Displaying reports 1521-1540 of 88640.Go to page Start 73 74 75 76 77 78 79 80 81 End
Reports until 13:49, Saturday 28 March 2026
H1 ISC (INS, IOO, ISC)
keita.kawabe@LIGO.ORG - posted 13:49, Saturday 28 March 2026 - last comment - 09:08, Monday 30 March 2026(89685)
Saturday morning: pico holder on the WFS sled for 50:50 splitter was loose (JennieW, Jason, Keita)

I've found that the pico mount for 50:50 BS on the REFL WFS sled in front of ASC REFL_A was loose and rotated counter-clockwise seen from the top by a huge amount (1st attachment, orange arrows show the direction of rotation). Our guess is that the BS mount was bumped when we were leaning into HAM1 from -Y door to work on the JAC output periscope. In general, it's hard to rotate the mount clockwise seen from the top even if the screw is not super tight (because the screw tends to be tightened), but it's easier to go counter-clockwise.

When this was found, the beam was hitting the +X-Y edge of the mirror, there was no clear reflection beam found so no beam on WFSA, but somehow the ugly transmission beam with lots of diffraction patterns was making it to WFSB.

We reverted the RM1 and RM2 bias sliders back to O4 level (RM1 PIT=-180, YAW=-57, RM2 PIT=890, YAW=-530) and I confirmed that the centering on the 2" lens was good. WFSA mount was screwed down tight to the post.

RM1 bias was adjusted further (RM1 PIT=-190, YAW=263) to roughly center the beam on WFSB.

At this point I looked at the beam on WFSA and it was still off mostly in YAW but there was also a large PIT offset. These were taken care of by adjusting the picos I've just screwed down.

I enabled the REFL WFS centering which worked right away. LSC REFL diodes are receiving almost equal amount of light. We'll have to make sure that the beam is not clipped on LSC diodes. Anyway, I relieved the ASC using RM sliders and ended up these numbers: RM1 PIT=-192, YAW=274, RM2 PIT=910, YAW=-532.

Images attached to this report
Non-image files attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 17:08, Saturday 28 March 2026 (89686)

Afternoon: Done with the REFL path

Making sure that the beam is centered-ish on the LSC sensors

We enabled the WFS DC centering, relieved the WFS output by RM sliders, disabled the WFS centering. Then scanned RM1 in PIT to find out where the LSC REFL A and B DC starts to fall off, and make an average position in terms of RM1 PIT offset ("plateau center"). In general the plateau center is not the same as WFS DC center.

Use the common pico for the REFL LSC sensors to make the plateau center come closer to the WFS DC center. See the 1st attachment.

Repeat the same thing for YAW. We noticed that LSC REFL B is not exactly the mirror image of REFL A, mostly horizontally, as you can see from the 2nd attachment. If they are, we expect both to start falling at the same time but they don't. To fix that we need to touch up the non-pico 50:50 splitter that steer half of the beam to REFL B, but we chose not to do it because the scan range you see here is huge, and beam will totally fall off of WFSB before LSC REFL A and B starts falling.

After all of these and minor tweaks here and there, we ended up with: RM1 P = -196, Y=281, RM2 P=910, Y=-490.

Tilting WFSA in YAW

I checked the beam position along the REFL path and unfortunately the reflection from WFSA was hitting the mirror mount. I tilted the WFS clockwise, paying attention NOT to change the optical path length significantly. After this, Jason and Jennie used pico to steer the beam back to the center of WFS. I confirmed that the WFSA reflection goes into the beam dump.

Final check

I rechecked the beam position along the REFL path. Nothing was grossly off-centered except for 1" mirrors and BS on the WFS sled (this was always the case).

On M2, RM1 and M5, the beam position looked OK though it was hard to say anything quantitatively. No picture for these.

1" lens for the LSC censors, 2" lens on the WFS sled as well as 1" lens on the WFS sled looked good.

Reflection of all LSC and ASC REFL sensors fall on the beam dumps.

Pictures will follow.

No POP check yet

I looked at the POP path too but it wasn't flashing and it was already 4PM so we gave up. We'll continue on Monday.

Images attached to this comment
keita.kawabe@LIGO.ORG - 22:17, Saturday 28 March 2026 (89687)
New_ASC_REFL_A_angle.jpg: After I rotated the ASC-REFL_A angle clockwise seen from the top by about 10 degrees or so.
 
