TITLE: 04/21 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
SEI_ENV state: MAINTENANCE
Wind: 4mph Gusts, 2mph 3min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.29 μm/s
QUICK SUMMARY:
BSC2 work continues, CDS work. HAM5 & 6 HEPI has been down overnight due to HEPI pump system issues noted yesterday.
Workstations were updated and rebooted. This was an OS packages update. Conda packages were not updated.
Madi, Camilla
Started ETMX Ring Heater test for HWS testing, turned up from 1W/segment to 2W/segment. Plan to leave until tomorrow.
TITLE: 04/20 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: Some BSC2 and cabling work done today. HAM5 and HAM6 HEPI are going to be down overnight since the HEPI pump is having issues.
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 14:40 | FAC | Kim, Nellie | LVEA | n | Tech clean | 15:32 |
| 15:09 | FAC | Randy | MidY | n | Rock cleanup on aisle MY | 16:10 |
| 15:51 | FAC | Nellie | HAM Shack | n | Tech clean | 16:34 |
| 16:01 | VAC | Jordan | LVEA | n | CP1 RGA work | 16:29 |
| 16:44 | SEI | Jim | LVEA | n | BSC2 work | 17:49 |
| 16:57 | IAS | Jason | LVEA | n | FARO scoping | 17:28 |
| 17:20 | FAC | Randy | LVEA | n | Just looking at HAM6 cleanroom | 18:07 |
| 17:21 | VAC | Jordan, Gerardo, Travis | LVEA | n | CP1 pump work | 18:44 |
| 17:34 | SEI | Mitchell | LVEA | n | BSC2 work | 17:49 |
| 19:10 | EPO | Jeff, +1 | LVEA | n | Tour | 20:17 |
| 19:47 | EE | Fil | MER | n | HAM6 cabling | 00:20 |
| 20:18 | VAC | Travis, Jordan, Gerardo | LVEA | n | CP1 pump work | 23:29 |
| 20:23 | CDS | Dave | EX | n | Reboot BRSX | 21:30 |
| 20:25 | FAC | Randy | LVEA | n | Looking at HAM6 some more | 21:43 |
| 20:29 | Jennie | Optics Lab | n | Looking for laptop | 20:40 | |
| 20:37 | Betsy | LVEA | n | Looking at HAM6 with Randy | 21:46 | |
| 20:44 | IAS | Jason | LVEA | n | Grabbing FARO parts | 20:57 |
| 21:10 | SEI | Jim | LVEA | n | Talking with Betsy / SEI work | 23:28 |
| 21:19 | TCS | Madi, Camilla, Maya | OpticsLab, LVEA | n | Grabbing parts | 21:51 |
| 21:30 | CDS | Dave | EY | n | Verifying 24V power supply | 22:02 |
| 21:44 | SEI | Mitchell | LVEA | n | BSC2 work | 23:28 |
| 21:52 | TCS | Camilla, Madi, Maya | EX | y(local) | HWS work | 23:06 |
| 22:12 | IAS | Jason, RyanC | LVEA | n | Setting up FARO for tomorrow | 22:40 |
| 22:22 | TCS | TJ | LVEA | n | Draining TCS chiller line | 23:19 |
| 23:15 | SUS | Ibrahim | LVEA | n | BBSS ExpreSS | 23:33 |
In prep for WP13185 to replace the flex lines on the table and the expansion joint, I've drained and vacuumed the TCSX table and then I just used gravity to drain most of the water in the supply line, enough so that the expansion joint is free of water.
WP13182 BRS-EX recovery
Jim, Patrick, Dave
I tried to reboot, but ended up power cycling, h1brsex. Patrick started the code and IOC. Since then it has been attempting to damp over the past few hours.
FAMIS 31134
The ISS diffracted power has drifted down a bit over the past few days, so I'll bring that back up towards 4%. No other major events of note.
Mon Apr 20 10:09:19 2026 INFO: Fill completed in 9min 16secs
TITLE: 04/20 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
SEI_ENV state: MAINTENANCE
Wind: 9mph Gusts, 5mph 3min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.19 μm/s
QUICK SUMMARY:
BSC2 work today
ETMY SWWD tripped at 01:17 PDT this morning due to an EQ. I have reset the DAC drives.
