignStatus_120111.pdf). In doing so we identified: - Some suspensions had a few of their magnets flipped - The BSFMs had some mistakes in their Basis Transformation Matrices - Some suspensions had there OSEMINF sat-amp compensation filters defined with more poles than zeros For the most part, we have now rectified these problems. I attach a new assessment. The remaining action items are - Fix H2 ITMY M0's F2 and F3 magnet sign; to fix when we swap out for real ITM Optic - Figure out what's wrong with X2 PR3's SD OSEM; currently under investigation.
In regards to the issue with PR3's SD OSEM: Because of the arrangement and tight quarters in HAM2, it has been decided (RODA pending!) that some suspensions have their OSEMs Sensor/Coils over on "opposite side" (as opposed to the "side"), i.e. -Trans side of the suspension -- see E1100109 for which is which. It turns out PR3 is one of those suspensions. We therefore expect the magnet polarity to be [N] not [S], which means the COILOUTF gain should be [+] not [-]. Once this COILOUTF gain was flipped, the Transverse damping loop closed and was stable, just the like the rest of the DOFs. Mystery solved! *achoo!*
I replaced a fuse in the break-out box for the dust monitor at location 7 in the LVEA. From a plot of H0:PEM-LVEA_DST7_MODE, it looks like it went out around January 28, 2012 at 17:15 UTC. Attached are plots of dust counts > .5 microns. Seems like high counts in the LVEA, including in the clean room around BSC8 (H0:PEM-LVEA_DST15_5).
The H2 PSL is back up and essentially where it was before the power outage. PMC and ISS are both locked, FSS is turned off due to work on servo board. We are getting 150W downstream of the PMC.
We had some issues getting the HV power supply working, it was determined that a zener diode on the PMC locking board blew, I have replaced that board with a spare.
List of relavent settings:
Note that the phase on the PMC LO will be reset to a different value when the function generator loses power. To set the phase, sweep the cavity by turning on the ramp in the medm screen. Look at the locking PD with a scope and adjust the phase until the sidebands disappear (there are LEMO cables labeled “SPARE” which run from inside the laser room to the racks outside). Then add 180 degrees to get the optimized phase.
For future reference:
Oscillator settings
Frontend settings
A. Effler, J. Garcia Transfer function measurements on the H2 ITMY R0 top mass were completed last night with the results attached. Damping loops on the M0 mass was ON as well as for the FMY M1 stage. The BSC ISI was UNlocked for this measurement as well. Resonances for this measurement are more prominent than the previous night's, which were incomplete due to the watchdogs tripping in the morning work hours. Initial observations indicate the DC scaling calibration factor is still slightly lower than the model with the exception of the Longitudinal and Yaw DoFs. The second Longitudinal mode is about ~0.2Hz lower than the model. The first Yaw mode is slightly higher (~0.05Hz) than the model prediction. Comparisons of this measurement with previous data from the LVEA test stand and Staging Building are still to come.
Attached are comparisons to the previous ITMY TFs (before chamber insertion). Not much has changed. Which is good.
The 'wire' model (for metal build, main chain) was mistakenly used for the reaction chain in the above plots. I attach new plots below with the proper model, made with the latest version of plotallquad_dtttfs.m. More importantly, I confirm AE and JG's assessment. Still looks good -- good enough to close up the chamber and see what we'll see!
I've often been confused by the functionality of the "Quick Search" and the "Search Tasks" shortcuts in the top middle of the aLOG home page. Rather, I understand the "Search Tasks" field -- I just think it's functionality is not needed -- and the "Quick search" is not reliable enough at returning good search results to be useful. More accurately, "Quick search" only searches for the entered text as a keyword. For example a search for "kissel" doesn't return anything written by me, only the one's to which I've added my name in the text. Another example, if I enter in an aLOG entry number "1933" it returns nothing. Conversely, I think the "Search" page has enough versatility that it returns exactly what I want, everytime. In summary I suggest: Removal of the "Search Tasks" possibility, and have the "Quick Search" use the user input as all possible search fields (Section/Task, Report ID, Date, Authors, and keywords), if we want to keep the quick search at all.
The quick search area should be more versatile now. The task selection is staying on the main page so that users can easily filter to only a specific task if needed.