1inLSC_vert.jpg shows the centering on the 1" lens for the LSC REFL. You cannot judge horizontal centering from this picture due to parallax but it looked OK to me.
 
2inOnSled_hor.jpg: Centering on the 2" lens on the REFL WFS sled. Horizontally it's somewhat to the left (+Y) but not too bad. You cannot judge vertical centering due to parallax but it looked OK. See the 1" lens on the sled.
 
1inchOnSled_hor.jpg and 1inchOnSled_vert.jpg: Centering on the 1" lens on the sled. Vertically it's quite good. Cannot tell much horizontally due to parallax, but it looked a bit off to +Y direction to me. Not too bad.
 
LSC_REFL_A_ghost.jpgLSC_REFL_B_ghost.jpgWFSA_ghost.jpgWFSB_ghost.jpg show that reflections from all REFL sensors go to corresponding beam dump.
 
(The last picture (POP_stiffener_base_touching_.jpg) shows how the POP periscope stiffener base touches the base of the REFL path BS that splits the beam to the in-vac and air path. So, this could have made a very minor change mostly in YAW. But anyway it's without doubt that the loose mount on the REFL WFS sled was the main cause of the trouble.)
Images attached to this comment
keita.kawabe@LIGO.ORG - 09:08, Monday 30 March 2026 (89690)

Correction: In the above alog text, "LSC_REFL_B_ghost.jpg" points to the picture for REFL_A ghost beam. This is the correct one: LSC_REFL_B_ghost.jpg.

LHO VE
david.barker@LIGO.ORG - posted 10:13, Saturday 28 March 2026 (89683)
Sat CP1 Fill

Sat Mar 28 10:07:24 2026 INFO: Fill completed in 7min 21secs

 

Images attached to this report
H1 CDS (CDS, OpsInfo)
keita.kawabe@LIGO.ORG - posted 09:23, Saturday 28 March 2026 - last comment - 09:21, Monday 30 March 2026(89682)
alarm handler is beeping about H1:PEM-CS_DUST_LAB2_300NM

We're getting <HIHI, MAJOR> alarm about H1:PEM-CS_DUST_LAB2_300NM_PCF, which I cannot trend using NDSCOPE, but I can plot H1:PEM-CS_DUST_LAB2_300NM_RAW and that number has been 30[V?] for the past 6 days.

I've silenced it but could somebody check?

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 10:21, Saturday 28 March 2026 (89684)

Not sure why this would be alarming today, it has been invalid for many weeks. The 0.3um raw counts went from 0 to 30 last Sunday 15:34 22mar2026, but it looks like the calibrated counts-per-cubic-foot has been NAN or INF for a very long time, which is why it cannot be trended with ndscope.

Perhaps the alarms system was recently restarted and DUST2 is regularly acknowledged on startup?

Images attached to this comment
ryan.crouch@LIGO.ORG - 09:21, Monday 30 March 2026 (89691)

This station (PCAL lab/LAB2) doesn't even have a device connected currently. There have been connections issues here for the past few months that I haven't tracked down yet. I'm not sure why it would have restarted to alarm, but I'll comment this station out of the ioc startup for now.

H1 ISC (SUS)
jenne.driggers@LIGO.ORG - posted 17:43, Friday 27 March 2026 (89679)
Alignment to get PRX flashes

[Jenne, Jennie, Keita, Jim, Sheila, Betsy]

Sum up: We're aligned pretty similarly to a reference time of [ 14 March 2026  02:20:00 UTC ], which is an "IR Found" time in between DRMI locks.  We're back in the same place on IM4 Trans QPD as well as both forward-going POP QPDs.  We have some mediocre PRX flashes.  We are back (in the REFL path) on the same 3 of 4 PDs that we were before re-venting, so we should be in a good place for the team to work on REFL path things tomorrow.

Some details:

The 4 attachments have vaguely descriptive names. The one from March 14th is my reference time, and then probably the other one that is useful to look at is the End of Day where I left thingsI'm not sure I was seeing PRX flashes at the very very end of the day, so PRM might need some adjustment in the morning. The other two attachments are showing where the suspensions were before I started, as well as showing that I first got back to the reference time osem positions

Images attached to this report
H1 IOO (ISC)
jennifer.wright@LIGO.ORG - posted 17:42, Friday 27 March 2026 - last comment - 17:44, Friday 27 March 2026(89680)
Re-aligned JAC IOT1 table with HAM1 in air

Jennie W, Keita, Betsy, Jason,

 

Betsy and Jason rolled up IOT1 this afternoon and Keita tweaked the alignment from the top periscope mirror to the DCPDs with the PSL output at 2W. He also re-aligned the trigger PD.

