jim.warner@LIGO.ORG - posted 16:47, Thursday 06 June 2024 - last comment - 16:56, Thursday 06 June 2024(78287)
High Voltage CPS testing at HAM7

On Tuesday, I tried running the HAM7 high voltage CPS electronics again. This time I was able to get all 6 sensors to the noise floor I expected. I replace another ethernet cable and one of the in-air triax cables, and I had to use one of the test cards the RichM used at MIT to get there though. After that I was able to turn the isolation loops on let the ISI run for a while when we were trying to lock. My initial measurement looked like the low frequency isolation was pretty bad, so I reverted to the old electronics that afternoon. 

Attached first 3 images compare the performance with the high voltage cps (refs) and the nominal cps (live traces). The high frequency performance is not really any better, and in RX and Y the below .3hz performance seems much worse. Z seems to be worse pretty much everywhere below 10hz. Maybe one of these sensors has a funny calibration or something? I will probably try again next week and see if I can measure that with some ISI tfs.

Fourth image is calibrated spectra for each individual hv cps. The H1 still had a high noise floor for this test. Another thing to try to fix next time.

 

Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 16:56, Thursday 06 June 2024 (78289)

The stage0 to stage1 passive tfs while isolated are a little less disappointing. Attached image are the same datasets, live is nominal, refs are with the hv cps. Still no clear improvement. 

Images attached to this comment
H1 SUS (PEM)
thomas.shaffer@LIGO.ORG - posted 16:40, Thursday 06 June 2024 (78288)
Accepted ITMY R0 offsets in SDF

Robert noticed that we have been set to the incorrect R0 offsets for ITMY for the last 36 days. I've saved what they should be in SDF so they will not be reverted.

Images attached to this report
LHO General
ryan.short@LIGO.ORG - posted 16:30, Thursday 06 June 2024 (78285)
Ops Day Shift Summary

TITLE: 06/06 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Commissioning
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Since my midshift report (alog78281), efforts have been focused on trying to find optimal alignment into the OMC with this new spot position in the OFI. Work on this is ongoing, with the current plan being to use picomotors to adjust the pointing onto PDs in HAM6 to then give more range with OM mirrors to direct light into the OMC.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
21:03 SAF HAZARD LVEA YES LVEA is Laser HAZARD Ongoing
14:40 FAC Tyler, Eric EY N Picking up cooler 15:40
15:14 FAC Karen Opt Lab N Technical cleaning 15:37
15:42 PEM Robert LVEA - Taking viewport pictures 17:03
16:40 FAC Kim H2 N Technical cleaning 16:51
17:00 SQZ Terry Opt Lab Local SHG work 19:36
17:08 ISC Sheila, TJ, Jennie CR N OFI spot move 20:08
17:36 PEM Robert LVEA - Turning on shaker amp 18:26
20:49 CAL Jeff CER N Checking electronics 20:51
21:26 SQZ Terry, Kar Meng Opt Lab Local SHG work Ongoing
23:12 CAL Francisco PCal Lab N Measurements Ongoing
H1 ISC
jennifer.wright@LIGO.ORG - posted 16:23, Thursday 06 June 2024 - last comment - 14:18, Monday 17 June 2024(78283)
Measuring Spot positions

Jennie W, Sheila

 

After our picoing efforts, at 22:50 UTC today Sheila and I measured the A2L gains of the SRM and SR2.

Sheila notched 31 Hz out of the SRCL control loop and I injected a line at this frequency first at SUS-SRM_M3_LOCK_P_EXC, then SUS-SRM_M3_LOCK_Y_EXC.

Sheila changed the A2L gains for the corresponding degree of freedom so that the line height was minimised in SRCL_IN1 and in DARM using the channels H1:SUS-SRM_M3_DRIVEALIGN_P2L_GAIN and H1:SUS-SRM_M3_DRIVEALIGN_Y2L_GAIN.

SRM spot position:

P2L -1 with precision of 1

Y2L 6 with a precision of 0.5

(not too sure of the precisions as I lost my alog halfway through (d'oh).