Sun Apr 19 10:06:23 2026 INFO: Fill completed in 6min 19secs
Sat Apr 18 10:08:36 2026 INFO: Fill completed in 8min 33secs
BSC6 annulus ion pump railed about 20 minutes ago, there appears to be a little communication to the main vacuum volume noted on PT410B, however it may just be electrical noise due to the change reported by PT410B, very small, its changes it is not replicated by other gauges, like PT424B. Nothing to do about it right now, but we will check the annulus system ASAP.
(Jordan V., Gerardo M.)
Last Tuesday, we did some troubleshooting of the annulus ion pump, we tried to induce for the short internal to the pump to let go by hitting the pump with an object, the pump only flashed once but no good results. Likely the pump is shot, this unit was a salvaged unit that was removed from another chamber, see aLOG 68433 for more info regarding the pump.
TITLE: 04/17 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: No further work was done regarding BSC2
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 14:51 | FAC | Nellie, Kim | LVEA | n | Tech clean, Nellie out 15:51 | 15:55 |
| 15:51 | FAC | Nellie | EY | n | Tech clean | 16:35 |
| 16:37 | PEM | Robert | EX | n | Grounding | 00:25 |
| 17:13 | PCAL | Tony, Erik | EX, EY | n | Trying to get PCAL laser back up | 17:57 |
| 18:29 | PCAL | Tony | PCAL Lab | y(local) | Fixing broken chassis | 19:09 |
| 18:33 | EPO | Jeff, friend | LVEA | n | Tour of LVEA | 19:35 |
| 18:38 | VAC | Jordan | LVEA | n | Turning HAM1 AUX carts off | 18:42 |
| 20:07 | EPO | Jeff, friend | LVEA | n | Tour | 21:26 |
| 22:02 | VAC | Jordan, Mitchell | LVEA | n | Decoupling AUX carts | 22:25 |
| 22:40 | VAC | Jordan | LVEA | n | Turning on RGA for CP1 | 22:54 |
Jennie W, Sheila D, Jenne D
Summary: JAC TRANS A LF calibration gives correct ratio now but PD is too noisy to give us sensible values when the light pipe is closed.
On Tuesday I changed the scaling filter in JAC-TRANS_A_LF to give us an output calibrated in W out of HAM1. The previous calculation (alog #89805) gave us too large a value.
When we checked the power on the JAC-TRANS PD in chamber (alog #89251) we did not monitor the JAC input power between successive measurements and since we had purge air on I think the input power drifted.
To avoid confusion I just took a time when the output power was measured in chamber and used the JAC TRANS LF OUTPUT power at that time to give the calibration into HAM2 W.
When we measured 96mW out of HAM1, the JAC-TRANS_A_LF filter bank was roughly calibrated into mW (using standard values for V/count, A/V, mW/A) but had not been calibrated using in-chamber measurements.
For 96mW we got 0.00120mW on JAC_TRANS_A_LF_OUTPUT on 2026/02/24 at 20:30:10 UTC, see image.
We want the output in W so the calibration factor is 80000/1000. This value is now in FM10 in the filter bank. Unfortunately the signal still is too nosiy to give a sensible looking value and so changes from -20 to 20 W.
Jenne and I put a 0.01Hz pole to low pass the noise in the FM7 filter, and also tuned the offset on the filter bank.
This channel still seems too noisy though (image) and changes between -0.5 W and 0.5.
So maybe we need a filter that goes down to lower frequency, or its just too noisy a PD to use as a JAC input power monitor.
I'm going to revisit this after the BSC2 work when we next have an opportunity to open the light pipe.
I accepted the sdf diffs for JAC-TRANS_A_LF - filters, offset and tramp.
I also turned off the FM1 and FM2 filters in JAC-L_SERVO as Jenne noticed the integrator hwas causing a large value at the output of the servo. Checking the 'DOWN' state in the JAC guardian these should not be on so I accepted them as off in sdf.
Today i noticed the output of the JAC SERVO wa son so i turned it off and sdfed it. This should also be the case when in DOWN.
Closes FAMIS 28639
Last checked in alog 89720
TCSX read 30.8 - nothing added
TCSY read 10.6 - nothing added\
No water in small cup under the slow leak
TCSX is reading HIGH PRESSURE FAULT. See attached picture.
Betsy and I locked and tagged out the CO2X chiller, it will remain off until all needed repairs and safety documents have been completed next week.