We can now lock JAC in air with the double door cover so we can have the light exiting the chamber but most of the door covered above and below where the beam exits.

I had to run the 'normalise signals guardian state after we closed the table and turned the light off, probably because the signal on REFL PD changed.

We left the light pipe closed and both doors covered at the end of the day.

NB: I accidentally pinged the HAM1 -Y side large table beam dump plate (the one that would catch the scatter from the JAC REFL viewport) with one of the small paper clips we use to hold the two door covers to each other. I will wipe this down tomorrow.

Comments related to this report
jennifer.wright@LIGO.ORG - 17:44, Friday 27 March 2026 (89681)

I turned the power down to 100mW at the end of the day using the rotation stage.

H1 ISC (SUS)
elenna.capote@LIGO.ORG - posted 17:34, Friday 27 March 2026 (89674)
OMCA and OMCB flashing, OMCB path needs more work

Today, Keita and Oli turned up the power to the fiber again so we are getting about 30 mW out. This is enough power that you can see the beam transmit the input coupler of the OMCs easily on a card. With the room lights dim, you can see flashing of the beam at both DCPDs with a card. Using the IR sensitive camera, we are able to fine tune the input alignment further.

The beam has been well-aligned to OMCA, and the flashing present at the DCPDs looks mostly like mode mismatch, which makes sense because we made no effort to mode-match our input beam to the OMCs.

The beam alignment to OMCB still needs some work. It looks predominantly like a yaw misalignment, so we need to make some finer adjustments of the upstream optics.

Since all of the BHSS optics have been pre-angled using the clocking tool, I followed the map on slide 86 of T2100190 to place the lenses, mirrors and beam dumps. Just by eye, I can tell we will need to fine tune this alignment further once we have laser beam. I was unable to place one beam dump because it goes roughly where we are pooling the OMCA cables. Once we can route and clamp those cables properly, there will be room for the dump. I haven't placed the QPDs yet, they are still bagged. I also placed the three head optics randomly, since they can't be placed until the OMC iris template is removed.

As a reminder, the order of operations here is to use the iris alignment template to get good input alignment to both OMCs. Then, this template is removed so we can use other precision templates to align the three optics that go at the head of the manta ray: mirrors A/B-R1 and 36-R1. Then, we can finish the rest of the beam paths to the OMC QPDs.

I took a phone video of OMCA flashing and managed to screengrab these two photos from the videos: DCPDA, DCPDB

 

Images attached to this report
H1 SQZ
sheila.dwyer@LIGO.ORG - posted 17:12, Friday 27 March 2026 (89675)
Looking at SQZ beam today, refl beam shape varies with ZM2 psams

I returned to the slider positions where we had good green and IR flashes in the filter cavity from Wed 89629.  Since this time of good flashes the H:M2 pico was reverted to 0 (it's setting during in chamber work), so the green flashes weren't as high as they were on Wed.  Screenshot attached here. In this alignment I went to SQZT7 and looked at the green reflection, it looks like an oval but is coming to the persicope.  (photo)

While at the table, I moved the picomotor for H:M2 X to -710, which is where Tony and I had put it 89596 and where we had both IR and green flashes good.  I could see the beam move on SQZT7, but it wasn't a big move, and after pico'ing it looks like we still have an oval beam making it to the table..  (photo, screenshot)  On the cameras we can clearly see cavity flashes in green but the IR flashes were off the camera, so I went to FCES and centered the flashes on the IR camera (using the seed).  I also went back to sqzt7 and aligned the FC green refl onto the diode.  

While I was there, I watched the green beam on a card while I moved suspensions to see how much freedom we have to move the suspensions and still get the green beam out of the viewport.  I could move ZM1,2,3 and FC1 by a couple hundred urad in both pitch and yaw (although not in either direction), so there is a reasonable adjustment range.  Moving these didn't impact the oblong shape of teh beam, so it doesn't seem like that is caused by clipping.  I then moved the ZM2 psams settings, and saw a big change in beam shape, it became more round looking on a card for strain guages of 4.15 or 5.15 (our usual setting is 3.15).  

I left the alignment back at the one that gave good flashes for IR and green (up to 10, is off in pitch), and with the CLF IR flashes are max 0.15 counts.  