 

We then did the same thing for SR2 by injecting in the M3_LOCK P and Y channels and changing the DRIVE ALIGN gains for SR2.

SR2 spot position was

Y2L 1.5 with precisioon of 0.25

P2L is 8 with precision of 1

 

The templates for these measurements are in: /ligo/home/jennifer.wright/Templates/Measure_spot_size_SR2.xml and Measure_spot_size_SRM.xml

 

Comments related to this report
jenne.driggers@LIGO.ORG - 16:06, Wednesday 12 June 2024 (78398)

Sheila and Keita have recently found and fixed sign convention errors.  Please see alog 78393 for the corrected interpretation of A2L gains to inferred spot positions.

keita.kawabe@LIGO.ORG - 14:18, Monday 17 June 2024 (78493)

At the request of Alena, here's a table of M1 OSEM P and Y INMON for the output optics right after A2L measurements (starting ~2024/Jun/06 23:10:00 UTC):

  XXX=PIT XXX=YAW
H1:SUS-SR3_M1_DAMP_XXX_INMON -171 -600
H1:SUS-SR2_M1_DAMP_XXX_INMON -25.9 +55.3
H1:SUS-SRM_M1_DAMP_XXX_INMON -280 -474

 

Images attached to this comment
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 16:16, Thursday 06 June 2024 (78286)
OPS Eve Shift Start

TITLE: 06/06 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Commissioning
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 13mph Gusts, 10mph 5min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.22 μm/s
QUICK SUMMARY:

IFO is at PREP_DC_READOUT_TRANSITION and is staying there for comissioning.

Comissioners are testing out alignment configurations with respect to the OFI spot trials

H1 ISC
jennifer.wright@LIGO.ORG - posted 14:25, Thursday 06 June 2024 - last comment - 15:30, Friday 28 June 2024(78282)
Measured OMC Scan before Moving spot in OFI Today

At the start of commissioning today, before we got to NLN and were having trouble locking the OMC, I took an OMC mode scan when we were locked at 10W.

We manually move the OMC_PZT2_OFFSET to its lowest slider value then run the template which does two 100s ramps of this PZT.

Posting for posterity.

Time of Scan = 15:23:43 UTC

200s length. See template in /ligo/home/jennifer.wright/Documents/OMC_scan/2024_06_06_OMC_Scan_15-23UTC.xml

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 15:30, Friday 28 June 2024 (78730)

Analysed mode scan and C02 mode for this full-lock scan but since we have mainly been doing single bounce to look at losses and all the sidebands were on, just posting it here for posterity. The mode-matching calculation has to take into account all the sidebands I think.

Analysis code is In OMCscan_nosidebands13.py and fit_two_peaks_no_sidebands13.py in /gitcommon/labutils/omc_scan/figures/2024-06-06 on the /dev branch (ie. it will not be on the workstation branch as this is the master branch).

Non-image files attached to this comment
H1 General
ryan.short@LIGO.ORG - posted 13:52, Thursday 06 June 2024 (78281)
Ops Day Mid Shift Report

Today has so far been busy with commissioning time, mostly in the effort of moving to a new spot on the OFI. We were able to get locked to NLN this morning after diagnosing the OMC locking issue (see alog78275), then after some commissioning tests, we purposefully unlocked the IFO to start the process of moving the OFI spot in single-bounce configuration.

As of about 20 minutes ago, we've successfully completed an initial alignment after finding the new spot (where SRM needed to move a LOT) and are locking, no issues so far up to ENGAGE_ASC.

H1 SEI
thomas.shaffer@LIGO.ORG - posted 12:37, Thursday 06 June 2024 (78280)
H1 ISI CPS Noise Spectra Check - Weekly

FAMIS25994

I didn't notice any higher frequency noise that was appreciably elevated.

Non-image files attached to this report
H1 ISC (ISC)
jennifer.wright@LIGO.ORG - posted 11:46, Thursday 06 June 2024 - last comment - 14:42, Friday 07 June 2024(78278)
Changed ISC_LOCK.py to remove ASC-AS_B offset

Jenne D,  Jennie W

Jenne noticed that our AS_A_YAW offset was on in the current state (PREP_FOR_LOCKING). We don't want this offset on currently as we are moving to a different spot on the OFI for which we need the picomotors, therefore we my need to rethink AS_A and B alignment offsets.