At 15:44UTC the chiller tripped off with a HIGH PRESSURE fault, as found by Ibrahim. The CR verbal machine also alerted Ryan with a "low flow" warning.
I went out to the LVEA inspecting the pipes and found that the outgoing CO2X line was disconnected at the quick-connect, image. There was no water on the floor, so the quick-connect did it's job. Maybe this was already loose and we need to add the the pipe inspection FAMIS to inspect these connections, but. there was also lot of work in that area today. I inspected and reconnected the quick-connect, it did not seem damaged.
At out weak point in the chiller lines at the flexible baffle in the pipe bridge, the plastic sheeting that had been attached to catch any spilled water had a small amount of water in it photo, but none leaking out. Expect this water would have been pushed out a small break in the rubber while the chiller was trying to push water around the incomplete lines before the high pressure sensor in the chiller tripped it off. We will inspect and probably repair this part next week. In the meantime with water dripping out of the pipe bridge connection, we are fine to leave the chiller locked out with water in the lines.
TJ, Camilla. Trending the pressure values doesn't tell us much. The chiller appears to go to the high pressure fault and trip off before the pressure channels register a high pressure, see attached. ANALOG1 PSI starts at 65 PSI and ANOLOG2 started at 26 PSI, after the chiller trips off, these increase to 6500 PSI (unphysicial) and 44 PSI. The chiller must have an additional pressure sensor it uses to trip off. Current settings are to trip off at 80 PSI, see FRS 30283.
Randy built a unistrut protective barrier around both CO2X and CO2Y water pipes to protect them from this happening in the future. Photos attached.
FRS37572 ISI HAM and BSC WD Plots Not Working
Ryan C, Erik, Dave:
The MEDM buttons to plot the latest ISI WD trips (see attached) have not been working for some time.
We found the issue was due to the wd_plot python code running in the conda environment had pydv issues, but if ran in native debian it worked.
Fixing the conda pydv issue would take significant resouces, so we chose to change the MEDM launcher to explicitly run local /usr/bin/python3.
The common MEDM file ISI_CUST_CHAMBER_WATCHDOG.adl in the isi/comon/medm/[hamisi, bscisi]/ directories were changed accordingly.
I tested with HAM2 and BSC3.
Thanks Dave, those plots are useful diagnostic tools.
I dropped a note in the LLO log so we can be sure this fix gets used at both sites
-Brian
From the previous alog post, Sheila measured a power drop across BL2 lens.
Here I modeled the loss of the 700um beam waist across the 2.5mm BL2 lens aperture.
The expression for the transmitted power obtained by modifying the Gaussian beam equation to include a displacement d2 = dx2 + dy2 is shown here.
And here's the contour plot for the loss (1-T) across the lens aperture for a beam transversing across various spots on the lens.
Me and Sheila are checking on the geometry of the setup.
The BM3 mirror mount IXM100.2 has a 100 pitch screw adjustment ( correspond to 100 turn per inch translation), and the beam appeared to be clipped symmetrically for 1/8 rev. For the mirror mount of 1.87" width, this correspond to a mirror rotation of 1.3mrad.
The distance between BM3 and BL2 is 148mm, from the mirror rotation, the IR beam angular deflection is 2.6mrad, this implies an aperture size of 0.4mm. Something is not right, and it's a mystery that requires more investigation.
The aperture stop installed is D2000339: 0.2" (5mm) diameter.
Naoki, Camilla, Sheila
We had another look at some signals that we could use to track or servo the SQZ angle.
We turned the ADF back on, making a line at 1.3kHz in DARM, and tuned the demod phase for the ADF so that the SQZ angle readout was 0 for the angle we've been using for observing in this lock. Camilla added a bandstop at 1.3kHz in the SQZ BLRMS4, sum and null channels and checked that the ADF isn't dominating those. We did a sweep of the CLF6 demod phase like this, with the dither amplitude reduced comapred to Wed (0.01 instead of 0.03 CLK gain).
In the first attachment we are using BLRMS 4 to demodulate for the noise lock signal, the second cursor shows that the zero crossing of the ADF SQZ angle and the noise lock both correspond roughly with the minimum of noise in the BLRM4 (1kHz BLRMs). This isn't the same phase as the one that minimizes the brown trace, a blrms centered at 350 Hz. This means we have a frequency dependence of the SQZ angle, so we should probably look into things like the SRCL offset that might be causing this.