Images attached to this report
H1 General
ryan.crouch@LIGO.ORG - posted 16:29, Friday 27 March 2026 (89665)
OPS Friday Day shift summary

TITLE: 03/27 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: The +/- Y doors were taken off HAM1 this afternoon alog89672, See Jims' alog89673 for the current CS seismic configuration. I'm not sure if the intercom is working, I didn't hear myself after plugging it in and trying the offices. THe LVEA is now is LASER HAZARD, as is the optics lab.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
14:28 FAC Kim, Nellie LVEA N Tech clean, Nellie out @ 15:12 15:19
14:58 FAC Randy LVEA N Prep for door removal 15:26
15:51 VAC Travis LVEA N Check HAM1 purge air 15:57
16:05 FAC Mitch LVEA N Cleanup in the Wbay 16:22
16:07 EE Fil, Marc LVEA N Cable pulling for moving electronics, HAM3/BSC2 17:09
16:22 VAC Nellie LVEA N Tech clean 17:11
16:36 VAC Gerardo, Randy LVEA N HAM1 door removal! Travis to join 18:33
16:41 FAC Jason Optics lab, LVEA LOCAL/N Property hunt 17:21
16:50 VAC Betsy, Mitch, Travis LVEA N HAM1 doors, Mitch Betsy out 1700 18:30
16:51 FAC Kim LVEA N Tech clean 17:44
17:02 SEI Jeff, Sina Optics lab LOCAL SPI work 18:07
17:03 BHD Oli Optics lab LOCAL BHSS work, magnet pol checks 20:16
17:12 EE Fil, Marc MidY N Cable search 19:53
17:32 OPS Ibrahim Optics lab N Put away parts before curtains 17:46
17:35 FAC Jason PCAL lab N Property inventory 17:39
17:38 OPS Betsy, Jim LVEA N Cleanroom move  by HAM1, Betsy out 1746 18:43
17:53 EE Thomas Vac Prep LOCAL In-Vac cables 19:19
18:32 SUS Ibrahim Vac-Prep LOCAL Joining Thomas 19:19
18:55 BHD Elenna, Keita Optics lab LOCAL BHSS, Keita out 19:37 20:16
19:20 SEI Jim, Betsy LVEA N LOCK HAM1 ISI 20:07
19:43 FAC Randy LVEA N -> Y Work on emod, handrails 22:46
19:49 SPI Jeff, Sina, Jason Optics lab/PSL encl LOCAL Measure fiber power, in lab checks then PSL enc checks, Jason out of PSL at 20:40 21:25
19:53 EE Fil, Marc LVEA N ->Y Cables, all around HAM3 and BSC2 23:10
19:56 SQZ Sheila LVEA LOCAL turn on SQZ laser, potential transition to hazard at SQZT7 20:06
20:11 EE Ibrahim, Thomas Vac Prep N In-Vac cabling 20:18
20:37 EE Ibrahim, Thomas, Sophie Vac Prep N In-Vac cable making, Ibrahim 23:15 Ongoing
20:20 ISC Betsy LVEA N Look into putting tables back 21:39
20:48 ISC Sheila LVEA N Talk to Fil and Marc, Turn on SQZ laser 21:03
21:02 TCS Sophie Vac prep N CHETA work 22:56
21:22 ISC Betsy Optics lab then LVEA N Parts/tools checks then LVEA 21:56
21:22 VAC Travis, Bram LVEA N Look at laser stop buttons 21:27
21:24 BHD Elenna Optics lab LOCAL Turn on the laser, cont BHD work Ongoing
21:31 OPS Sheila LVEA N -> Y Laser Hazard transition 21:47
21:58 SQZ Sheila LVEA Y SQZT7 work, alignment 22:15
21:59 OPS Jenne LVEA Y Look for Fil/Marc 22:14
22:06 ISC Keita, Jennie, Betsy LVEA Y HAM1 work Ongoing
22:08 SEI Jeff LVEA N Find  part 22:28
22:09 ISC Jason LVEA Y Join HAM1 team 22:21
22:38 SQZ Sheila LVEA then FCES Y SQZT7 check before FCES 23:06
23:21 SQZ Sheila LVEA Y SQZ7 checks Ongoing
23:29 SEI Jim LVEA Y Unlock HAM3 HEPI Ongoing
Images attached to this report
H1 PSL
ryan.crouch@LIGO.ORG - posted 15:52, Friday 27 March 2026 (89678)
PSL Status Report - Weekly