We turned this offset and the other AS_A and AS_B oiffsets off.

 

This is the only one I could find set in the ISC_LOCK guardian so I commented this line out.

        elif self.counter==9 and self.timer['LoopShapeRamp']:
            #ezca['ASC-AS_A_DC_YAW_OFFSET'] = -0.15 J Wright commented this out on 6th June 2024.
            ezca.switch('ASC-AS_A_DC_YAW', 'OFFSET', 'ON')

I reloaded the ISC_LOCK guardian.

Non-image files attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 14:42, Friday 07 June 2024 (78306)

I just reverted this as Sheila pointed out this offset was put in in here in March to remove some of our DHARD coupling to DARM. This was last checked by Sheila in May.

LHO VE
david.barker@LIGO.ORG - posted 10:38, Thursday 06 June 2024 (78276)
Thu CP1 Fill

Thu Jun 06 10:09:03 2024 INFO: Fill completed in 9min 0secs

Gerardo confirmed a good fill curbside.

Images attached to this report
H1 CAL
preet.baxi@LIGO.ORG - posted 08:47, Thursday 06 June 2024 - last comment - 13:10, Tuesday 11 June 2024(78254)
Characterization of the new 524KHz Analog OMC Anti-Alias Chassis - S2300162

P. Baxi , J. Kissel 

Finished characterizing (Channels 1 to 4) of the new 524KHz Analog OMC Anti-Alias ChassisS2300162

                       BodePlot_Channel_1_4_freq_start_102.4KHz_Freq_stop_1KHz_Steps_100.pdf  (raw data - available in DCC S2300162)

              - We see the frequency response has minimal phase impact in the gravitational wave band, with deg of phase loss at 1000 Hz.

                       ASD_SR785_Channel_1_4_Freq_span_4to32K_32to256K_128to1024K.pdf  (raw data - available in DCC S2300162)

              - We see the noise of each of these 4 representative ADC channels is around ~20 nV/rtHz

                      

 

Non-image files attached to this report
Comments related to this report
preet.baxi@LIGO.ORG - 13:10, Tuesday 11 June 2024 (78370)

For the bode: We see the frequency response has minimal phase impact in the gravitational wave band, with 1.5 deg of phase loss at 1000 Hz. 

LHO General
ryan.short@LIGO.ORG - posted 07:35, Thursday 06 June 2024 - last comment - 09:55, Thursday 06 June 2024(78274)
Ops Day Shift Start

TITLE: 06/06 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 4mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.26 μm/s
QUICK SUMMARY: H1 has been unlocked since 10:21 UTC this morning. It's been able to relock up to PREP_DC_READOUT_TRANSITION, but the OMC is not locked; OMC_LOCK is stuck in the SET_WHITENING state with a notification reading "DCPD gains not equal after whitening switch -- yikes!" (I've not seen this before; will start troubleshooting)

Comments related to this report
ryan.short@LIGO.ORG - 09:55, Thursday 06 June 2024 (78275)ISC, OpsInfo

Looking into why the OMC couldn't lock this morning, it seems to me like the sequence of events went as follows (all times in UTC):