Naoki then turned the ADF phase with the SQZ angle set to minize BLRMS3, as shown in the second screenshot from Camilla (at the begining here you can see that Naoki set the ADF phase so SQZ ANG was zero for CLF demod phase of 145). We repeated the sweep and see that we have lower SNR for the noise lock using this lower frequency BLRMS, and that the ADF as it is now wouldn't make a good error signal for this sqz angle because it doesn't really go negative.
Detchar: We are planning to leave the ADF on over the weekend, which will create a line at 1.3kHz. We are hoping to use this to track changes in the squeezing angle over time.
The ADF calculated SQZ angle (H1:SQZ-ADF_OMC_TRANS_SQZ_ANG) seems to follow our SQZ BLRMs over the weekend, plot attached. Unsure why the SQZ is different lock to lock, i.e. sometimes changes over first 6 hours (1 day ago) and is sometimes is stable (-12 hours ago).
The Wensday work that is referenced above is described with helpful detail in 74223
Varun, Stefan
We measured the (one-way) SRC Gouy phase to be 28.97deg +-0.11deg (statistical) +-0.43deg (systematic).
We used a method pioneered in KAGRA (klog 9241 and 9246). A similar method was used for the PRC Gouy phase measurement(alog 52504).
Plots:
- Plot 1: SRC scan across 6 fringes. Blue: AS_C_SUM, Red: AS_C_SEG2 (both derived from IOP channels). Magenta: round trip Gouy phase fit from the two results.
- Plot 2: Raw data strech of 6 best fringes.
Method:
- Align BS and SRM (in SRY) carefully (especially BS)
- Misalign SRM by 70 urad (we did pos pitch)
- Let SRMI freely swing (i.e. SRM and BS aligned, PRC misaligned)
- SRCL1 and MICH1 gains 0 (we don't need the control loops)
- Put an offset in SRCL2_OFFSET (we used -1e6cts)
- Turn off all whitening on ASC-AS_C
- We want the LSC trigger to fire and stay latched (and apply the offset) when ASC-AS_C reaches the darkest possible spot.
This corresponds to the BS in the correct position, increasing the finesse of the SRC to its maximum.
- We achieved that with a negative trigger matrix element, a slightly positive enable threshold (since AS_C swing slightly negative), and a very negative disable threshold
We used: H1:LSC-TRIG_MTRX_4_14 =-100, H1:LSC-SRCL_TRIG_THRESH_ON = 0.015, H1:LSC-SRCL_TRIG_THRESH_OFF = -100
- To read the data we need the highest possible data rate. So in addition to H1:ASC-AS_C_NSUM_OUT we recorded the IOP channels
H1:IOP-ASC0_MADC6_TP_CH9, H1:IOP-ASC0_MADC6_TP_CH10, H1:IOP-ASC0_MADC6_TP_CH11, H1:IOP-ASC0_MADC6_TP_CH12
- Those IOP channels are before the fixed anti-whitening filter. So we did the anti-whitening filter in diaggui ('zpk([40],[0.4],1,"n")').
- The 4 quadrants of ASC-AS_C saw different combinations of the higher order modes.
- We used SEG4 (CH9) for finding the 00 modes.
- We used SEG2 (CH11) and the sum of all 4 segments for 10 modes (red and blue traces respecively).
- We chose a bank of 6 fringes with very low AS_DC values - those have the highest finesse.
- A spline fit was used to interpolate the time to degrees in-between 00 peaks.
- We calculated the mean from the red and blue traces separately. Their mean is reported as final value.
- We calculated the standard deviation of the mean from the red and blue traces separately. The maximum of the two errors is reported as statistical error.
- The difference between the red and blue result is reported as systematic error.
The code and raw data is found in /ligo/home/controls/sballmer/20191022.
Should add that the SR3 ring heater, as well as the ITM TCS system, was turned on during the measurement. Snap shots are attached.
The time of measurement was GPS 1255829128, Oct 23 2019 01:25:10 UTC.
At the time of this measurement, CO2 heating was central. The CO2 FLIP1MON was 0 and FLIP2MON was 1. Comparing this to the current values and the MEDM that says we are currently using annular heating, I believe this means that these measurements were done with central CO2.