Closes FAMIS39757

Laser Status:
    NPRO output power is 1.83W
    AMP1 output power is 70.39W
    AMP2 output power is 138.8W
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 20 days, 23 hr 13 minutes
    Reflected power = 26.87W
    Transmitted power = 104.4W
    PowerSum = 131.3W

FSS:
    It has been locked for 0 days 2 hr and 21 min
    TPD[V] = 0.4974V

ISS:
    The diffracted power is around 3.9%
    Last saturation event was 0 days 2 hours and 25 minutes ago


Possible Issues:
    PMC reflected power is high

H1 SEI
ryan.crouch@LIGO.ORG - posted 15:47, Friday 27 March 2026 (89677)
H1 ISI CPS Noise Spectra Check - Weekly (FAMIS #39341)

Closes FAMIS39341

As expected from the vent and HEPI's being locked on HAMs 1,2,3 and BSC 2,3 the plots look noisier. I don't see any glaring issues with the other chambers.

Non-image files attached to this report
H1 SUS (CDS)
jenne.driggers@LIGO.ORG - posted 15:33, Friday 27 March 2026 (89676)
HAM1 DAC Kill also affected HAM2 suspensions

[Jenne, DaveB, RyanC, Jason, Betsy, Jennie, Keita, Fil, Marc]

Several folks prepping to align optics for HAM1 work noticed at about the same time that several HAM1 and HAM2 suspensions were not behaving as expected.  Eg, I tried misaligning and aligning PR3, but nothing happened (I did this because the sliders hadn't changed, but the OSEM readbacks had changed).  The team prepping to go into chamber similarly noticed that there were no counts going out to the IOP models.

Fil and Marc had a quick look at the sat amps for the HAM2 suspensions (just in case, even though no one had been near them), and everything looked fine.

Upon going back to the control room, I finally noticed on the CDS overview screen that IOPSUSH12 had its DAC Kill light illuminated. After some clicking around, we found the dac kill watchdog screen, and hit RESET.  However, PM1's RMS was still larger than threshold, because HAM1 is at air and the purge air is on, so its undamped motion was too high.  I clicked BYPASS on PM1, which let signal go through to damp PM1.  PM1 quieted down, and I un-bypassed it by hitting RESET again. 

At this point, it seems like all of the suspensions on HAMs 1 and 2 are again behaving as expected.

X1 SUS (EPO, SUS)
ryan.crouch@LIGO.ORG - posted 13:50, Friday 27 March 2026 (89651)
HXDS - OMA2, OMA3 triples lab assembly and test stand M1 actuation measurements

Rahul, RyanC

Over the past few weeks we've prepped the two basic HXDS assembly (structure sn 035, alog88828 - OMA3 with B2-03 optic, and structure sn 029, alog87998 - OMA2 with A1-03 optic) that we received from LLO. Rahul and Camille assembled these PSAMs at CIT. We attached the wires using the provided HDS Bottom Mass Wire Plate ASSY D2000395 setting the top blade height first using the blade height gauge then feeding the wire down, and the bottom mass (PSAM) with its fixture assembly plate. The wire clamps really wanted to clock when I was doing the final turn, turn and a half on their fasteners so I had to tighten them very slowly. For OMA2 we had some trouble trying to remove the intermediate mass fixture plate D2000230 to replace it with the BOSEM Cam Adj Mtg ASSY as there was a stuck dowel pin. The fixture plate was a V1 version which did not have the emergency release feature that the later version has. We brute forced it out without damaging any components or wires but the intermediate mass shifted a decent amount despite my efforts to keep it still. OMA3 did not have this issue as it was as the V2 plate.

Once we had OMA2 fully assembled, we balanced it then brought it over to the test stand and connected the four OSEMs to the top stage. This first suite of phase 1a measurements was done without the M2 PSAMs cable connected.

The OLV counts were as follows (we later swapped LR and UR, which is reflected below) for OMA2:

UL: 28460 UR: 31900

LL: 29720 LR: 30860

The data/templates live in: /ligo/svncommon/SusSVN/sus/trunk/HXDS/X1/OMA2/SAGM1/Data

and the result pdfs are in: /ligo/svncommon/SusSVN/sus/trunk/HXDS/X1/OMA2/SAGM1/Results

I ran 3 sets of measurements on Wed with a bw of 0.02. The first measurement 2026-03-25_1730 showed some weirdness with LR and lots of cross coupling from Length into Pitch and a bit into Yaw, the shapes were a little different, as well along with the magnitude but that's probably from the coil driver gain. Rahul then swapped the LR and UR OSEMs and double checked the centering, and I ran another measurement 2026-03-25_1900, which looked about the same. Rahul then went and improved the balancing as best he could (the masses were pitching a bit) and I ran a final measurement 2026-03-25_2130.