  • 12:17:27 - ISC_LOCK, at the end of DARM_OFFSET, requests OMC_LOCK to READY_W_NO_WHITENING (to lock the OMC)
  • 12:17:31 - OMC_LOCK enters the SET_WHITENING state
  • 12:17:32 - OMC_LOCK ramps entirely to DCPDA to switch whitening on DCPDB
    • DCPD_A0 gain goes to 2 while DCPD_B0 gain goes to 0
  • 12:17:36 - With the DCPDB whitening set, OMC_LOCK now ramps entirely to DCPDB to switch whitening on DCPDA
  • 12:17:37 - DCPD_A0 gain goes to 0 while DCPD_B0 gain goes to 2
    • While waiting for gains to ramp, ISC_LOCK (now in PREP_DC_READOUT_TRANSITION) sees the OMC is not ready and requests OMC_LOCK to DOWN so the OMC can try locking again
  • 12:17:38 - OMC_LOCK is redirected to DOWN before all the whitening change steps are finished; DCPD gains are not reset
  • 12:17:43 - OMC_LOCK reenters the SET_WHITENING state
  • 12:17:44 - OMC_LOCK ramps entirely to DCPDA to switch whitening on DCPDB
    • DCPD_B0 gain goes to 0
  • 12:17:48 - OMC_LOCK ramps entirely to DCPDB to switch whitening on DCPDA
    • DCPD_B0 gain goes to 4 (since the value the agin is found at is simply doubled)
  • 12:17:52 - OMC_LOCK sets DCPD gains back to what they were at the start of the state
    • DCPD_A0 gain goes to 0 while DCPD_B0 gain goes to 2
  • 12:17:53 - OMC_LOCK enters the PREP_OMC_SCAN state to continue locking the OMC
  • OMC_LOCK is unsuccessful in locking the OMC, it cycles back through DOWN and tries to relock
  • 12:19:29 - Back in SET_WHITENING, this time OMC_LOCK sees that whitening is already off and that the gains of the DCPDs are not the same
    • OMC_LOCK shows a notifcation that says "DCPD gains not equal after whitening switch -- yikes!" and stops

H1 sat in this position for several hours (owl operator wasn't notified, separate issue) until Sheila manually set the DCPD gains to what they should be (1), then the OMC was able to lock without issue and the IFO followed suit.

I believe this interaction happened because now that the DARM_OFFSET state has been moved to after all the ASC states, the OMC doesn't have as long to lock before we reach PREP_DC_READOUT_TRANSITION, and the OMC_LOCK DOWN state request happened in a bad spot. If we want to keep DARM_OFFSET where it is, we might want to change ISC_LOCK to give OMC_LOCK more time to lock the OMC before having it try again.

LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 01:14, Thursday 06 June 2024 (78273)
OPS Eve Shift Summary

Ibrahim, Ryan S, Jenne, Sheila, TJ

TITLE: 06/06 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Observing at 146Mpc
INCOMING OPERATOR: TJ
SHIFT SUMMARY:

IFO is in NLN and OBSERVING (after a 24 hour outage) as of 07:59 UTC (cutting it real close H1...)

For the majority of the shift, I was unable to lock - here’s why:

  1. DARM_OFFSET: 

    • Issue: IFO loses lock at DARM offset at which point PR Gain oscillates very vigorously and plummets. If we can manually turn the offset off in time, it recovers and we don’t lose lock.
    • Fix: ASC Loops were not fully converged at the time of the PRG plummet and so we moved the states around (Ryan S, Jenne, Ibrahim) and put ENGAGE_SOFT_LOOPS and ASC states before DARM_OFFSET, which seemed to have fix the issue owing to the less noisy IFO. However, this means that the OMC locks sequentially instead of simultaneously, adding time to lock
    • Cause: Largely unknown but seemingly related to an alignment issue that Jenne thinks has to do with ASC Yaw.
    • Other Troubleshooting/Relevant Alogs: We have seen that DHARD_Y gets spiky and oscillates in the seconds leading up to SRC1_Y and SRC2_Y engaging right before the lockloss. Alog 78251, Alog 78258, Alog 78259, Alog 78265, Alog 78267
  2. ACQUIRE_DRMI_1F:

    • Issue: IFO loses lock here due to very poor pop19 and pop90 flashes, not even making it past PRMI or MICH_FRINGES at times. 
    • Fix: Turning on an 0.6 Offset in SRC1_Y seems to help get past this state but interestingly, this was not required during the last lock attempt post-initial alignment so maybe just initial alignment?
    • Cause: Unknown but thought to be due to more SRC alignment issues.
    • Other Troubleshooting/Relevant Alogs: Alog 78267, Alog 78265
  3. LOCKING_ARMS_GREEN: 