Overall the Transfer Functions look ok but there's still room for improvement. 

On Thursday Rahul balanced and set up OMA3 on the test stand. Unfortunately while manipulating OMA2 one of the bottom M2 stage wires broke, so we will have to replace it and remeasure it next week.

The OLV counts were as follows for OMA3:

UL: 27030 UR: 32100

LL: 27500 LR: 28900

The data/templates live in: /ligo/svncommon/SusSVN/sus/trunk/HXDS/X1/OMA3/SAGM1/Data

and the result pdfs are in: /ligo/svncommon/SusSVN/sus/trunk/HXDS/X1/OMA3/SAGM1/Results

This set of transfer functions, 2026-03-26_1900 also looks ok and similar to OMA2, see the comparison plot of OMA2 and OMA3. Both measurements shows some weirdness with LR that is not seen in an OSEM noise spectra that we're still looking into.

The next steps are to repair the bottom wire on OMA2, attach PSAMs cables, remeasure then fine tune both OMA2 and OMA3.

Tagging EPO for Corey for the pictures. I'm realizing I only have pictures of OMA3, I might comment with more photos on Monday of OMA2.

Images attached to this report
Non-image files attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 13:20, Friday 27 March 2026 (89672)
HAM1 Both Doors are Removed

(Travis, Randy, Fil, Betsy, Marc, Jim, Gerardo)

Both doors for HAM1 are off, +Y and -Y. 
    - Y Door we had one of the O-rings get stuck to the door, used the tool to set the O-ring back into its groove without a problem.  No further issues.
    + Y door, we took a peek at the O-rings and they looked suspiciously good all over, even with a small gap, but then out of the sudden one of the O-rings jumped out and dropped to the floor.  So, the +Y door has the flange protectors over a single O-ring.

Purge air was turned up.  A photo of the dew point is attached, this as measured this morning by Travis.

A big "thank you" for those that helped with the cleanroom moves.

Images attached to this report
H1 SEI (OpsInfo)
jim.warner@LIGO.ORG - posted 13:13, Friday 27 March 2026 (89673)
SEI status for vent Mar 27

As of right now, HAM1-2-3 and BSC 2 & 3 HEPI are locked, and after doors came off HAM1 ISI is also locked. HEPIs are all bypassed in the model (by setting ISO master gain to 0). HAM1 should just be left tripped for the work in that chamber, HAM 2 &3 and the BSCs should be left with damping loops engaged, unless there is a need for better alignment to the IFO.

H1 SQZ
camilla.compton@LIGO.ORG - posted 16:40, Wednesday 25 March 2026 - last comment - 13:36, Friday 27 March 2026(89645)
SQZ into IFO, Vented alignment not making it to HAM6 or SQZT7.

Sheila, Camilla.

We decided to pause on looking for the FC alignment and look at SQZ alignment into the IFO. To check our irises and possibly look for scattering shelves. 

Locked SEED_DITHER with 75mW into fiber. Moved ZM3 to get OPO_IR PD up to max 0.9mW. Opened BeamDiv and after aligning SRM, mis-aligning SR2, could see beams on AS_A,B and C. AS_A and B had a similar size ~60 to reported before (e.g. 86965), but there was less power on AS_C: 4e-4 rather than the previous 7e-4.

We used Ryan's DRMI time from 89573 to put OM1, OM2 and SRM back. The with the SQZ SEED injected, we tried to use SQZ centering loops to center on AS_A and AS_B, this didn't work quite as expected. ZM4 kept saturating although these loops had worked Janaury 19th, before moves were done on the VOPO to help with clipping e.g. 88966.

Then went back to alignment on Jan 31st with 88976 with ZM1,2,3, 4,5,6 FC1. And took OM1,2,SRM to the January 19th positions. However here we saw no light on SQZT7 OPO IR PD, HAM7 WFs or in HAM6 on the AS_A,B,C diodes. This was done at 23:10 UTC. Comparisons of osems attached. The sliders to get these osems values are fairly different, see attached, so I also tried going to these slider values, that gave me some light on SQZT7 IR PD, HAM7 WFS and FCES with SEED injected, see attached. HAM6 AS diodes saw nothing. 

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 10:58, Thursday 26 March 2026 (89657)

Sophie, Camilla

This morning, starting from this vented slider alignments, I randomly moved ZM1 until we had IR and green light on the FCES PDs. I misaligned FC1 and FC2 to single pass light. Then I could move ZM3 in pitch from P,Y (-324,-200) to (-506, -200) to get max light on IR FCES  PD. In both these positions, we get scattered light on the camera and light on both IR and green PDs. This makes me think the best position in pit h is in the middle of these, ZM3 (-415, 200). We tried to repeat this excise in yaw but only could get the one edge. Even after movingZM1 in YAW in both directions, I couldn't find the other side of the yaw clipped beam. Attached is the sliders when we thought we were centered in pitches and clipping the coating on one side in yaw. 

Images attached to this comment
sheila.dwyer@LIGO.ORG - 13:36, Friday 27 March 2026 (89658)

We need 4 things for the squeezer to operate:

  • Filter cavity green sensing (FCGS) reaching SQZT7
  • IR reflected off filter cavity making it back through VIP without clipping (IR on SQZT7)
  • IR flashes in filter cavity
  • Green flashes in filter cavity

Continuing from where Sophie and Camilla left off, I confirmed that the green I see on the FCES green camera is a single pass beam by misaligning both FC1 and FC2, they do not move this beam, and also confirmed that it's green by toggling the fiber input.  I moved ZM3 to bring this to center on the camera, as we have not disturbed this FCES green path since O4.  Then I did a raster on FC1 to get the IR seed reflected onto the SQZT7 IR PD (screenshot).  

After moving FC2, I could see some green and IR on the FCES cameras.  For IR, this is not the single pass beam but some reflections off FC1, for green we still see the single pass beam and a second beam coming close to the single pass beam, but it seems to have a very different beam size.  screenshot.  After this I went to SQZT7 and saw that the FCGS refl beam is making it to the table and aligned it onto the diode.  Now we have 2.5 of the 4 things we need for the squeezer to work: green on SQZT7, IR on SQZT7, and the green single pass is heading towards FC2 well, which is part of what we need to get green flashes.  

Sophie and I went to FCES.  We tried moving FC2 to see what happened on the green transmission, we can see the beam reflected off FC2 leaks around the edge of the FC2 coating for some alignments, but that we can overlap them for some alignments, but do not see cavity flashes. 

Camilla moved ZM3 back to the alignment where she found the green centered between the two places where it goes over the edge of the coating.  This was a simlar situation in green.  

In IR, we did see a single pass beam on the camera when Camilla steered ZM3.  In 986 ZM3 is in the position where Camilla thinks IR is centered on FC2, you can see two green spots, and to the right and slightly below a very faint IR beam (I had to clean the laptop screen, turn the brigthness up to see it).  In   988 ZM3 was at -506 P, -200 Y, the negative pitch clipping, you can see the brighter IR beam where it is transmitting through the uncoated part of FC2.  990 is ZM3 -324 P, -200 Y, going over the lower edge of the coating.  994 is with FC1 + FC2 in the aligned guardian state, moving FC2 until we saw something on the IR camera.  996 is for FC2 359 P, 28 Y, there was light on the IR diode, but the photo is dark.  

Images attached to this comment
H1 ISC (SUS)
oli.patane@LIGO.ORG - posted 10:17, Tuesday 24 March 2026 - last comment - 11:05, Friday 27 March 2026(89615)
BHSS / BHD work Mar 23: OMCB almost done, OMCA ready for alignment!

Elenna, Oli

Summary: OMCB has its DCPDs installed and most of its stoppers (barring one that needs to be retapped + helicoiled). OMCA has been aligned according to the OMCA / OMCB install template and is ready for laser alignment. Cabling for both OMCs should be good to go.

After last week's issue of the OMCB DCPD A (TRANS) missing a diode hole (89583), we were able to clear up the issue and fix it (89606). Elenna then installed the DCPD B (REFL) diode. Note that once again there were tiny metal shavings below where the DCPD's were installed, which we picked up with a qtip that we wet with iso. Once we had both installed (pic), we installed the rest of the stoppers around OMCB. The only one we didn't install is one of the vertical stops. This is because there was an issue with the helicoil last week, and we ended up having to remove it. We think this might be due to an issue with the threads, so we are waiting on a 3/8"-24 tap to retap the hole before installing a new helicoil. Because we don't have this stop yet, it doesn't really make sense to fasten on the butter dish, but we might just sit it on today to hinder dust accumulation.