    • Issue: IFO cannot lock green arms for longer than 5 minutes. ALSX is the problem here, where it gets extremely fuzzy and oscillates with a huge amplitude (between 1.2 and 0.8 on its scale). The fuzziness is indicative of a weird comb interference. During this time, it usually loses lock but in the times that it doesn’t, TRX degrades slowly and randomly and then recovers slightly. While this is happening, PLL and PDH fault alarms go up and the state prompts me to “check crystal frequency”
    • Fix: With Jenne, we went into the ALSX Crystal Polarization screen and found that the number was 15 when it is usually supposed to be below 9. Playing with it manually, we were able to get it to 11 by playing with the sliders, which allowed us to get past ALS. I don’t know if this mattered though because after our most recent initial alignment, the value read: 9.
    • Cause: Unknown
    • Other Troubleshooting/Relevant Alogs: None
  4. VIOLINS:

    • Issue: Violins are extremely high, to the point that the primary is occulted by the DARM legend on Nuc30
    • Fix: Damp them violins.
    • Cause: All the above, though it’s interesting that they maintained their magnitude despite initial alignments and hours of non-movement during non-violin exciting periods.
    • Other Troubleshooting/Relevant Alogs: There was a lockloss when violins got sufficiently damp to go to OMC_WHITENING but when there is no wind, there is still pain - and we lost lock minutes before that happened.

On top of these, we have also had a few totally spontaneous lock losses which do not have tracks as egregious as the three above. Moreover, running INITIAL_ALIGNMENT seems to randomize what issue I face in that I either will be acquainted with a totally new problem or face no problems.

A summary of controls moved (with permission):

SDF Diffs Attached (All Seismic Related)

LOG:

None

Images attached to this report
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 21:57, Wednesday 05 June 2024 (78272)
OPS Eve Midshift Update

IFO is still LOCKING (I know)

While violins were damping to turn on whitening (pretty steadily too), we lost lock unexpectedly. DRMI could not lock - PRMI could not lock and MICH Fringes could not lock PRMI. So now I'm doing an initial alignment.

H1 AOS (SYS)
alena.ananyeva@LIGO.ORG - posted 18:45, Wednesday 05 June 2024 (78268)
Reproducing SRC alignment at LHO in Zemax
Attaching a set of slides summarizing the on-going efforts to reproduce the beam path from SR3 through the OFI. Two sets of the beam spot position measurement on SR2 and SR3 were used to "calibrate" the slider moves and to simulate the beam position for other alignment configurations using Zemax. The slides are posted to the dcc https://dcc.ligo.org/E2400189-x0
The aLogs with the beam spot position measurements:
beam spot 1 measurement  aLog77443 
beam spot 2 measurement aLog78119
Images attached to this report
Non-image files attached to this report
H1 ISC
jenne.driggers@LIGO.ORG - posted 18:43, Wednesday 05 June 2024 - last comment - 18:47, Wednesday 05 June 2024(78267)
Attempting to relock, maybe yaw alignment issues?

[Jenne, Sheila, Jennie, RyanS, Ibrahim]

The interferometer has remained fussy all day.  I suspect we may have some alignment issue somewhere, since the AS_AIR camera looks a little bit lobe-y and wrong in yaw.  Twice now, if we've gotten past the DARM_OFFSET state and convinced the OMC to lock (which it will do with the autolocker once the darm offset is at the full value of 9e-5), we're able to at least start powering up. 

My hypothesis of why we might be needing this alignment offset, is that increasing the DARM offset is letting more carrier light go to the AS port, and the the AS WFS loops are seeing that and responding poorly.  So, this alignment offset helps kepe them in the right spot.  That's not a very satisfying story though.

Earlier, I had tried moving SR3 by +/- 1urad on the slider in yaw (reverting afterward to the previous slider val), but that didn't really help.  I also tried offsets in AS_C and AS_A pointing, but neither of those had the same improvement in the AS camera or POP90.  They would improve momentarily, but after the rest of the ASC caught up, we still had the lobe-y look on the AS camera.