We also realized that the PZT cables were going to be way too short to reach the cable pylons for both OMCA and OMCB, so we very carefully opened the peak cable wraps on the OMCs and took out what we think should be enough. On OMCB we noticed that the little copper wrap for one of the PZTs was not right above the PZT as it should be, but instead was on the other side of the peak cable wrap and was only wrapped loosly around one of the cables. Because of this we just removed the wrap. The cables right above the PZT are looking fine and don't look like they need anything to hold them together.

I had been confused about the OMCA / OMCB install template for the past few days, since installing it was putting OMCB further back in its slot than OMCA, but I was able to confirm yesterday in eDrawings that OMCB is actually supposed to be an extra 1 mm away from the front of its slot as compared to OMCA, so we moved OMCA into position and bolted it down.

I also went in and installed the magnet mounts for the OSEM magnets (without the magnets), as well as the big mass on the back of the BHSS.

 

Installed serial numbers:

OMCB (SN 105)

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 08:46, Thursday 26 March 2026 (89652)EPO

Tagging For EPO photos.

elenna.capote@LIGO.ORG - 11:05, Friday 27 March 2026 (89671)

The naming convention listed above is incorrect. Please refer to alog 89669 for correct information.

H1 SUS (EPO, ISC)
oli.patane@LIGO.ORG - posted 18:07, Wednesday 18 March 2026 - last comment - 11:05, Friday 27 March 2026(89562)
BHSS / BHD work Mar 18: Input alignment finished and OMCA in cradle!

Elenna, Keita, Oli

Input alignment:

Today Elenna and I finished fine-tuning the alignment into the BHSS, so now our beams pass through both sets of irises on the alignment tooling plate, so they should be well aligned to go into the OMCs nicely.

OMCA in cradle:

We also opened up what is going to be OMCA (SN 103) and placed it in its cradle. We followed the installation instructions and placed the OMC 1mm away from the three installed horizontal stops, but then we noticed that lining the short end up against the side horizontal stop led to the OMC overhanging the cradle on that side by a few mm. We're not sure if these end stops are also supposed to be further inwards than 1mm (the instructions just specify that they (the horizontal stops) will be touching the OMC 1mm in), but if we placed the other end stop in with the way that the OMC is currently positioned, that other stop would have to screw in at least 3 or 4 mm to meet up with the OMC. We will add photos of this tomorrow.

Installing DCPDs onto OMCA:

Keita and Elenna (who was chained to the optics table via grounding strap) very carefully trimmed the DCPD wires for D1-20 and C1-12 and placed them in the DCPD mounts for DCPD A (which is what we think TRANS is), and DCPD B (which is what we think REFL is), respectively. Faceplate SN 10 was installed onto DCPD A, and faceplate SN 06 was installed onto DCPD B to hold the PDs in place. It was very hard to get the length of the DCPD pins right, and getting the faceplates on and secured. There seems to be a small gap in between the faceplates and the DCPD housing (pic1, pic2), but LLO has the same thing happening so it seems okay. 

Checking pins

After we used a multimeter to check that the case pins followed D220027. According to D220027 the case pins should be going to pins 6 and 9, but we found that they went to what we thought were pins 1 and 4 when we were looking at the prongs. Later we found that the DB9 pin layout is read the other way, so that means that it was actually pins 2 and 5. This still doesn't line up with D220027, but looking at the schematic for the cable D2300119, it turns out that the wires for the Cases for the two DCPDs are shifted so that the Cases do actually go to pin 2 and pin 5, so they're wired correctly! It's just that D220027 is wired up incorrectly and appearantly the cable pins are read with the prongs facing away from you for some reason.

OMCA is looking pretty good so far! (pic3, pic4)

Summary of installed serial numbers:

OMCA (SN 103)

Images attached to this report
Comments related to this report
oli.patane@LIGO.ORG - 09:17, Thursday 19 March 2026 (89564)

Here are photos of the OMC overhang when the OMC is 1mm away from the horizontal stop (set screw is in by 1mm), and photos showing the other end.

overhang side

overhang top

underhang side

underhang top

Images attached to this comment
oli.patane@LIGO.ORG - 10:10, Thursday 19 March 2026 (89565)

other angle

Images attached to this comment
elenna.capote@LIGO.ORG - 11:05, Friday 27 March 2026 (89670)

The naming designation listed above for DCPDA/B is incorrect. Please refer to alog 89669 for correct information.