Jennie also noted earlier (and it happened at least both of these last two power ups) that we're saturating the OMC suspension when we power up (before moving spots).  But, after all the ASCs catch up, it's no longer saturated.  So, we're close to the edge of what the OMC suspension can do, but it seems surviveable.

Attached are screenshots at CHECK_DRMI_ASC of the AS air camera, without and then with my hokey offset of 0.6 in SRC1 yaw.  You can see that with the offset in place, the camera image is much more centered and nice looking.


For posterity, here are the mattermost versions of the notes, starting with this message:

Jennifer Wright, 4:17 PM: We managed to get past DARM offset because Jenne slowly increased the DARM offset up to its nominal value. The OMC managed to lock but was starting to saturate during power up, we made it through to Power 25W but may be we want to relieve the OMC sus alignment during our next commissioning period so we have more range.
Jenne Driggers, 5:18 PM: Ibrahim is seeing that SRC1 and SRC2 respond to the DARM offset, and it seems like they are pulling things to be misaligned and that's why we're losing PR gain. I don't really have a great solution for this though. I've tried putting in some offsets into AS_C, but when we engage the DARM offset, the same bad behavior happens. I also tried putting in yaw offsets in AS_A yaw, but that time the bad behavior happened fast enough that we lost lock before we could turn off the darm offset
Jenne Driggers, 5:30 PM: It still feels like we have an alignment fiasco going on. The AS_AIR camera looks yaw-ish when we have DRMI locked. When DRMI ASC is engaged, I can move SR3, and before SRC1 and SRC2 can catch up, the AS camera looks round again, and the POP90 buildup goes down while POP18 goes up. But, once SRC1 and SRC2 catch up, we're back to the same ol' story.
Jenne Driggers, 6:21 PM: Ugh. Putting in a SRC1 Y offset of +0.6 makes the AS Air camera better, makes POP90 go down and POP18 go up (not much POP18 though). And, when we do DARM offset, we don't see the PR gain drop. Once the OMC was locked, I slowly removed the offset, and we're currently powering up
Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 18:47, Wednesday 05 June 2024 (78269)

There may be some additional low frequency noise, which could be due to whatever is causing our alignment woes.  At this point, I'm going to let Ibrahim bring down the violin modes and put us in Observing, and hopefully we'll be able to address this tomorrow during commissioning.

H1 ISC
jennifer.wright@LIGO.ORG - posted 13:52, Wednesday 05 June 2024 - last comment - 12:33, Thursday 06 June 2024(78260)
GUARDIAN changes to remove bounce mode of BS

Jennie W, Jenne D, Sheila

 

The other day I designed a new notch to get rid of a bump we had been seeing at 17.75Hz. We think this is the bounce mode of the BS coupling into the MICH loop. The notch for this in LSC-MICH has not been updated since the BRDs were installed I think, and there is no notch for this mode in the ASC-MICH_P and ASC_MICH_Y loops.

 

Changes made to ISC_DRMI.py

Added Bounce-roll notch (FM8) to MICH2 (line 457) in DRMI_LOCKED state and took out line turning on notch at 15Hz (line 456).

#ezca.get_LIGOFilter('LSC-MICH2').switch_on('FM10') # Turn on the BounceRoll FM10 in LSC FM

ezca.get_LIGOFilter('LSC-MICH2').switch_on('FM8') # Turn on the BounceRoll FM8 in LSC FM 05/06/2024 J Wright

 

Added bounce-roll notch to PREP_PRMI_ASC in FM MICH_P and MICH_Y in FM9. These were added to lines 690 and 692.

ezca.get_LIGOFilter('ASC-MICH_P').only_on('INPUT','FM3','FM9','FM10','OUTPUT', 'DECIMATION') #SED increased gain by 20db (removing FM1) Oct 28 2019. added in BR for damped mode at 17Hz 05/06/2024 J Wright
 
ezca.get_LIGOFilter('ASC-MICH_Y').only_on('INPUT','FM3','FM4','FM7','FM9','FM10','OUTPUT', 'DECIMATION')#change LP from LP6 to ELP15 SED DDB March 17 2019, added in BR for damped mode at 17Hz 05/06/2024 J Wright
 

Added these same filters in PREP_DRMI_ASC state in FM MICH_P and MICH_Y in FM9 in lines 920 and 922.

ezca.get_LIGOFilter('ASC-MICH_P').only_on('INPUT', 'FM1','FM3','FM9','FM10','OUTPUT', 'DECIMATION')# added in tuned BR for 17Hz mode J Wright 05/06/2024.
 
ezca.get_LIGOFilter('ASC-MICH_Y').only_on('INPUT', 'FM1','FM3','FM4','FM7','FM9','FM10','OUTPUT', 'DECIMATION')#change LP from LP6 to ELP15 SED DDB March 17 2019, added in tuned BR for 17Hz mode J Wright 05/06/2024.
 
Next time we lock I will see if these remove our noise bump at 17.75 Hz.
 
I loaded the ISC_DRMI guardian when we had just dropped to down.
Comments related to this report
jennifer.wright@LIGO.ORG - 14:11, Wednesday 05 June 2024 (78261)

Just undid these changes in case they prevent us locking DRMI as we keep falling put from ENGAGE DRMI ASC in ISC_LOCK guardian.

jennifer.wright@LIGO.ORG - 12:33, Thursday 06 June 2024 (78279)

I turned these filters on when we got to NLN at the start of today's commissioning period. Since they didn't unlock us and seem to possibly make the noise at 17.7Hz better in DARM I have put them back in ISC_DRMI as detailed above and reloaded it. I will post a high resolution spectra next time we get a good lock stetch.

H1 SEI
jim.warner@LIGO.ORG - posted 19:55, Monday 03 June 2024 - last comment - 20:05, Wednesday 05 June 2024(78226)
Some extra St2 isolation loops and St1-2 sensor correction running on BSCs temporarily

Wind was brutal today, so I was messing with the controls on the BSCs to try to help. Mostly just running RX/RY sensor correction from St1 to St2 on the BSCs. Sensor correction is focused on improving the ~.4hz motion and doesn't really affect the motion outside that. I don't think this really helped, winds calmed down after I left at 5:30 and Ryan is almost all the way NLN now, so I will leave the extra sensor correction running for tonight. I'll accept SDFs for tonight, then turn it back of during maintenance tomorrow.

Wind and lock losses make a comparison challenging, but the blrms trends show the (log scale) improvement ETMY sees when it's turned on over .3-1hz. Top trace is the St2 RX GS13 blrms, middle trace is the gain for the sensor correction, sc is on when it's 1, off at 0. Bottom trace are the .3-1hz ground blrms. For the las 3-ish hours the St2 RX motion is lower with the sensor correction on even when the ground is moving more in this frequency band.

Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 20:05, Wednesday 05 June 2024 (78270)

These changes were reverted today, in case they were causing issues with locking. ASC was engaged at the time, I didn't see any effect turning this stuff off, so I think they were probably benign.

H1 SEI
jim.warner@LIGO.ORG - posted 15:06, Monday 03 June 2024 - last comment - 20:06, Wednesday 05 June 2024(78216)
Changed ETM ISI cps diff filters to try to help with the wind, doesn't seem to have helped.

With the difficulties locking, I tried changing some of the CPS diff filters to get the end station ISI to follow the corner ISI. The CPS diff uses a band pass filter to reduce the differential motion between the ISI, but this filter can inject motion at the microseism. Microseism is low, I'm guessing the arms are struggling more with the below .1hz motion. In the attached image the filters we had been using, are the red trace, blue is what I switched to. It kind of seemed to help, but I think we lost lock after I left the room. I'll leave this running for now, going to try a couple other things.

Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 20:06, Wednesday 05 June 2024 (78271)

These changes were reverted today, in case they were causing issues with locking. ASC was engaged at the time, I didn't see any effect reverting to the old configuration, so I think they were probably benign.

Displaying reports 8721-8740 of 84727.Go to page Start 433 434 435 436 437 438 439 440 441 End